PERCUVANCE PERCUTANEOUS SURGICAL SYSTEM

K153063 · Teleflexmedical, Inc. · GEI · Apr 8, 2016 · General, Plastic Surgery

Device Facts

Record IDK153063
Device NamePERCUVANCE PERCUTANEOUS SURGICAL SYSTEM
ApplicantTeleflexmedical, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateApr 8, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery.

Device Story

Micro-laparoscopic platform for hospital use; comprises reusable handles (ratcheted/non-ratcheted) and interchangeable shafts/tool tips. Surgeon manually operates device to penetrate soft tissue; exchanges tool tips (scissors, graspers, dissectors, cautery, clip applier) extracorporeally through a trocar. Seal bridges maintain insufflation during exchange. Cautery tips apply monopolar RF energy for cutting/coagulation; clip applier delivers Hem-o-lok ligating clips. Provides surgeons with multifunctional instrumentation via smaller incisions; enhances surgical precision and tissue management.

Clinical Evidence

Bench testing only. Included biocompatibility (ISO 10993-1), functional/force/dimensional verification, and electrosurgical safety (AAMI/ANSI ES60601-1, IEC 60601-2-2). Side-by-side bench testing compared thermal effects and clip performance against predicates. Porcine model usability study confirmed equivalent performance in trocar compatibility, clip security, and tissue manipulation.

Technological Characteristics

Active accessory (1000V peak). Materials: Stainless steel, PEEK, PET, PC, Nylon, Silicone, Cyrolite. Components: Reusable handles, single-patient-use shafts/tips. Connectivity: Standalone mechanical/electrosurgical. Sterilization: Gamma (single-use), Steam (reusable handles). Standards: IEC 60601-1:2005, IEC 60601-2-2, ISO 10993-1, ISO 8600 series.

Indications for Use

Indicated for patients undergoing laparoscopic surgery requiring soft tissue penetration, grasping, manipulation, cutting, cauterization, or delivery of Hem-o-lok ligating clips. Contraindicated for contraceptive coagulation of fallopian tissue and use when endoscopic techniques are generally contraindicated.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 8, 2016 Teleflex Medical, Inc. Ms. Holly Hallock Senior Regulatory Affairs Specialist 3015 Carrington Mill Blvd Morrisville, North Carolina 27560 Re: K153063 Trade/Device Name: Percuvance Percutaneous Surgical System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI, GCJ, GDO Dated: March 4, 2016 Received: March 7, 2016 Dear Ms. Hallock: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Jennifer R. Stevenson -A For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K153063 Device Name Percuvance(TM) Percutaneous Surgical System Indications for Use (Describe) The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery. Type of Use (Select one or both, as applicable) | <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | |-------------------------------------------------------------|------------------------------------------------------------| |-------------------------------------------------------------|------------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) SUMMARY - K153063 # Percuvance™ Percutaneous Surgical System # A. Name, Address, Phone, and Fax Number of Applicant Teleflex Medical, Incorporated 3015 Carrington Mill Boulevard Morrisville, NC 27560 USA 919-433-4918 Phone: Fax: 919-433-4996 #### B. Contact Person Holly Hallock Senior Regulatory Affairs Specialist # C. Date Prepared April 7th, 2016 ## D. Device Name Trade Name Percuvance™ Percutaneous Surgical System #### Common Name | Primary: | Electrosurgical, Cutting and Coagulation and Accessories | |----------|----------------------------------------------------------| Secondary: Laparoscope, General and Plastic Surgery Applier, Surgical, Clip #### Classification Regulation | Primary: | 21 CFR 878.4400 | |----------|-----------------| |----------|-----------------| | Secondary: | 21 CFR 876.1500 | |------------|-----------------| | | 21 CFR 878.4800 | #### Product Code Primary: GEI {4}------------------------------------------------ Secondary: GCJ GDO #### Classification