DIGITEX SUTURE DELIVERY SYSTEM, DIGITEXPOLYPROPYLENE SUTURE CARTRIDGE SIZE 0, WITH NEEDLE AND WITHOUT NEEDLE, DIGITEX

K120284 · Coloplast A/S · NEW · May 23, 2012 · General, Plastic Surgery

Device Facts

Record IDK120284
Device NameDIGITEX SUTURE DELIVERY SYSTEM, DIGITEXPOLYPROPYLENE SUTURE CARTRIDGE SIZE 0, WITH NEEDLE AND WITHOUT NEEDLE, DIGITEX
ApplicantColoplast A/S
Product CodeNEW · General, Plastic Surgery
Decision DateMay 23, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4840
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Digitex Delivery Device is a sterile, disposable device intended to deliver a suture to the operative site. The device assists in suturing by passing a needle through the tissue to capture the suture, and suture ligation and knot-tying by holding the suture. The Digitex Suture Cartridge holds sterile suture indicated for soft tissue approximation.

Device Story

Digitex Suture Delivery System (SDS) consists of a delivery device and interchangeable suture cartridges; designed for physician use in surgical settings where direct visualization is limited or anatomical access is difficult. Delivery device features an adjustable needle housing angle to facilitate precise suture placement. Device functions by passing a needle through tissue to capture suture, assisting in ligation and knot-tying. Cartridges are pre-loaded with non-absorbable polypropylene, absorbable polydioxanone, or absorbable polyglycolic acid coated sutures. System aids surgeons in consistent suture placement, potentially reducing operative time and improving procedural outcomes in difficult-to-reach anatomical locations.

Clinical Evidence

No clinical data. Evidence consists of bench testing and cadaveric laboratory simulated use testing. Biocompatibility profile supported by testing conducted for original predicate (K093112) and additional evaluation for new absorbable materials.

Technological Characteristics

Sterile, disposable surgical instrument. Materials include non-absorbable polypropylene, absorbable polydioxanone, and polyglycolic acid coated sutures. Features adjustable needle housing angle. Sterilization per ISO 11135; biocompatibility per ISO 10993-7. Mechanical delivery system; no software or electronic components.

Indications for Use

Indicated for soft tissue approximation in patients requiring surgical suturing.

Regulatory Classification

Identification

An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 7. 510(K) SUMMARY ## MAY 2 3 2012 | 510(k) Owner/Submitter | Coloplast A/S; Division: Coloplast Corp<br>Holtedam 1, Humlebaek 3050 - Denmark<br>FDA Establishment Registration Number: 2125050 | |------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact | Janell A. Colley<br>Coloplast Corp<br>1601 West River Road North<br>Minneapolis, Minnesota 55411 USA | | Date Prepared | 30 January 2012 | | Common name/<br>Classification<br>(current/proposed) | Delivery Device: surgical instrument, 878.4800, Instrument, ligature passing and knot<br>tying<br>Current Suture Cartridge: surgical suture, 878.5010; Nonabsorbable polypropylene<br>surgical suture<br>Proposed Suture Cartridge | | Proprietary Name | Digitex Suture Delivery System (SDS); Digitex Delivery Device, Digitex Suture<br>Cartridge | | Predicate Devices | Digitex System: Digitex Suture Delivery System: K093112<br>Proposed Digitex Suture Cartridge predicates: Deknatel Deklene II suture, K930738;<br>Monodek PDO suture, K030212; Bondek PGA suture, K991191 & K992088 | | Device Description | The Digitex SDS is composed of delivery device and suture cartridge and is designed for use by the physician to<br>facilitate the consistent placement of suture when direct visualization is not possible and/or the anatomical<br>location is difficult to reach. The shaft of the delivery device has been designed to allow for adjusting the angle of<br>the needle housing to further facilitate suture placement in the desired location. The suture cartridges are<br>provided pre-loaded with non-absorbable polypropylene, absorbable polydioxanone, or absorbable polyglycolic<br>acid coated suture. | | Intended Use | The Digitex Delivery Device is a sterile, disposable device intended to deliver a suture to the operative site. The<br>device assists in suturing by passing a needle through the tissue to capture the suture, and suture ligation and knot-<br>tying by holding the suture.<br><br>The Digitex Suture Cartridge holds sterile suture indicated for soft tissue approximation. | | Technological Characteristics Compared to Predicate | The Digitex Suture Delivery System has the same intended use and fundamental scientific technology as the<br>predicates Digitex Suture Delivery System: K093112, Deknatel Deklene II suture, K930738; Monodek PDO suture,<br>K030212; and Bondek PGA suture, K991191 & K992088. | | Summary Of The Nonclinical Tests Submitted | The modifications to the Digitex Suture Cartridges, specifically, adding suture cartridges with polydioxanone and<br>polyglycolic acid coated absorbable suture, were evaluated via design verification testing and simulated use tests in<br>a cadaveric laboratory. These tests confirmed that the Digitex Suture Cartridges, as modified, meet the established<br>design specifications and are substantially equivalent to the predicate polypropylene Digitex Suture Cartridges. The<br>predicate Digitex Suture Cartridges were subjected to biocompatibility testing to support the original 510k (K093112)<br>and acceptable results from this testing apply to the proposed absorbable Digitex Suture Cartridges; additional<br>evaluation and testing was conducted and acceptable results from this testing further demonstrate the acceptable<br>biocompatibility profile of the proposed absorbable Digitex Suture Cartridges. The sterilization method and sterility<br>assurance lever (SAL) have not changed; the change to the contract sterilization facility and modified sterile barrier<br>have been evaluated and meet the requirements of ISO 11135 and ISO 10993-7. | | Summary Of Clinical Tests Submitted (As Applicable) | Not applicable | Conclusions Drawn From The Nonclinical Tests The Digitex Suture Delivery System is substantially equivalent to the predicates. ાં જ {1}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure with outstretched arms, resembling a bird in flight. The figure is positioned within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper portion of the circle. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Coloplast A/S % Ms. Janell Colley Regulatory Affairs Manager 1601 West River Road North Minneapolis, Minnesota 55411 # 2 3 2012 Re: K120284 Trade/Device Name: Digitex Suture Delivery System (SDS) Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly (glycolide/L-lactide) surgical suture Regulatory Class: Class II Product Code: NEW,GAM Dated: April 20 . 2012 Received: April 23, 2012 Dear Ms. Colley We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may · publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {2}------------------------------------------------ ## Page 2 - Ms. Janell Colley or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Eyaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### 6. STATEMENT OF INDICATIONS FOR USE . 510(k) Number (if known): K120284 Device Name: Digitex Suture Delivery System (SDS) Indications for Use: The Digitex Delivery Device is a sterile, disposable device intended to deliver a suture to the operative site. The device assists in suturing by passing a needle through the tissue to capture the suture, and suture ligation and knot-tying by holding the suture. The Digitex Suture Cartridge holds sterile suture indicated for soft tissue approximation. | Prescription Use | X | |------------------|---| |------------------|---| Over-The-Counter Use (Part 21 CFR 801 Subpart D) AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Keane for MM (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K120284 Coloplast Digitex Suture Delivery System Traditional 510(k) Digitex Absorbable Suture Cartridges Page 23 of 315
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...