ArtiSential Bipolar Fenestrated Forceps

K190909 · Livsmed, Inc. · GEI · Feb 13, 2020 · General, Plastic Surgery

Device Facts

Record IDK190909
Device NameArtiSential Bipolar Fenestrated Forceps
ApplicantLivsmed, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateFeb 13, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ArtiSential Bipolar Fenestrated Forceps are invasive instruments which is used with electrosurgical generator. They are intended to be used during laparoscopic surgical procedures for incision/coagulation (to electrocauterize, grasp and dissect tissue) during surgery.

Device Story

Single-use, sterile, invasive laparoscopic instrument; used with electrosurgical generator to grasp, dissect, and coagulate tissue. Device features jaw, 8mm diameter shaft, and grip with control ring. Operator manipulates grip to open/close jaw, bend jaw (up/down/left/right ±80°), and rotate jaw 360°. Principle of operation: bipolar radiofrequency (RF) energy applied to electrode tip; bioimpedance generates heat to incise and coagulate tissue. Used by surgeons in laparoscopic procedures. Output is mechanical tissue manipulation and thermal coagulation. Benefits include precise tissue handling and electrosurgical control during minimally invasive surgery.

Clinical Evidence

Bench testing only. Evaluated for biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-18, 60601-2-2), and mechanical performance (tensile strength, dimensional accuracy, operational functionality). All tests passed.

Technological Characteristics

Bipolar electrosurgical instrument; 8mm shaft diameter; 411.4mm length. Materials: stainless steel (electrode), polyetherimide (insulation), glass fiber (shaft). Energy: RF. Articulation: ±80° pitch/yaw, 360° rotation. Sterilization: EO (ISO 11135). Connectivity: dual-pin connector for electrosurgical unit.

