K243330 · Lyma Life , Ltd. · OHS · Feb 21, 2025 · General, Plastic Surgery
Device Facts
Record ID
K243330
Device Name
LYMA Laser PRO
Applicant
Lyma Life , Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Feb 21, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The LYMA Laser PRO is an Over-the-Counter (OTC) light based device intended for the use in treating face wrinkles.
Device Story
Compact, handheld low-level laser therapy (LLLT) device for wrinkle reduction; utilizes three 808nm laser diodes and three red LEDs (620nm) to deliver light energy to the skin; includes optical lenses to diffuse light across a 20.4 cm2 treatment window; operated by the user in a home environment; device status indicated by a white LED; powered by a rechargeable 3.6V/5000mA Li-ion battery; treatment regimen involves 3 minutes per area daily for 8 weeks, followed by 3 minutes per area 3 times per week; provides non-invasive light therapy to potentially improve skin appearance.
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through bench testing, including biocompatibility (ISO 10993-1), electrical/mechanical/thermal safety (IEC 60601 series), laser safety (IEC 60825-1), and electromagnetic compatibility (IEC 60601-1-2).
Technological Characteristics
Handheld laser device; 3x 808nm (±2nm) laser diodes, 3x 620nm red LEDs, 1x white status LED; 20.4 cm2 treatment area; 12.8 J/cm2 dose; 1400 mW radiant power; 71 mW/cm2 irradiance; Class 3R laser; rechargeable 3.6V/5000mA Li-ion battery; conforms to IEC 60601-1, IEC 60601-1-6, IEC 60601-1-11, IEC 60601-2-22, and IEC 60825-1.
Indications for Use
Indicated for the treatment of face wrinkles in an Over-the-Counter (OTC) setting.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 21, 2025
LYMA Life Ltd. % Lucie Dalet Principal Consultant Ram+ 2790 Mosside Boulevard Suite 800 Monroeville, Pennsylvania 15146
Re: K243330
Trade/Device Name: LYMA Laser PRO Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: October 23, 2024 Received: October 24, 2024
Dear Lucie Dalet:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by TANISHA TANISHA L. HITHE -S L. HITHE -○ Date: 2025.02.21
13:58:41 -05'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243330
Device Name LYMA Laser PRO
Indications for Use (Describe)
The LYMA Laser PRO is an Over-the-Counter (OTC) light based device intended for the use in treating face wrinkles.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# L Y M A
## 510(k) SUMMARY
DATE PREPARED
February 18, 2025
#### MANUFACTURER AND 510(k) OWNER
LYMA Life Ltd. 2nd Floor Edison House, 223-231 Old Marylebone Road, London, United Kingdom Telephone: Official Contact: Lucy Goff, Founder
#### REPRESENTATIVE/CONSULTANT
Lucie Dalet, Ph.D. Dulciana Chan, MSE Allison Komiyama, Ph.D., RAC RQM+ Telephone: +1 (412) 816-8253 Email: ldalet@rqmplus.com, dchan@rqmplus.com, akomiyama@rqmplus.com
#### DEVICE INFORMATION
| Proprietary Name/Trade Name: | LYMA Laser PRO |
|------------------------------|------------------------------------------------|
| Common Name: | Light Based Over The Counter Wrinkle Reduction |
| Regulation Number: | 21 CFR 878.4810 |
| Class: | II |
| Product Code: | OHS |
| Premarket Review: | OHT4 / DHT4A – General Surgery Devices |
| Review Panel: | General & Plastic Surgery |
#### PREDICATE DEVICE IDENTIFICATION
The LYMA Laser PRO is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary<br>Predicate |
|---------------|-----------------------------------------------------------------------------------|----------------------|
| K210823 | LYMA Laser / LYMA Life Ltd. | ✓ |
| K230042 | Q-Rejuvalight Pro Facewear (Model: P19-0023) / Light<br>Tree Ventures Europe B.V. | |
The predicate devices have not been subject to a design related recall.
#### DEVICE DESCRIPTION
The LYMA Laser PRO is a compact, handheld laser device intended to provide low-level laser light therapy (LLLT) for the treatment of wrinkles. The LYMA Laser PRO system is composed of a handheld casing enclosing three low-level, 808nm laser diodes, and three red light-emitting diodes (LED). The device also includes one white LED that inform the user about device status during operation, and optical lenses that diffuse the laser light across a larger treatment window.
{5}------------------------------------------------
The LYMA Laser PRO includes a rechargeable battery and is provided with a power cable.
