K983058 · Ceram Optec, Inc. · GEX · Oct 21, 1998 · General, Plastic Surgery
Device Facts
Record ID
K983058
Device Name
CERALAS D LASER SYSTEM, MODEL # CERALAS D15
Applicant
Ceram Optec, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 21, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Ceralas D Laser System has been cleared as a surgical instrument intended for incision, excision, hemostasis, coagulation, and vaporization of soft tissue in open and closed endoscopic procedures in general surgery, urology, gynecology, neurosurgery, gastroenterology, plastic surgery, dermatology, and otolaryngology. Specific indications for the previously cleared Ceralas D Laser System include tonsillectomy, thyroidectomy, vocal cord polypectomy, hemiglossectomy, tracheal stenosis, neck dissection, and oral cavity lesions. The company is expanding the Ceralas D Laser System's indications for use to include the following applications on intraoral and extraoral soft tissue (including marginal and interdental gingiva and epithelial lining of free gingiva): frenectomy; frenotomy; biopsy; operculectomy; implant recovery; gingivectomy; gingivoplasty; gingival troughing; crown lengthening; hemostasis of donor site; removal of granulation tissue; laser assisted flap surgery; debridement of diseased epithelial lining; incisions and draining of abscesses; tissue retraction for impressions: papillectomy: vestibuloplasty; excision of lesions; exposure of unerupted/partially erupted teeth: leukoplakia; removal of hyperplastic tissues; treatment of aphthous ulcers; and sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket).
Device Story
Ceralas D Laser System is a compact, self-contained surgical laser using GaAlAs semiconductor diodes to generate 980nm near-infrared radiation. Device includes cooling module, optics module with beamsplitter, control panel, and RFI-shielded power supply. Laser energy is delivered via interchangeable fiber optic cables with SMA 905 connectors. Operated by clinicians in surgical or dental settings for incision, excision, hemostasis, coagulation, and vaporization of soft tissue. Aiming beam (635nm) assists targeting. Output power ranges 1-15 Watts. Device provides precise tissue interaction for various oral and general surgical applications, facilitating procedures like gingivectomy and lesion excision.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
GaAlAs semiconductor diode laser; 980 ± 30nm wavelength; 635nm aiming beam; 1-15W power output. Modular design: cooling module (thermoelectric coolers, heat sink, fans), optics module, control electronics. Fiber optic delivery via SMA 905 connector; quartz fiber core. RFI-shielded power supply.
Indications for Use
Indicated for intraoral and extraoral soft tissue surgical procedures including frenectomy, biopsy, gingivectomy, crown lengthening, and sulcular debridement in patients requiring soft tissue management.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Ceralas Diode Laser System (Model D15) (K951775, K964497)
Premier Laser Systems' Aurora Diode Dental Laser (K954316, K974586)
American Dental Technologies Pulsemaster® 1000 ST Dental Diode Laser System (K972325)
American Dental Technologies Pulsemaster® 1000 Nd:YAG Laser (K961269, K961136)
Related Devices
K964497 — CERALAS DIODE LASER SYSTEM (CERALAS D15) · Ceram Optec, Inc. · Feb 6, 1997
K023547 — DIOLASE 980 D LASER SYSTEM · American Dental Technologies, Inc. · Apr 15, 2003
K072262 — GENTLERAY 980 DIODE LASER SYSTEM · Kavo America · Oct 26, 2007
K191349 — 980nm Diode Laser Therapy Device · Shanghai Apolo Medical Technology Co., Ltd. · Aug 16, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
983058
OCT 21 1998
### 510(k) Summary Ceralas Diode Laser System
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Submitted by Regulatory Counsel for: CeramOptec. Inc. 515 Shaker Road East Longmeadow, Massachusetts 01028 (413) 525-0600 Phone: Facsimile: (413) 525-0611
Contact Person: Carol Morello, V.M.D. Date prepared: September 1, 1998
### Name of Device and Name/Address of Sponsor
Ceralas Diode Laser System (Model D15) CeramOptec, Inc. 515 Shaker Road East Longmeadow, MA 01028
### Classification Name
Surgical laser
#### Predicate Device
Ceralas Diode Laser System (Model D15) Premier Laser Systems' Aurora Diode Dental Laser American Dental Technologies Pulsemaster® 1000 ST Dental Diode Laser System American Dental Technologies Pulsemaster® 1000 Nd:YAG Laser
#### Intended Use
There have been no significant changes to the Ceralas D Laser System since is was cleared by FDA (K951775, K964497). The Ceralas D Laser System has been cleared as a surgical instrument intended for incision, excision, hemostasis, coagulation, and vaporization of soft tissue in open and closed endoscopic procedures in general surgery, urology, gynecology, neurosurgery, gastroenterology,
{1}------------------------------------------------
plastic surgery, dermatology, and otolaryngology. Specific indications for the previously cleared Ceralas D Laser System include tonsillectomy, thyroidectomy, vocal cord polypectomy, hemiglossectomy, tracheal stenosis, neck dissection, and oral cavity lesions.
