MODIFICATION TO BONE ANCHOR AND KIT (BAK)

K994313 · Axya Medical, Inc. · GAW · Jan 13, 2000 · General, Plastic Surgery

Device Facts

Record IDK994313
Device NameMODIFICATION TO BONE ANCHOR AND KIT (BAK)
ApplicantAxya Medical, Inc.
Product CodeGAW · General, Plastic Surgery
Decision DateJan 13, 2000
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.5010
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Bone Anchor System and Kit is indicated for securing soft tissue structures to bone anchors in surgical procedures where bone anchors are indicated. Examples of these procedures include: Shoulder procedures (e.g. Rotator cuff repair) Knee procedures (e.g. Patellar ligament / tendon avulsion repair) Ankle procedures (e.g. Achilles tendon reconstruction and repair) Foot procedures (e.g. Hallux valgus reconstruction) Elbow procedures (e.g. Biceps tendon reattachment) Bladder neck suspension) (female urinary incontinence due to urethral hypermobility) The BAK is indicated for use in both traditional open surgery and endoscopic surgical procedures.

Device Story

System comprises electronic control module, reusable welding activator, and reusable welding sleeve; used with bone anchors and USP polypropylene monofilament suture. Operates by ultrasonic energy to weld/fuse suture loops, replacing manual knot-tying. Used in OR by surgeons for soft tissue-to-bone attachment. Ultrasonic generator remains isolated from tissue; sleeve prevents heated suture contact with patient. Welded sutures provide holding strength equivalent to conventional knots; verified by bench testing (knot strength, cyclic fatigue, elongation). Benefits include secure fixation without thermal injury risk or chemical additives.

Clinical Evidence

No clinical trials; evidence based on bench testing and animal studies. Bench testing: knot strength exceeded USP requirements; cyclic testing confirmed fatigue/elongation resistance. Animal study (New Zealand rabbits): compared welded vs. manual knots in skin and bowel; no histopathologic differences or safety concerns observed. Ultrasonic welding demonstrated equivalent efficacy and safety to traditional knotting.

Technological Characteristics

System includes electronic control module, reusable welding activator, and sleeve. Ultrasonic energy source used for friction-based polymer welding. Materials have established medical history. Non-sterile delivery for reusable components; suture/anchors sterilized by manufacturer. No flux or welding rods used. Connectivity: integral connector cable between activator and control module.

Indications for Use

Indicated for securing soft tissue to bone anchors in open or endoscopic surgical procedures, including rotator cuff repair, patellar ligament/tendon avulsion repair, Achilles tendon reconstruction, hallux valgus reconstruction, biceps tendon reattachment, and bladder neck suspension for female urinary incontinence.

