K011950 · Galil Medical , Ltd. · GEH · Jan 29, 2002 · General, Plastic Surgery
Device Facts
Record ID
K011950
Device Name
SEEDNET, MODEL FP6T5, FP5T5, FP5T3
Applicant
Galil Medical , Ltd.
Product Code
GEH · General, Plastic Surgery
Decision Date
Jan 29, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SeedNet™ System is intended for cryogenic destruction of tissue during surgical procedures. It is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology, thoracic surgery, ENT, gynecology, oncology, proctology, and urology. The system is designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts. The SeedNet System has the following specific indications: Urology (ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia "BPH") Oncology (ablation of cancerous or malignant tissue, and ablation of benign tumors, and palliative intervention) Dermatology (ablation or freezing of skin cancers and other cutaneous disorders) Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia) General surgery (destruction of warts or lesions, palliation of tumors of the oral cavity, rectum and skin, and ablation of leukoplakia of the mouth, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas small hemanglomas, mucocele cysts, multiple warts, plantar warts, hemorroids, anal fissures, perianal conylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemanglomas, and recurrent cancerous lesions) Thoracic surgery (ablation of arrhythmic cardiac tissue, and ablation of cancerous lesions) Proctology (ablation of benign or malignant growths of the anus or rectum, and ablation of hemorrhoids)
Device Story
SeedNet™ System is a cryosurgical unit utilizing extreme cold to destroy targeted tissue. Device inputs include user-defined parameters via computer interface; system delivers cryotherapy through specialized probes (including 2mm, 3mm, and flexible variants). Operated by physicians in surgical settings; system provides visual feedback via computer screens to monitor procedure. Output is localized tissue necrosis. Benefits include minimally invasive destruction of cancerous or benign lesions across multiple medical specialties. Device modifications include expanded probe options and updated display interfaces compared to predicate systems.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and design similarities to cleared predicate devices.
Technological Characteristics
Cryosurgical unit; includes 2mm, 3mm, and flexible probes. System utilizes extreme cold for tissue ablation. Connectivity includes computer interface for monitoring. Software algorithm is unchanged from previous version; modifications limited to hardware accessories and display screens.
Indications for Use
Indicated for patients requiring tissue ablation across general surgery, dermatology, neurology, thoracic surgery, ENT, gynecology, oncology, proctology, and urology. Specific applications include prostate cancer, BPH, malignant/benign tumors, skin cancers, cutaneous disorders, female genital dysplasia, cardiac arrhythmias, and various lesions/cysts.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
# p.1/2 Galil Medical Proprietary Information
# SECTION 8-510(K) SUMMARY
# 510(k) Summary
Galil Medical - SeedNet™ System
JAN 2 9 2002
510(k) Number
Kolly50
Company Name:
# Galil Medical Ltd.
#### Contact Person:
Dr. Roni Zvuloni, Director of IP & Regulatory Affairs Telephone: +972-4-959 10 80 +972-4-959 10 77 Fax:
Trade Proprietary Name:
SeedNet™.
Classification Name:
## CRYOSURGICAL UNIT
Classification:
GEH
{1}------------------------------------------------
#### Predicate Devices:
- 1. SeedNet™
- 2. CRYO-HIT™
- 3. Cryo-Mono with TUF Probe
#### Indication for Use:
The modified SeedNet™ System is intended for cryogenic destruction of tissue during surgical procedures. is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology, thoracic surgery, ENT, gynecology, oncology, proctology, and urology. The system is designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.
The modified SeedNet has the following specific indications:
Urology (ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia "BPH")
Oncology (ablation of cancerous or malignant tissue, and ablation of benign tumors, and palliative intervention)
Dermatology (ablation or freezing of skin cancers and other cutaneous disorders)
Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia)
General surgery (destruction of warts or lesions, palliation of tumors of the oral cavity, rectum and skin, and ablation of leukoplakia of the mouth, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas small hemanglomas, mucocele cysts, multiple warts, plantar warts, hemorroids, anal fissures, perianal conylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemanglomas, and recurrent cancerous lesions)
Thoracic surgery (ablation of arrhythmic cardiac tissue, and ablation of cancerous lesions)
Proctology (ablation of benign or malignant growths of the anus or rectum, and ablation of hemorrhoids)
{2}------------------------------------------------
#### Technological Characteristics:
The Galil Medical's SeedNet™ System is a modification of Galil Medical LTD's cleared SeedNet™ System with STPS (K010991). The SeedNet™ System is the exact same device as the SeedNet™ except for the following modifications to its technological characteristics:
Addition of 2 and 3 mm probes which are included in the previously cleared the Cryo-Hit™, to which the SeedNet™ is a modification.
Addition of flexible probes as an accessory to the SeedNet™
The systems computer screens, but not its software algorithm have slightly been modified.
Substantial Equivalence
The modified SeedNet has the same intend use as the cleared SeedNet, Cryo-Hit, and Cryo-Mono with TUF Probe, the same general and specific indications as the Cryo-Hit, the same principles of operation as the cleared SeedNet, and very similar technological characteristics as the cleared SeedNet, Cryo-Hit and Cryo-Mono with TUF Probe. The minor difference in the diameter of the modified SeedNet's and Cryo-Mono's flexible probes and the modified SeedNet's computer screens does not raise any new questions of safety or effectiveness. Thus, the SeedNet™ System is substantially equivalent to these predicate devices.
{3}------------------------------------------------
FEB 2 1 2008
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Galil Medical Ltd. c/o Mr. Jonathan S. Kahan Hogin & Hartson L.L.P. 555 Thirteenth Street, NW Washington, DC 20004-1109
Re: K011950
Trade/Device Name: SeedNet™ System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: II (two) Product Code: OCL. GEH Dated: October 30, 2001 Received: October 31, 2001
Dear Mr. Kahan:
This letter corrects our substantially equivalent letter of January 29, 2002.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Jonathan S. Kahan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
> Sincerelv yours,
Q.M.a
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
JAN 2 9 2002
Koll1950 510(k) Number (if known):
Device Name:
SeedNet™ System
Indications for Use:
The SeedNet™ System is intended for cryogenic destruction of tissue during surgical procedures. It is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology, thoracic surgery, ENT, gynecology, oncology, proctology, and urology. The system is designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.
The SeedNet System has the following specific indications:
Urology (ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia "BPH")
Oncology (ablation of cancerous or malignant tissue, and ablation of benign tumors, and palliative intervention)
Dermatology (ablation or freezing of skin cancers and other cutaneous disorders)
Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia)
General surgery (destruction of warts or lesions, palliation of tumors of the oral cavity, rectum and skin, and ablation of leukoplakia of the mouth, angiomas,
{6}------------------------------------------------
sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas small hemanglomas, mucocele cysts, multiple warts, plantar warts, hemorroids, anal fissures, perianal conylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemanglomas, and recurrent cancerous lesions)
Thoracic surgery (ablation of arrhythmic cardiac tissue, and ablation of cancerous lesions)
Proctology (ablation of benign or malignant growths of the anus or rectum, and ablation of hemorrhoids)
## (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-off) Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices
510(k) Number
Prescription Use 1 (Per 21 CFR 801.109)
Over the Counter Use
Muriam C. Provost
OR
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number_Koll 950