CONTOUR NECKLIFT THREADS

K050247 · Surgical Specialties Corp · GAW · Mar 4, 2005 · General, Plastic Surgery

Device Facts

Record IDK050247
Device NameCONTOUR NECKLIFT THREADS
ApplicantSurgical Specialties Corp
Product CodeGAW · General, Plastic Surgery
Decision DateMar 4, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.5010
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Contour Necklift Thread™ is indicated for use in Necklift Surgery. The Necklift Contour Threads are specifically indicated for use to fixate and elevate the subdermis to the deep fascia of the retromastoid area.

Device Story

Contour Necklift Threads are sterile, nonabsorbable, polypropylene surgical strands (USP size 2-0). Device features 25cm length with 10cm unidirectional barbed section; 7-inch straight needle at distal end; 26mm curved needle at proximal end. Used by surgeons in necklift procedures to fixate and elevate subdermis to deep fascia of retromastoid area. Device provides mechanical fixation via barbs, replacing or augmenting traditional suture techniques. Single-use, sterile supply.

Clinical Evidence

Bench testing only. Physical testing conducted to USP 27 standards for tensile strength and force required to remove after implantation. Biocompatibility testing performed per ISO 10993 for permanent implantation.

Technological Characteristics

Material: Polypropylene (USP size 2-0). Design: 25cm strand with 10cm unidirectional barbed section. Needles: 7-inch straight (distal), 26mm curved (proximal). Sterilization: Sterile, single-use. Biocompatibility: ISO 10993 compliant.

Indications for Use

Indicated for use in necklift surgery to fixate and elevate the subdermis to the deep fascia of the retromastoid area.