Name | Primary: | Electrosurgical Cutting and Coagulation Device and Accessories | |-----------------|-------------------------------------------------------------------------| | Secondary: | Endoscope and Accessories<br>Manual Surgical Instrument for General Use | | Classification: | Class II | Panel General & Plastic Surgery #### E. Device Description Teleflex Medical's next generation Percuvance™ Percutaneous Surgical System is a micro-laparoscopic platform that comprises fourteen (14) unique components, which are combined in various configurations to create a multifunctional set of instruments for laparoscopic procedures to be performed in a hospital setting. In accordance with IEC 60601-1:2005 (A1:2012), this next generation Percuvance™ system is classified as an active accessory with a rated accessory voltage of 1000V peak- System components include two reusable Handles (Ratcheted and Non-Ratcheted), which are manipulated by the surgeon and connect to a Shaft. The Shaft, which is available in two lengths (29 cm and 36 cm), affords various Tool Tips (or End Effectors) to be attached in order to perform basic surgical manipulation. The next generation Percuvance™ Handles, Shafts, and Tool Tips are not compatible or interchangeable with components from other percutaneous systems, including those of the existing Percuvance™ Percutaneous Surgical System, which was cleared under 510(k) K143299. Initial access to the surgical site is achieved with the Introducer Tool Tip attached to the Shaft. Once inside the patient, the Introducer Tool Tip is extracorporealized through a pre-inserted, central trocar and is then exchanged for one of the other Tool Tips, which include Scissors. Gripper Grasper, Johans Grasper, Maryland Dissector, Hook Cautery, Spatula Cautery and Clip Applier. No ligating clips are provided with the Percuvance™ Percutaneous Surgical System; however, the Clip Applier Tool Tip is compatible with Teleflex Medical's M/L Hem-o-lok® ligating clip (SKU 544230). {5}------------------------------------------------ Finally, the Seal Bridge, which is available in two sizes (5 mm and 12 mm), is used to protect the trocar seal and to aid in maintaining insufflation when it used in conjunction with a trocar to facilitate the extracorporeal exchange of Tool Tips. | Model Number | Component<br>Description | Reusability<br>Status | Sterility Status | Intended to<br>Apply<br>Monopolar<br>Energy? | |--------------|---------------------------------------|------------------------|---------------------------------------------------------------------------------|------------------------------------------------------------------------| | PCVINT3 | Introducer Tool Tip | Single-<br>Patient-Use | Provided Sterile -<br>Gamma Irradiated | No | | PCVSC5 | Scissors Tool Tip | Single-<br>Patient-Use | Provided Sterile -<br>Gamma Irradiated | Yes | | PCVGG5 | Traumatic Gripper<br>Grasper Tool Tip | Single-<br>Patient-Use | Provided Sterile -<br>Gamma Irradiated | Yes | | PCVJG5 | Atraumatic Johans<br>Grasper Tool Tip | Single-<br>Patient-Use | Provided Sterile -<br>Gamma Irradiated | Yes | | PCVMD5 | Maryland Dissector<br>Tool Tip | Single-<br>Patient-Use | Provided Sterile -<br>Gamma Irradiated | Yes | | PCVHK5 | Hook Cautery Tool<br>Tip | Single-<br>Patient-Use | Provided Sterile -<br>Gamma Irradiated | Yes | | PCVSPT5 | Spatula Cautery<br>Tool Tip | Single-<br>Patient-Use | Provided Sterile -<br>Gamma Irradiated | Yes | | PCVHCA5 | Clip Applier Tool<br>Tip | Single-<br>Patient-Use | Provided Sterile -<br>Gamma Irradiated | No | | PCVSH3 | 29 cm Shaft | Single-<br>Patient-Use | Provided Sterile -<br>Gamma Irradiated | Yes | | PCVSHL3 | 36 cm Shaft | Single-<br>Patient-Use | Provided Sterile -<br>Gamma Irradiated | Yes | | PCVNRH | Non-Ratcheted<br>Handle | Reusable | Provided Non-<br>Sterile - Intended<br>to be Steam<br>Sterilized by<br>Hospital | Yes | | PCVRH | Ratcheted Handle | Reusable | Provided Non-<br>Sterile - Intended<br>to be Steam<br>Sterilized by<br>Hospital | Yes | | PCVSB5 | 5 mm Seal Bridge | Single-<br>Patient-Use | Provided Sterile -<br>Gamma