K201109 · Shanghai Apolo Medical Technology Co., Ltd. · GEX · Jul 15, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201109
Device Name
CO2 Laser Therapy System
Applicant
Shanghai Apolo Medical Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CO2 Laser Therapy System is used for body soft tissue vaporization and coagulation in dermatology and plastic surgery, general surgery, gynecology.
Device Story
CO2 Laser Therapy System generates 10,600nm wavelength laser energy; energy absorbed by tissue water causing boiling and evaporation (vaporization). Adjacent tissue absorbs heat causing coagulation (hemostasis) and thermal stimulation of deep skin layers (neocollagenesis). Device used in dermatology, plastic surgery, general surgery, and gynecology. Operated by clinicians. System includes RF sealed-off CO2 laser, aiming beam, and footswitch controls. Output power ranges 1-55W depending on model. User interface via LCD. Device provides fractional, normal, and vaginal operational modes. Clinical benefit includes tissue ablation and coagulation for surgical and aesthetic procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and validation studies confirming compliance with electrical safety and electromagnetic compatibility standards.
Technological Characteristics
RF sealed-off CO2 laser; 10,600nm wavelength; Class IV laser. Output power 1-55W; pulse energy 1-300mJ/dot. Operational modes: fractional, normal, vaginal (CW, Single, Pulse, S. pulse, U. pulse). Aiming beam: <2mW/650nm semiconductor laser. User interface: LCD. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22.
Indications for Use
Indicated for body soft tissue vaporization and coagulation in dermatology, plastic surgery, general surgery, and gynecology.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
July 15, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
Shanghai Apolo Medical Technology Co., Ltd. % Shelley Li Director Shanghai Landlink Medical Information Technology Co., Ltd. Room 703, 705, Building 1, West Guangzhong Road 555 Jingan District, Shanghai, Shanghai 200071, China
Re: K201109
Trade/Device Name: CO2 Laser Therapy System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 16, 2020 Received: April 27, 2020
Dear Shelley Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K201109
Device Name CO2 Laser Therapy System
Indications for Use (Describe)
The CO2 Laser Therapy System is used for body soft tissue vaporization in dermatology and plastic surgery, general surgery, gynecology.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt;"> <span style="font-family:Arial;"> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt;"> <span style="font-family:Arial;"> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K201109-510(k) summary
#### l Submitter
Shanghai Apolo Medical Technology Co., Ltd. 4F, Building A, No.388, Yindu Road, Xuhui District, Shanghai 200231, China
Establishment Registration Number: 3007120647
Contact person: Felix Li Position: Regulatory Affairs Phone: +86-138 4919 0618 Fax: +86-21-34622840 E-mail: liqiang@apolo.com.cn
Preparation date: 2020-07-09
#### II Proposed Device
| Trade Name of Device: | CO2 Laser Therapy System |
|-----------------------|-----------------------------------|
| Common name: | Powered Laser Surgical Instrument |
| Regulation Number: | 21 CFR 878.4810 |
| Regulatory Class: | Class II |
| Product code: | GEX |
| Review Panel | General & Plastic Surgery |
### III Predicate Devices
| a | 510(k) Number: | K162169 |
|-----------------|-----------------------------------|-----------------------------------|
| | Trade name: | EdgeOne CO2 Laser |
| | Common name: | Powered Laser Surgical Instrument |
| | Classification: | Class II |
| | Product Code: | GEX |
| | Manufacturer | Jeisys Medical Inc. |
| b | 510(k) Number: | K133895 |
| 510(k) Number: | KT33895 | |
| Trade name: | DEKA SmartXide2 Laser System | |
| Common name: | Powered Laser Surgical Instrument | |
| Classification: | Class II | |
| Product Code: | GEX | |
| Manufacturer | Electronic Engineering S.p.A. | |
{4}------------------------------------------------
#### IV Device description
The CO2 Laser Therapy Systems generate a 10,600nm wavelength, which is absorbed by water in the tissue. The laser energy heats up the water until it reaches a boiling point causing the evaporation of the affected tissue. Some heat is absorbed by tissue adjacent to the ablated target area, causing tissue coagulation which induces hemostasis (the cessation of bleeding) as well as thermal stimulation of deep skin layers, which induces fibroblast stimulation and neocollagenesis (the formation of new collagen).
