K250165 · Wontech Co., Ltd. · GEX · Jul 3, 2025 · General, Plastic Surgery

Device Facts

Record IDK250165
Device NamePastelle
ApplicantWontech Co., Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateJul 3, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Pastelle laser system is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 1064nm in nanosecond mode, including microbeam handpieces: - Tattoo removal: dark ink (black, blue, and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation - Treatment of Common Nevi - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of melasma 1064nm in Genesis (long-pulse) mode: - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne vulgaris 532nm in nanosecond mode, including microbeam handpieces (nominal delivered energy of 595 nm and 660 nm with optional dye handpieces): - Tattoo removal: light ink (red, tan, purple, orange, sky blue, green) - Removal of Epidermal Pigmented Lesions - Removal of Minor Vascular Lesions including but not limited to telangiectasias - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of Lentigines - Treatment of Café-au-Lait - Treatment of Seborrheic Keratoses - Treatment of Post Inflammatory Hyperpigmentation - Treatment of Becker's Nevi, Freckles, and Nevi spilus

Device Story

Pastelle is a Q-switched Nd:YAG laser system for aesthetic, cosmetic, and surgical procedures. Device comprises Nd:YAG laser head, power supply, cooling system, and delivery system. Laser cavity utilizes two Nd:YAG media and two high-intensity xenon flash lamps. Operates in 1064nm (nanosecond and long-pulse Genesis modes) and 532nm (nanosecond mode) wavelengths; optional dye handpieces provide 595nm and 660nm. Operated by clinicians in dermatology/surgical settings. Laser energy delivered via articulated arm to various handpieces (zoom, collimated, microbeam). Output affects soft tissue via photothermal/photomechanical interaction for incision, excision, ablation, vaporization, coagulation, and hemostasis. Benefits include targeted treatment of pigmented lesions, vascular lesions, tattoos, and skin resurfacing.

Clinical Evidence

No clinical data. Bench testing only.

Technological Characteristics

Nd:YAG laser medium; 1064nm and 532nm wavelengths (expandable to 595nm/660nm via dye handpieces). Energy source: xenon flash lamps. Cooling: water-cooled housing. Delivery: articulated arm with zoom/microbeam handpieces. Materials: Aluminum powder (handpiece tip). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-22, IEC 60825-1.

