PicoWay Laser System

K220853 · Candela Corporation · GEX · Oct 13, 2022 · General, Plastic Surgery

Device Facts

Record IDK220853
Device NamePicoWay Laser System
ApplicantCandela Corporation
Product CodeGEX · General, Plastic Surgery
Decision DateOct 13, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K220853 · Oct 13, 2022PicoWay Laser SystemCandela CorporationPeer-reviewed clinical literature; Retrospective clinical studies; Prospective open-label clinical trialsThe sponsor conducted a systematic literature review of 9 peer-reviewed articles and identified an additional investigator-initiated prospective study to support the safety and performance of the PicoWay Laser System for the treatment of benign pigmented lesions (melasma, CALMs, nevus of Ota, and lentigines).Literature Review; Retrospective Study; Benign Pigmented Lesions; Melasma; Nevus of Ota

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Systematic Literature Review (9 articles); Randomized controlled, prospective open-label, and retrospective evaluator-blinded studies; Study Period: January 1, 2015 to May 31, 2022262 subjects (Fitzpatrick Skin Types I-V) with benign pigmented lesions; Sample Size: 262; Number of Sites: Multiple (USA, UK, Thailand, China, Israel)Not applicable for this studySafety and efficacy for treatment of benign pigmented lesions
Investigator-initiated prospective, open-label clinical study (Bernstein, M.D.); Prospective, open-label clinical study; Follow-up/Duration: 3-month follow-upSubjects with melasma; Number of Sites: 1 (Main Line Center Surgery for Laser Surgery, Ardmore, PA)BaselineMean mMASI scores, investigator and subject satisfaction

Indications for Use

The PicoWay laser system is indicated for the following at the specified wavelength: 532 nm: Removal of tattoos for Fitzpatrick Skin Types I-III to treat the following tattoo colors: red, yellow and orange. 730 nm: Removal of tattoos for Fitzpatrick Skin Types II-IV to treat the following tattoo colors: green and blue. 785 nm: Removal of tattoos for Fitzpatrick Skin Types II-IV to treat the following tattoo colors: green and blue. 1064 nm: Removal of tattoos for all Fitzpatrick Skin Types to treat the following tattoo colors: black, brown, green, blue and purple. The PicoWay laser system is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. The Resolve handpiece (1064 nm) is also indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V and for treatment of Melasma for Fitzpatrick Skin Types I-IV. The Resolve handpieces (532 nm HE, 532 nm, and 1064 nm) are also indicated for the treatment of wrinkles in Fitzpatrick Skin Types I-IV. The Resolve Fusion handpiece (532 nm) is indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. The PicoWay laser system is indicated for the following at the specified wavelengths: 532 nm: Treatment of Melasma for Fitzpatrick Skin Types I-IV. Treatment of café au lait macules (CALMs) for Fitzpatrick Skin Types I-IV. Treatment of Lentigines for Fitzpatrick Skin Types I-IV. 730 nm: Treatment of Lentigines for Fitzpatrick Skin Types I-IV. 1064 nm: Treatment of Melasma for Fitzpatrick Skin Types I-IV. Treatment of Nevus of Ota for Fitzpatrick Skin Types III-IV.

Device Story

Solid-state laser system; delivers 532 nm, 730 nm, 785 nm, or 1064 nm wavelengths; pulse duration 240-500 ps. Uses 755 nm Alexandrite laser to pump 1064/532 nm wavelengths. Output delivered via articulated arm with Zoom, Resolve, or Resolve Fusion handpieces. Resolve handpieces deliver 10x10 array of focused spots; Resolve Fusion delivers focused and ring spots. Used in clinical settings by physicians/trained professionals. Output energy targets chromophores in skin for tattoo removal, lesion treatment, scar reduction, and skin rejuvenation. Clinical decision-making based on visual assessment of skin condition and Fitzpatrick type. Benefits include non-invasive treatment of pigmented lesions and aesthetic skin concerns.

