K093009 · Dr. Mach GmbH & Co. KG · FSY · Apr 1, 2010 · General, Plastic Surgery
Device Facts
Record ID
K093009
Device Name
MACH LED SC
Applicant
Dr. Mach GmbH & Co. KG
Product Code
FSY · General, Plastic Surgery
Decision Date
Apr 1, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The Mach LED SC lighting system is designed for illuminating an examination area and surgical field at the hospital and doctor's practice. The surgical light Mach LED SC is intended to illuminate the surgical field and the patient.
Device Story
Mach LED SC is a ceiling-mounted surgical lighting system; provides illumination for surgical fields and patients. System comprises lamp housing, LED modules, mechanical suspension arms, and sterilizable handle sleeves. Operated via keypad on lamp head or wall-mounted unit. Features include light field merging and intensity control. Accessories include camera modules, monitor yokes, instrument trays, and battery backup. Used in hospitals and clinics by medical staff. Enhances visualization of surgical site to support clinical decision-making and patient care.
Clinical Evidence
No clinical data is required for this device classification submission. Bench testing performed to verify compliance with EN 60601-1, EN 60601-2-42, EN 55015, and EN 61000-4.
Technological Characteristics
LED-based surgical lighting; ceiling-mounted suspension with 360-degree horizontal rotation and vertical articulation. Components include lamp housing, LED modules (one white LED per module), and electrical/mechanical assemblies. Connectivity includes optional network interface for camera modules. Power source 24V DC battery backup support available. Sterilizable handle sleeves provided.
Indications for Use
Indicated for illumination of the surgical field and patient during examinations or surgical procedures in hospital or clinical settings.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
STERIS Corporation POWERLED SURGICAL LIGHT SYSTEM (K070442)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: This image is a document titled "5.0 Traditional 510(k) Summary" from Dr. Mach. It includes the submitter's information, such as the company name, address, and contact person, Rainer Adams, along with their phone and fax numbers. The document also lists the preparation date as July 13, 2009, the trade name as Mach LED SC, the regulation number as 21 CFR 878.4580, the common name as Surgical Lamp, and the regulatory class as Class II. The document is dated APR - 1 2010.
#### Device Description:
Product Code:
---
The surgical light Mach LED SC is intended to illuminate the surgical field and the patient. This light system can be added to the ceiling mounted suspension system supporting the horizontal arms and spring arms. The horizontal arms can be rotated horizontally with 360°, the spring arms can be rotated horizontally with 360° and moved vertically with 50° downwards and 45° upwards. The light system is operated by a keypad on the lamp head or, by special request of the customer, by a keypad on the wall.
The Mach LED SC consists of lamp housing, LED modules, optical -/electrical and mechanical components, one sterilizable handle sleeve as well as the cables.
One LED-module consists of one white LED.
The surgical light Mach LED SC will be market with merging of light fields and light intensity control. Colour temperature adjustment not available.
Available accessories for the Mach LED SC lighting systems are as follows:
- Camera module
- Remote control for camera module
- Remote control with network interface for camera module
- Single monitor yoke for flat panel monitors
FSY
- Double monitor yoke for flat panel monitors
- Instrument trays
- Trays for CRT monitors
- 24V DC battery backup support
- Low profile wall control unit
- Integrated laser pointer
- Sterilizable handle sleeves
File: 05_510(k)_Summary_SC_rev01
Page 1 of 2
{1}------------------------------------------------
D 2 of C
Image /page/1/Picture/1 description: The image shows the text "Dr. Mach Medizinische Leuchten". The text is in a bold, sans-serif font. The words "Dr. Mach" are on the top line, and the words "Medizinische Leuchten" are on the bottom line. The text is black on a white background.
- *Technik*
Image /page/1/Picture/3 description: The image shows a surgical light fixture. The fixture has three circular lights arranged in a triangular pattern. The lights are attached to a curved arm that extends from the ceiling.
#### Intended Use:
The Mach LED SC lighting system is designed for illuminating an examination area and surgical field at the hospital and doctor's practice.
### Indications for Use:
The surgical light Mach LED SC is intended to illuminate the surgical field and the patient.
### Predicate Devices:
Harmony LED-1 Surgical Lighting System (K072072), STERIS Corporation POWERLED SURGICAL LIGHT SYSTEM (K070442), MAQUET S.A.
### Substantial Equivalence:
The Mach LED SC is substantially equivalent to the surgical light Harmony LED-1 Surgical Lighting System and the POWERLED SURGICAL LIGHT SYSTEM. Any difference that exists between the Mach LED SC and both predicate devices has no negative effect on safety or effectiveness and actually enhances the usefulness in the choosen application.
## Technological Characteristics Comparison:
Technological Characteristics of Subject Device are based on the same LED (light-emitting diode) technology as the Predicate Device. For detailed comparison of all functionalities of the subject device and the predicate devices refer to Chapt. 12: Substantial Equivalent discussion.
# Discussion according non-clinical performance data testing:
Performance testing was conducted to verify that Mach LED SC meet the requirements for Medical Electrical Equipment as defined in EN 60601-1, EN 60601-2-42, EN 55015 and EN 61000-4.
# Discussion according clinical performance data testing:
No clinical data is required for this device classification submission.
# Test Conclusions of non-clinical and clinical performance data:
Testing was performed according to external procedures. Performance testing and validation were done at TÜV SÜD Product Service and mikes-testingpartners GmbH. Please refer to Chapter 18 and Appendic C.
Dr. Hans-Jorg Kemper General Manager
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized depiction of an eagle with its wings spread, symbolizing protection and care. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", indicating the department's name and national affiliation. The seal is presented in black and white, emphasizing its official and governmental nature.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
APR - 1 2010
Dr. Mach GmbH & Co. KG % Rainer Adams Technical Director Flossmannstrasse 28 85560 Ebersberg, Germany
Re: K093009
Trade/Device Name: Mach LED SC Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: FSY Dated: March 10, 2010 Received: March 30, 2010
Dear Rainer Adams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Rainer Adams
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm four the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use 510(k) Number (if known): K093009 Device Name: Mach LED SC
Indications for Use:
The surgical light Mach LED SC is intended to illuminate the surgical field and the patient.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R. Ogden for mxm
(Division Sign 66)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K093009
04_Indications_for_Use_Statement_SC rev02.doc