Regulatory Classification

Identification

Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # SUMMARY OF SAFETY AND EFFECTIVENESS #### General Company Information - Axva Medical, Inc. Name: 100 Cummings Center Address: Suite 444C Beverly, MA 01915 - (978) 232 9997 Telephone: (978) 232 - 9998 Fax: ## General Device Information - Product Name: Bone Anchor System and Kit (BAK) Classification: Accessory to an Endoscope - Class II ## Predicate Devices Axya Suturing and Ligating System (ASLS) [510(k) K980988] Axya Suture Welding System and Kit (SWK) [510(k) K983108] ## Description The Axya Bone Anchor System consists of an electronic control module and a reusable welding activator with an integral connector cable. The Kit (BAK) consists of a reusable suture welding sleeve packaged together with appropriate and currently approved bone anchors and USP polypropylene monofilament suture. The Kit is intended for use with the reusable system components in order to complete the suture welding process for the attachment of soft tissues to bone anchors in surgical procedures where bone anchors are indicated {1}------------------------------------------------ ## Intended Use The Bone Anchor System and Kit is indicated for securing soft tissue structures to bone anchors in surgical procedures where bone anchors are indicated. Examples of these procedures include: | Shoulder procedures | (e.g. Rotator cuff repair) | |--------------------------|-------------------------------------------------------------| | Knee procedures | (e.g. Patellar ligament / tendon avulsion repair) | | Ankle procedures | (e.g. Achilles tendon reconstruction and repair) | | Foot procedures | (e.g. Hallux valgus reconstruction) | | Elbow procedures | (e.g. Biceps tendon reattachment) | | Bladder neck suspension) | (female urinary incontinence due to urethral hypermobility) | The BAK is indicated for use in both traditional open surgery and endoscopic surgical procedures. # Substantial Equivalence This submission supports the position that the Axya Bone Anchor System and Kit (BAK) which includes USP size 2 polypropylene suture (manufactured by United States Surgical or Sharpoint) and a reusable welding sleeve is substantially equivalent to the original Bone Anchor System and Kit [501(k) K992405], and to the Axya Suturing and Ligating System (ASLS) [510(k) K980988], and the Axya Suture Welding System and Kit (SWK) [510(k) K983108); and that it is appropriate for its intended application. Suture applicators and suture placement devices which may be used in both endoscopic or traditional open surgical procedures have been classified under 21 CFR 876.1500. These devices are indicated for the placement of sutures to close either traumatic or surgically produced wounds. The BAK is fabricated from materials with a substantial history of use in medical devices. Both the BAK and the predicate devices secure the suture loop with an ultrasonic weld. The 510(k) Notice for the predicate ASLS system contains summaries of both in vivo and in vitro studies which were conducted to evaluate the safety, efficacy and appropriateness of the suture welding system. These data are applicable to the Bone Anchor System because the design and components of the welding mechanism are the same as those used in the BAK, ASLS and the SWK that were previously cleared for marketing. Data were presented which demonstrate that sutures placed by means of the suture welding process exhibit "knot strength" characteristics substantially above the USP requirements for the various USP suture materials tested. These tests confirm that the polypropylene monofilament sutures placed with the ultrasonic suture welding technology are equivalent in holding strength (efficacy) to sutures placed with conventional knotting techniques. Data presented in the original Notice for the Bone Anchor System and Kit confirm that USP size 2 polypropylene monofilament sutures secured by a weld formed with the AxvaWeld Tips exceed the USP knot strength requirements. Data were presented that demonstrate that strength requirements are met regardless of the angle of applied tension between the suture {2}------------------------------------------------ and the bone anchor stem. Suture fatigue and elongation over the functional life were evaluated in bench-test models and the welded sutures have been found to meet the suggested criteria following cyclic testing. The safety of the suture delivery system was previously evaluated by placing both welded sutures and manually knotted sutures in the dorsal skin and in bowel tissue of New Zealand rabbits [510(k) K980988]. Positive (electrocautery contact) and negative (normal tissue) controls were used in the bowel study. In both animal models there was no histopathologic change seen at the suture implant sites where the suture was sealed with ultrasonic energy. The investigators concluded that there was no significant difference in safety of efficacy between the traditional method of suture placement and the technique which includes replacement of manual knot-tying with suture welding. The design of the AxyaWeld Sleeve also prevents the heated area of suture from coming into direct contact with tissues. The ultrasonic energy source used to weld and secure the suture loop is the same generator used for the Automatic Suturing and Ligating System and that used for the Suture Welding System and Kit, and is similar to the energy source used in the UltraCision Harmonic Scalpel [510(k) K895252]. Because of design features of the BAK, no portion of the ultrasonic generator comes into contact with human tissues. Because of this, there is virtually no risk of causing a thermal injury to the patient. The suture material is heated and welded by friction and the weld is formed by melting and fusing the polymer. No "flux" or "welding rod" is employed and no new chemical entities are introduced or produced in the welding process. Axya Bone Anchor System and Kit with a reusable sleeve is provided non-sterile. The suture material and bone anchors are sterilized by the original manufacturer as dictated by the type of suture included with a given product configuration. The sterility processes, the manufacturing process, and the packaging process are validated by the manufacturer. Axya Medical, Inc. believes that the information provided establishes that similar legally marketed and pre-enactment devices have been used for the same types of clinical applications as the Axya Bone Anchor System and Kit. The materials from which the Axya device is fabricated have an established history of use in medical applications, and the specific materials used by Axya have been thoroughly tested in accordance with applicable FDA guidelines. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three lines representing its wings. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 1 3 2000 Mr. Howard L. Schrayer Axya Medical, Inc. 100 Cummings Center Suite 444C Beverly, Massachusetts 01915 Re: K994313 Trade Name: Axya Bone Anchor System and Kit (BAK) Regulatory Class: II Product Code: GAW and MBI Dated: December 21, 1999 Received: December 22, 1999 Dear Mr. Schrayer: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that. through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a {4}------------------------------------------------ #### Page 2 - Mr. Howard L. Schrayer legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Tunnell J. Vayor Sames E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K494313 510(k) Number (if known): Device Name: Axya, Bone Anchor System and Kit, (BAK) Indications For Use: The Bone Anchor System and Kit is indicated for securing soft tissue structures to bone anchors in surgical procedures where bone anchors are indicated. Examples of these procedures include: | Shoulder procedures | (e.g. Rotator cuff repair) | |-------------------------|-------------------------------------------------------------| | Knee procedures | (e.g. Patellar ligament / tendon avulsion repair) | | Ankle procedures | (e.g. Achilles tendon reconstruction and repair) | | Foot procedures | (e.g. Hallux valgus reconstruction) | | Elbow procedures | (e.g. Biceps tendon reattachment) | | Bladder neck suspension | (female urinary incontinence due to urethral hypermobility) | The BAK is indicated for use in both traditional open surgery and endoscopic surgical procedures. #### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Ressell L. Jogan (Division Sign-Off) Dr. sion of Leneral Restorative Devices 16994313 STU(K) Number_ Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...