Skin Care Beauty Mask (Model: MJ-06)

K220168 · Shenzhen Kaiyan Medical CO , Ltd. · OHS · May 17, 2022 · General, Plastic Surgery

Device Facts

Record IDK220168
Device NameSkin Care Beauty Mask (Model: MJ-06)
ApplicantShenzhen Kaiyan Medical CO , Ltd.
Product CodeOHS · General, Plastic Surgery
Decision DateMay 17, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Skin Care Beauty Mask (Model: MJ-06) emits energy in the red and blue region of the spectrum, specifically indicated to treat full face wrinkles and/or mild to moderate acne.

Device Story

Facemask-shaped device applies light to face skin surface; utilizes 150 blue LEDs (465nm) for mild to moderate inflammatory acne and 150 red LEDs (640nm) for wrinkle improvement. Irradiance is 30mW/cm2. User-operated in home environment (OTC). Device features selectable 5-minute or 10-minute treatment timers with automatic shutdown. Benefits include non-invasive cosmetic treatment of acne and wrinkles.

Clinical Evidence

Bench testing only. Evaluated for safety and performance per ANSI AAMI ES60601-1, IEC 60601-1-11, IEC 60601-2-57, IEC 60601-1-2, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation/sensitization).

Technological Characteristics

Facemask form factor; 300 total LEDs (150 blue @ 465nm, 150 red @ 640nm); 30mW/cm2 irradiance. Powered by external adapter (100-240V AC input, 5V DC output). Complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-57. Biocompatibility per ISO 10993-5 and 10993-10.

