TruView Light Cable, TruView Lateral Retractor Light Cable, Bifurcated Light Cable With Universal End
K200874 · Innovasis, Inc. · FST · Jun 30, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200874
Device Name
TruView Light Cable, TruView Lateral Retractor Light Cable, Bifurcated Light Cable With Universal End
Applicant
Innovasis, Inc.
Product Code
FST · General, Plastic Surgery
Decision Date
Jun 30, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Intended Use
The TruView Light Cable is intended to provide surgical site illumination from a fiber optic light source.
Device Story
TruView Light Cable is a fiber optic surgical illumination device. It transmits light from an external fiber optic light source to a surgical site via borosilicate glass optical fibers. The device is used in an operating room setting by surgeons or surgical staff to provide visualization during procedures. It connects to compatible surgical retractors. The device does not perform data processing or automated analysis; it functions as a passive light transmission conduit. Clinical benefit is derived from improved surgical site visibility, facilitating surgical performance.
Clinical Evidence
Bench testing only. Performance testing included a fit test for retractor compatibility, a collateral heating test to ensure user safety (burn prevention), and an intensity test to ensure patient safety and adequate illumination. The device met all acceptance criteria.
Technological Characteristics
Fiber optic light cable consisting of borosilicate glass optical fibers in a silicone sheath with surgical stainless-steel couplings (ASTM F899). Sterilization trays consist of Anodized 5052 Aluminum, Nylon, Stainless Steel, Polypropylene, and RADEL (ASTM D6394 SP031). Passive light transmission; no electronic or software components.
Indications for Use
Indicated for providing surgical site illumination during surgical procedures requiring fiber optic light sources.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
K991208 — ENDOSCOPIC FIBEROPTIC CABLE · Isolux America · Jun 23, 1999
K980166 — LIGHTSOURCE OR ILLUMINATOR · Cuda Products Co. · Apr 1, 1998
K042034 — NUVASIVE MAXCESS LIGHT GUIDE · Nuvasive, Inc. · Oct 26, 2004
K972225 — UFR FIBEROPTIC CABLE · Universal Fiberoptic Repair, Inc. · Mar 18, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
June 30, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Innovasis, Inc. Marshall Mccarty Director QA/RA 614 East 3900 South Salt Lake City, Utah 84107
Re: K200874
Trade/Device Name: TruView Light Cable, TruView Lateral Retractor Light Cable, Bifurcated Light Cable With Universal End Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical Lamp Regulatory Class: Class II Product Code: FST Dated: March 31, 2020 Received: April 1, 2020
Dear Marshall Mccarty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name TruView™ Light Cable
Indications for Use (Describe)
The TruView Light Cable is intended to provide surgical site illumination from a fiber optic light source.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Innovasis. The logo is black and features the word "INNOVASIS" in all caps. Below the word is the phrase "INNOVATE / INVOLVE / INVENT" in a smaller font.
510(k)
#### 5.0 510(k) Summary Report:
TruView™ Lateral Retractor Light Cable
| Company: | Innovasis, Inc. |
|----------|--------------------------|
| | 614 E. 3900 South |
| | Salt Lake City, UT 84107 |
- Contact: Marshall C. McCarty Phone: (801) 261-2236 x8012 mmccarty@innovasis.com
- Trade Name: TruView Light Cable
- Common Name: Fiber Optic Light Cable
- Classification: Requlation No.: 21 CFR 878.4580 Class 2 Product Code: FST Review Panel: General & Plastic Surgery
- Primary Predicate: K042034 NuVasive, Inc. MaXcess Light Guide This predicate has not been subject to a design-related recall.
Additional Predicate:
K901035
Cuda Products Co./Sunoptic Technologies Fiberoptic Light Cable
- Device Description: The TruView Light Cable is a fiber optic surgical light designed to be compatible with a variety of fiber optic light sources.
- Performance Data: (Non-clinical)—The TruView Light Cable was subjected to three (3) performance tests that indicated it performs in accordance with its intended use. Testing included a Fit Test to show that it is compatible with the retractor; a Collateral Heating Test to show that the cable, when used with the identified light source, does not pose a risk to the surgeon or operating room users (burn/injury to user); and an Intensity Test to show that the cable, when used with the identified light source and the worst-case retractor blade configuration, does not pose a risk to the patient (burn/patient injury).
The subject TruView Light Cable met the acceptance criteria for the testing performed.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Innovasis. The logo is black and features the word "INNOVASIS" in all caps. Below the word is the phrase "INNOVATE / INVOLVE / INVENT" in a smaller font. The logo is simple and modern.
Materials: The TruView Light Guide contains borosilicate glass optical fibers wrapped in a silicone sheath and has surgical stainless-steel couplings per ASTM F899.
> The TruView sterilization trays are comprised of Anodized 5052 Aluminum and have components made of Nylon, Stainless Steel, Polypropylene, and RADEL per ASTM D6394 SP031.
- Indications for Use: The TruView Light Cable is intended to provide surgical site illumination from a fiber optic light source.
## Comparison of Technological Characteristics with the Predicate Devices:
The TruView Light Cable has been subjected to risk analysis, engineering analysis and testing to recognized standards and has been shown to be substantially equivalent to the predicate devices, NuVasive MaXcess Light Guide (K042034) and Sunoptic Fiberoptic Light Cable (K901035).
- Technology is substantially equivalent. -
- Design is substantially equivalent. —
- -Size is substantially equivalent.
- Indications for use (intended use) are substantially ー equivalent.
- Materials (biocompatibility profile) are substantially equivalent.
- Conclusion: The overall technology characteristics and mechanical performance data lead to the conclusion that the subject device is substantially equivalent to legally marketed predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.