### V Indication for use
The CO2 Laser Therapy System is used for body soft tissue vaporization and coagulation in dermatology and plastic surgery, general surgery, gynecology.
| Item | Proposed device | Predicate device<br>(K162169) | Predicate device<br>(K133895) | |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Product name | CO₂ Laser Therapy System | EdgeOne CO₂ Laser | DEKA SmartXide² Laser System | |
| Product Code | GEX | GEX | GEX | |
| Regulati on No. | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | |
| Class | Class II | Class II | Class II | |
| Indicatio n for use | The CO₂ Laser Therapy System is used for body soft tissue vaporization and coagulation in dermatology and plastic surgery, general surgery, gynecology. | It is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynaecology, neurosurgery, orthopaedics, general and thorasic surgery (including open and | It is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynaecology, neurosurgery, orthopaedics, general and thorasic | |
| | | | endoscopic), dental surgery | (including open and |
| | | | and oral surgery and genitourinary surgery. | endoscopic), dental and oral surgery and |
| | | | The use with the scanning unit is | genitourinary surgery. The use |
| | | | indication for ablative skin resurfacing. | with the scanning unit is indication for ablative skin |
| | | | | resurfacing. |
| Laser<br>Type | RF Sealed-off CO2 | | CO2 | CO2 |
| Light<br>Delivery<br>system | | | Articulated arm | Articulated arm |
| CO2<br>Laser<br>wavelen<br>gth | 10600nm | | 10600nm | 10600nm |
| Aiming<br>beam<br>wavelen<br>gth | < 2mw /650nm<br>/Semiconductor Laser<br>LD | | Diode laser(Red) 655<br>+/-<br>10nm, Max 1mW | Diode laser (Red)<br>635nm,<br>4mW max |
| Laser<br>Controls | Footswitch | | Footswitch | Footswitch |
| Output<br>power | HS-411: 1~35W<br>HS-411A: 1~55W | | 30W | 60W |
| Pulse<br>energy | 1~300mJ/dot | | 1-300mJ | Unknown |
| Pulse<br>duration | CW | - | 1-1000ms | 1-2000us |
| | Single | 10~500ms | | |
| | Pulse | On time:<br>5~500ms<br>Off time:<br>1-500ms | | |
| | S.<br>Pulse | On time: 1~4ms<br>Off time:<br>1-100ms | | |
VI Comparison of technological characteristics with the predicate devices
{5}------------------------------------------------
{6}------------------------------------------------
| | U<br>puls<br>e | On time:<br>0.1~0.9ms<br>Off time:<br>1-100ms | | | | | | |
|-----------------------------|----------------------------------------------------------------------------------------|-----------------------------------------------|--|---------------------------------------------------------------------------------------------|----------------------|--|---------------------------------------------------------------------------|--|
| Spot<br>size | 150um (fractional) | | | 120um, 350um, 800um | | | 125μm, 155μm,<br>267μm,<br>325μm, 489μm,<br>530μm | |
| Scan<br>area<br>size | 2x2mm~20x20mm | | | 15mmx15mm | | | 15mmx15mm | |
| Operatio<br>nal<br>mode | Fractional mode,<br>normal mode and vaginal<br>(CW, Single, Pulse, S. pulse, U. pulse) | | | Fractional mode,<br>normal mode (CW, Pulse,<br>Single Pulse, Repeat, Group<br>pulse, Ultra) | | | Fractional CW, SP,<br>DP,<br>HP, UP<br>Normal<br>Interlaced<br>SmartTrack | |
| User<br>interface | LCD | color<br>Touchscreen | | LCD | color<br>Touchscreen | | LCD color<br>Touchscreen | |
| Laser<br>classific<br>ation | Class IV | | | Class IV | | | Class IV | |
| Software | Yes | | | Yes | | | Yes | |
### VII Non-Clinical Testing
A battery of tests have been performed to verity that the proposed device met all design specification. The test result demonstrated that the proposed device complies with the following standards:
# Electrical safety and electromagnetic compatibility
IEC 60601-1: 2005+corr.1:2006+Corr.2.2007+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC 60825-1:2014 Safety of Laser products-Part 1: Equipment classification and requirements
IEC 60601-2-22:2007(third edition)+A1:2012 for use in conjunction with IEC
{7}------------------------------------------------
60601-1:2005 (third edition)+A1:2012 Medical electrical equipment - Part 2-22: Particular requirements for the safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
## VIII Clinical Testing
It is not applicable.
# IX Conclusion
Base on the performance testing and validation studies that the subject device is substantially equivalent to the predicate device.