Clinical Evidence

Systematic literature review of 9 peer-reviewed studies (n=262 subjects) and one investigator-initiated prospective study. Indications included melasma, CALMs, Nevus of Ota, and lentigines. Results showed statistically significant improvement in mMASI scores for melasma (p<0.05) and high investigator/subject satisfaction (>85%). No device-related adverse events reported.

Technological Characteristics

Solid-state laser; wavelengths 532/730/785/1064 nm; pulse duration 240-500 ps. Materials: Aluminum (patient-contacting). Delivery: Articulated arm with Zoom, Resolve, or Resolve Fusion handpieces. Power: 200-240 VAC, 50/60 Hz. User interface: Touchscreen GUI. Standalone system.

Indications for Use

Indicated for tattoo removal (Fitzpatrick I-IV, specific colors), benign pigmented lesions (Fitzpatrick I-IV), acne scars (Fitzpatrick II-V), wrinkles (Fitzpatrick I-IV), melasma (Fitzpatrick I-IV), café au lait macules (Fitzpatrick I-IV), lentigines (Fitzpatrick I-IV), and Nevus of Ota (Fitzpatrick III-IV).

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 13, 2022 Candela Corporation Danielle Gibboney Sr. Regulatory Affairs Specialist 251 Locke Drive Marlborough, Massachusetts 01752 Re: K220853 Trade/Device Name: PicoWay Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 8, 2022 Received: September 8, 2022 Dear Danielle Gibboney: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K220853 Device Name PicoWay Laser System #### Indications for Use (Describe) The PicoWay laser system is indicated for the following at the specified wavelength: - · 532 mm: Removal of tattoos for Fitzpatrick Skin Types I-III to treat the following tattoo colors: red, yellow and orange. - · 730 nm: Removal of tattoos for Fitzpatrick Skin Types II-IV to treat the following tattoo colors: green and blue. - · 785 mm: Removal of tattoos for Fitzpatrick Skin Types II-IV to treat the following tattoo colors: green and blue. · 1064 nm: Removal of tattoos for all Fitzpatrick Skin Types to treat the following tattoo colors: black, brown, green, blue and purple. The Pico Way laser system is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. The Resolve handpiece (1064 nm) is also indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V and for treatment of Melasma for Fitzpatrick Skin Types I-IV. The Resolve handpieces (532 nm, and 1064 nm) are also indicated for the treatment of wrinkles in Fitzpatrick Skin Types I-IV. The Resolve Fusion handpiece (532 nm) is indicated for benign pigmented for Fitzpatrick Skin Types I-IV. The PicoWay laser system is indicated for the following at the specified wavelengths: 532 nm : · Treatment of Melasma for Fitzpatrick Skin Types I-IV. · Treatment of café au lait macules (CALMs) for Fitzpatrick Skin Types I-IV. - · Treatment of Lentigines for Fitzpatrick Skin Types I-IV. 730 nm: · Treatment of Lentigines for Fitzpatrick Skin Types I-IV. 1064 nm: - · Treatment of Melasma for Fitzpatrick Skin Types I-IV. - · Treatment of Nevus of Ota for Fitzpatrick Skin Types III-IV. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image is a logo for Candela. The logo features a purple abstract shape on the left, followed by the word "CANDELA" in black, sans-serif font. Below the word "CANDELA" is the tagline "Science. Results. Trust." in a smaller, purple, sans-serif font. The logo is clean and modern, with a focus on the company's name and values. # 510(k) SUMMARY K220853 # Picoway Laser System This summary of 510(k) submitted in accordance with the requirements of 21 CFR 807.92. ## 1. Date Prepared Остовек 12, 2022 #### 2. APPLICANT NAME Candela Corporation 251 Locke Drive Marlborough MA 01752 USA #### 3. Official Correspondent Danielle Gibboney Sr. Regulatory Affairs Specialist