Indications for Use

Indicated for electrosurgical coagulation, dissection, and grasping of tissue during laparoscopic and general surgical procedures.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 13, 2020 Livsmed Inc. % Do Hyun Kim CEO BT Solutions, Inc. Unit 904, Eonju-ro 86-gil 5, Gangnam-gu Seoul, Korea 06210 Re: K190909 Trade/Device Name: ArtiSential Bipolar Fenestrated Forceps (Model ABF01-L, ABF01-F) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 21, 2020 Received: January 22, 2020 Dear Do Hyun Kim: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K190909 Device Name ArtiSential Bipolar Fenestrated Forceps (Model name: ABF01-L, ABF01-F) Indications for Use (Describe) Indications for Use include electrosurgical coagulation, dissection, and grasping of tissue during the performance of laparoscopic and general surgical procedures. X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 6. 510(k) Summary ### 1. General Information | Applicant/Submitter: | Livsmed Inc. | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address: | #304, D-dong, 700, Pangyo-ro, Bundang-gu, 13516<br>Seongnam-si, Gyeonggi-do, Republic of Korea<br>Tel) +82-70-4282-7652<br>Fax) +82-31-706-3211 | | Contact Person: | Do Hyun Kim, BT Solutions, Inc. | | Address: | Unit 904, Eonju-ro 86-gil 5, Gangnam-gu, 06210<br>Seoul, Republic of Korea<br>Tel) +82-2-538-9140<br>Fax) +82-2-539-9140<br>Email) smanager@btsolutions.co.kr | | Preparation Date: | 01-02-2020 | ### 2. Device Name and Code | Device Trade Name: | ArtiSential Bipolar Fenestrated Forceps<br>(Model name: ABF01-L, ABF01-F) | |----------------------|---------------------------------------------------------------------------| | Common Name: | Electrosurgical Instruments | | Classification Name: | Electrosurgical, cutting & coagulation & accessories | | Product Code: | GEI | | Regulation Number: | 21 CFR 878.4400 | | Classification: | Class II | | Review Panel: | General & Plastic Surgery | ### 3. Predicate Devices ArtiSential Bipolar Fenestrated Forceps is substantially equivalent to the following devices | Table 3.1 Predicate device | | | |----------------------------|-------------|---------------| | Applicant | Device Name | 510(k) Number | {4}------------------------------------------------ ### ArtiSential Bipolar Fenestrated Forceps 510(k) Summary | Gyrus Medical, Inc. | Everest Bipolar MACRO, | K031078 | |---------------------|----------------------------|---------| | | MICRO&MOLLY Forceps Gyrus | | | | Bipolar MACRO, MICRO&MOLLY | | | | Forceps | | #### Table 3.2 Reference Predicate device | Applicant | Device Name | 510(k) Number | |---------------------------|-------------------------------------------------------------------------------------------------|---------------| | Ethicon Endo-Surgery, LLC | ENSEAL® G2 Articulating Tissue<br>Sealers | K122797 | | Intuitive Surgical, Inc. | Intuitive Surgical Da Vinci Surgical | K050369 | | Intuitive Surgical, Inc. | da Vinci® SpTM Surgical System,<br>Model SP999, EndoWrist® SpTM<br>Instruments, and Accessories | K131962* | * This reference device is referred, of which articulating parts are the same to those of the ArtiSential Bipolar Fenestrated Forceps. This device is used in robotic surgery, but articulating function is very similar. #### 4. Device Description This product is sterile, single-use, invasive instrument that used in laparoscopic surgery. There are two models, one for ABF01-F and the other for lock functions, ABF01-L. The two models are same except for lock function. This product is a specific component, but not the entire electrosurgical device. The device is not intended to be marketed with multiple components, accessories, and as part of a system. #### 5. Indications / Intended Use #### 5.1 Intended use The ArtiSential Bipolar Fenestrated Forceps are invasive instruments which is used with electrosurgical generator. They are intended to be used during laparoscopic surgical procedures for incision/coagulation (to electrocauterize, grasp and dissect tissue) during surgery. #### 5.2 Indications for use Indications for use include electrosurgical coagulation, dissection, and grasping of tissue during the performance of laparoscopic and general surgical procedures. #### 6. Technical Characteristics in Comparison to Predicate and Reference Devices ArtiSential Bipolar Fenestrated Forceps is substantially equivalent to the following legally marketed predicate device. Table 6.1 Predicate Device {5}------------------------------------------------ ### ArtiSential Bipolar Fenestrated Forceps 510(k) Summary | | Proposed Device | Predicate Device | |---------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------| | 510(K) Number | Not Available | K031078 | | Manufacturer | LivsMed, Inc. | Gyrus Medical, Inc. | | Device Name | ArtiSential Bipolar Fenestrated Forceps | Everest Bipolar MACRO,<br>MICRO&MOLLY Forceps Gyrus Bipolar<br>MACRO, MICRO&MOLLY