Indications for Use

Indicated for patients requiring soft tissue incision, excision, ablation, or vaporization for general dermatology and surgery. Specific applications include tattoo removal (various ink colors), removal of Nevus of Ota, hair removal, treatment of common nevi, acne scars, wrinkles, melasma, inflammatory acne vulgaris, epidermal pigmented lesions, minor vascular lesions (telangiectasias), lentigines, café-au-lait, seborrheic keratoses, post-inflammatory hyperpigmentation, Becker's nevi, freckles, and nevi spilus.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION Build Correspondence Convert to PDF July 3, 2025 WONTECH Co., Ltd. Hyun Yoon Team Leader 64 Techono 8-ro, Yuseong-gu Daejeon, 34028 Korea, South Re: K250165 Trade/Device Name: Pastelle Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: June 5, 2025 Received: June 5, 2025 Dear Hyun Yoon: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250165 - Hyun Yoon Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250165 - Hyun Yoon Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MARK MACIOS -S Digitally signed by MARK MACIOS -S Date: 2025.07.03 09:49:40 -04'00" for Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 510(k) Number (if known) K250165 Device Name Pastelle Indications for Use (Describe) The Pastelle laser system is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 1064nm in nanosecond mode, including microbeam handpieces: - Tattoo removal: dark ink (black, blue, and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation - Treatment of Common Nevi - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of melasma 1064nm in Genesis (long-pulse) mode: - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne vulgaris 532nm in nanosecond mode, including microbeam handpieces (nominal delivered energy of 595 nm and 660 nm with optional dye handpieces): - Tattoo removal: light ink (red, tan, purple, orange, sky blue, green) - Removal of Epidermal Pigmented Lesions - Removal of Minor Vascular Lesions including but not limited to telangiectasias - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of Lentigines - Treatment of Café-au-Lait - Treatment of Seborrheic Keratoses - Treatment of Post Inflammatory Hyperpigmentation - Treatment of Becker's Nevi, Freckles, and Nevi spilus Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. FORM FDA 3881 (6/20) Page 1 of 2 PISC Publishing Services (301) 443-6740 {4} This section applies only to requirements of the Paperwork Reduction Act of 1995. **DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (6/20) Page 2 of 2 {5} W WONTECH 64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491 # 510(k) Summary [As required by 21 CFR 807.92] # K250165 ## 1. Date Prepared [21 CFR 807.92(a)(a)] June 27, 2025 ## 2. Submitter's Information &amp; Contact Person [21 CFR 807.92(a)(1)] - Name of Manufacturer: WON TECH Co., Ltd. - Address: 64 Techno 8-ro, Yuseong-gu, Daejeon, 34028, Republic of Korea - Contact Name: Hyun Sik Yoon - Telephone No.: +82-10-6750-5346 - Fax No.: +82-70-7836-0110 - Email Address: yoonhs21@wtlaser.com ## 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] Common name: Q-switched Nd:YAG Laser System Trade name: Pastelle | Classification Description | 21 CFR Section | Product Code | | --- | --- | --- | | Powered Laser Surgical Instrument | 878.4810 | GEX | As stated in 21 CFR, parts 878.4810, this generic types of devices has been classified as Class II. Page 1 / 6 {6} WONTECH 64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491 ## 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)] The identified predicate devices within this submission are shown as follow: ### #1 Predicate device - 510(k) Number: K123293 - Applicant: WON TECH Co., Ltd. - Classification Name: Powered Laser Surgical Instrument - Trade Name: Pastelle ### #2 Predicate device - 510(k) Number: K241529 - Applicant: WON TECH Co., Ltd. - Classification Name: Powered Laser Surgical Instrument - Trade Name: Pastelle Pro ### #3 Predicate device - 510(k) Number: K213569 - Applicant: Lutronic Corporation - Classification Name: Powered Laser Surgical Instrument - Trade Name: HOLLYWOOD SPECTRA Laser System ## 5. Description of the Device [21 CFR 807.92(a)(4)] The Pastelle laser system consists of an Nd:YAG laser head, a power supply, a cooling system, a delivery system and other electrical components. The laser head contains two Nd:YAG laser medium, and two high-intensity xenon flash lamps enclosed together into the water cooling housing and two reflected mirrors fixed, in the special adjustable holders composed the laser cavity. ## 6. Indications for Use [21 CFR 807.92(a)(5)] The Pastelle system is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 1064nm in nanosecond mode, including microbeam handpieces: - Tattoo removal: dark ink (black, blue, and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation - Treatment of Common Nevi {7} WONTECH 64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491 - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of melasma 1064nm in Genesis (long-pulse) mode: - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne vulgaris 532nm in nanosecond mode, including microbeam handpieces (nominal delivered energy of 595 nm and 660 nm with optional dye handpieces): - Tattoo removal: light ink (red, tan, purple, orange, sky blue, green) - Removal of Epidermal Pigmented Lesions - Removal of Minor Vascular Lesions including but not limited to telangiectasias - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of Lentigines - Treatment of Café-au-Lait - Treatment of Seborrheic Keratoses - Treatment of Post Inflammatory Hyperpigmentation - Treatment of Becker's Nevi, Freckles, and Nevi spilus Page 3 / 5 {8} WONTECH 64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491 # 7. Determination of Substantial Equivalence [21 CFR 807.92(a)(6) and 21 CFR 807.92(b)] There are no significant differences between the Pastelle and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and technical characteristics. | | Proposed Device | Predicate Device #1 | Predicate Device #2 | Predicate Device #3 | SE Decision | | --- | --- | --- | --- | --- | --- | | K Number | - | K123293 | K241529 | K213569 | - | | Manufacturer | WONTECH Co., Ltd. | WONTECH Co., Ltd. | WONTECH Co., Ltd. | Lutronic Corporation | - | | Model | Pastelle | Pastelle | Pastelle Pro | HOLLYWOOD SPECTRA Laser System | - | | Product Code | GEX | GEX | GEX | GEX | Same | | Intended Use | The Pastelle system is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 1064nm in nanosecond mode, including microbeam handpieces: - Tattoo removal: dark ink (black, blue, and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation | Pastelle Q-Switched Nd:YAG Laser System is indicated for the Incision, Excision, Ablation and Vaporization of Soft Tissue for General Dermatology, Dermatologic and General Surgical Procedures for Coagulation and Hemostasis. | The Pastelle Pro system is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 1064nm in nanosecond mode, including microbeam handpieces: - Tattoo removal: dark ink (black, blue, and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation | The HOLLYWOOD SPECTRA System is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 1064nm in nanosecond mode, including microbeam handpieces: - Tattoo removal: dark ink (black, blue, and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation | Same as Predicate Device #2 and 3 | {9} WONTECH 64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491 | | - Treatment of Common Nevi - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of melasma 1064nm in Genesis (long-pulse) mode: - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne vulgaris 532nm in nanosecond mode, including microbeam handpieces (nominal delivered energy of 595 nm and 660 nm with optional dye handpieces): - Tattoo removal: light ink (red, tan, purple, orange, sky blue, green) - Removal of Epidermal Pigmented Lesions - Removal of Minor Vascular Lesions including but not limited to telangiectasias - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of Lentigines - Treatment of Café-au-Lait - Treatment of Seborrheic Keratoses - Treatment of Post Inflammatory Hyperpigmentation - Treatment of Becker's Nevi, Freckles, and Nevi spilus | | - Treatment of Common Nevi - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of melasma 1064nm in Genesis (long-pulse) mode: - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne vulgaris 532nm in nanosecond mode, including microbeam handpieces (nominal delivered energy of 595 nm and 660 nm with optional dye handpieces): - Tattoo removal: light ink (red, tan, purple, orange, sky blue, green) - Removal of Epidermal Pigmented Lesions - Removal of Minor Vascular Lesions including but not limited to telangiectasias - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of Lentigines - Treatment of Café-au-Lait - Treatment of Seborrheic Keratoses - Treatment of Post Inflammatory Hyperpigmentation - Treatment of Becker's Nevi, Freckles, and Nevi spilus | | | | --- | --- | --- | --- | --- | --- | Page 2 / 5 {10} WONTECH 64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491 | | | | Freckles, and Nevi spilus | | | | --- | --- | --- | --- | --- | --- | | Principle/Method of Operation | The Pastelle laser system consists of an Nd:YAG laser head, a power supply, a cooling system, a delivery system and other electrical components. The laser head contains two Nd:YAG laser medium, and two high-intensity xenon flash lamps enclosed together into the water cooling housing and two reflected mirrors fixed ,in the special adjustable holders composed the laser cavity. | The Pastelle laser system consists of an Nd:YAG Laser head, a Power supply and a cooling system. The laser head contains two Nd:YAG laser medium, and two high-intensity xenon flash lamps enclosed together into the water cooling housing and two reflected mirrors fixed ,in the special adjustable holders composed the laser cavity. | The Pastelle Pro laser system consists of an Nd:YAG laser head, a power supply, a cooling system, a delivery system and other electrical components. The laser head contains two Nd:YAG laser medium, and two high-intensity xenon flash lamps enclosed together into the water cooling housing and two reflected mirrors fixed ,in the special adjustable holders composed the laser cavity. | The HOLLYWOOD SPECTRA Laser System contains a Nd:YAG (Neodymium-doped Yttrium Aluminum Garnet) resonator which generates Q-switched and/or pulsed laser sources at the nominal wavelength of 1064 nm and 532 nm using KTP. The outputs of each laser generator and the aiming