Irradiated | N/A – Not<br>connected to<br>the Handle and<br>Shaft | | | | | | configuration | | PCVSB12 | 12 mm Seal Bridge | Single-<br>Patient-Use | Provided Sterile –<br>Gamma Irradiated | N/A – Not<br>connected to<br>the Handle and<br>Shaft<br>configuration | All model numbers and associated descriptions of components included in the next generation Percuvance™ Percutaneous Surgical System are identified below. {6}------------------------------------------------ #### F. Indications for Use The following are indications for use of the next generation Percuvance™ Percutaneous Surgical System: > The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-olok ligating clips to soft tissue during laparoscopic surgery. These indications for use were primarily generated from identical indications for use identified from the predicate Percuvance™ Percutaneous Surgical System (K143299): > The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate soft tissue to access certain areas of the human abdomen and used to grasp, hold and manipulate tissue during laparoscopic surgery. The only differences in these indications for use statements are that the next generation Percuvance™ Percutaneous Surgical System introduces the abilities to "cut, cauterize" and "deliver Hem-o-lok® ligating clips," which were generated from the predicate MiniLap® MiniPolar Electrocautery Instruments and the reference Hem-o-lok® Ligating Clip System, respectively. The following are indications for use of the predicate MiniLap® MiniPolar Electrocautery Instruments (K083754): The MiniPolar Instruments are used to cut and cauterize soft tissue. The following are indications for use of the reference Hem-o-lok® Ligating Clip System: > Weck Hem-o-lok® Endoscopic Ligating Clip Appliers are indicated for use as delivery devices for Hem-o-lok® Ligating Clips. {7}------------------------------------------------ These additions to the next generation Percuvance™ Percutaneous Surgical System's indications for use are not critical, as they were generated from predicate and reference devices that have already been cleared by the FDA. ### G. Contraindications The following are contraindications of the next generation Percuvance™ Percutaneous Surgical System: > The monopolar active Tool Tips are not intended for contraceptive coagulation of fallopian tissue, but may be used to achieve hemostasis following transection of the fallopian tube. Do not resterilize single patient use components. These components are provided sterile and are intended for use in a single procedure. Discard after use. The device is not intended for use when endoscopic techniques are generally contraindicated. The following are contraindications of the M/L Hem-o-lok® ligating clip (SKU 544230), which is an accessory to the Percuvance™M system: > Hem-o-lok® ligating clips are not intended for use as a fallopian contraceptive tubal occlusion device. Hem-o-lok® ligating clips are contraindicated for use in ligating the renal artery during laparoscopic donor nephrectomies. # H. Substantial Equivalence Teleflex Medical's next generation Percuvance™ Percutaneous Surgical System is substantially equivalent to the following predicate systems. | Predicate Device | Manufacturer | 510(k) No. | Date Cleared | |--------------------------------------------------|---------------------------|------------|-------------------| | Percuvance™ Percutaneous<br>Surgical System | Teleflex Medical,<br>Inc. | K143299 | January 21, 2015 | | MiniLap® MiniPolar<br>Electrocautery Instruments | Teleflex Medical,<br>Inc. | K083754 | November 20, 2009 | Additionally, the technology of Teleflex Medical's Percuvance™ Clip Applier Tool Tip is substantially equivalent to that of the M/L Hem-o-lok® manual, endoscopic clip appliers cleared in 510(k) K133202. Like the M/L Hem-o-lok® manual, endoscopic clip appliers, the Percuvance™ Clip Applier Tool Tip is compatible with Teleflex Medical's M/L Hem-o-lok® ligating clip (SKU 544230), which was also cleared in 510(k) K133202. {8}------------------------------------------------ | Reference (Accessory) Device | Manufacturer | 