Forceps | | Clearance Date: | N/A | 05/16/2003 | | Classification /<br>Regulation | Class 2 / 878.4400 | Class 2 / 878.4400 | | Product Code | GEI | GEI | | Intended for | Prescription Use | Prescription Use | | Indications for<br>Use | Electrosurgical coagulation, dissection, and<br>grasping of tissue during the performance of<br>laparoscopic and general surgical<br>procedures. | Electrosurgical coagulation, dissection, and<br>grasping of tissue during the performance of<br>laparoscopic and general surgical<br>procedures. | | Principles of<br>operation | This product is a single-use instrument used<br>in electrosurgical units to hold soft tissues or<br>coagulate and make an incision (tissue<br>dissection) during general laparoscopic<br>surgery, which uses the principle of<br>applying high-frequency currents from the<br>electrode to the human body to generate<br>heat by bioimpedance when radio frequency<br>(RF) energy from the electrosurgical unit<br>applies an electric current to the electrode<br>part, and using the generated heat to incise<br>cellular tissues and cause coagulation.<br>It is composed of a jaw, Φ8 diameter shaft,<br>grip (including a control ring), and<br>electrosurgical unit connection electrode<br>connector.<br>During a procedure with this product, the<br>jaw opens if the control ring opens, and jaw<br>closes if the control ring closes. In addition,<br>the jaw is also bent up, down, left and right<br>within a range of ±80° or more by moving<br>the grip up, down, left and right, and the jaw<br>can also turn 360° when rotating the grip. | Unknown | | Energy Type | Radiofrequency | Radiofrequency | | Electrode type<br>(monopolar or<br>bipolar) | Bipolar | Bipolar | | Physical<br>dimensions and<br>design (size,<br>length,<br>connector type) | Shaft diameter: 8mm<br>Length: 411.4mm<br>Connector type: Dual-pin connector type | Shaft diameter: 5mm<br>Length: 450mm<br>Connector type: Dual-pin connector type | | Rated voltage | 200 Vp | Unknown | | Materials<br>(electrode) | Stainless steel | Unknown | | Materials<br>(insulation) | Polyetherimide | Unknown | | Materials<br>(Shaft) | Glass fiber | Unknown | | Rotation | Pitch:±80° or more, Yaw:±80° or more<br>and Open-Close | Unknown | | Sterilization | EO | EO | {6}------------------------------------------------ # ArtiSential Bipolar Fenestrated Forceps 510(k) Summary ### Table 6.2 Reference Predicate Device | | Proposed Device | Reference Predicate<br>Device 1 | Reference Predicate Device<br>2 | |----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(K)<br>Number | Not Available | K122797 | K050369 | | Manufacturer | LivsMed, Inc. | Ethicon Endo-Surgery, LLC | Intuitive Surgical, Inc. | | Device Name | ArtiSential Bipolar<br>Fenestrated Forceps | ENSEAL® G2 Articulating<br>Tissue Sealers | Intuitive Surgical Da Vinci<br>Surgical System, Model<br>IS2000 | | Clearance<br>Date: | N/A | 01/25/2013 | 04/29/2005 | | Classification /<br>Regulation | Class 2 / 878.4400 | Class 2 / 878.4400<br>Class 2/ 884.4120 | Class 2/ 876.1500 | | Product Code | GEI | GEI, HGI | NAY | | Intended for<br>Indications for<br>Use | Prescription Use<br>Electrosurgical coagulation,<br>dissection, and grasping of<br>tissue during the<br>performance of laparoscopic<br>and general surgical<br>procedures. | Prescription Use<br>The ENSEAL® G2<br>Articulating Tissue Sealers<br>are indicated for bipolar<br>coagulation and mechanical<br>transection of tissue during<br>laparoscopic and open<br>procedures. The devices are<br>bipolar electrosurgical<br>instruments for use with the<br>Generator GEN11<br>(GEN11). They are intended<br>for use during open or<br>laparoscopic, general and<br>gynecological surgery to cut<br>and seal<br>vessels, and to cut, grasp<br>and dissect<br>tissue during surgery.<br>Indications for use include<br>open and laparoscopic,<br>general and gynecological<br>surgical procedures<br>(including urologic,<br>thoracic, plastic and<br>reconstructive, bowel<br>resections, hysterectomies,<br>cholecystectomies, gall<br>bladder procedures, Nissen | Prescription Use<br>The Intuitive Surgical' da<br>Vinci Surgical System is<br>intended to assist in the<br>accurate control of Intuitive<br>Surgical Endoscopic<br>Instruments including rigid<br>endoscopes, blunt and sharp<br>endoscopic dissectors,<br>scissors, scalpels, ultrasonic<br>shears, forceps/pick-ups,<br>needle holders, endoscopic<br>retractors, stabilizers,<br>electrocautery and<br>accessories for endoscopic<br>manipulation of tissue,<br>including grasping, cutting,<br>blunt and sharp dissection,<br>approximation, ligation,<br>electrocautery, suturing, and<br>delivery and placement of<br>microwave ablation probes<br>and accessories during<br>urologic surgical procedures,<br>general laparoscopic surgical<br>procedures, gynecologic<br>laparoscopic surgical<br>procedures, general non-<br>cardiovascular thoracoscopic | | | | fundoplication, adhesiolysis,<br>oophorectomies, etc.), or<br>any procedure where vessel<br>ligation (cutting and<br>sealing), tissue grasping,<br>and dissection is performed.