Candela Corporation 251 Locke Drive Marlborough MA 01752 USA Phone: 617-904-3820 Email: danielleg@candelamedical.com - 4. Product Information Name of Device: Picoway Laser System Common/Usual Name: Powered Laser Surgical Instrument Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology (per 21 CFR Part 878.4810) Device Classification: Class II (per 21 CFR Part 878.4810) Product Code: GEX {4}------------------------------------------------ # 5. Legally Marketed Predicate Device for Claimed Equivalence: Predicate Device: Picoway Laser System (K191685) # 6. Device Description: The PicoWay Laser System is a solid-state laser capable of delivering energy at wavelengths of 1064 nm, 532 nm, or 730 nm, at extremely short duration in the range of 240-500 (ps). The laser system contains one 755 nm (Alexandrite) laser head which is used to 'pump' (create) the 1064 / 532nm wavelengths. The outputs of the two lasers are optically combined on the laser rail so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system which can output either the 532nm or 1064nm wavelengths. The 532nm or 1064nm output energy is delivered to the skin through an Articulated Arm and Zoom Handpiece (HP) delivery system. There are also 3 hand pieces (Resolve), one for 1064nm and two for 532nm (low energy and high energy), which deliver a 10 x 10 array of focused spots to the skin and 1 Axicon hand piece (Resolve Fusion) for 532nm, which delivers a 10 x 10 array of focused and ring spots to the skin. The following handpieces are cleared to be used with Picoway Laser System Zoom Handpieces: ● - 532 nm - 1064 nm Handpieces: - 730 nm ● - 785 nm Resolve Handpieces: - 532 nm ● - 532 nm HE ● - . 1064 nm Resolve Fusion Handpieces: - 532 nm ● - . 1064 nm #### 7. Intended Use and Indications for Use: The PicoWay laser system is indicated for the following at the specified wavelength: - 532 nm: Removal of tattoos for Fitzpatrick Skin Types I-III to treat the following tattoo ● colors: red, yellow and orange. - 730 nm: Removal of tattoos for Fitzpatrick Skin Types II-IV to treat the following . tattoo colors: green and blue. - . 785 nm: Removal of tattoos for Fitzpatrick Skin Types II-IV to treat the following tattoo colors: green and blue. - 1064 nm: Removal of tattoos for all Fitzpatrick Skin Types to treat the following tattoo . colors: black, brown, green, blue and purple. {5}------------------------------------------------ The PicoWay laser system is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. The Resolve handpiece (1064 nm) is also indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V and for treatment of Melasma for Fitzpatrick Skin Types I-IV. The Resolve handpieces (532 nm HE, 532 nm, and 1064 nm) are also indicated for the treatment of wrinkles in Fitzpatrick Skin Types I-IV. The Resolve Fusion handpiece (532 nm) is indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. #### The PicoWay laser system is indicated for the following at the specified wavelengths: #### 532 nm: - Treatment of Melasma for Fitzpatrick Skin Types I-IV. 0 - Treatment of café au lait macules (CALMs) for Fitzpatrick Skin Types I-IV. 0 - Treatment of Lentigines for Fitzpatrick Skin Types I-IV. 0 #### 730 nm: - o Treatment of Lentigines for Fitzpatrick Skin Types I-IV. #### 1064 nm: - Treatment of Melasma for Fitzpatrick Skin Types I-IV. o - Treatment of Nevus of Ota for Fitzpatrick Skin Types III-IV. O #### 8. TECHNOLOGICAL COMPARISON: The subject device Picoway Laser System is substantially equivalent and identical in the design, function, and intended use to the Picoway Laser System (K191685). The difference between the subject Picoway Laser System and its predicate is the additional indications for benign pigmented lesions including but