K252308 · Atop Lighting Co., Ltd. · OHS · Nov 13, 2025 · General, Plastic Surgery
Device Facts
Record ID
K252308
Device Name
LED Beauty Mask (RLBXX)
Applicant
Atop Lighting Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Nov 13, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LED Beauty Mask is an over-the-counter device that is intended for use in the treatment of full-face wrinkles.
Device Story
The LED Beauty Mask is a wearable, silicone-based mask designed for over-the-counter, home-use wrinkle reduction. It utilizes 128 light-emitting diodes (LEDs) to deliver red (630nm) and near-infrared (830nm) light therapy to the face. The device is operated by the user via a controller, with a prescribed treatment regimen of 9 minutes daily, 5 days per week, for 8 weeks. By applying low-energy light to the skin, the device aims to provide therapeutic effects for wrinkle treatment. The mask is constructed from medical-grade silicone and includes an elastic headband for comfortable, long-term wear. The system is powered by a rechargeable lithium-ion battery.
Clinical Evidence
No clinical data was performed or submitted for this device. Substantial equivalence is supported by non-clinical bench testing, including electrical safety, electromagnetic compatibility, and biocompatibility assessments.
Technological Characteristics
Light-based wrinkle reduction device using 128 LEDs (630nm/830nm). Materials: medical-grade silicone (mask), PP (controller/adaptor). Power: 100-240V AC input, 5V 1A output, lithium-ion battery. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use), ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization), ISO 10993-23 (irritation).
Indications for Use
Indicated for the treatment of full-face wrinkles in adults.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 13, 2025
ATOP Lighting Co., Ltd.
Tse Adrian
Registration Representative
4th. Floor, Building C, No. 3, Jian'an Road
Tangwei Community, Fuhai Street, Baoan District
Shenzhen City, Guangdong province 518110
China
Re: K252308
Trade/Device Name: LED Beauty Mask (RLBXX)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS
Dated: July 24, 2025
Received: July 24, 2025
Dear Tse Adrian:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252308 - Tse Adrian
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252308 - Tse Adrian
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tanisha Hithe
Digitally signed by
Tanisha Hithe
Date: 2025.11.13
20:20:39 -05'00"
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252308 | ? |
| Please provide the device trade name(s). | | ? |
| LED Beauty Mask (RLBXX) | | |
| Please provide your Indications for Use below. | | ? |
| The LED Beauty Mask is an over-the-counter device that is intended for use in the treatment of full-face wrinkles. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
atop Lighting
ATOP Lighting Co., Ltd.
# 510(k) Summary
# K252308
## Submitter Information
- Company: ATOP Lighting Co., Ltd.
- Address: 4th. Floor, Building C, No. 3, Jian'an Road, Tangwei Community, Fuhai Street, Baoan District, Shenzhen City
- Phone: +086-15915873605
- Contact: Tse Adrian, Registration representative
- Mail box: certificate@atophort.com
## Device Information
- Trade/Device Name: LED Beauty Mask
- Model: RLBXX
- Common Name: Light-Based Over-The-Counter Wrinkle Reduction
- Classification regulation:
- Regulation number: 21 CFR 878.4810
- Regulation Description: Laser surgical instrument for use in general and plastic surgery and in dermatology
- Review Panel: General & Plastic Surgery
- Product Code: OHS
- Device Class: Class II
- Submission number: # K252308
## Subject Device Description
This product is based on low-energy red light therapy (LED Phototherapy), which achieves through precise wavelength control:
Wavelength selection: Using 630 red light and 830nm near-infrared light to treat wrinkle for 25-50 years women
The main body of the mask is made of medical-grade silicone, with a total weight of $\leqslant 300\mathrm{g}$, and an elastic headband, which can be worn for a long time without pressure.
## Predicate Device Information
- Predicate Device: LG PRA.L DERMA LED MASK
- Manufacturer: LG Electronics, Inc.
- 510(k) number: K183671
Page 1
{5}
atop Lighting
ATOP Lighting Co., Ltd.
Indication of use:
The LG PRA.L DERMA LED MASK is an over the counter device that is intended for the use in the treatment of full face wrinkles.
Page 2
{6}
atop Lighting
ATOP Lighting Co., Ltd.
# Indications for Use - Subject device
The LED Beauty Mask is an over-the-counter device that is intended for use in the treatment of full-face wrinkles.
Comparison of Technological Characteristics with the Predicate Device
| Comparison Items | Subject Device:LED Beauty Mask | Predicate Device 1:LG PRA.L DERMA LEDMASK(K183671) |
| --- | --- | --- |
| Classification & Intended Use | | |
| Classification | OHSClass II21 CFR 878.4810 | OHSClass II21 CFR 878.4810 |
| Intended use | The LED Beauty Mask is an over-the-counter device that is intended for use in the treatment of full-face wrinkles. | The LG PRA.L DERMA LEDMASK is an over the counter device that is intended for the use in the treatment of full face wrinkles. |
| Location for use | Face | Face |
| Patient Population | For use in adults | For use in adults |
| OTC or prescription | Over-the-Counter | OTC |
| Classification name | Light based over the counter wrinkle reduction | Light based over the counter wrinkle reduction |
| Comparison Statement | The subject device enjoys the same classification and intended use with the predicate device. | |
| Technological Characteristics | | |
| --- | --- | --- |
| Principle / Method of Operation | The subject device adopts light emitting diodes | The LG PRA.L DERMA LED MASK is a device which |
{7}
atop Lighting
ATOP Lighting Co., Ltd.
