IPL Home Use Hair Removal Device (SJ15)

K240093 · Shenzhen Shuge Medical Beauty Devices Co., Ltd. · OHT · Apr 24, 2024 · General, Plastic Surgery

Device Facts

Record IDK240093
Device NameIPL Home Use Hair Removal Device (SJ15)
ApplicantShenzhen Shuge Medical Beauty Devices Co., Ltd.
Product CodeOHT · General, Plastic Surgery
Decision DateApr 24, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

IPL Home Use Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Device Story

Over-the-counter, home-use, single-person device for hair reduction; utilizes Intense Pulsed Light (IPL) via Xenon Arc Flashlamp; emits light wavelengths 550-1200nm; light energy absorbed by melanin in hair shaft; heat disables hair follicle to prevent regrowth; operated by user; device powered by external adapter (100-240V); output is light pulse; benefits patient through long-term hair reduction.

Clinical Evidence

No clinical data. Bench testing only. Device verified against ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62471, and IEC 60601-2-83. Biocompatibility testing performed per ISO 10993-5, ISO 10993-10, and ISO 10993-23.

Technological Characteristics

Energy source: Xenon Arc Flashlamp; Wavelength: 550-1200nm; Energy density: 2.7-4.4 J/cm^2; Spot size: 3.15 cm^2; Power: 100-240V, 50/60Hz adapter; Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62471, IEC 60601-2-83; Biocompatibility: ISO 10993-5, -10, -23.

Indications for Use

Indicated for removal of unwanted hair and permanent reduction in hair regrowth in home-use settings.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. April 24, 2024 Shenzhen Shuge Medical Beauty Devices Co., Ltd. % Yvonne Liu Registration Specialist Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90 Qianhai Road Shenzhen, Guangdong 518100 China Re: K240093 Trade/Device Name: IPL Home Use Hair Removal Device (SJ15) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHT Dated: January 12, 2024 Received: January 12, 2024 Dear Yvonne Liu: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). {2}------------------------------------------------ Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Tanisha Digitally signed by Tanisha L. L. Hithe - Hithe -S S Date: 2024.04.24 15:49:25 -04'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Submission Number (if known) K240093 Device Name IPL Home Use Hair Removal Device (SJ15) #### Indications for Use (Describe) IPL Home Use Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ | 510(k) Summary | | | Prepared on: 2024-04-22 | |------------------------------------|-------------------------------------------------------------------------------------------------------------------|--------------|-------------------------| | Contact Details | | | 21 CFR 807.92(a)(1) | | Applicant Name | Shenzhen Shuge Medical Beauty Devices Co., Ltd | | | | Applicant Address | 6th Floor, Building B, Jingang Science Park, Qiaotou Community, Baoan<br>District Shenzhen Guangdong 518103 China | | | | Applicant Contact Telephone | 86-15712144494 | | | | Applicant Contact | Mr. Shengbiao Lin | | | | Applicant Contact Email | sales@yeahone.net | | | | Correspondent Name | Feiying Drug & Medical Consulting Technical Service Group | | | | Correspondent Address | Rm 2401 Zhenye International Business Center, No. 3101-90,<br>Qianhai Road Shenzhen Guangdong 518100 China | | | | Correspondent Contact Telephone | 86-17780639776 | | | | Correspondent Contact | Ms. Yvonne Liu | | | | Correspondent Contact Email | lq1399@feiying-china.com | | | | Device Name | | | 21 CFR 807.92(a)(2) | | Device Trade Name | IPL Home Use Hair Removal Device (SJ15) | | | | Common Name | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology | | | | Classification Name | Light Based Over-The-Counter Hair Removal | | | | Regulation Number | 878.4810 | | | | Product Code(s) | OHT | | | | Legally Marketed Predicate Devices | | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | | K231717 | IPL Home Use Hair Removal Device | OHT | | | K221002 | IPL Hair Removal Device | OHT | | | Device Description Summary | | | 21 CFR 807.92(a)(4) | IPL Home Use Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. IPL Home Use Hair Removal Device (Model:515) is an over-the-counter, home-use and single-person-use device for hair reduction by {5}------------------------------------------------ using Intense Pulsed Light (IPL)to emit a specific wavelength of the light ranging from 550-1200m and delivers to the skin. The device is designed to help break the cycle of hair re-growth. Light energy is transferred through the skin's surface and is absorbed by melanin present in the hair shaft. The absorbed light energy (below the surface of the surface of the skin), which disables the har follicle preventing further growth, so as to achieve effective hair removal. # Intended Use/Indications for Use IPL Home Use Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. # Indications for Use Comparison The subject device and predicate devices have the same indications for use, which is used for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 3, 6, and 12 months after the completion of a treatment regime. # Technological Comparison The technical characteristic of PL Home Use Hair Removal Device (5/15) are substantially equivalent to the predicate devices in the following aspects: 1) the same intended use, mode of action; 2) the same source energy and power supply: supplied by external adapter with 100-240V~, 50/60Hz; 3) the same light source: Intense Pulsed Light; 4) the same energy medium: Xenon Arc Flashlamp; 5) similar wavelength: The subject device's wavelength is 500-1200m while the primary predicate device is 550-1100mm, 600-1100mm and 640-1100nm but the wavelength of secondary predicate device is 550-1200nm. The difference between the subject device and the predicate devices mainly includes the following: 1) different spot size of the subject device is 3.15 cm^2 while that of the primary predicate device is 3cm^2, 3.3cm^2 and 4cm^2. The spot size is related to light intensity and since the difference in light intensity is not significant, so this difference will not raise any safety or effectiveness issue. 2) different energy density: the energy density of the subject device is 2.7-4.4 /cm^2 while that of the primary predicate device is 2-5 J/ cm^2 and 2-6cm^2. The energy density of subject device is within the range of the minimum value of the predicate devices, and they all comply with IEC 60601-2-83 and IEC 62471 requirements, so this difference will not raise any safety or effectiveness issue. Thus the technology comparison supports a decision of substantial equivalence to the predicate devices. #### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Non-clinical testing have been conducted to verify that the PL Home Use Hair Removal Device meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicatedevices. The testing results demonstrate that the subject device complies with the following standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 Consolidated Text)(Including Amendment 2(2012)] Medical electrical equipments for basic safety and essential performance - Amendment 2. IEC 60601-1-2 Edition 4.1 2020-09 Consolidated Version, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests. IEC 60601-1-11 Edition 2.1 2020-07 Consolidated Version, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Sequiremnts for medical electrical equipment and medical electrical systems used in the home healthcare environment. IEC 62471 First edition 2006-07, Photobilogical safety of lamps and lamp systems. IEC 60601-2-83, Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment The device has been tested for biocompatibility, it complies with the following standards: 21 CFR 807.92(a)(5) ## 21 CFR 807.92(a)(5) ## 21 CFR 807.92(a)(6) {6}------------------------------------------------ ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices, Part 5: Tests for in vitro cytotoxicity, ISO 10993-10 Fourth edition 2021-11, Biological evaluation of medical devices, Part 10: Tests for skin sensitization ISO 10993-23 First edition 2021-01,Biological evaluation of medical devices, Part 23 Tests for irritation We've also conducted: Software verification and validation test according to the FDA "Content of Premarket Submission for Device Software Functions" The clinical test is not applicable, there's no clinical data. The subject device and predicate device have indications for use and similar technological characteristics. The subject device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...