not limited to: Nevus of Ota, café au lait macules (CALMs), melasma, and lentigines that this Premarket Notification is proposing. The expanded indications between the subject device and its predicate does not raise any new concerns of safety or effectiveness of the device. Thus, based on the information presented in this Premarket Notification, Picoway Laser System is substantially equivalent to its predicate Picowav Laser System (K191685). Please refer to specification comparison tables in Table 1 and Table 2 for comparisons between intended use/indications for use, and technological & biological characteristic comparison below. {6}------------------------------------------------ | Table 1: Intended/Indication for use comparison table. | | |--------------------------------------------------------|--| |--------------------------------------------------------|--| | Name of Device:<br>510(k)<br>Product Code<br>Device Class | Picoway Laser System<br>Proposed<br>K220853<br>GEX<br>Class II | Picoway Laser System<br>Predicate<br>K191685<br>GEX<br>Class II | | |-----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------| | Intended use /<br>Indications: | The PicoWay laser system is indicated for the following at the specified wavelength:<br>532 nm: Removal of tattoos for Fitzpatrick Skin Types I-III to treat the following tattoo colors: red, yellow and orange.730 nm: Removal of tattoos for Fitzpatrick Skin Types II-IV to treat the following tattoo colors: green and blue.785 nm: Removal of tattoos for Fitzpatrick Skin Types II-IV to treat the following tattoo colors: green and blue.1064 nm: Removal of tattoos for all Fitzpatrick Skin Types to treat the following tattoo colors: black, brown, green, blue and purple. | The PicoWay laser system is indicated for the following at the specified wavelength:<br>532 nm: Removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors:<br>red, yellow and orange.730 nm: Removal of tattoos for Fitzpatrick skin types II-IV to treat the following tattoo colors:<br>green and blue.785 nm: Removal of tattoos for Fitzpatrick skin types II-IV to treat the following tattoo colors:<br>green and blue.1064 nm: Removal of tattoos for all skin types (Fitzpatrick I-VI) to treat the following tattoo<br>colors: black, brown, green, blue and purple. | | | | The PicoWay laser system is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. | The PicoWay laser system is also indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. | | | | The Resolve handpiece (1064 nm) is also indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V and for treatment of Melasma for Fitzpatrick Skin Types I-IV. | The Resolve handpiece (1064 nm) is also indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V. | | | | The Resolve handpieces (532 nm HE, 532 nm, and 1064 nm) are also indicated for the treatment of wrinkles in Fitzpatrick Skin Types I-IV. | The Resolve handpieces (532 nm HE, 532 nm, 1064 nm) are also indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV. | | | | The Resolve Fusion handpiece (532 nm) is indicated for benign pigmented lesions removal for Fitzpatrick Skin Types I-IV. | The Resolve Fusion handpiece (1064 nm) is indicated for the treatment of wrinkles as well as benign pigmented lesions in Fitzpatrick Skin Types I-IV. | | | | The PicoWay laser system with Zoom Handpieces is indicated for the following at the specified wavelengths: | The Resolve Fusion handpiece (532 nm) is indicated for the treatment | | | | 532 nm: | of benign pigmented lesions in Fitzpatrick Skin Types I-IV. | | | | Treatment of Melasma for Fitzpatrick Skin Types I-IV. Treatment of café au lait macules (CALMs) for Fitzpatrick Skin Types I-IV. Treatment of Lentigines for Fitzpatrick Skin Types