ISOLATOR LINEAR PEN

K130521 · AtriCure, Inc. · OCL · Apr 5, 2013 · General, Plastic Surgery

Device Facts

Record IDK130521
Device NameISOLATOR LINEAR PEN
ApplicantAtriCure, Inc.
Product CodeOCL · General, Plastic Surgery
Decision DateApr 5, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Isolator® linear pen is a sterile, single use electrosurgery device intended to ablate cardiac tissue during cardiac surgery using radiofrequency (RF) energy when connected directly to the ASU or ASB in Ablation mode. The Isolator® linear pen may be used for temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external cardiac pacemaker or recording device.

Device Story

Isolator Linear Pen is a sterile, single-use electrosurgical instrument; used during cardiac surgery. Input: bipolar radiofrequency (RF) energy from AtriCure Ablation and Sensing Unit (ASU) or ASU Source Switch (ASB). Operation: operator controls RF energy delivery via footswitch to ablate cardiac tissue; device also functions as a surgical tool for temporary cardiac pacing, sensing, recording, and stimulation when connected to external auxiliary equipment. Output: thermal ablation of cardiac tissue or electrical signals for cardiac monitoring/pacing. Used in surgical settings by clinicians. Benefits: enables precise cardiac tissue ablation and intraoperative arrhythmia evaluation.

Clinical Evidence

Bench testing only. Testing conducted per 21 CFR 820.30 and AtriCure's Quality System to verify conformance to design controls and specifications. Biocompatibility testing conducted in accordance with ISO 10993-1.

Technological Characteristics

Electrosurgical device; bipolar RF energy source; single-use; sterile. Materials tested for biocompatibility per ISO 10993-1. System components include Pen, ASU, ASB, and footswitch.

Indications for Use

Indicated for patients undergoing cardiac surgery requiring ablation of cardiac tissue or evaluation of cardiac arrhythmias via temporary pacing, sensing, recording, or stimulation.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K130521 ## 510(k) Summary # APR 0 5 2013 General Information | Classification | Class 2 | |---------------------|--------------------------------------------------------------------------------------------------------| | Trade name | Isolator® Linear Pen | | Classification Name | Surgical Device, For Ablation Of Cardiac Tissue<br>(21 CFR 878.4400, Product Code OCL) | | Manufacturer | AtriCure, Inc.<br>6217 Centre Park Dr.<br>West Chester, OH 45069<br>P: 513-755-4100<br>F: 513-755-4108 | | Contact | Rebecca Walters, RAC<br>Regulatory Affairs Manager | Date of Submission February 27, 2013 #### Intended Use The Isolator® linear pen is a sterile, single use electrosurgery device intended to ablate cardiac tissue during cardiac surgery using radiofrequency (RF) energy when connected directly to the ASU or ASB in Ablation mode. The Isolator® linear pen may be used for temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external cardiac pacemaker or recording device. #### Cleared Device The device proposed for modification in this submission is the Isolator® linear pen cleared via 510(k) K100501 on June 18, 2010. #### Device Description The Isolator™ linear pen System is comprised of the AtriCure® Ablation and Sensing Unit (ASU), Isolator™ linear pen (Pen), Footswitch, ASU Source Switch(ASB). The Pen is a single patient use electrosurgical instrument designed for use only with the ASU and ASB. The Pen is used to ablate cardiac tissues and as a surgical pacing and mapping tool. When the Pen is connected to the ASU, the ASU provides the bipolar radiofrequency (RF) energy flowing between both electrodes of the Pen. The Operator controls the application of this RF energy by pressing the Footswitch. When the Pen is connected to an auxiliary pace, sense, or stimulation device, the Pen is designed to provide temporary pacing or monitoring. #### Materials All materials in the modified Isolator® linear pen are suitable for their intended use. Testing was conducted on all patient contacting materials in accordance with ISO 10993-1 to ensure appropriate biocompatibility of all appropriate materials. Isolator® linear pen {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "AtriCure" in a stylized font. The "C" in "AtriCure" is designed with a circle around it, and there is a registered trademark symbol after the word. The font is bold and black, and the background is white. ### Testing Testing per 21 CFR 820.30 and AtriCure's Quality System was performed to verify the modified Isolator linear pen conformance to design controls and specification. Testing determined that the modified Isolator® linear pen conformed to design controls and product specifications. ### Summary of Equivalence The modified Isolator® linear pen proposed in this submission is considered substantially equivalent to the Isolator® linear pen cleared via K100501. The indications for use, basic overall function, and materials used are substantially equivalent. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 April 5, 2013 Atricure, Inc. Rebecca Walters 6217 Centre Park Drive West Chester, OH 45069 US Re: K130521 > Trade/Device Name: Isolator linear pen Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: OCL Dated: March 5, 2013 Received: March 7, 2013 Dear Ms. Walters: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, Misbranding by reference to premarket notification (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htmfor the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Owen P. Faris -S for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ # Indications for Use 510(k) Number (if known)_K130521 Device Name: Isolator® linear pen Indications for Use: - The Isolator® linear pen is intended to ablate cardiac tissue during cardiac surgery using . radiofrequency (RF) energy when connected directly to the ASU or to the ASU Source Switch (ASB) in Ablation mode. - The Isolator® linear pen may be used for temporary cardiac pacing, sensing, recording, . and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external cardiac pacemaker or recording device. Prescription Use × (Part 21 CRF 801 Subpart D) AND/OR Over-The-Counter Use (21 CRF 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Owen P. Faris -S 2013.04.05 13:35:26 -04'00' Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...