K130596 · Aesculap, Inc. · GEI · Mar 22, 2013 · General, Plastic Surgery
Device Facts
Record ID
K130596
Device Name
CAIMAN SEAL AND CUT TECHNOLOGY
Applicant
Aesculap, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 22, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Caiman Seal and Cut Technology consists of dedicated bipolar electrosurgical instruments intended for use in general surgery and gynecologic surgical procedures where ligation and division of vessels is desired. The instruments create a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue. The Caiman 12 Plus (44cm) and the Caiman 5 are indicated for laparoscopic procedures and the Caiman 12 Plus (24cm) is indicated for open procedures. The indications for use include general surgical procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, bowel resection, and oophorectomy etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument. Caiman Seal and Cut Technology has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.
Device Story
System comprises Lektrafuse RF generator and sterile, single-use bipolar electrosurgical instruments; used in open and laparoscopic surgery. Physician operates device to grasp, seal, and divide tissue/vessels. Device applies bipolar RF energy to create vessel seal; integrated cutting blade divides tissue. Output is physical tissue ligation and division. Benefits include efficient vessel management and reduced surgical time. Used in OR by surgeons.
Clinical Evidence
Bench testing only.
Technological Characteristics
Bipolar electrosurgical system; RF energy source; single-use instruments; laparoscopic and open form factors; manual actuation of cutting blade.
Indications for Use
Indicated for general, gynecologic, urologic, vascular, thoracic, and thoracoscopic surgical procedures requiring vessel ligation and tissue division. Suitable for vessels up to 7mm and tissue bundles fitting within device jaws. Contraindicated for tubal sterilization or tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Aragon Surgical RF System Laparoscopic (L2) Instrument (K090306)
Aragon Surgical RF System Teleo Instrument (K093075)
{0}------------------------------------------------
# K130596
# MAR 2 2 2013
Page 1 of 2
5
#### 510(k) SUMMARY (as required by 21 CFR 807.92) C.
## Caiman® Seal and Cut Technology March 4, 2013
| COMPANY: | Aesculap ®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
|----------------------|---------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Denise R. Adams<br>610-984-9076 (phone)<br>610-791-6882 (fax) |
| TRADE NAME: | Caiman Seal and Cut Technology |
| COMMON NAME: | Electrosurgical, Cutting & Coagulation & Accessories |
| CLASSIFICATION NAME: | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| REGULATION NUMBER: | 21 CFR 878.4400 |
| PRODUCT CODE: | GEI |
### SUBSTANTIAL EQUIVALENCE
Caiman Seal and Cut Technology is substantially equivalent to the Aragon Surgical RF System Laparoscopic (L2) Instrument cleared via K090306, the Aragon Surgical RF System Teleo Instrument cleared via K093075 and the Aragon Surgical RF System- 5mm Laparoscopic Instrument cleared via K110824.
### DEVICE DESCRIPTION
Caiman Seal and Cut Technology consists of the Lektrafuse RF Generator and the sterile, single use Caiman devices. These devices are capable of vessel sealing, blunt dissection, grasping and dividing tissue enclosed within its jaws during open and laparoscopic procedures. The devices are designed to be used with the dedicated Lektrafuse RF Generator and create vessel ligation by the application of bipolar electrical RF energy and tissue division with a cutting blade.
{1}------------------------------------------------
Page 2 of 2
റ്
## INDICATIONS FOR USE
Caiman Seal and Cut Technology consists of dedicated bipolar electrosurgical instruments intended for use in general surgery and gynecologic surgical procedures where ligation and division of vessels is desired. The instruments create a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.
30596
The Caiman 12 Plus (44cm) and the Caiman 5 are indicated for laparoscopic procedures and the Caiman 12 Plus (24cm) is indicated for open procedures. The indications for use include general surgical procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, bowel resection, and oophorectomy etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument.
Caiman Seal and Cut Technology has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.
#### TECHNOLOGICAL CHARACTERISTICS (compared to predicates)
The modifications made to the Caiman Seal and Cut Technology system do not affect the fundamental scientific technology. The design, materials, and principal of operation have not changed for these devices. The modifications made to these devices do not raise any new issues of safety and effectiveness.
### PERFORMANCE DATA
Bench testing was performed on the modified devices and found them to be substantially equivalent to the predicate devices.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a stylized emblem. The emblem consists of a stylized eagle or bird-like figure with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 22, 2013
Aesculap, Incorporated % Ms. Denise Adams Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K130596 Trade/Device Name: Caiman® Seal and Cut Technology Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: March 04, 2013 Received: March 07, 2013
Dear Ms. Adams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Peter D. Rumm-S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT B. Device Name: Caiman® Seal and Cut Technology
#### Indications for Use:
Caiman Seal and Cut Technology consists of dedicated bipolar electrosurgical instruments intended for use in general surgery and gynecologic surgical procedures where ligation and division of vessels is desired. The instruments create a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.
The Caiman 12 Plus (44cm) and the Caiman 5 are indicated for laparoscopic procedures and the Caiman 12 Plus (24cm) is indicated for open procedures. The indications for use include general surgical procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, bowel resection, and oophorectomy etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument.
Caiman Seal and Cut Technology has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.
Prescription Use
(per 21 CFR 801 Subpart D)
(per 21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE (IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Long H. Chen-A |
|----------------|
|----------------|
| Digitally signed by Long H. Chen-A |
|--------------------------------------|
| ON: c=US, o=U.S. Government, ou=HHS, |
| o=FDA, ou=People, cn=Long H. Chen |
| 6.9.2342.19200300.100.1.1=1300369056 |
| Date: 2013.03.22 06:50:49 -04'00' |
for MXM(Division Sign-Off)
Division of Surgical Devices
510(k) Number ________________________________________________________________________________________________________________________________________________________________