Hee HR Mini

K170269 · Oriental Inspiration Limited · OHT · May 3, 2017 · General, Plastic Surgery

Device Facts

Record IDK170269
Device NameHee HR Mini
ApplicantOriental Inspiration Limited
Product CodeOHT · General, Plastic Surgery
Decision DateMay 3, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

HEE HR Mini is an over-the-counter device intended for the removal of unwanted hair. The HEE HR Mini is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of treatment regimen.

Device Story

HEE HR Mini is a handheld pulsed light hair removal device for OTC use. It utilizes selective photothermolysis; optical energy heats hair follicles to interrupt growth cycles. Device includes lamp, skin proximity sensor, and skin color sensor. Input: user skin contact and skin tone detection. Operation: device triggers pulse only when in full contact with skin and if skin tone is within safe range (up to Fitzpatrick 4). Output: light pulse. User interface includes on/off button, flash button, intensity buttons, and LED indicators. Powered by external supply. Benefits: long-term hair reduction. Healthcare provider role: none; device is for patient self-use.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993-1), electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57, IEC 62471), and EMC (IEC 60601-1-2).

Technological Characteristics

Handheld pulsed light device; 3 cm² spot size; 475-1200 nm spectrum; 5 J/cm² max fluence; 8 user-selectable energy levels. Materials: ABS/PC plastic shell, optical glass filter. Safety features: skin proximity and pigmentation sensors. Power: 100-240 VAC. Standards: IEC 60601-1, 60601-1-11, 60601-2-57, 62471, 60601-1-2.