510(k) No. | Date Cleared | |------------------------------------|------------------------|------------|------------------| | Hem-o-lok® Ligating Clip<br>System | Teleflex Medical, Inc. | K133202 | December 3, 2013 | #### I. Comparison to Predicate Devices Teleflex Medical's next generation Percuvance™ Percutaneous Surgical System is substantially equivalent to the predicate systems with regards to technology, intended use, indications for use and functional characteristics. One predicate system is Teleflex Medical's existing Percuvance™ Percutaneous Surgical System, which is legally marketed under 510(k) K143299. Like the next generation Percuvance™ system, the existing Percuvance™ system is a microlaparoscopic system comprised of a multifunctional set of instruments that reduces the need for ports during laparoscopic procedures. Both systems are indicated for the penetration of soft tissue to access certain areas of the abdomen and used to grasp, hold and manipulate tissue. Like this predicate system, the next generation Percuvance™ system utilizes an Introducer Tool Tip, Shaft and Handle to percutaneously introduce the system into the patient's surgical site. After introduction, both systems are capable of exchanging the Introducer Tool Tip for a variety of other Tool Tips in order to perform the surgical steps of the laparoscopic procedure. Each Percuvance™ system also includes a Seal Bridge, which is used to protect the trocar seal and to aid in maintaining insufflation when used in conjunction with a trocar to facilitate the extracorporeal exchange of Tool Tips. The next generation Percuvance™ Percutaneous Surgical System combines the features of the existing Percuvance™ System with the electrosurgical features of Teleflex Medical's MiniLap® MiniPolar Electrocautery Instruments, which are legally marketed under 510(k) K083754. The MiniLap® system, which is also a predicate device, is a percutaneous system that allows surgeons to apply RF energy through a variety of monopolar instruments. Like the predicate MiniLap® system, the next generation Percuvance™ system includes a Hook Cautery Tool Tip and a Spatula Cautery Tool Tip for the application of monopolar RF energy. A manual Clip Applier Tool Tip has also been added to the next generation Percuvance™ system to afford users the ability to ligate. While ligation clips are not provided with the next generation Percuvance™ system, the manual Clip Applier Tool Tip is compatible with M/L Hem-o-lok® ligating clip (SKU 544230), currently marketed by Teleflex Medical under 510(k) K133202. The similarities and differences of the next generation Percuvance™ system compared to the predicate devices are presented in the following table. {9}------------------------------------------------ | Comparative<br>Characteristics | Next<br>Generation<br>Percuvance™<br>System | Percuvance™M<br>System | MiniLap®<br>MiniPolar<br>Electrocautery<br>Instruments | Hem-o-lok®<br>Ligating<br>Clip System | Explanation<br>of<br>Differences | |--------------------------------|------------------------------------------------|------------------------------------|---------------------------------------------------------------------------------------------------------------|---------------------------------------|-----------------------------------------| | | K153063 | Predicate -<br>K143299 | Predicate -<br>K083754 | Reference -<br>K133202 | | | Handle | Yes | Yes | Yes, thumb grip<br>Handle pre-<br>assembled to<br>Shaft and<br>Instrument Tip<br>as one finished<br>component | N/A | Equivalent to<br>K143299 and<br>K083754 | | Shaft | Yes (2.9mm) | Yes (3mm) | Yes (2.4mm) | N/A | Equivalent to<br>K143299 and<br>K083754 | | Introducer Tip | Yes | Yes | Yes, pre-<br>assembled to<br>Shaft and<br>Instrument Tip<br>as one finished<br>component | N/A | Equivalent to<br>K143299 and<br>K083754 | | Tool Tips | Yes (5mm) | Yes (5mm) | Yes, but not<br>exchangeable;<br>Tips pre-<br>assembled to<br>Shaft | N/A | Equivalent to<br>K143299 and<br>K083754 | | Tool Tip<br>Attachment | Snap<br>connection to<br>Shaft over<br>detents | Threaded to<br>Shaft | Pre-assembled | N/A | Equivalent to<br>K143299 | | Graspers | Yes | Yes | No | N/A | Equivalent to<br>K143299 | | Maryland<br>Dissectors | Yes | Yes | No | N/A | Equivalent to<br>K143299 | | Scissors | Yes | Yes | No | N/A | Equivalent to<br>K143299 | | Cautery Tips | Yes (Spatula<br>and Hook) | No | Yes (Spatula,<br>Hook, and<br>Conical) | N/A | Equivalent to<br>K083754 | | Clip Applier | Yes | No | No | Yes | Equivalent to<br>K133202 | | Seal Bridges | Yes | Yes | No | N/A | Equivalent to<br>K143299 | | Energy | Monopolar | No | Monopolar | N/A | Equivalent to<br>K083754 | | Exchange<br>Technology | Exchanged<br>extra-<br>corporeally | Exchanged<br>extra-<br>corporeally | Instrument Tips<br>are pre-<br>assembled to<br>Shaft and not<br>exchangeable | N/A | Equivalent to<br>K143299 | {10}------------------------------------------------ # J. Materials Patient contacting materials of the next generation Percuvance™ Percutaneous Surgical System have been evaluated in accordance with ISO 10993-1:2009, FDA Bluebook Memorandum G95-1 and FDA Draft Guidance: Use of International Standard ISO 10993. | Model<br>Number | Component<br>Description | Materials | Type of Contact | |-----------------|-----------------------------------|---------------------------------------------------|------------------------------------------------------------------------------| | PCVINT3 | Introducer Tool<br>Tip | Stainless Steel | External Communicating Device<br>with Tissue Contact for Limited<br>Duration | | PCVSC5 | Scissors Tool<br>Tip | Stainless Steel<br>PEEK<br>White Ink<br>Lubricant | External Communicating Device<br>with Tissue Contact for Limited<br>Duration | | PCVGG5 | Gripper Grasper<br>Tool Tip | Stainless Steel<br>PEEK<br>White Ink | External Communicating Device<br>with Tissue Contact for Limited<br>Duration | | PCVJG5 | Johans Grasper<br>Tool Tip | Stainless Steel<br>PEEK<br>White Ink | External Communicating Device<br>with Tissue Contact for Limited<br>Duration | | PCVMD5 | Maryland<br>Dissector Tool<br>Tip | Stainless Steel<br>PEEK<br>White Ink | External Communicating Device<br>with Tissue Contact for Limited<br>Duration | | PVCHK5 | Hook Cautery<br>Tool Tip | Stainless Steel<br>PEEK<br>White Ink | External Communicating Device<br>with Tissue Contact for Limited<br>Duration | | PCVSPT5 | Spatula Cautery<br>Tool Tip | Stainless Steel<br>PEEK<br>White Ink | External Communicating Device<br>with Tissue Contact for Limited<br>Duration | | PCVHCA5 | Clip Applier<br>Tool Tip | Stainless Steel<br>PEEK<br>White Ink | External Communicating Device<br>with Tissue Contact for Limited<br>Duration | | PCVSH3 | 29cm Shaft | Stainless Steel<br>PEEK<br>PET<br>PC | External Communicating Device<br>with Tissue Contact for Limited<br>Duration | {11}------------------------------------------------ | | | Green Ink | | |---------|-------------------------|---------------------------------------------------------|------------------------------------------------------------------------------| | PCVSHL3 | 36cm Shaft | Stainless Steel<br>PEEK<br>PET<br>PC<br>Green Ink | External Communicating Device<br>with Tissue Contact for Limited<br>Duration | | PCVNRH | Non-Ratcheted<br>Handle | Stainless Steel<br>PEEK<br>Nylon<br>Medical Grade Epoxy | Surface Device with Skin Contact<br>for Limited Duration | | PCVRH | Ratcheted<br>Handle | Stainless Steel<br>PEEK<br>Nylon<br>Medical Grade Epoxy | Surface Device with Skin Contact<br>for Limited Duration | | PCVSB5 | 5mm Seal<br>Bridge | Silicone<br>Cyrolite | External Communicating Device<br>with Tissue Contact for Limited<br>Duration | | PCVSB12 | 12mm Seal<br>Bridge | Silicone<br>Cyrolite | External Communicating Device<br>with Tissue Contact for Limited<br>Duration | # K. Technological Characteristics A comparison of the technological characteristics of Teleflex Medical's next generation Percuvance™ Percutaneous Surgical System and the predicate systems has been performed. The results of this comparison demonstrate that the proposed system utilizes substantially equivalent technology as the predicate systems. Both predicate systems and the