K033172 · Candela Corp. · GEX · Oct 30, 2003 · General, Plastic Surgery
Device Facts
Record ID
K033172
Device Name
CANDELA GENTLEYAG FAMILY OF LASER SYSTEMS
Applicant
Candela Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 30, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The lasers are indicted on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions. such as but not limited to port wine stains, hemoangioma, warts, telangiectasia, rosacea, Venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.
Device Story
Flashlamp-excited Nd:YAG laser system; emits 1064 nm pulsed energy. Targets hair or lesions via selective photothermolysis; pulse duration controlled to match thermal relaxation time of target, minimizing collateral thermal damage. Components include console, control/display panel, lens-coupled optical fiber handpiece with aiming beam, footswitch/handswitch, remote interlock, and optional cooling device. Operated by clinicians in dermatology/surgery settings. Output used for photocoagulation, hemostasis, and tissue ablation. Benefits include permanent hair reduction, lesion clearance, and scar improvement.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics, operating principles, and mechanism of action compared to predicate devices.
Technological Characteristics
Flashlamp-excited Nd:YAG laser; 1064 nm wavelength. Includes console, control electronics, lens-coupled optical fiber handpiece, aiming beam, footswitch/handswitch, and optional cooling device. Complies with 21 CFR 1000.1040, UL 2601, and EN 60601-1-2.
Indications for Use
Indicated for hair removal (all Fitzpatrick skin types I-VI, including tanned), treatment of PFB, photocoagulation/hemostasis of vascular/pigmented lesions (e.g., port wine stains, hemangioma, warts, telangiectasia, rosacea, venous lake, leg/spider veins), soft tissue coagulation/hemostasis, treatment of benign pigmented lesions (e.g., lentigos, cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos), reduction of red pigmentation in hypertrophic/keloid scars, and treatment of wrinkles.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K033172 1 of 3
Image /page/0/Picture/1 description: The image shows the Candela company logo, which is a stylized letter C. The text "CANDELA" is printed to the right of the logo. The date "OCT 3 0 2003" is printed above the text "510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS".
This 510(k) Summary of Safety and Effectiveness for the Candela GentleYAG Family of Laser Systems is being submitted in accordance with the requirements of the SMDA 1990, 21 CFR 807.92 and follows the guidance concerning the organization and content of a 510(k) summary.
| I. General Information<br>Applicant: | Candela Corporation |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Address: | 530 Boston Post Road<br>Wayland, MA 01778-1886 |
| Contact Person: | Lorraine Calzetta Patrovic<br>Manager, Regulatory Affairs |
| Date Prepared: | August , 2003 |
| II. Names<br>Device Trade Name: | GentleYAG Family of Laser Systems |
| Common Name | Dermatology Laser |
| Classification: | Class II (21 CFR § 878.4810 Laser Surgical<br>Instrument for use in General and Plastic<br>Surgery and in Dermatology) |
## III. Predicate Devices
Candela GentleYAG a.k.a Long Pulse Nd:YAG Laser (K010104, K0022923. K022951, K023193), Lyra I (K020021,K003765, K003147) Altus CoolGlide Family of Lasers (K023954, K010834,K003765)
### IV. Product Description
The Candela GentleYAG Family of Laser Systems is a flashlamp-excited, Nd: YAG(Neodymium-doped Yttrium Aluminum Garnet) laser. Pulsed laser energy at a nominal wavelength of 1064 nanometers (nm) is used. This wavelength causes maximum energy absorption by the target (hair or lesion) and minimum absorption by surrounding skin structures. In addition, the laser pulse duration is controlled to be equal to or shorter than the thermal relaxation time of the target, to minimize heat transfer to surrounding tissues.
The GentleYAG consists of the following main components: a laser system console (including software and control electronics) a control and display panel a lens-coupled, user replacement optical fiber handpiece with aiming beam a footswitch or handswitch a remote interlock connector an optional cooling device
{1}------------------------------------------------
K033172 2 of 3
## V. Indications for Use
The GentleYAG Family of Laser Systems is intended for:
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The lasers are indicted on all skin types Fitzpatrick I-VI including tanned skin.