The company is expanding the Ceralas D Laser System's indications for use to include the following applications on intraoral and extraoral soft tissue (including marginal and interdental gingiva and epithelial lining of free gingiva): frenectomy; frenotomy; biopsy; operculectomy; implant recovery; gingivectomy; gingivoplasty; gingival troughing; crown lengthening; hemostasis of donor site; removal of granulation tissue; laser assisted flap surgery; debridement of diseased epithelial lining; incisions and draining of abscesses; tissue retraction for impressions: papillectomy: vestibuloplasty; excision of lesions; exposure of unerupted/partially erupted teeth: leukoplakia; removal of hyperplastic tissues; treatment of aphthous ulcers; and sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket).
As explained below, the Company's Ceralas D Laser System is substantially equivalent to the previously cleared Ceralas D Laser System (K951775, K964497), Premier Laser Systems' Aurora Diode Dental Laser (K954316, K974586). American Dental Technologies' Pulsemaster® 1000 ST Dental Diode Laser System (K972325), and American Dental Technologies' Pulsemaster® 1000 Nd:YAG Laser (K961269, K961136).
# Technological Characteristics and Substantial Equivalence
As stated above, there have been no significant changes to the Ceralas D Laser System since it was cleared by FDA. The Ceralas D Laser System is a complete self-contained compact surgical laser that utilizes gallium aluminum arsenide (GaAlAs) semiconductor diodes to generate near-infrared laser radiation. The laser employs a modular design comprised of: (1) a laser diode; (2) a cooling module containing the laser diode, thermoelectric coolers, heat sink fins and fans; (3) an optics module containing a beamsplitter for control of optical power; (4) front control and display panel; and (5) RFI-shielded, transformer power supply and control electronics. Interchangeable fiber optic delivery systems are coupled to the laser via an SMA 905 connector to deliver laser radiation to the target tissues. The diode laser is enclosed in a rugged, factory aligned, environmentally protective module.
The wavelength for the Ceralas D Laser System is 980 ± 30nm, and its laser aiming beam wavelength is 635nm+10nm. The power output ranges from 1-15 Watts. The delivery systems for the Ceralas D Laser System consist of an optical fiber fitted with an SMA 905 connector at the proximal end. The optical fiber is composed of quartz fiber core with a coaxially mounted protective sheath.
{2}------------------------------------------------
Moreover, the optical fiber delivery system has been previously cleared for dental use (K942182).
CeramOptec has made no significant changes to the hardware or software of the Ceralas D Laser since it was cleared by FDA. Likewise, CeramOptec has made no significant changes to its software development process since its previous 510(k) submission for the Ceralas D Laser System. Additionally, the software verification and validation were conducted in accordance with FDA's Guidance Document, "Reviewer Guidance For Computer Controlled Medical Devices Undergoing 510(k) Review."
The principal technological difference between the Ceralas Diode Laser and Premier Laser System's Aurora Diode Dental Laser is that the devices operate at different wavelengths. The Ceralas D Laser's wavelength is 980 ±30mm and the Aurora Diode Dental Laser's wavelength is 805-820mm. However, it is not believed that the difference between the devices' wavelength raise new questions of safety and effectiveness.
The Ceralas D Laser is also similar to American Dental Technologies' Pulsemaster® 1000 ST Diode Dental Laser. Both devices are similar in function and intended use. and are diode lasers which operate in a continuous or pulsed mode with wavelength ranges similar to the Premier Laser System's Aurora Diode Dental Laser. In summary, although there are minor differences between the Ceralas D Laser System and its predicate devices, these differences do not raise new questions of safety and efficacy.
# Performance Data
None required.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, rendered in black.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 21 1998
CeramOptec, Inc. c/o Jonathan S. Kahan Hogan and Hartson Columbia Square 555 Thirteenth Street, N.W. Washington, D.C. 20004-1109
K983058 Re: Ceralas Diode Laser System(Model D15) Trade Name: Regulatory Class: II Product Code: GEX 1998 Dated: September 01, Received: September 01, 1998
Dear Mr. Kahan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਖੇ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Paqe 2 - Mr. Jonathan S. Kahan
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K993058
Ceralas D Laser System Device Name:
### Indications For Use:
The Ceralas D Laser System is indicated for the following applications on intraoral The Cordial D 2002 vincluding marginal and interdental gingiva and epithelial ana one free gingiva): frenectomy; frenotomy; biopsy; operculectomy; implant Hillig of recogning; gingivoplasty; gingival troughing; crown lengthening; I covery, ging: rootens), grag-vel of granulation tissue; laser assisted flap surgery; debridement of diseased epithelial lining; incisions and draining of abscesses; tissue retraction for impressions; papillectomy; vestibuloplasty; excision of lesions; exposure of unerupted/partially erupted teeth; leukoplakia; removal of hyperplastic tissues; treatment of aphthous ulcers; and sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket).
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEE
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Sign-Off | |
|--------------------------------|---------|
| of General Restorative Devices | |
| 510(k) Number | K983058 |
**Prescription Use**
(Per 21 CFR 801.109)
OR
**Over-The-Counter Use**
(**Optional Format 1-2-96**)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.