Indications for Use

Indicated for individuals with full face wrinkles and/or mild to moderate acne.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. May 17, 2022 Shenzhen Kaiyan Medical Co Ltd Alain Dijkstra CEO 40A Fuxin Road Fuyong Subdistrict BaoAn District, Shenzhen, Guangdong 518101 China Re: K220168 Trade/Device Name: Skin Care Beauty Mask (Model: MJ-06) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS, OLP Dated: January 17, 2022 Received: January 20, 2022 Dear Alain Dijkstra: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K220168 Device Name Skin Care Beauty Mask (Model: MJ-06) Indications for Use (Describe) The Skin Care Beauty Mask (Model: MI-06) emits energy in the region of the spectrum, specifically indicated to treat full face wrinkles and/or mild to moderate acne. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary of K220168 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 878.4810. ## 1. Submitter's Information Company Name: SHENZHEN KAIYAN MEDICAL CO LTD Address: 40A Fuxin Road Fuyong Subdistrict BaoAn District Shenzhen, Guangdong, China Contact Person (including title): Alain Dijkstra (CEO) Tel: +86 755 82129361 E-mail: regulation@kaiyanmedical.com ### Distributor Company Name: GODDESS'S MESSENGER LTD Address: 1624 Market St Ste 226, #79525 Denver CO 80202 USA Phone: +86 17310376031 ### Application Correspondent: Contact Person: Alain Dijkstra Company: SHENZHEN KAIYAN MEDICAL CO LTD Address: 40A Fuxin Road Fuyong Subdistrict BaoAn District Shenzhen, Guangdong, China Tel: +86 755 82129361 Email: regulation@kaiyanmedical.com ### 2. Subject Device Information Classification Name: Over-The-Counter Powered Light Based Laser For Acne (OLP), Light Based Over-The Counter Wrinkle Reduction (OHS) Trade Name: Skin Care Beauty Mask Model Name: MJ-06 Review Panel: General & Plastic Surgery Product Code: OHS, OLP Regulation Number: 878.4810 Regulatory Class: II ### 3. Predicate Device Information Sponsor: SHENZHEN KAIYAN MEDICAL CO LTD {4}------------------------------------------------ ## K220168 Trade Name: Aduro Light Beauty Mask, MK-660, MK-66USBO, MK-66USBA, MK-66USBB, MK-020, MK-02A, MK-02B Classification Name: Over-The-Counter Powered Light Based Laser For Acne (OLP), Light Based Over-The Counter Wrinkle Reduction (OHS) 510(K) Number: K202390 Review Panel: General & Plastic Surgery Product Code: OHS, OLP Regulation Number: 878.4810 Regulation Class: II ### 4. Device Description The Skin Care Beauty Mask (Model: MJ-06) is a facemask-shaped device, which makes use of specific light spectral characteristics and directly applies light onto the face skin surface. The Skin Care Beauty Mask can emit two different kinds of wavelengths. The blue light (465nm) is intended to help reduce the appearance of mild to moderate inflammatory acne. Red light (640nm) is intended to improve the appearance of wrinkles. There are 150 LED lights with a center wavelength of 465nm and 150 LED lights with a center wavelength of 640nm, and the energy provided by each mode is 30mw/cm2 . The user wears the mask on their face for the treatment, and the device will shut down automatically after finishing a 5-minute treatment or 10-minute treatment selected. ### 5. Intended Use / Indications for Use The Skin Care Beauty Mask (Model: MJ-06) emits energy in the red and blue region of the spectrum, specifically indicated to treat full face wrinkles and/or mild to moderate acne. ### 6. Test Summary Skin Care Beauty Mask (Model: MJ-06) has been evaluated the safety and performance by lab bench testing as following: | Standards No. | Standard Title | Version | Date | |------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|------------| | ANSI AAMI<br>ES60601-1 | Medical electrical equipment - Part 1:<br>General requirements for basic safety and<br>essential performance | 2005/(R)2012 and<br>A1:2012,<br>C1:2009/(R)2012 and<br>A2:2010/(R)2012 | 07/09/2014 | | Standards No. | Standard Title | Version | Date | | IEC 60601-1-11 | Medical electrical equipment - Part 1-11:<br>General requirements for basic safety and<br>essential performance - Collateral<br>Standard: Requirements for medical<br>electrical equipment and medical electrical<br>systems used in the home healthcare<br>environment | Edition 2.0 2015-01 | 06/27/2016 | | IEC 60601-2-57 | Medical Electrical Equipment - Part 2-57:<br>Particular requirements for the basic safety<br>and essential performance of non-laser<br>light source equipment intended for<br>therapeutic, diagnostic, monitoring and<br>cosmetic/aesthetic use | Edition 1.0 2011-01 | 03/16/2012 | | IEC 60601-1-2 | Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and<br>essential performance - Collateral<br>Standard: Electromagnetic disturbances -<br>Requirements and tests | Edition 4.0 2014-02 | 09/17/2018 | | ISO 10993-5 | Biological evaluation of medical devices -<br>Part 5: Tests for in vitro cytotoxicity | Third edition 2009-<br>06-01 | 12/23/2016 | | ISO 10993-10 | Biological evaluation of medical devices -<br>Part 10: Tests for irritation and skin<br>sensitization | Third Edition 2010-<br>08-01 | 07/26/2016 | {5}------------------------------------------------ # K220168 ## 7. Comparison to predicate device and conclusion The subject device is very similar in design principle, indications for use, functions, materials, and applicable performance, when compared to the above models of the K202390 predicate. Differences {6}------------------------------------------------ ## K220168 between the subject device and predicate do not raise new types of questions with regard to the safety and effectiveness of the subject device for the proposed indications for use. | Elements of<br>Comparison | Subject Device | Predicate Device | Remark | |---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | Company | SHENZHEN KAIYAN MEDICAL CO LTD | SHENZHEN KAIYAN MEDICAL CO LTD | -- | | Trade Name | Skin Care Beauty Mask | Aduro Light Beauty Mask | -- | | Classification<br>Name | Over-The-Counter Powered<br>Light Based Laser For Acne (OLP), Light Based Over-The<br>Counter Wrinkle Reduction (OHS) | Over-The-Counter Powered Light<br>Based Laser For Acne (OLP), Light<br>Based OverThe Counter Wrinkle<br>Reduction (OHS) | -- | | 510(k) Number | K220168 | K202390 | -- | | Product Code | OHS, OLP | OHS, OLP | SE | | Intended Use /<br>Indications for Use | The Skin Care Beauty Mask<br>(Model: MJ-06) emits energy in<br>the red and blue region of the<br>spectrum, specifically indicated<br>to treat full face wrinkles and/or<br>mild to moderate acne. | For MK-660, MK-66USBO, MK-<br>020:<br>The Aduro Light Beauty Mask<br>(Model: MK-660, MK-66USBO,<br>MK-02O) emits energy in the red<br>and blue region of the spectrum,<br>specifically indicated to treat full<br>face wrinkles and/or mild to<br>moderate acne. | SE | | Wavelengths | Red: 640nm±10nm<br>Blue: 465nm±10nm | Red: 640nm±10nm<br>Blue: 465nm±10nm | SE | | Power supply | Adapter:<br>Input: 100-240Va.c., 50/60Hz,<br>0.5A<br>Output: 5.0Vd.c., 2.0A | For model MK-660, MK-66USBO:<br>3.7Vdc 2600mAh Lithium battery<br>Adapter for charging only:<br>Input: 100- 240Vac;<br>Output: 5Vdc, 2A.<br>For model MK-020: 3.7Vdc<br>2600mAh Lithium battery | SE | | Irradiance source | LEDs | LEDs | SE | | Treatment time | 5 or 10 minutes/day, 3 times per | 10minutes/day, 3 times per week | SE | | Elements of<br>Comparison | Subject Device | Predicate Device | Remark | | | week | | | | Power Density<br>(mw/cm²) | 30mW/cm² | 30mW/cm² | SE | | Location for Use | Face | Face | SE | | Environment of<br>Use | OTC | OTC | SE | | Safety and EMC | IEC 60601-1<br>IEC 60601-1-2<br>IE C60601-1-11<br>IEC 60601-2-57 | IEC 60601-1<br>IEC 60601-1-2<br>IE C60601-1-11<br>IEC 60601-2-57 | SE | | Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | SE | {7}------------------------------------------------ #### 8. Conclusion: The proposed device utilizes the same technological characteristics as the above listed model configurations of K202390, and is also for use in the treatment of mild to moderate acne and treatment of wrinkles. After an analysis of the proposed device's design, functionality, safety, materials, and performance testing results, the sponsor believes that the device's technology does not raise new types of questions regarding safety and efficacy of the subject device for the proposed indications, and performance testing supports that the device can be used safely and effectively. The subject device Skin Care Beauty Mask (Model: MJ-06) is considered to substantially equivalent to the above-specified devices of K202390. ### 9. Date prepared: May 2022
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...