WAVELIGHT FS200 LASER SYSTEM MODEL: FS200

K101006 · Alcon Research, Ltd. · GEX · Oct 21, 2010 · General, Plastic Surgery

Device Facts

Record IDK101006
Device NameWAVELIGHT FS200 LASER SYSTEM MODEL: FS200
ApplicantAlcon Research, Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateOct 21, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The WaveLight® FS200 Laser System is an ophthalmic surgical laser indicated for use: - In the creation of a corneal flap in patients undergoing LASIK surgery or other surgery or treatment requiring initial lamellar resection of the cornea. - In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments. - In the creation of a lamellar cut/resection of the cornea for lamellar keratoplasty. - In the creation of a penetrating cut/incision for penetrating keratoplasty and for corneal harvesting.

Device Story

Stationary scanning-spot femtosecond laser system; used in refractive surgery. Device creates corneal flaps, tunnels for ring segments, or lamellar/penetrating incisions. Operates via femtosecond laser pulses; higher repetition rate and shorter pulse duration than predicate lasers. Used in clinical settings by ophthalmic surgeons. Provides precise corneal tissue resection as an alternative to mechanical microkeratomes. Benefits include improved accuracy and quality of corneal cuts for various ophthalmic procedures.

Clinical Evidence

Bench testing only. Performance testing conducted per FDA guidance for keratome and replacement keratome blades to evaluate accuracy, precision, and quality of corneal cuts. Results demonstrated performance equivalent to predicate laser and microkeratome systems.

Technological Characteristics

Stationary scanning-spot femtosecond laser system. Energy source: femtosecond laser. Features higher pulse repetition rate and shorter pulse duration compared to predicate lasers. Utilizes common materials similar to predicate devices. Intended for ophthalmic surgical use.

Indications for Use

Indicated for patients undergoing LASIK, corneal ring segment placement, lamellar keratoplasty, or corneal harvesting requiring lamellar or penetrating corneal resection.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## WaveLight® FS200 Laser System OCT 2 1 2010 K10joo6 #### 5. 510(K) SUMMARY #### 510(k) Summary This summary document is being prepared in accordance with section 21 CFR 807.92(c). The submitter of the 510(k) is: Michael Buenger Associate Director, Regulatory Affairs Alcon Research, Ltd. 6201 South Freeway Fort Worth, TX 76134 Phone: (817) 551-6810 Fax: (817) 551-4630 Date Summary Revised: September 30, 2010 Device Subject to this 510(k): WaveLight® FS200 Laser System Trade Name: Common Name: Laser Powered Laser Surgical Instrument Classification Name: #### 1. Predicate Devices: The legally marketed device(s) to which we are claiming substantial equivalence are: | 510(k) Number | Device | |---------------|---------------------------| | K060372 | FS Laser System | | K032910 | Carriazo Pendular Keratom | | K073404 | iFS Laser System | #### 2. Device Description: The WaveLight® FS200 Laser System is a stationary scanning-spot femtosecond laser system used in refractive surgery. {1}------------------------------------------------ ### WaveLight® FS200 Laser System #### 3. Indications for Use: Indications for Use: The WaveLight® FS200 Laser System is an ophthalmic surgical laser indicated for use: - · In the creation of a corneal flap in patients undergoing LASIK surgery or other surgery or treatment requiring initial lamellar resection of the cornea. - · In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments. - · In the creation of a lamellar cut/resection of the cornea for lamellar keratoplasty. - · In the creation of a penetrating cut/incision for penetrating keratoplasty and for corneal harvesting. # 4. Statement of how the Technological Characteristics of the Device compare to those of the Predicate or legally marketed Device The technological characteristics of the WaveLight® FS200 Laser System are very similar to those of the predicate laser devices. The WaveLight® FS200 Laser System and the predicate laser devices are femtosecond lasers which work in the same mode at similar frequencies with similar spot sizes. The WaveLight® FS200 Laser System works at a slightly higher pulse repetition rate and shorter pulse duration. The WaveLight® FS200 Laser System and the predicate laser devices use similar common materials and the same energy source. Therefore, the technological characteristics of the WaveLight® FS200 Laser System are substantially equivalent those of the predicate laser devices cleared under K060372 and K073404. The technological characteristics of the WaveLight® FS200 Laser System and the Carriazo Pendular Keratome (K032910) are different in that the predicate device is a mechanical keratome. The Carriazo Pendular Keratome was used as a predicate device for the purpose of flap cutting comparison. The WaveLight® FS200 is considered as an alternative to mechanical microkeratomes. {2}------------------------------------------------ # WaveLight® FS200 Laser System ### 5. Brief Summary of Nonclinical Test and Results: The WaveLight® FS200 Laser System has undergone testing and is in compliance with the applicable safety standards. Performance tests according to the FDA Guidance "Keratome and Replacement Keratome Blades Premarket Notification [(510(k)] Submissions'' were conducted to collect data on the accuracy, precision and quality of the cuts achieved with the WaveLight® FS200 Laser System . The WaveLight® FS200 Laser System and the accessory Patient Interface were found to perform equivalently to the predicate laser and microkeratome and their patient interfaces for the creation of corneal cuts. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized symbol resembling three human figures connected at the arms, positioned to the right. Encircling this symbol is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Alcon Research, Ltd. c/o Mr. Michael Buenger Associate Director, Regulatory Affairs 6201 South Freeway Forth Worth, TX 76134 OCT 2 1 2010 Re: K101006 Trade/Device Name: WaveLight® FS200 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Powered Laser Surgical Instrument Regulatory Class: Class II Product Code: GEX, HNO Dated: October 1, 2010 Received: October 6, 2010 Dear Mr. Buenger: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Kasia Alexander Image /page/4/Picture/7 description: The image shows a handwritten word "for" in cursive, positioned to the left of three lines of text. The text on the right consists of the letters "M", "D", and "O", each followed by a colon. The text is vertically aligned, with the word "for" appearing to be an annotation or label associated with the text on the right. Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ KIDIOOG OCT 2-1-2010 WaveLight® FS200 Laser System ### 4. INDICATIONS FOR USE STATEMENT 510(k) Number: K101006 WaveLight® FS200 Laser System Device Name: Indications for Use: The WaveLight® FS200 Laser System is an ophthalmic surgical laser indicated for use: - · In the creation of a corneal flap in patients undergoing LASIK surgery or other surgery or treatment requiring initial lamellar resection of the cornea. - · In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments. - · In the creation of a lamellar cut/resection of the cornea for lamellar keratoplasty. - · In the creation of a penetrating cut/incision for penetrating keratoplasty and for corneal harvesting. X Prescription Use _ AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Dr. Callan Page 16 Of 1890 (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K101006 17
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...