Indications for Use

Indicated for removal of unwanted hair and permanent reduction in hair regrowth in over-the-counter users. Contraindicated for skin tones darker than Fitzpatrick scale 4.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 3, 2017 Oriental Inspiration Limited Andy Wu Director Unit 607, The Lakeside 2, West Wing, No. 10 Science Park West Avenue Hong Kong, CN Re: K170269 Trade/Device Name: Hee Hr Mini Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: October 17, 2016 Received: January 27, 2017 Dear Andy Wu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing. {1}------------------------------------------------ with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K170269 Device Name HEE HR Mini ### Indications for Use (Describe) HEE HR Mini is an over-the-counter device intended for the removal of unwanted hair. The HEE HR Mini is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of treatment regimen. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Department of Health and Human Services Centre of Device and Radiological Health Office of Device Evaluation Traditional 510(k) section # 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION ### as required by section 21 CFR 807.92 ## 1. Submitter of 510(K): | Company Name: | Oriental Inspiration Limited | |-------------------|--------------------------------------------------------------------------------------------------------------------------| | Address: | Unit 607, The Lakeside 2, West Wing, No. 10 Science Park West<br>Avenue, Hong Kong Science Park, Shatin, N.T. Hong Kong. | | Contact person: | Mr. Andy Wu | | TEL: | +(852) 2687 3496 | | FAX: | +(852) 2687 4148 | | E-mail: | andy_wu@netop.com.hk | | Date of Prepared: | 10/20/2016 | #### 2. Proposed Device and code: | Device Trade Name: | HEE HR Mini | |--------------------|---------------------------| | Regulation Medical | General & Plastic Surgery | | Specialty | | | Product Code: | OHT | | Regulation number | 878.4810 | | Device Class | 2 | #### Predicate Device: 3. | 510(K) | Trade or Proprietary or Model Name | Manufacturer | |---------|---------------------------------------------|-----------------------| | K140381 | sensiLight (sensiLight and sensiLight Plus) | EL Global Trade Ltd. | | K141242 | Glide device | Home skinovations Ltd | #### 4. Description of Proposed Device: The HEE HR Mini devices are pulsed light hair removal device. Light-based hair removal is based on the theory of selective photothermolysis in which optical energy is used to disable hair growth. The HR Mini devices are composed of a hand held applicator and an external power supply. The spot size (treatment area) in the HR Mini devices is 3 cm². The device contains a lamp, a skin proximity {4}------------------------------------------------ sensor and a skin color sensor to detect appropriate skin tones. If the HR Mini is not properly applied (in full contact with the skin) or user skin tone is too dark/tanned, the HR Mini will not trigger a pulse. The device has an on/off button, skin color sensor, light exit window, flash button, led indicator, two intensity button, socket for battery charging. #### 5. Intended for Use HEE HR Mini is an over-the-counter device intended for the removal of unwanted hair. The HEE HR Mini is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of treatment regimen. #### 6. Technical and Performance The following table compares the device to the predicate device with basic technological | characteristics. | |------------------| |------------------| | Parameter | New device | Predicate device | Predicate device | Remark | |--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------| | 510(k)<br>Number | (to be assigned) | K140381 | K141242 | -- | | Device Name<br>and Model | HEE HR Mini | sensiLight (sensiLight and<br>sensiLight Plus) | Glide device | -- | | Manufacturer | Oriental Inspiration<br>Limited | EL Global Trade Ltd. | Home skinovations Ltd | -- | | Regulation<br>Number | 878.4810 | 878.4810 | 878.4810 | SE | | Intend of use | HEE HR Mini is an<br>over-the-counter device<br>intended for the removal<br>of unwanted hair. The<br>HEE HR Mini is also<br>intended for permanent<br>reduction in hair<br>regrowth, defined as a<br>long-term, stable<br>reduction in the number<br>of hairs regrowing when<br>measured at 6, 9, and 12<br>months after the<br>completion of treatment<br>regimen. | The sensiLight is an over<br>the counter device<br>intended for the removal<br>of unwanted hair. The<br>sensiLight is also intended<br>for permanent reduction<br>in hair regrowth, defined<br>as a long-term, stable<br>reduction in the number<br>of hairs re-growing when<br>measured at 6, 9, and 12<br>months after the<br>completion of treatment<br>regimen. | The Glide device is an over<br>the counter device<br>intended for the removal o<br>unwanted hair. The Glide<br>device is also intended for<br>permanent reduction in<br>hair regrowth, defined as a<br>long-term, stable reduction<br>in the number of hairs<br>re-growing when measured<br>at 6, 9, and 12 months afte<br>the completion of<br>treatment regimen. | SE | | Operation | Photothermolysis. Light | Photothermolysis. Light | Photothermolysis. Light | SE | | Principle | energy is absorbed, heats<br>the hair follicles and<br>interrupts the growth cycle. | energy is absorbed, heats<br>the hair follicles and<br>interrupts the growth cycle. | energy is absorbed, heats<br>the hair follicles and<br>interrupts the growth cycle. | | | Technology | Home Pulsed Light | Home Pulsed Light | Home Pulsed Light | SE | | Comparison<br>of treatment<br>regime | • The 1-4 treatments<br>will be approximately<br>two weeks apart<br>• Followed by 5-7<br>treatments plan four<br>week time intervals.