RD-1200H NORMAL MODE RUBY LASER SYSTEM
Device Facts
| Record ID | K955612 |
|---|---|
| Device Name | RD-1200H NORMAL MODE RUBY LASER SYSTEM |
| Applicant | Spectrum Medical Technologies, Inc. |
| Product Code | GEX · General, Plastic Surgery |
| Decision Date | Jul 9, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4810 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Spectrum Epilaser® has several indications for use based on the selective absorption of natural chromophores, particularly melanin and tattoos. The Ruby laser wavelength of 694.3 nanometers and the pulse duration of the normal mode system are well absorbed by these pigments. Indications for use include the treatment of benign epidermal pigmented lesions such as lentigines and café au lait and tattoos (blue and black).
Device Story
Normal mode ruby laser system; 694.3 nm wavelength; 10-40 J/cm² fluence; 200 μsec to 3 msec pulse duration; 5-12 mm spot size. Delivery via articulating arm and cooled contact handpiece. Targets melanin and tattoo pigments (blue/black) via selective photothermolysis. Used by clinicians in dermatology settings for lesion treatment and tattoo removal. Output is laser energy delivered to skin; clinical benefit is removal or reduction of pigmented lesions and tattoos.
Clinical Evidence
Evidence provided via in vitro, in vivo, and open literature clinical results demonstrating efficacy of normal mode ruby laser for pigmented lesion treatment and tattoo removal.
Technological Characteristics
Normal mode ruby laser; 694.3 nm wavelength; 10-40 J/cm² fluence; 200 μsec to 3 msec pulse duration; 5-12 mm spot size; articulating arm delivery; cooled contact handpiece.
Indications for Use
Indicated for treatment of benign epidermal pigmented lesions (lentigines, café au lait) and removal of blue and black tattoos.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
- pre-enactment normal mode ruby laser devices
- post-enactment devices with substantially equivalent mechanisms of action