K253219 · Shenzhen Leaflife Technology Co., Ltd. · GEX · Nov 25, 2025 · General, Plastic Surgery
Device Facts
Record ID
K253219
Device Name
LeafPure-Echo (LM-LPNdA)
Applicant
Shenzhen Leaflife Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 25, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LeafPure-Echo is intended for use in surgical and aesthetic application in the medical dermatology and general and plastic surgery as follows: 1064nm wavelength: Removal of tattoos on all skin type (Fitzpatrick skin types I-VI) with the following tattoo colours: black, brown, green, blue and purple. Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. 532nm wavelength: Removal of tattoos on Fitzpatrick skin types I-III with the following tattoo colours: red, yellow and orange. Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV.
Device Story
LeafPure-Echo is a laser surgical instrument for dermatology and plastic surgery. Device emits 1064nm and 532nm wavelengths to target specific tattoo pigments and benign pigmented lesions. Operated by clinicians in medical settings. Laser energy interacts with target chromophores to facilitate removal or treatment. Device provides aesthetic and surgical benefits by clearing unwanted pigment. No AI/ML or complex software processing described.
Technological Characteristics
Laser surgical instrument; 1064nm and 532nm wavelengths; Class II device (21 CFR 878.4810).
Indications for Use
Indicated for tattoo removal and treatment of benign pigmented lesions. 1064nm wavelength for Fitzpatrick skin types I-VI (black, brown, green, blue, purple tattoos) and types I-IV (pigmented lesions). 532nm wavelength for Fitzpatrick skin types I-III (red, yellow, orange tattoos) and types I-IV (pigmented lesions).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Related Devices
K183392 — PicoLO Nd:YAG Picosecond Laser System · Laseroptek Co., Ltd. · Jan 4, 2019
K041011 — ULTRALIGHT II ND:YAG LASER SYSTEM · Sandstone Medical Technologies, LLC · Dec 14, 2004
K083272 — SKINCLEAR Q-SWITCHED ND:YAG LASER · Global USA Distribution, LLC · Dec 16, 2008
K960772 — PHOTODERM PL. · Esc Medical Systems , Ltd. · Sep 11, 1996
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 25, 2025
Shenzhen Leaflife Technology Co., Ltd.
% Asher Zeng
Regulatory Affairs Manager
AestheticCert Compliance LLC
11/F, Block A, Gaoke Innovation Centre, Guangqiao Road
Guangming District
Shenzhen, Guangdong 518000
China
Re: K253219
Trade/Device Name: LeafPure-Echo (LM-LPNdA)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: September 26, 2025
Received: September 29, 2025
Dear Asher Zeng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253219 - Asher Zeng
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253219 - Asher Zeng
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YAN FU-S
Digitally signed by YAN FU-S
Date: 2025.11.25 13:44:52
-05'00"
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253219 | |
| Device Name LeafPure-Echo (LM-LPNdA) | |
| Indications for Use (Describe) The LeafPure-Echo is intended for use in surgical and aesthetic application in the medical dermatology and general and plastic surgery as follows: 1064nm wavelength: Removal of tattoos on all skin type (Fitzpatrick skin types I-VI) with the following tattoo colours: black, brown, green, blue and purple. Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. 532nm wavelength: Removal of tattoos on Fitzpatrick skin types I-III with the following tattoo colours: red, yellow and orange. Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.