I-IV. | | | | | | | 730 nm: | | Treatment of Lentigines for Fitzpatrick Skin Types I-IV. | | | | | | 1064 nm: | | | | | Treatment of Melasma for Fitzpatrick Skin Types I-IV. Treatment of Nevus of Ota for Fitzpatrick Skin Types III-IV. | | | | | Similarities/Differences | | Identical to Predicate device, but with expanded indications identified in <b>BOLD</b> . | {7}------------------------------------------------ {8}------------------------------------------------ | General<br>Specifications | Picoway Laser System<br>Subject Device<br>Not Assigned | | | | Picoway Laser System<br>Predicate<br>K191685 | | | | |---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|------------------------------------|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|-------------------------------|---------------------------------------------------------------------------| | Wavelength | IDENTICAL<br>532 nm | IDENTICAL<br>730 nm | IDENTICAL<br>785 nm | IDENTICAL<br>1064 nm | IDENTICAL<br>532 nm | IDENTICAL<br>730 nm | IDENTICAL<br>785 nm | IDENTICAL<br>1064 nm | | | IDENTICAL | IDENTICAL | IDENTICAL | IDENTICAL | IDENTICAL | IDENTICAL | IDENTICAL | IDENTICAL | | Laser Type | Resolve: &<br>Resolve<br>Fusion<br>Double Nd:<br>Yag<br>Zoom:<br>Titanium<br>Sapphire | Handpiece:<br>Titanium<br>Sapphire | Handpiece:<br>Titanium<br>Sapphire | Resolve & Resolve Fusion:<br>Nd: Yag Frequency<br>Zoom: Titanium Sapphire | Resolve: &<br>Resolve<br>Fusion<br>Double Nd:<br>Yag | Zoom:<br>Titanium<br>Sapphire | Zoom:<br>Titanium<br>Sapphire | Resolve & Resolve Fusion:<br>Nd: Yag Frequency<br>Zoom: Titanium Sapphire | | Repetition Rate | IDENTICAL | | | | IDENTICAL | | | | | (Hz) | Single 1, 2, 2, 4, 5, 6, 7, 8, 9, 10 Hz | | | | Single 1, 2, 2, 4, 5, 6, 7, 8, 9, 10 Hz<br>IDENTICAL | | | | | Fluence Ranges | IDENTICAL<br>Zoom Handpiece:<br>2 mm: 1-00-6.25 J/cm²<br>3 mm: 0.040-2.80 J/cm²<br>4 mm: 0.30-1.60 J/cm²<br>5 mm: 0.20-1.00 J/cm2<br>6 mm: 0.20-0.72 J/cm2<br>7 mm: 0.26-0.52 J/cm2<br>8 mm: 0.20- 0.40 J/cm2 | | | | Zoom Handpiece:<br>2 mm: 1-00-6.25 J/cm²<br>3 mm: 0.040-2.80 J/cm²<br>4 mm: 0.30-1.60 J/cm²<br>5 mm: 0.20-1.00 J/cm2<br>6 mm: 0.20-0.72 J/cm2<br>7 mm: 0.26-0.52 J/cm2<br>8 mm: 0.20- 0.40 J/cm2 | | | | # Table 2: Technological & Biological specification comparison {9}------------------------------------------------ | General<br>Specifications | Picoway Laser System<br>Subject Device<br>Not Assigned | Picoway Laser System<br>Predicate<br>K191685 | |---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | 9 mm: 0.16-0.32 J/cm2<br>10 mm: 0.13-0.25 J/cm2<br>Resolve High Energy Handpiece:<br>0.20-1.50 mJ/µbeam<br>Resolve Low Energy Handpiece:<br>0.16-0.30 mJ/µbeam<br>Resolve Fusion Handpiece:<br>0.20-0.70<br>IDENTICAL | 9 mm: 0.16-0.32 J/cm2<br>10 mm: 0.13-0.25 J/cm2<br>Resolve High Energy Handpiece:<br>0.20-1.50 mJ/µbeam<br>Resolve Low Energy Handpiece:<br>0.16-0.30 mJ/µbeam<br>Resolve Fusion Handpiece:<br>0.20-0.70<br>IDENTICAL | | Pulse Duration | 240-500 picosecond<br>IDENTICAL | 240-500 picosecond<br>IDENTICAL | | Spot Size (mm) | Zoom:<br>2 to 10 in increments of 1mm<br>Resolve:<br>6 mm x 6 mm array in 10 x 10 matrix<br>Resolve Fusion:<br>6 mm x 6 mm array in 10 x 10 matrix | Zoom:<br>2 to 10 in increments of 1mm<br>Resolve:<br>6 mm x 6 mm array in 10 x 10 matrix<br>Resolve Fusion:<br>6 mm x 6 mm array in 10 x 10 matrix | | BIOLOGICAL CHARACTERISTICS | | | | Patient<br>Contacting<br>Material | IDENTICAL<br>Aluminum | IDENTICAL<br>Aluminum | | SYSTEMS | | | | Delivery system | IDENTICAL<br>Articulated arm with dedicate handpiece | IDENTICAL<br>Articulated arm with dedicate handpiece | | User Interface | IDENTICAL<br>Touchscreen with GUI | IDENTICAL<br>Touchscreen with GUI | | Electrical Power | IDENTICAL | IDENTICAL | | General<br>Specifications | Picoway Laser System<br>Subject Device<br>Not Assigned | Picoway Laser System<br>Predicate<br>K191685 | | | 200-240 VAC, 50/60 HZ, 4600 VA single | 200-240 VAC, 50/60 HZ, 4600 VA single | | Physical<br>Dimensions<br>/Weight (Console) | IDENTICAL<br>275 lb (125 kg) | IDENTICAL<br>275 lb (125 kg) | {10}------------------------------------------------ {11}------------------------------------------------ ## 9. Performance Data: ### Performance Testing: Bench: The performance testing of the subject PicoWay Laser System is based on the established testing previous cleared under PicoWay Laster System predicate (K191685). There are no changes in the design therefore the subject Picoway Laser System is based on the established performance testing of the device's predicate. ## Performance Testing-Clinical A systematic literature search, using PubMed, Embase and Cochrane databases, was conducted to identify peer-reviewed articles published during January 1, 2015 to May 31, 2022 in which the PicoWay Laser System was used to treat benign pigmented lesions, with any of the PicoWay Laser Handpieces at 532 nm, 730 nm, 785 nm, and 1064 nm wavelengths. A combination of MeSH terms and free-text searches for the device (picoway laser, picosecond laser) and for pigmented lesions and specific benign pigmented lesions nevus of Ota, CALM (macules) and/or melasma, were included in the search criteria. A supplemental search in ClinicalTrials.gov was performed using similar search terms to identify clinical trials with results relevant to the clinical safety and performance of the Picoway Laser System. The Literature Search methodology and findings are described in detail within the Clinical Literature Summary below. Nine articles were identified that reported on randomized controlled, or prospective, open label, evaluator-blinded clinical trials or retrospective evaluatorblinded studies that treated at least 10 individuals in each study using the PicoWay Laser System. The studies were conducted in the USA, the UK, Asia (Thailand, China) and the Middle East (Israel) and included a total of 262 subjects (228 females, 34 males; Fitzpatrick Skin Type (FST) I-V) treated for benign pigmented lesions, including nevus of Ota (n=20), café au lait macules (n=32), melasma (n=70) and lentigines (n=140), using the PicoWay Laser System. Melasma was treated with the PicoWay Zoom handpieces (532 nm and/or 1064 nm) and the PicoWay Resolve 1064 nm handpiece alone or combined with 12 weeks of daily application of 4% hydroquinone cream. CALMs were treated with the Zoom 532 nm handpiece; nevus of Ota was treated using the PicoWay Zoom 1064 nm handpiece, and lentigines were treated with the PicoWay 730 nm handpiece and the Zoom 532 nm handpiece. The articles are identified in table 3 below. | Article | Title | Indication | PicoWay Laser<br>Handpiece (s) | |---------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1 | Wong CSM, Chan MWM, Shek SYN, Yeung<br>CK & Chan HHL. Fractional 1064 nm<br>picosecond laser in treatment of melasma<br>and skin rejuvenation in Asians, a | Melasma | Melasma: (Resolve<br>1064 nm laser) | | Article | Title | Indication | PicoWay Laser<br>Handpiece (s) | | 1 | prospective study. Lasers Surg Med<br>2021;53(8):1032-1042. | | | | 2 | Kung KY, Shek SYN, Yeung CK, & Chan<br>HHL. Evaluation of the safety and efficacy of<br>the dual wavelength picosecond laser for the<br>treatment of benign pigmented lesions in<br>Asians. Lasers Surg Med 2019;51(1):14-22. | Melasma;<br>CALMs | Melasma: (Zoom 1064<br>nm laser)<br>CALM: (Zoom 532 nm<br>laser) | | 3 | Chalermchai T, Rummaneethorn P. Effects of<br>a fractional picosecond 1,064 nm laser for the<br>treatment of dermal and mixed type<br>melasma. J Cosmet Laser Ther. 2018<br>Jun;20(3):134-139.