Regulatory Classification

Identification

Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAR 4 - 2005 Image /page/0/Picture/1 description: The image shows the logo for Surgical Specialties Corporation. The logo is black and white and features the company name in a stylized font. Above the logo is the text 'K050247 P. 1/3'. ## 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS SUBSTANTIAL EQUIVALENCY | Submitter: | Surgical Specialties Corporation | |----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address: | 100 Dennis Drive<br>Reading, PA 19606 | | Telephone: | 610 404 1000, ext. 2231 | | Contact Person: | Elizabeth Lazaro<br>Regulatory Affairs Specialist | | Date Prepared: | January 31, 2005 | | Name of Device: | Contour Necklift Threads™ | | Common / Usual<br>Classification Name: | GAW<br>Suture, Non Absorbable, Synthetic, Polypropylene | | Predicate Device: | Contour Midface Threads™ K041593<br>Contour Browlift Threads™ K042856<br>Prolene™ Sutures N16374 | | Indications For Use: | The Contour Necklift Thread™ is indicated for use<br>in Necklift Surgery. The Necklift Contour Threads<br>are specifically indicated for use to fixate and<br>elevate the subdermis to the deep fascia of the<br>retromastoid area. | Contour Necklift Threads™ Surgical Specialties Corporation Image /page/0/Picture/5 description: The image shows the word "LOOK" in black font, with the two O's being perfect circles. To the right of the K is a black circle with white swirls inside. Underneath the word "LOOK" is the phrase "Precisely what you need" in a smaller font. Image /page/0/Picture/7 description: The image shows the logo for Sharpoint, a company that specializes in surgical tools. The logo features the word "Sharpoint" in a stylized font, with the registered trademark symbol next to it. To the right of the word is a black triangle pointing to the right. Below the word "Sharpoint" is the tagline "The Surgeon's Edge" in a smaller font. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Surgical Specialties Corporation. The logo features the words "Surgical Specialties" in a stylized font, with a black triangle to the right. Below the main text, it says "Corporation" and "Incision & Healing Technologies" in smaller fonts. There is also a handwritten "K050247" and "P2/3" at the top of the image. ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (continued) | Device Description | The Contour Necklift Threads™ are a clear, nonabsorbable, sterile, surgical strand of polypropylene. The base product is USP size 2-0 polypropylene suture. The 25centimeter length thread incorporates a unidirectional barbed section 10 centimeters long from the distal end. The Threads will have a 7inch straight needle attached to the distal end and a 26 mm curved needle attached to the proximal end. The threads are supplied sterile for single use. | |--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Technological Characteristics: | The Polypropylene material used for the Contour Necklift Threads™ is commonly used in medical applications and has been proven to be biocompatible. Bench and animal evaluations have demonstrated the device to be safe and effective. These characteristics are the same for Surgical Specialties Corporation's Contour Threads as presented in the predicate 510 (k)'s. There is no change in chemistry, material or composition. | | Performance Data | Physical testing was conducted on the thread to USP 27 tensile strength, force required to remove after implantation and Biocompatibility for permanent implantation, ISO 10993. | Image /page/1/Picture/4 description: The image shows the word "LOOK" in bold, black letters. To the right of the word is a black circle with white swirls inside. Below the word "LOOK" is the phrase "Precisely what you need" in a smaller font. The letters are all capitalized except for the "o" in "what", the "y" in "you", and the "n" in "need". 100 Dennis Drive · Reading, PA · 19606, U.S.A. 610 404-1000 • 800 523-33332 • Fax: 610 404-4010 www.surgicalspecialties.com Image /page/1/Picture/6 description: The image contains the word "Sharpoint" in a stylized font, with the registered trademark symbol to the right of the word. Below "Sharpoint" is the phrase "The Surgeon's Edge" in a smaller font. To the right of the word "Sharpoint" is a black triangle pointing to the right. Below the text is the number 9 in a handwritten style. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image contains the handwritten text "K050247 P3/3" at the top. Below this, the logo for "Surgical Specialties Corporation Incision & Healing Technologies" is displayed. The logo features a stylized font for "Surgical Specialties" with a solid black triangle pointing to the right. Substantial Equivalence The Contour Necklift Thread™ is equivalent in the intended use to the predicate devices,The Contour Midface and Browlift Threads. The Contour Necklift Threads uses the same suture material as the predicate devices, Contour Midface and Browlift Threads. The use in different anatomical sites for the Necklift, Brow lift and Midface applications does not change the safety and effectiveness. The Contour Necklift Threads also have a predicate device in the use of the PROLENE suture for what has been established as a safe and effective Gold Standard Surgical Procedure for Necklift in the last 2 decades, as described in this submission. The substantial equivalence determination is equivalent for these indications. Image /page/2/Picture/4 description: The image shows the word "LOOK" in bold, sans-serif font. To the right of the word "LOOK" is a circular graphic with a spiral design inside. Below the word "LOOK" is a line of smaller text that is difficult to read due to its size and resolution. The overall design appears to be a logo or branding element. 100 Dennis Drive · Reading, PA 19606, U.S.A. 610 404-1000 · 800 523-3332 · Fax: 610 404-4010 www.surgicalspecialties.com Image /page/2/Picture/6 description: The image shows the word "Sharpoint" in a stylized font, with a registered trademark symbol next to it. To the right of the word is a black triangle pointing to the right. Below the word "Sharpoint" is the phrase "The Surgeon's Edge" in a smaller font. The number 10 is written in the bottom right corner of the image. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, facing to the left. The eagle is depicted in a simple, black and white design. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 4 - 2005 Ms. Elizabeth Lazaro Regulatory Affairs Specialist Surgical Specialties Corporation 100 Dennis Drive Reading, Pennsylvania 19606 Re: K050247 Trade/Device Name: The Contour Necklift Threads™ Regulation Number: 21 CFR 878.5010 Regulation Name: Non-absorbable synthetic polypropylene suture Regulatory Class: II Product Code: GAW Dated: January 31, 2005 Received: February 2, 2005 Dear Ms. Lazaro: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your became 37 (4) pe device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encroours for regardy and the Medical Device Amendments, or to comments provision to May 20, 1978, the excordance with the provisions of the Federal Food, Drug, devices that have been recurse in assess approval of a premarket approval application (PMA). and Cosmetic rice (11ct) that as nover subject to the general controls provisions of the Act. The T ou may, therefore, manel the Act include requirements for annual registration, listing of general controls profitive of ractice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (500 a0070) als. Existing major regulations affecting your device can may be subject to subli adaterial egulations, Title 21, Parts 800 to 898. In addition, FDA may be found in the Out of Founts concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean I least of advised that I Drivised or our device complies with other requirements of the Act mat 1 127 has made a aoth regulations administered by other Federal agencies. You must of any rederal statutes and regarants ancluding, but not limited to: registration and listing (21 comply with an the Her 812 CFR Part 801); good manufacturing practice requirements as set CITN Fatt 807), labeling (21 CFR Part 820); and if applicable, the electromic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Ms. Elizabeth Lazaro This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin makemig your averal equivalence of your device to a legally premarket notification. The PDA miding of sacolantial of your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please s and the many of the comments of the cases the recultion antitled If you desire specific advice for your ac not (240) 276-0115. Also, please note the regulation entitled, contact the Office of Complanes at (210) 215 Part 807.97). You may obtain " Misoranding by reference to premantee notifican under the Act from the Division of Snall other general International and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Consumer Absociates of the Manufactures html. Sincerely yours. Muriam C Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Ko50247 ## Indications for Use Image /page/5/Picture/2 description: The image shows the logo for Surgical Specialties Corporation. The logo is in black and white and features the company name in a stylized font. The word "Surgical" is stacked on top of the word "Specialties", and there is a black triangle to the right of the word "Specialties". Below the company name, the words "Corporation" and "Incision & Healing Technologies" are written in a smaller font. 510(k) Number (if known): Device Name: The Contour Necklift Threads™ Indications For Use: The Contour Necklift Threads™ are indicated for use in Necklift surgery. The Contour Necklift Threads - are incollected for use to fixate and The Contour Necklift Throuds - and Subdermis to the retromastoid area. Muriam C Phurst (Division Sign-Division of General, Restorative, and Neurological Devices \$10(k) Number K650247 Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Image /page/5/Picture/15 description: The image shows the text "Page 1 of 1". The text is in a simple, sans-serif font and is positioned in the upper left corner of the image. The number "1" is written after the word "of". Image /page/5/Picture/16 description: The image shows the word "LOOK" in bold, black letters. To the right of the word is a black circle with a white swirl design inside. Below the word "LOOK" is the phrase "Precisely what you need" in a smaller, lighter font. The logo appears to be for a company or product named "LOOK". 100 Dennis Drive · Reading, PA 19606, U.S.A. 610 404-1000 - 800 523-3332 - Fax: 610 404-4010 www.surgicalspecialties.com Sharpoint® The Surgeon's Edge
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...