next generation Percuvance™ Percutaneous Surgical System offer alternatives to traditional laparoscopic procedures, resulting in smaller incision sites to the abdomen. Like the existing Percuvance™ predicate system (K143299), the next generation Percuvance™ Percutaneous Surgical System includes a reusable Handle that connectors to a variety of interchangeable instrument tips, including Graspers, Scissors and Dissectors, which are used for the manipulation of soft tissue. Like the MiniLap® MiniPolar predicate system (K083754), the next generation Percuvance™ Percutaneous Surgical System includes Hook Cautery and Spatula Cautery instrument tips to apply monopolar RF energy. The next generation Percuvance™ Percutaneous Surgical System also includes a Clip Applier Tool Tip, which utilizes technology equivalent to that of the M/L Hem-olok® manual, endoscopic clip applier cleared in 510(k) K133202. Like the M/L Hemo-lok® manual, endoscopic clip appliers, the Percuvance™ Clip Applier Tool Tip is {12}------------------------------------------------ compatible with Teleflex Medical's M/L Hem-o-lok® ligating clip (SKU 544230). which was also cleared in 510(k) K133202. # L. Performance Data Comprehensive bench testing has been successfully completed on Teleflex's Medical next generation Percuvance™ Percutaneous Surgical System. Similar to the predicate devices, the next generation Percuvance™ System was subjected to a variety of tests, including biocompatibility evaluation, functional verification, dimensional verification and force verification. Resulting data found that Teleflex Medical's next generation Percuvance™ Percutaneous Surgical System performed equivalently or better than the predicate systems. Like the predicate Percuvance™ Percutaneous Surgical System (K143299), Teleflex Medical's next generation Percuvance™ Percutaneous Surgical System was evaluated for biocompatibility according to ISO 10993-1:2009, FDA Bluebook Memorandum G95-1 and FDA Draft Guidance: Use of International Standard ISO 10993. The following tests were performed on both the predicate Percuvance™ Percutaneous Surgical System (K143299) and the next generation Percuvance™ Percutaneous Surgical System. All tests were successful. | Model<br>Number | Component<br>Description | Type of Contact | Tests Conducted | Results | |-----------------|-----------------------------------|---------------------------------------------------------------------------------|----------------------------------------------------------------------|---------| | PCVINT3 | Introducer Tool<br>Tip | External<br>Communicating Device<br>with Tissue Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization<br>Systemic Toxicity | Pass | | PCVSC5 | Scissors Tool<br>Tip | External<br>Communicating Device<br>with Tissue Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization<br>Systemic Toxicity | Pass | | PCVGG5 | Gripper Grasper<br>Tool Tip | External<br>Communicating Device<br>with Tissue Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization<br>Systemic Toxicity | Pass | | PCVJG5 | Johans Grasper<br>Tool Tip | External<br>Communicating Device<br>with Tissue Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization<br>Systemic Toxicity | Pass | | PCVMD5 | Maryland<br>Dissector Tool<br>Tip | External<br>Communicating Device<br>with Tissue Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization<br>Systemic Toxicity | Pass | | PVCHK5 | Hook Cautery<br>Tool Tip | External<br>Communicating Device<br>with Tissue Contact for | Cytotoxicity<br>Intracutaneous<br>Sensitization | Pass | {13}------------------------------------------------ | | | Limited Duration | Systemic Toxicity | | |---------|-----------------------------|---------------------------------------------------------------------------------|----------------------------------------------------------------------|------| | PCVSPT5 | Spatula Cautery<br>Tool Tip | External<br>Communicating Device<br>with Tissue Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization<br>Systemic Toxicity | Pass | | PCVHCA5 | Clip Applier<br>Tool Tip | External<br>Communicating