K990902 · Medical Laser Technologies , Ltd. · GEX · May 27, 1999 · General, Plastic Surgery
Device Facts
Record ID
K990902
Device Name
MODIFICATION OF MLT R694 RUBY LASER SYSTEM
Applicant
Medical Laser Technologies , Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 27, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The MLT R694 Ruby Laser System (operating in the Q-Switched mode) is indicated for cutting, vaporization, or ablation of soft tissue. This includes tatoo removal and treatment of benign pigmented lesions. The MLT R694 Ruby Laser System is restricted to sale to or use by licensed professionals in the United States
Device Story
MLT R694 Ruby Laser System emits coherent light at 694 nm; operates in normal/long pulse or Q-Switched mode. Q-Switched mode provides shorter pulse duration for soft tissue cutting, vaporization, or ablation; tattoo removal; treatment of benign pigmented lesions. Used in clinical settings by licensed professionals. Output allows precise tissue interaction for dermatological and surgical applications. Device modification adds Q-Switched capability to existing ruby laser platform.
Clinical Evidence
No clinical data. Bench testing only; substantial equivalence based on comparison of specifications and intended use to predicate devices.
Technological Characteristics
Ruby laser; 694 nm wavelength; Q-Switched and normal/long pulse modes. Class II laser surgical instrument (21 CFR 878.4810).
Indications for Use
Indicated for cutting, vaporization, or ablation of soft tissue, including tattoo removal and treatment of benign pigmented lesions in patients requiring such procedures. Restricted to use by licensed professionals.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
LaseAway Long Pulse and Q-Switched Ruby Laser
MM-Ruby Laser
Spectrum RD 1200
Candela AlexLAZR
Related Devices
K991285 — LASER SYSTEM RUBYSTAR WITH Q-SWITCH AND NORMAL MODE · Aesculap-Meditec North America · Jul 12, 1999
K040433 — SINON · Wavelight Laser Technologie AG · May 19, 2004
K150997 — RUBY STAR · Bison Medical Co, Ltd. · Dec 22, 2015
K050218 — WAVELIGHT SINON · Wavelight Laser Technologie AG · Feb 15, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) SUMMARY MEDICAL LASER TECHNOLOGIES LTD MLT R694 RUBY LASER SYSTEM Q-SWITCHED MODE
This 510(k) summary of safety and effectiveness for the Medical Laser Technologies Ltd. MLT R694 Ruby Laser System operating in the O-Switched Mode is submitted in accordance with the requirements of SMDA 1990 and follows Office of Device Evaluation guidance concerning the organization and content of a 510(k) summary.
MEDICAL LASER TECHNOLOGIES LTD Applicant:
- Address: Unit 15 Belleknowes Industrial Estate Inverkeithing Fife KY11 1HZ United Kingdom
- Contact Person: David Hamilton . Managing Director
- Telephone: 011 44 1383 411555 011 44 1383 411666 Facsimile:
- Preparation Date: March 1999
- MLT R694 Ruby Laser System Device Name:
- Common Name: Ruby Laser, Normal Pulse or Q-Switched Mode
Classification Laser surgical instrument for use in general and plastic surgery and in Name: dermatology (see: 21 CFR 878.4810). Product Code: GEX. Panel: 79
- Legally marketed LaseAway Long Pulse and Q-Switched Ruby Laser, MM-Ruby Laser; Spectrum RD 1200; Candela AlexLAZR.
- Description: The MLT R694 laser operates as a normal pulse or as a O-Switched rubv laser which emits a beam of coherent light at 694 nanometers. In the Q-Switched mode the beam has a much shorter pulse duration than in the normal or long pulse mode. The specifications for the normal or long pulse mode remain the same as described in K980187.
{1}------------------------------------------------
- The MLT R694 Ruby Laser System in the Q-Switched mode is indicated for Intended Use: cutting, vaporization, or ablation of soft tissue. This includes tatoo removal and treatment of benign pigmented lesions.
The intended uses of the MLT R 694 ruby operating in normal or long pulse mode remain the same as described in K980187.
- Comparison to: The specifications of the MLT R694 laser in the O-Switched mode are the same as or very similar to those of legally marketed lasers such as the LaseAway Long Pulse and Q-Switched Ruby Laser, the MM-Ruby Laser; the Spectrum RD 1200, and the Candela AlexLAZR
- Performance Data: None. The specifications and intended uses of the MLT R694 ruby system laser operating in the Q-Switched mode are the same or very similar to those of the claimed predicate devices.
Because of this, performance data were not required.
- CONCLUSION: When operating in the Q-Switched mode the MLT R694 is substantially equivalent to legally marketed predicate devices, e.g., ruby and alexandrite lasers, operating in the Q-Switched mode.
The MLT R694 ruby laser system, operating in the normal mode, has been found substantially equivalent to legally marketed devices (K980187); the characteristics, specifications, and intended uses of the MLT R 694 operating in this mode are not affected by the modification to allow use in the Q-Switched mode.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle. The text is written in all capital letters.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 7 1999
Mr. David Hamilton Managing Director Medical Laser Technology, Ltd. Unit 15 Belleknowes Industrial Estate Inverkeithing Fife KY11 1HZ United Kingdom
Re: K990902
> Trade Name: MLT R694 Ruby Laser System Regulatory Class: II Product Code: GEX Dated: March 16, 1999 Received: March 18, 1999
Dear Mr. Hamilton:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register, Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
## Page 2 - Mr. David Hamilton
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## APPENDIX B
## INDICATIONS FOR USE STATEMENT
510(K) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: MLT R694 RUBY LASER SYSTEM ______________________________________________________________________________________________________________________________________
Indications For Use Statement:
The MLT R694 Ruby Laser System (operating in the Q-Switched mode) is indicated for cutting, vaporization, or ablation of soft tissue. This includes tatoo removal and treatment of benign pigmented lesions.
The MLT R694 Ruby Laser System is restricted to sale to or use by licensed professionals in the United States
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Prescription Use | X |
|----------------------|---|
| (Per 21 CFR 801.109) | |
OR
| Over-The-Counter Use | |
|----------------------|--|
|----------------------|--|
| (Division Sign-Off) | |
|-----------------------------------------|---------|
| Division of General Restorative Devices | |
| 510(k) Number | K990902 |
Concurrence of CDRH, Office of Device Evaluation (ODE)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.