<br>The devices can be used on<br>vessels up to (and including)<br>7mm and tissue and/or<br>vascular bundles as large as<br>will fit in the jaws of the<br>instruments. | surgical procedures, and<br>thoracoscopically assisted<br>cardiotomy procedures. The<br>system can also be<br>employed, with adjunctive<br>mediastinotomy to perform<br>coronary anastomosis during<br>cardiac revascularization. It<br>is intended for use by trained<br>physicians in an operating<br>room environment in<br>accordance with the<br>representative specific<br>procedures set forth in the<br>Professional Instructions for<br>Use. | | Principles of<br>operation | This product is a single-use<br>instrument used in<br>electrosurgical units to hold<br>soft tissues or coagulate and<br>make an incision (tissue<br>dissection) during general<br>laparoscopic surgery, which<br>uses the principle of<br>applying high-frequency<br>currents from the electrode<br>to the human body to<br>generate heat by<br>bioimpedance when radio<br>frequency (RF) energy from<br>the electrosurgical unit<br>applies an electric current to<br>the electrode part, and using<br>the generated heat to incise<br>cellular tissues and cause<br>coagulation.<br>It is composed of a jaw, Φ8<br>diameter shaft, grip<br>(including a control ring),<br>and electrosurgical unit<br>connection electrode<br>connector.<br>During a procedure with<br>this product, the jaw opens<br>if the control ring opens,<br>and jaw closes if the control<br>ring closes. In addition, the<br>jaw is also bent up, down,<br>left and right within a range<br>of ±80° or more by moving<br>the grip up, down, left and<br>right, and the jaw can also | The ENSEAL G2<br>Articulating Tissue<br>Sealers have a 5 mm<br>diameter shaft<br>and are available in two<br>shafts<br>lengths: 35 and 45cm; and<br>two jaw types: 3mm curved<br>jaw and 5mm straight jaw.<br>The jaws are normally in the<br>opened position and can be<br>partially or fully closed by<br>squeezing<br>the closing handle. The jaws<br>have atraumnatic teeth for<br>grasping and holding<br>targeted tissue when<br>clamped. The handles of the<br>Subject ENSEAL G2<br>Articulating Tissue Sealers<br>have an ergonomic interface<br>for the user. Like the<br>Predicate, the Subject<br>devices include an<br>integrated energy activation<br>button. The devices use a<br>combination of the<br>application of bipolar<br>electrosurgical energy in<br>conjunction with the I-<br>BLADE knife, to compress,<br>coagulate, and transect<br>tissue. | This product is a reusable<br>foot control electro-surgical<br>electrode equipped with a<br>product mounted on a<br>cleared product like da Vinci<br>S Surgical system or da<br>Vinci Si Surgical System and<br>used by connecting it to an<br>electrosurgical machine with<br>a cable. It consists of a<br>housing, which is attached to<br>the robotic arm of the da<br>vinci system, and shaft<br>which the bipolar energy is<br>transferred and an electrode<br>(tip) that coagulates,<br>removes, dissects, or<br>destroys tissues using the<br>energy. The principles<br>applied to the development<br>of this product are as<br>follows.<br>When high-frequency energy<br>from an electrosurgical<br>machine is energized on the<br>electrodes through cables,<br>housings, and shafts, high-<br>frequency currents from<br>electrodes are applied to the<br>human body to generate heat<br>due to bio-resistance, and the<br>heat generated is used to<br>generate incision and<br>coagulation of tissue. This<br>product is a medical device<br>that can be reused by | {7}------------------------------------------------ # ArtiSential Bipolar Fenestrated Forceps 510(k) Summary {8}------------------------------------------------ | ArtiSential Bipolar Fenestrated Forceps | |-----------------------------------------| | 510(k) Summary | | | turn 360° when rotating the grip. | | cleaning and sterilizing after use. | |---------------------------------------------------------------------------|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------| | Energy Type | Radiofrequency | Radiofrequency | Radiofrequency | | Electrode type<br>(monopolar or<br>bipolar) | Bipolar | Bipolar | Bipolar | | Physical<br>dimensions and<br>design (size,<br>length,<br>connector type) | Shaft diameter: 8mm<br>Length: 411.4mm<br>Connector type: Dual-pin<br>connector type | Shaft diameter: 5mm<br>Length: 350mm, 450mm<br>Connector type: Unknown | Shaft diameter: 8mm<br>Total length: 511mm<br>Connector type: Dual-pin<br>connector type | | Rated voltage | 200 Vp | Unknown | 384 Vp | | Materials<br>(electrode) | Stainless steel | Stainless steel | Stainless steel | | Materials<br>(insulation) | Polyetherimide | Unknown | Polyetherimide / ETFE<br>(Ethylenetetrafluoroethylene) | | Materials<br>(Shaft) | Glass fiber | Unknown | Glass fiber | | Rotation | Pitch:±80° or more,<br>Yaw:±80° or more and<br>Open-Close | Articulates approximately<br>50~65° to<br>the left and the right of<br>center and<br>Open-Close | Pitch: ±70°<br>Yaw: ±90° | | Sterilization | EO | EO | Moist Heat or Steam<br>Sterilization | ### 7. Performance Data #### 7.1 Biocompatibility The device has been evaluated for its biological safety according to ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process". Following endpoints have been assessed during the evaluation: - Cytotoxicity - . Sensitization - Intracutaneous reactivity - Acute systemic toxicity #### 7.2 Electrical Safety The ArtiSential Bipolar Fenestrated Forceps has been tested according to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-18 and IEC 60601-2-2. The test setup included: - Active accessory insulation - Active accessory hf leakage ● - Active accessory hf dielectric strength ● - Active accessory mains frequency dielectric strength ● - Feedthrough test ● {9}------------------------------------------------ The device had passed all performed tests. 7.3 Sterilization ArtiSential Bipolar Fenestrated Forceps is provided sterile, intended to be single-use. This product is EO-Sterilization in accordance with ISO-11135. ### 7.4 Shelf life The proposed expiration date is 2 years from the manufacturing date. The real-time testing will be performed to confirm the shelf-life for 3 years. ### 7.5 Performance test | Test clause and Test requirement | Test specification | Results- Remarks | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | 1. Appearance<br>There should be no defects in the<br>appearance of the product and there<br>should be no problem in use. | Visual observation | No crack, stain or<br>no substances on<br>the surface of the<br>product | | 2. Dimension<br>It should be within ± 5% of the indicated<br>value of the dimensional term. | Measure by Vernier calipers<br>and dial gauge etc. | Pass<br>Refer to<br>[Measurement<br>data] on 65 page<br>at attachment #3 | | 3. Operational test<br>The jaw must be smoothly opened and<br>closed and free from jamming, the jaw<br>and hub can be bent up, down, left, and<br>right a range of above ±80° and are<br>capable of 360° rotation. | Adjusting open and close the<br>jaw by manipulating the grip<br>and control ring, and measure<br>the angle at bending and<br>turning. | The jaw and hub<br>are bent up, down,<br>left and right<br>within above 80°,<br>and can rotate<br>360°. | | 4. Tensile strength<br>The jaw and shaft connections shall not<br>be damaged from pulling of 20 N. | Hold the jaw and shaft<br>connections respectively and<br>apply a force of 20 N using<br>Push pull gauge. | No damage to the<br>connection when<br>applying a force of<br>20N | | 5. Feedthrough test<br>Electricity should be transmitted<br>between the electrode tip and the<br>connector. | Electrical conduction between<br>the electrode tip and the<br>connector is tested using a DMM<br>(digital multi meter). | The<br>resistance<br>value between the<br>electrode tip and<br>the connector is<br>less than 1Ω | The device had passed all performed tests. {10}------------------------------------------------ Based on these performance characteristics, the results demonstrate that the performance requirements were met, the device performs as intended and that the subject device has substantially equivalent performance characteristics to the predicate and reference predicate devices. #### 8. Substantial Equivalence The ArtiSential Bipolar Fenestrated Forceps indication for use is similar to the predicate device (K031078). The energy type, electrode type, sterilization as well as similar physical characteristics are the same. Although there are some minor differences with each product, these differences between the ArtiSential Bipolar Fenestrated Forceps and the predicate device do not raise new or different questions of safety and efficacy. There is no new technology and no difference that would raise new or different questions of safety or efficacy. #### 9. Conclusions In conclusion, the comparison carried out covers all products, sizes, and the entire intended purpose of the device under evaluation. The subject device which is the Disposable Laparoscopic Instruments - ArtiSential Bipolar Fenestrated Forceps - is considerably similar to the predicate and reference devices in principles of operation, technological characteristics, as well as performance characteristics. The testing was conducted to evaluate the performance of subject device in comparison to the predicate and reference devices. Results of validation and verification activities in design control that include testing/certification to designated standards and performance testing of the devices has demonstrated substantial equivalence of the subject device to the predicate in terms of safety and effectiveness for requested intended use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...