### INDICATIONS FOR USE
The LYMA Laser PRO is an Over-the-Counter (OTC) light based device intended for the use in treating face wrinkles.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
| | Subject Device | Primary Predicate Device | Secondary Predicate Device |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | LYMA Life, Ltd. | LYMA Life, Ltd. | Light Tree Ventures Europe<br>B.V. |
| | LYMA Laser PRO<br>K243330 | LYMA Laser<br>K210823 | Q-Rejuvalight Pro Facewear<br>K230042 |
| Indications for<br>Use | The LYMA Laser PRO is an<br>Over-the-Counter (OTC) light<br>based device intended for the use<br>in treating face wrinkles. | The LYMA Laser is an Over-<br>the-Counter (OTC) light based<br>device intended for the use in<br>treating full-face wrinkles. | The Q-Rejuvalight Pro Facewear<br>(Model: P19-0023) is an Over-<br>the-Counter (OTC) device<br>intended for treatment of<br>wrinkles and mild to moderate<br>inflammatory acne. |
| Product Code /<br>Regulation | OHS / 21 CFR 878.4810 | OHS / 21 CFR 878.4810 | OHS, OLP / 21 CFR 878.4810 |
| Regulation<br>Description | Laser surgical instrument for use<br>in general and plastic surgery<br>and in dermatology | Same | Same |
| Anatomical Sites | Full face | Same | Same |
| Environment of<br>Use | OTC | Same | Same |
| Energy source | Laser diodes, LEDs | Same | LEDs |
| Laser Class | Class 3R | Class 1 | N/A |
| Wavelength | Laser diodes: 808 nm ± 2 nm<br>Red LEDs: 620 nm | Same | Wrinkles: 605nm, 630nm,<br>660nm, and 880nm<br>Acne: 415nm and 630nm |
| Output Mode | Continuous | Same | N/A |
| Treatment area | 20.4 cm2 | 8.55 cm2 | For wrinkles: 140 cm2 |
| Treatment Dose | 12.8 J/cm2 | 11.3 J/cm2 | 12.6 J/cm2 |
| Radiant Power | 1400 mW | 500 mW | Unknown |
| Irradiance | 71 mW/cm2 | 62.5 mW/cm2 | Total for wrinkles: 70 mW/cm2 |
| Power supply | 3.6 V/ 5000 mA rechargeable Li-<br>ion battery (non-removable) | 3.6V / 3500 mA rechargeable Li-<br>ion battery (removable) | 5V, 50/60Hz, 2A Li-ion Polymer<br>Battery: 3.7V, 600mAh, 2.22Wh |
| Treatment<br>regimen | 3 minutes per treatment area,<br>every day for 8 weeks<br>After 8 weeks, 3 minutes per<br>treatment area, 3 times per week. | Same | 3 minutes per treatment |
The subject device has the same intended use and Indications for Use statement as the primary predicate device. The LYMA Laser PRO uses a similar design and identical materials as the device cleared in K210823. Both the subject and primary predicate devices are handheld devices using laser diodes to output infrared light at a wavelength of 808 ± 2 nm. The subject device is
{6}------------------------------------------------
# L Y M A
able to treat a larger area at a higher irradiance level than the primary predicate device. However, the output irradiance is similar to the secondary predicate device cleared in K230042. The technological differences do not raise new types of questions of safety and effectiveness. The different technological characteristics of the subject device have undergone testing to ensure the device is as safe and effective as the predicates.
#### SUMMARY OF NON-CLINICAL TESTING
The following tests were performed to demonstrate safety based on current industry standards:
Biocompatibility: Biocompatibility assessment was conducted per ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
Electrical, Mechanical, and Thermal Safety: The subject device was tested and shown to conform with the following safety and performance standards:
- IEC 60601-1:2005/AMD1:2012/AMD2:2020 Medical electrical equipment Part 1: -General requirements for basic safety and essential performance
- -IEC 60601-1-6:2010/AMD1:2013/AMD:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- -IEC 60601-1-11:2015/AMD1:2020 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- -IEC 60601-2-22:2019 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
Laser Safety: The subject device was tested and shown to conform with IEC 60825-1:2014 Safety of laser products - Part 1: Equipment classification and requirements
Electromagnetic Compatibility: The subject device was tested and shown to conform with IEC 60601-1-2:2014/AMD1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
#### SUMMARY OF CLINICAL TESTING
Clinical testing was not necessary to demonstrate substantial equivalence.
## CONCLUSION
Based on the testing performed, including electrical, EMC, and laser safety testing, and biocompatibility evaluation, it can be concluded that the subject device does not raise different questions of safety or effectiveness compared to the predicate devices. The identical indications for use and materials and the similar technological characteristics and performance characteristics for the proposed LYMA Laser PRO are assessed to be substantially equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.