Device<br>with Tissue Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization<br>Systemic Toxicity | Pass | | PCVSH3 | 29cm Shaft | External<br>Communicating Device<br>with Tissue Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization<br>Systemic Toxicity | Pass | | PCVSHL3 | 36cm Shaft | External<br>Communicating Device<br>with Tissue Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization<br>Systemic Toxicity | Pass | | PCVNRH | Non-Ratcheted<br>Handle | Surface Device with<br>Skin Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization | Pass | | PCVRH | Ratcheted<br>Handle | Surface Device with<br>Skin Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization | Pass | | PCVSB5 | 5mm Seal<br>Bridge | External<br>Communicating Device<br>with Tissue Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization<br>Systemic Toxicity | Pass | | PCVSB12 | 12mm Seal<br>Bridge | External<br>Communicating Device<br>with Tissue Contact for<br>Limited Duration | Cytotoxicity<br>Intracutaneous<br>Sensitization<br>Systemic Toxicity | Pass | Design verification testing for the next generation Percuvance™ Percutaneous Surgical System consisted of functional testing on the Handle, Shaft, Tool Tips and Seal Bridges. The Handles were exposed to repeated actuations, as well as durability studies. The Shafts underwent durability and strength testing, while the Tool Tips were evaluated for grip strength and tissue retention. The Clip Applier Tool Tip was tested for compatibility with the accessory ligating clip, and the Cautery Tool Tips were exposed to a variety of electrosurgical performance tests. Additionally, the Seal Bridges were exposed to insufflation leak prevention testing, and the entire system was evaluated for compatibility. All design verification testing results were acceptable. {14}------------------------------------------------ Additional testing was conducted to compare the next generation Percuvance™ Percutaneous Surgical System's functionality to that of the predicate devices. Sideby-side bench testing compared the Clip Applier Tool Tip (SKU PCVHCA5) to the existing Hem-o-lok® Manual Endoscopic Clip Applier (SKU 544965, K133202). Tests evaluated trocar compatibility, clip formation and clip retention. The results of these tests demonstrate that trocar compatibility is equivalent and that there is not an increased incidence in poorly formed clips or poor clip retention when compared to the reference Clip Appliers. Side-by-side testing was also conducted to compare the thermal effects of the next generation Percuvance™ Percutaneous Surgical System to that of the predicate devices in accordance with FDA Draft Guidance: Submissions for Electrosurgical Devices for General Surgery. This testing compared the thermal effects of the Maryland Dissector Tool Tip (SKU PCVMD5) to the thermal effects of the existing Pilling® Marvland Dissector (SKU 728014. K040855). Additionally, the Spatula Cautery Tool Tip (SKU PCVSPT5) and the Hook Cautery Tool Tip (SKU PCVHK5) were evaluated against the predicate MiniLap® MiniPolar Straight Spatula (SKU ECMS300, K083754) and L-Hook Probe (SKU ECMH300, K083754), respectively. As part of this testing, thermal effects were evaluated in three ex vivo soft tissue samples (chicken breast, beef liver and beef kidney) at minimum, default and maximum intensity settings for each type of activation (cut and coagulation). Thermal effects were measured and the dimensions (length, width and depth) of each thermal effect were compared to evaluate the thermal effects of each Tool Tip. Results of this testing concluded that the thermal effects of the next generation Percuvance™ Percutaneous Surgical System are equivalent to, or better than, the thermal effects of existing electrosurgical devices. Design and usability validation testing of the next generation Percuvance™ Percutaneous Surgical System consisted of surgeons assembling the system, performing laparoscopic procedure steps in a porcine model and disassembling the system. The surgeons were guided without interference by a study director. Surgeons then