| | (LED) in the red (630nm±5nm) + infrared (830nm±5nm) spectrum to irradiate on the face to realize its therapeutic effect. | allows emission of LED light in the RED (637 nm) and IR (854 nm) spectrum on to the face. |
| --- | --- | --- |
| Wavelength | 630nm, 830nm | 637nm, 854nm |
| Material | Mask frame: Silicone
Eye shade: Silicone
Controller frame: PP
Adaptor frame: PP | Mask frame: PC
Eye frame: PP
Eye shield: Silicone
Viewing shield: ABS plastic
Controller frame: PP
Adaptor frame: PP
Cradle frame: PP |
| Dimensions (mm) | Mask Frame: 391 x 207 x 6
Controller: 95 x 55 x 19
Adaptor: 56 x 35 x 21.5 | LED Mask: 178.8 x 225 x 89
Controller: 40 x 30 x 110
Adaptor: 87x 44 x 28
Cradle: 180 x 229 x 167.8 |
| Net weight | LED Mask: 164 g
Controller: 66 g
Adaptor: 60 g | LED Mask: 225 g
Controller: 84 g
Adaptor: 75 g
Cradle: 380 g |
| Power supply | Input: 100 -240 V
Frequency: 50 / 60 Hz
Output: 5V 1A | Input: 100 -240 V
Frequency: 50 / 60 Hz
Output: 5V 1A |
| Light source | Light Emitting Diodes | Light Emitting Diodes |
Page 4
{8}
atop Lighting
ATOP Lighting Co., Ltd.
| | (LED) | (LED) |
| --- | --- | --- |
| Number of LEDs | Total 128 LEDs (64 of them are for 630 nm and the rest 64 of them are for 830 nm) | Total 160 LEDs (80 of them are for 637 nm and the rest 80 of them are for 854 nm) |
| LED power | 25 mW/cm2 total | 25 mW/cm2 total |
| Electronic power | Input: AC 100V ~ 240V Frequency: 50Hz/60Hz Output: 5V 1A | Input: AC 100V ~ 240V Frequency: 50Hz/60Hz Output: 5V 2A |
| Shape design | Mask type | Mask type |
| Battery | Lithium-ion | Lithium-ion |
| Treatment time | 9 minutes daily 5 days per week for 8 weeks | 9 minutes daily 5 days per week for 8 weeks |
| Comparison statement | The subject device has almost the same technological characteristics with the predicate device, and their slight difference in wavelength, material and Dimensions will not affect the core usage of the devices or validated by relevant standard evaluation or not bringing new safety and effectiveness concerns, thus will not affecting the substantial equivalence comparison. their differences are not affecting the core usage of the devices The efficiency of the subject device was able to reach an equivalence level of the predicate device. | |
| Patient-Contacting Materials | Silicone rubber | Not know | Difference |
| --- | --- | --- | --- |
| Electrical Safety | Verified according | IEC 60601-1 | Identical |
{9}
atop Lighting
ATOP Lighting Co., Ltd.
| | to IEC 60601-1 | | |
| --- | --- | --- | --- |
| EMC | Verified according to IEC 60601-1-2 | IEC 60601-1-2 | Identical |
| Biocompatibility | Verified according to ISO 10993-5 and ISO 10993-10 | Not know | Difference |
| Comparison Statement | The substantial equivalence of the subject device has been evaluated according to the FDA-recognized standards. | | |
# 5.6.1 Comparison summary
First, the subject device enjoys identical classification and intended use with the predicate device, which forms the foundation of their substantial equivalence.
Secondly, most technological characteristics have substantial equivalence differences is below:
| Difference item |
| --- |
| Principle / Method of Operation |
| Wavelength |
| Material |
| Dimensions |
| Net weight |
| Number of LEDs |
| Patient-Contacting Materials |
| Biocompatibility |
# Discussion of Tests Performed
1. Clinical Test
{10}
atop Lighting
ATOP Lighting Co., Ltd.
Clinical testing was not performed for the subject device as part of the submission.
# 2. Non-Clinical Tests
The subject device was tested/analyzed according to the following standards in order to ensure substantial equivalence:
| | Standard Designation Number | Title of Standard |
| --- | --- | --- |
| Electrical Safety | IEC 60601-1: 2020 | Medical electrical equipment Part 1: General requirements for basic safety and essential performance |
| Electromagnetic Compatibility | IEC 60601-1-2: 2020 | Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests |
| Home environment use requirement | IEC 60601-1-11: 2020 | Medical electrical equipment -Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical |
{11}
atop Lighting
ATOP Lighting Co., Ltd.
| | | electrical systems used in the home healthcare environment |
| --- | --- | --- |
| Biocompatibility | ISO 10993-5: 2009 | Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity |
| | ISO 10993-10: 2021 | Biological evaluation of medical devices – Part 10: Test for skin sensitization |
| | ISO 10993-23: 2021 | Biological evaluation of medical devices—Part 23: Tests for irritation |
# Conclusion
From the above analysis, it is proper to conclude that the subject device will be as safe and effective for usage as the listed predicate devices, K183671 that have already been on the U.S. market.
Date the Summary was prepared:
Oct. 08, 2025