<br>*article included treatement 4% hydroquinone | Melasma | Melasma: (Resolve<br>1064 nm laser) | | 4 | Mehrabi JN, Friedman O, Al-Niaimi F & Artzi,<br>O. Retrospective photographic review of<br>nontattoo indications treated by picosecond<br>laser. J Cosmet Dermatol 2020;<br>Mar;19(3):612-621. | Melasma;<br>CALMs;<br>Nevus of Ota | Melasma: (Resolve<br>1064 nm laser and<br>Zoom 1064 nm or 532<br>nm lasers)<br>CALM: (Zoom 532 nm<br>laser)<br>Nevus of Ota: (Zoom<br>1064 nm laser)<br><br>Lentigines:<br>(Zoom 532 nm laser) | | 5 | Yang H, Guo L, Jia G, Gong X, Wu Q, Zeng<br>R, Zhang M, Ding H, Fang F, Zheng H, Liu X,<br>Ge Y, Yang Y, Lin T. Treatment of nevus of<br>Ota with 1064 nm picosecond Nd:YAG laser:<br>A retrospective study. Dermatol Ther 2021;<br>Nov;34(6):e15152. | Nevus of Ota | Nevus of Ota: (Zoom<br>1064 nm laser) | | 6 | Artzi O, Mehrabi JN, Koren A, Niv R,<br>Lapidoth M, Levi A. Picosecond 532-nm<br>neodymium-doped yttrium aluminium garnet<br>laser-a novel and promising modality for the<br>treatment of café-au-lait macules. Lasers<br>Med Sci 2018 May;33(4):693-697. | CALMs | CALM: (Zoom 532 nm<br>laser) | | 7 | Kauvar ANB, Sun R, Bhawan J, Singh G,<br>Ugonabo N, Feng H, Schomacker K.<br>Treatment of facial and non-facial lentigines<br>with a 730 nm picosecond titanium: Sapphire<br>laser is safe and effective. Lasers Surg Med.<br>2022 Jan;54(1):89-97. | Lentigines | Lentigines (730 nm<br>laser) | | Article | Title | Indication | PicoWay Laser<br>Handpiece (s) | | 8 | Vachiramon V, Namasondhi A, Anuntrangsee<br>T, Jurairattanaporn N. Randomized,<br>evaluator-blinded comparative study of a<br>potassium titanyl phosphate (KTP) 532-nm<br>picosecond laser and an alexandrite 755-nm<br>picosecond laser for the treatment of solar<br>lentigines in Asians. J Cosmet Dermatol.<br>2022 Feb 7. | Lentigines | Lentigines (Zoom 532<br>nm laser) | | 9 | Vachiramon V, lamsumang W, Triyangkulsri<br>K. Q-switched double frequency Nd:YAG<br>532-nm nanosecond laser vs. double<br>frequency Nd:YAG 532-nm picosecond laser<br>for the treatment of solar lentigines in Asians.<br>Lasers Med Sci. 2018 Dec;33(9):1941-1947. | Lentigines | Lentigines (Zoom 532<br>nm laser) | Table 3. Identification of Clinical Articles in PicoWay literature search analysis {12}------------------------------------------------ {13}------------------------------------------------ Additionally, Candela identified an investigator-initiated prospective, open-label clinical study that had been conducted by Principal Investigator Eric F. Bernstein, M.D. of Main Line Center Surgery for Laser Surgery in Ardmore, PA. In this study, treatment of melasma with the PicoWay® device using the 1064nm Resolve handpiece resulted in statistically significant improvement (p<0.05) in mean mMASI scores at 3-month follow-up visits in comparison to baseline. Majority of the subjects (≥70 %) demonstrated improvement in the appearance of melasma at both followup visits. Investigator and Subject satisfaction ratings with treatment outcome were remarkably high (> 85%). Treatments were well tolerated, and there were no device-related adverse events. # 10. SUBSTANTIAL EQUIVALENCE COMPARISON When comparing the subject Picoway Laser System to the predicate device Picoway Laser System (K191685) the additional indications for use does not raise any new issues of safety and effectiveness. There are no technological changes between the subject and predicate device. The Picoway Laser System is substantially equivalent, in terms of technological characteristics, performance, and intended use to the predicate device Picoway Laser System (K191685) as they are identical.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...