Photocoagulation and hemostasis of pigmented and vascular lesions. such as but not limited to port wine stains, hemoangioma, warts, telangiectasia, rosacea. Venus lake leg veins and spider veins
Coagulation and hemostasis of soft tissue.
Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles
### VI. Rationale for Substantial Equivalence
The Candela GentleYAG Family of Laser Systems shares the same indications for use, operating principles, mechanism of action, technological features as the predicate devices and therefore is substantially equivalent to the predicate devices.
## VII. Safety and Effectiveness Information
The Candela GentleYAG Laser Family of Laser Systems is equivalent to the Candela GentleYAG (aka Candela Long Pulse Nd: YAG Laser) cleared under K010104, K022923,K022951,K023193 the Lyra I ( K020021,K003765, K003147) and the Altus CoolGlide Family of Lasers (K023954, K010834,K003765) in technological features, and equivalent in intended use and therefore the risks and benefits are comparable to the predicate devices. Candela Certifies that this device is in compliance with the laser performance standards required as per 21 CFR 1000.1040. The device conforms to the UL 2601 Electrical Safety Standard and with the Harmonized Standard EN 60601-1-2, Part 2 established by the European Community. We therefore believe that there are no new questions of safety or effectiveness raised by the introduction of the Gentle Y AG Family of Laser Systems.
#### VIII. Conclusion
It is the opinion of Candela Corporation that the GentleYAG Family of Laser Systems is substantially equivalent to the predicate devices based on operating principles mechanism of action, and for intended use.
{2}------------------------------------------------
K03 3172 3 of
# Statement of Substantial Equivalence
The purpose of this 510(k) submission is to obtain clearance to market the Candela Gentle YAG Family of Laser Systems.
The Candela GentleYAG Family of Laser Systems presented in this submission, is a modification of the Candela GentleY AG Laser System a.k.a Candela Long Pulse Nd: YAG previously cleared under K010104, K022951, K022923, K023193.
The modifications incorporate features of the currently marketed predicate devices, including expanded pulse durations, pulse frequencies, spot sizes, energy and indications for use. The Table of Comparative Features (included as Exhibit D) shows the features of the devices, including their similarities.
As indicated in the table, the GentleYAG Famly of Laser Systems is equivalent to the currently marketed predicate devices, the Laserscope Lyra I Nd: Y AG Laser (cleared under K020021, K0037565, K 0031747), and the Altus Cool Glide Family of Lasers (cleared under K023954, K010834, K003765), based on operating principles, mechanism of action, and shares similar functional features such as wavelength, pulse duration, pulse frequency, energy, spot sizes, and indications for use. Therefore, we do not believe that introduction of the modified GentleY AG Family of Laser Systems raises any new issues of safety or effectiveness.
In conclusion, Candela believes that the Gentle YAG family of Laser Systems is substantially equivalent to the currently marketed predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the top half of the circle. Inside the circle is a stylized image of an eagle with its wings spread.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 3 0 2003
Ms. Lorraine Calzetta Patrovic Manager, Regulatory Affairs Candela Corporation 530 Boston Post Road Wayland, Massachusetts 01778-1886
Re: K033172
Trade/Device Name: Candela GentleYAG Family of Laser Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: September 26, 2003 Received: September 30, 2003
Dear Ms. Patrovic:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Lorraine Calzetta Patrovic
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Candela. The logo consists of a stylized flame on the left and the word "CANDELA" on the right. The flame is black and white, and the word "CANDELA" is in a simple, sans-serif font.
# INDICATIONS FOR USE STATEMENT 033172 510(k) Number :
Candela GentleYAG Family of Laser Systems Device Name:
Indications For Use:
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The lasers are indicted on all skin types Fitzpatrick I-VI including tanned skin.
Photocoagulation and hemostasis of pigmented and vascular lesions. such as but not limited to port wine stains, hemoangioma, warts, telangiectasia, rosacea, Venus lake, leg veins and spider veins.
Coagulation and hemostasis of soft tissue.
Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
iriam C. Provost
Division of General, Restorative and Neurological Devices
KO33172 510(k) Number Prescription Use Over-The-Counter Use OK (Per 21 CFR 801.109)
228