<br>• Treatments 8 and<br>after: Treat as needed<br>until decided results are<br>achieved | • The 1-4 treatments will<br>be approximately two<br>weeks apart<br>• Followed by 5-7<br>treatments plan four week<br>time intervals.<br>• Maintenance will be<br>needed from time to time<br>if growth is still visible | • The 1-4 treatments<br>will<br>be approximately two<br>weeks apart<br>• Followed by 5-7<br>treatments plan four<br>week time intervals.<br>• Treatments 8 and<br>after: Treat as needed<br>until decided results are<br>achieved | SE | | Safety | IEC 60601-1:2005<br>IEC 60601-1-11:2010<br>IEC 60601-2-57:2011<br>IEC 62471:2006<br>IEC 60601-1-2:2013 | IEC 60601-1:2005<br>IEC 60601-1-11:2010<br>IEC 60601-2-57:2011<br>IEC 62471:2006<br>IEC 60601-1-2:2013 | IEC 60601-1:2005<br>IEC 60601-2-57:2011<br>IEC 60601-1-2:2013<br>IEC 62471:2006 | SE | | Performance Features | | | | | | Optical<br>aperture | 3 [cm²] | 3 [cm²] | 2.7 [cm²] | Same as<br>K140381, and<br>different with<br>K141242,<br>Note 1 | | Emitted<br>Spectrum | 475-1200 [nm] | 475-1200 [nm] | 475-1200 [nm] | SE | | Max energy<br>level<br>(fluence) | 5 [joules/cm²] | 5 [joules/cm²] | 5 [joules/cm²] | SE | | User<br>selectable<br>Energy levels | 8 | 3 | 5 | Different<br>with<br>K141242, and<br>K140381,<br>Note 1 | | Single pulse<br>duration | 500-800 [uSec] | 500-800 [uSec] | 500-800 [uSec] | SE | | Use<br>Classification | OTC | OTC | OTC | SE | | | 100-240 VAC, 50-60Hz | 110-240 VAC, 50-60Hz | 100-240 VAC, 50-60Hz | Note 1 | | | 0.8A | to 19VDC, 1.8A | 2A | | | Safety Features | | | | | | Close contact<br>detector | Prevent the activation of<br>the device while not in<br>close contact with the<br>skin. | Prevent the activation of<br>the device while not in<br>close contact with the<br>skin. | Prevent the activation of<br>the device while not in<br>close contact with the<br>skin. | SE | | Pigmentation<br>level detector | Optical - does not allow<br>activation on dark skin | Optical - does not allow<br>activation on dark skin | Optical - does not allow<br>activation on dark skin | SE | | Darkest skin<br>allowed | Fitzpatrick 4 | Fitzpatrick 4 | Fitzpatrick 4 | SE | | Materials | Top, Side and Bottom<br>shell - ABS/PC plastic<br>(Acrylonitrile butadiene<br>styrene/PolyCarbonate) | Top, Side and Bottom shell -<br>ABS/PC plastic (Acrylonitrile<br>butadiene<br>styrene/PolyCarbonate) | Device outer body -<br>ABS/PC plastic<br>(Acrylonitrile butadiene<br>styrene/PolyCarbonate) | SE | | | Optical Filter - Color<br>glass which purpose is of<br>an optical long pass<br>filter tested for<br>biocompatibility | Optical Filter - Color glass<br>which purpose is of an<br>optical long pass filter tested<br>for biocompatibility | Optical Filter - Color glass<br>which purpose is of an<br>optical long pass filter<br>tested for<br>biocompatibility | | | Overall Design | The HEE HR Mini device<br>is light based hair<br>removal system<br>composed of a<br>hand-held applicator. | The sensiLight device is<br>light based hair removal<br>system composed of a<br>hand-held applicator. | The Glide device is light<br>based hair removal<br>system composed of a<br>hand held applicator. | SE | {5}------------------------------------------------ {6}------------------------------------------------ # Note 1 Although it is a little different from the predicate devices, it will not affect the main function and the intended use of the device. They all also comply with safety and performance requirements. So the differences will not raise any safety or effectiveness issue. #### 7. PERFORMANCE DATA The testing for HEE HR Mini included performance, software, electrical safety, EMC, biocompatibility and bench. HEE HR Mini passed all testing in support of the substantial equivalence determination: #### 7.1. Biocompatibility testing The biocompatibility evaluation for the HEE HR Mini was conducted in accordance with International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation {7}------------------------------------------------ and Testing Within a Risk Management Process." As dictated by the application and duration of contact with the intact skin, the enclosure of testing included the following tests: - Cytotoxicity - . Sensitization - Irritation #### 7.2. Electrical safety and electromagnetic compatibility Electrical safety, performance and EMC testing were conducted on the HEE HR Mini. The device with compliance with the IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57 and IEC 62471 standards for safety and the IEC 60601-1-2 standard for EMC. # 7.3. Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern. #### 8. Conclusions: The proposed device has the same intended use and similar characteristics as the predicate device, the sensiLight (sensiLight and sensiLight Plus) (K140381) and Glide device (K141242). Meanwhile, performance testing, bench testing, and safety report documentation supplied in this submission demonstrates that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Based on performance testing, the proposed device is Substantially Equivalent (SE) to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...