K133895 · El.En S.P.A. · GEX · Sep 5, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133895
Device Name
DEKA SMARTXIDE LASER SYSTEM
Applicant
El.En S.P.A.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 5, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The SmartXide² laser is indicated for incision, excision, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynaecology, neurosurgery, orthopaedics, general and thorasic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.
Device Story
DEKA SmartXide2 is a 10600 nm CO2 laser system (max 60W) for soft tissue surgery and skin resurfacing. Laser energy is delivered via an articulated arm and distal accessory; activation is controlled by a footswitch. Used in clinical settings (OR, clinic) by physicians. Device transforms electrical input into laser energy for tissue interaction (incision, excision, ablation, coagulation). Benefits include precise tissue management across multiple surgical specialties. System is a floor-standing unit (95 kg).
Clinical Evidence
No human clinical data provided. Bench testing and an animal study on sheep skin were conducted to compare thermal lesion and ablation dimensions (depth and width) against predicate devices.
Technological Characteristics
10600 nm CO2 laser; 60W max power; articulated arm delivery system; footswitch activation; 100-120 Vac 15A, 50-60 Hz power requirement; dimensions 210 x 59 x 56 cm; weight 95 kg.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissues across various medical specialties including dermatology, plastic surgery, podiatry, ENT, gynecology, neurosurgery, orthopedics, general/thoracic surgery, dental/oral surgery, and genitourinary surgery. Also indicated for ablative skin resurfacing when used with a scanning unit.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
K240497 — Smartxide Tetra Pro · El.En S.P.A. · Mar 12, 2024
K113504 — DEKA SMARTXIDE AND DELIVERY ACCESSORIES · El.En Electronic Engineering Spa · Feb 1, 2013
K180193 — DEKA SMARTXIDE Family (Smartxide Touch, Smartxide Punto) · El.En Electronic Engineering Spa · Feb 21, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 5, 2014
Electronic Engineering S.p.A. Mr. Paolo Peruzzi Regulatory Affairs Manager 50041 Calenzano (FI) Italy
Re: K133895 Trade/Device Name: DEKA SmartXide2 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology. Regulatory Class: Class II Product Code: GEX Dated: August 4, 2014 Received: August 6, 2014
Dear Mr. Peruzzi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K133895
Device Name DEKA Smartxide2
#### Indications for Use (Describe)
It is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynaecology, neurosurgery, orthopaedics, general and thorasic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.
### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Neil R Ogden -S 2014.09.04 14:57:31 -04'00
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
| Submitter: | El.En. S.p.A.<br>Via Baldanzese, 17<br>50041 Calenzano (FI), Italy |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Paolo Peruzzi<br>Regulatory Affairs Manager<br>Phone: +39-055-882-6807<br>Email: p.peruzzi@elen.it |
| Date Summary Prepared: | December 18, 2013 |
| Device Trade Name: | DEKA SmartXide² Laser System |
| Common Name: | Medical Laser System |
| Classification Name: | Instrument, Surgical, Powered, Laser<br>79-GEX<br>21 CFR 878.4810 |
| Equivalent Device:<br>Affirm | DEKA SmartXide² Laser System (K113504) and Cynosure<br>CO2 Laser (K081424) |
| Device Description: | DEKA SmartXide² is a laser with a wavelength of 10600 nm having a maximal power of 60 Watts. The laser energy is delivered to the treatment area via an articulated arm and a delivery accessory connected to its distal end.<br><br>Laser activation is by footswitch. Overall weight of the laser is 95 Kg, and the size is 210 x 59 x 56 cm (H x W x D).<br><br>Electrical requirement is 100-120 Vac 15A, 50-60 Hz, single phase |
| Intended Use: | The SmartXide² laser is indicated for incision, excision, vaporization and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynaecology, neurosurgery, orthopaedics, general and thorasic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing. |
| | Comparison: The SmartXide² laser has the same indications for use, the same principle of operation, the same wavelength and the same pulse energy range as the predicate devices. |
| Nonclinical Performance Data: | none |
| Clinical Performance Data: | none |
{4}------------------------------------------------
| Animal Study Data : | Animal study on sheep's skin was conducted to show the<br>comparative effects of Smartxide laser and Affinity lasers on<br>thermal lesion depth and width as well as ablation depth and<br>width. |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion:<br>intended | The SmartXide 2 laser is a safe and effective device for the |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.