completed a device usability questionnaire to evaluate the performance of the system. All design and usability validation testing results were acceptable. During the porcine usability study, the surgeons also used both the Clip Applier Tool Tip (SKU PCVHCA5) and the existing Hem-o-lok® Manual Endoscopic Clip Applier (SKU 544965, K133202) to apply ligation in the porcine model. After ligation was applied, surgeons used a quantifiable score system to compare the compatibility with the trocar, clip security in the instrument during deployment, clip security in the instrument during tissue manipulation. feedback on closure, and lack of trauma during tissue manipulation and deployment. Results of this comparison conclude that the Clip Applier Tool Tip performs as well as the reference Hem-olok® Manual Endoscopic Clip Applier. Like the predicate MiniLap® MiniPolar Electrocautery Instruments (K083754), the next generation Percuvance™ Percutaneous Surgical System is classified as an active {15}------------------------------------------------ accessory, in accordance with AAMI/ANSI ES60601-1-:2005 (A1:2012). Based on this classification, the Percuvance™ System has been evaluated by a third party test house for electrosurgical safety and usability in accordance with the following applicable collateral and particular standards. All electrical safety and usability testing results were acceptable. - AAMI/ANSI ES60601-1:2005 (A1:2012) Medical Electrical Equipment Part 1: General Requirements For Basic Safety And Essential Performance (200 edition and 3.1 edition) - AAMI/ANSI/IEC 60601-1-2:2014 Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety And Essential Performance --Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests - IEC 60601-1-6:2010 (A1:2013) Medical Electrical Equipment -- Part 1-6: General Requirements For Basic Safety And Essential Performance --Collaterial Standard: Usability - AAMI/ANSI/IEC 60601-2-2:2009 Medical Electrical Equipment Part 2-2: Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgery Equipment And High Frequency Surgical Accessories - IEC 60601-2-18:2009 Medical Electrical Equipment Part 2-18: Particular Requirements For The Basic Safety And Essential Performance Of Endoscopic Equipment - ISO 8600-1:2013 Optics And Photonics -- Medical Endoscopes And Endotherapy Devices -- Part 1: General Requirements - ISO 8600-3:1997 (A1:2003) Optics And Optical Instruments Medical Endoscopes And Endoscopic Accessories - Part 3: Determination Of Field Of View And Direction Of View Of Endoscopes With Optics - ISO 8600-4:2014 Optics And Optical Instruments -- Medical Endoscopes And Certain Accessories -- Part 4: Determination Of Maximum Width Of Insertion Portion - IEC 62366:2007 (A1:2014) Medical devices Application of usability engineering to medical devices - Edition 1.0 Based on its intended use and active accessory classification, the next generation Percuvance™ System has been tested for high frequency dielectric strength and mains frequency dielectric strength. Additionally, the following electromagnetic {16}------------------------------------------------ | Emissions Testing (IEC 60601-1-2:2014) | | |-----------------------------------------|-----------| | Standard and Test Description | Results | | CISPR 11, Radiated Emissions (Class A) | Compliant | | CISPR 11, Conducted Emissions (Class A) | Compliant | | Immunity Testing (IEC 60601-1-2:2014) | | | Standard and Test Description | Results | | IEC 61000-4-3, Radiated Immunity | Compliant | | IEC 61000-4-6, Conducted Immunity | Compliant | compatibility (EMC) testing was conducted on the next generation Percuvance™ System. #### M. Conclusion Based upon the testing and research presented throughout the submission and in this 510(k) Summary, Teleflex Medical's next generation Percuvance™ Percutaneous Surgical System is substantially equivalent in to the predicate devices cleared to market via 510(k) K143299 and K083754. The new design of the Percuvance™ Percutaneous Surgical System does not introduce any new issues of safety and effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...