beam (655 nm) are delivered by articulated arm to a fixed (collimated), or focusing variable (zoom) spot handpiece, or a dual focused dots microbeam handpiece, or a 585nm/650nm dye laser converter handpiece. The dye handpieces convert the KTP 532 nm wavelength beam into a 585 nm or 650 nm wavelengths, correspondingly. | Same as Predicate Device #1 and #2 (Q-switched, Nd:YAG) | | Laser Material | Nd:YAG | Nd:YAG | Nd:YAG | Nd:YAG | Same | | | Proposed Device | Predicate Device #1 | Predicate Device #2 | Predicate Device #3 | SE Decision | | --- | --- | --- | --- | --- | --- | | Radiation diameter | Zoom: 2mm ~ 10mm DOE: 5mm x 5mm ~ 7mm x 7mm | Zoom: 2mm ~ 10mm | Zoom: 2mm ~ 10mm Collimation: 7mm MLA: 3mm ~ 8mm DOE: 5mm x 5mm ~ 7mm x 7mm Dye: 3mm | 1064 Zoom: 1 ~ 7mm (Q-Switch, Spectra) / 2 ~ 7mm (Q-PTP, Q-3, Q-4) 532 Zoom: 0.8 ~ 6mm Collimation: 3 ~ 8mm MDF: 1064(4~8mm) / 532(3.4~6.9mm) | Same (The range of specification of submitted device is covered by all predicate | {11} WONTECH 64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491 | | | | | Dye: 2~5mm | devices' feature) | | --- | --- | --- | --- | --- | --- | | Laser output power | Genesis mode: 100 ~ 3,500 mJ 1064 nm mode: 80 ~ 1,300 mJ 1064 nm PTP mode: 80 ~ 1,600 mJ 532 nm mode: 10 ~ 500 mJ 1064 nm DOE mode: 80 ~ 900 mJ | Genesis mode: 100 ~ 2,000 mJ 1064 nm mode: 80 ~ 1,300 mJ 532 nm mode: 10 ~ 500 mJ | Genesis mode: 100 ~ 5,000 mJ 1064 nm mode: 50 ~ 1,200 mJ 1064 nm PTP mode: 600 ~ 2,000 mJ 1064 nm Triple mode: 1,000 ~ 1400 mJ 1064 nm 2x PTP mode: 1,000 ~ 1400 mJ 532 nm mode: 30 ~ 500 mJ 1064 nm DOE mode: 60 ~ 900 mJ 595 nm Dye mode: 30 ~ 250 mJ | Spectra: Up to 1500mJ 1064nm: Up to 1200mJ 1064nm PTP: Up to 1400mJ 1064nm Q-3, Q-4: Up to 1400mJ 532nm: Up to 400mJ | Same (The range of specification of submitted device is covered by all predicate devices' feature) | | Pulse width | Genesis mode: 80~480μs PTP mode: 4~48ns 1064, 532 mode: 4~48ns | Genesis mode: 250~350μs 1064 mode: 4~48ns 532 mode: 4~48ns | 532nm/1064nm mode: 5ns~12ns 1064 PTP/Triple/2X PTP: Up to 20ns Genesis mode: 80~300μs | Normal: 5~10ns 1064 Q-3, Q-4: 10~20ns Spectra mode: 190μs | Same (The range of specification of submitted device is covered by all predicate devices' feature) | | Pulse repetition rate | Genesis mode: 1 ~ 10 Hz 1064 nm mode: 1 ~ 10 Hz 1064 nm PTP mode: 1 ~ 10 Hz 532 nm mode: 1 ~ 10 Hz 1064 nm DOE mode: 1 ~ 10 Hz | Genesis mode: 1 ~ 10 Hz 1064 nm mode: 1 ~ 10 Hz 532 nm mode: 1 ~ 10 Hz | Genesis mode: 1 ~ 10 Hz 1064 nm mode: 1 ~ 10 Hz 1064 nm PTP mode: 1 ~ 10 Hz 1064 nm Tripple mode: 1 ~ 10 Hz 1064 nm 2x PTP mode: 1 ~ 10 Hz 532 nm mode: 1 ~ 10 Hz 1064 nm DOE mode: 1 ~ 10 Hz 595 nm Dye mode: 1 ~ 5 Hz 660 nm Dye mode: 1 ~ 2 Hz | 1064, 532nm: Up to 10Hz | Same (The range of specification of submitted device is covered by all predicate devices' feature) | Page 4 / 5 {12} WONTECH 64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491 # Non-Clinical Test Summary [21 CFR 807.92(b)(1)] ## 1) Electrical Safety, Electromagnetic Compatibility Testing Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards: | Standard (Edition) | Standard Title | | --- | --- | | IEC 60601-1:2005/AMD2:2020 | Medical electrical equipment — Part 1: General requirements for basic safety and essential performance — Amendment 2 | | IEC 60601-1-2:2014+A1:2020 | Medical electrical equipment — Part 1-2: General requirements for basic safety and essential performance — Collateral Standard: Electromagnetic disturbances — Requirements and tests | | IEC 60601-1-6:2010/AMD2:2020 | Medical electrical equipment — Part 1-6: General requirements for basic safety and essential performance — Collateral Standard: Usability — Amendment 2 | | IEC 60601-2-22:2019 | Medical electrical equipment — Part 2-22: Particular requirements for the basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment | | IEC 60825-1:2014 (Third Edition) | Safety of laser products - Part 1: Equipment classification and requirements | ## 2) Software Validation The PASTELLE contains BASIC documentation level software. Software was designed and developed according to a software development process and was verified and validated. The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005. ## 3) Biocompatibility | Part | Material | Patient Contact | Duration of Contact by ISO 10993-1 | Bio-compatibility | | --- | --- | --- | --- | --- | | Handpiece Tip | Aluminium Powder (Cas No. 7429-90-5) | Intact Skin | Limited (< 24 hours) | Yes | {13} WONTECH 64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491 ## 4) Performance Testing The performance of the PASTELLE has been defined as follows. ### Treatment Laser - Genesis mode (1064nm): Up to 3500mJ (80μs to 480μs) - PTP mode (1064nm): Up to 1600mJ (4ns to 48ns) - 1064nm mode: Up to 1300mJ (4ns to 48ns) - 532nm mode: Up to 500mJ (4ns to 48ns) ### Aiming Beam - Wavelength: 635nm - Output power: &lt; 5mW ## Clinical Test Summary [21 CFR 807.92(b)(2)] No clinical studies were considered necessary and performed. ## Conclusion [21 CFR 807.92(b)(3)] In according with the Federal Food &amp; drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification WON TECH Co., Ltd. concludes that the PASTELLE is substantially equivalent to predicate devices as described herein. Page 2 / 5
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...