K253605 · Ningbo Medsun Medical Co., Ltd. · FMK · Feb 20, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253605
Device Name
Safety Heel Lancet
Applicant
Ningbo Medsun Medical Co., Ltd.
Product Code
FMK · General, Plastic Surgery
Decision Date
Feb 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4850
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Safety Heel Lancet is a single use device with a sterile blade that is used by medical professionals to obtain a blood sample from the heel of neonates and infants.
Device Story
Safety Heel Lancet is a single-use, sterile, mechanical blood sampling device; used by medical professionals to obtain capillary blood from the heel of neonates and infants. Device consists of a housing, trigger mechanism, spring, and stainless steel blade. Operation involves manual activation of a trigger button, which ejects the blade via a spring-loaded mechanical structure to perform the incision; the blade then automatically and permanently retracts into the housing to prevent reuse and accidental sharps injury. Device is provided sterile via Gamma irradiation. Clinical benefit includes safe, controlled blood sampling with integrated sharps injury protection. Healthcare providers use the device to facilitate diagnostic blood collection; the resulting sample is used for clinical decision-making.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing including dimension, blade sharpness, cutting depth/width, hardness, and retractability. Sharps injury protection mechanism validated via clinical simulated use testing per ISO 23908:2011. Biocompatibility testing (ISO 10993-5, -10, -23, -11) and pyrogen testing (USP <151>) confirmed safety.
Indicated for neonates and infants requiring capillary blood collection from the heel. Prescription use only.
Regulatory Classification
Identification
The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.
Special Controls
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
*Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
*Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
*Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class III (premarket approval).(3)
*Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 20, 2026
Ningbo Medsun Medical Co., Ltd.
Ping Liu
Regulation Affairs Manager
# 298 Huangjipu Rd., Jiangbei,
Ningbo, Zhejiang 315031
China
Re: K253605
Trade/Device Name: Safety Heel Lancet
Regulation Number: 21 CFR 878.4850
Regulation Name: Blood Lancets
Regulatory Class: Class II
Product Code: FMK
Dated: January 23, 2026
Received: January 23, 2026
Dear Ping Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253605 - Ping Liu
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253605 - Ping Liu
Page 3
Sincerely,
Colin K. Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2026.02.20
09:31:10 -05'00'
Colin Kejing Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253605 | ? |
| Please provide the device trade name(s). | | ? |
| Safety Heel Lancet | | |
| Please provide your Indications for Use below. | | ? |
| The Safety Heel Lancet is a single use device with a sterile blade that is used by medical professionals to obtain a blood sample from the heel of neonates and infants. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Submission Documents-Safety Heel Lancet
K253605
# 510(k) Summary
## 1. Date Prepared
February 19, 2026
## 2. Submitter
Ningbo Medsun Medical Co., Ltd.
No. 298 Huangjipu Road, Jiangbei, 315031, Ningbo, P.R.China
Contact Person: Liu Ping, Regulation Affairs Manager
Tel: +86-574-86301887 / Fax: +86-574-86301887
Date Prepared: February 19, 2026
## 3. Device
Trade Name: Safety Heel Lancet
Common Name: Blood Lancets
Classification Name: Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature
Regulation Number: 21 CFR 878.4850
Regulatory Class: II
Product Code: FMK
Review Panel: General & Plastic Surgery
## 4. Predicate Device
Manufacturer: SteriLance Medical (Suzhou) Inc.
Device Name: Heel Incision Safety Lancet
510(k) Number: K210745
## 5. Device Description
Safety Heel Lancet, could be divided into Model D and Model E according to the shape of the trigger lock. The difference of Model D and Model E is that trigger lock of Model D doesn't need to be broke before use, trigger lock of Model E need to be broke before use.
Safety Heel Lancet is composed of housing, trigger lock, trigger button, blade, blade holder, slider, side cover and spring. The material of blade is X65Cr13 stainless steel, the
{5}
510(k) Submission Documents-Safety Heel Lancet
K253605
material of spring is 304 stainless steel, the other plastic parts are ABS.The blade is ejected and retracted by a mechanical structure to realize blood sampling.There is an integrated automatic inactivation system of the device to prevent device reuse. After the protection system is activated, the blade will be permanently retracted into the housing and cannot be restored to the activated state to avoid accidental injury to personnel.Therefore, the device cannot be re-used.It has sharps injury protection feature that it can reduce the occurrence of accidental needlesticks.
Safety Heel Lancet is sterilized by Gamma ray(e-beam) and for single use.It is designed to be used by medical professionals to obtain a blood sample from the heel of neonates and infants.
Safety Heel Lancet is sterile and non-toxic. The device is intended for prescription (Rx) only.
## 6.Indications for Use
The Safety Heel Lancet is a single use device with a sterile blade that is used by medical professionals to obtain a blood sample from the heel of neonates and infants.
## 7.Comparison of Technological Characteristics With the Predicate Device
| Description | Subject Device (Model D) | Subject Device (Model E) | Predicate Device (K210745) | Remark |
| --- | --- | --- | --- | --- |
| Product Code | FMK | FMK | FMK | Same |
| Regulation Number | 21CFR 878.4850 | 21CFR 878.4850 | 21CFR 878.4850 | Same |
| Indications for use | The Safety Heel Lancet is a single use device with a sterile blade that is used by medical professionals to obtain a blood sample from the heel of neonates and infants. | The Safety Heel Lancet is a single use device with a sterile blade that is used by medical professionals to obtain a blood sample from the heel of neonates and infants. | Heel Incision Safety Lancet is intended for the collection of capillary blood from the heel of newborn, preemie, and toddler. The lancet has equipped with safety protection features. | Similar (Note 1) |
| Prescription/ over-the counter use | Prescription use | Prescription use | Prescription use | Same |
| Target patient populations | Newborns and infants | Newborns and infants | Newborn, preemie,and toddler | Similar (Note 2) |
{6}
510(k) Submission Documents-Safety Heel Lancet
K253605
| Sterilization | Irradiation Sterilization SAL:10^{-6} | Irradiation Sterilization SAL:10^{-6} | Irradiation sterilized, SAL: 10^{-6} | Same |
| --- | --- | --- | --- | --- |
| Biocompatibility | Meet the requirements of ISO 10993 series standards | Meet the requirements of ISO 10993 series standards | Meet the requirements of ISO 10993 series standards | Same |
| Safety protection features | Yes | Yes | Yes | Same |
| Reuse or single use | Single use | Single use | Single use | Same |
| Composition | Housing, trigger lock, trigger button, blade, blade holder, slider, side cover and spring | Housing, trigger lock, trigger button, blade, blade holder, slider, side cover and spring | 1.Triggering button; 2.Safety button/Screw button;3.Spring;4.Cam 5.Lancet core;6.Shell 7.Blade;8.Swing arm | Similar (Note 3) |
| Materials | Blade:X65Cr13 stainless steel Spring:304 stainless steel Other plastic parts:ABS | Blade:X65Cr13 stainless steel Spring:304 stainless steel Other plastic parts:ABS | Blade: 304 stainless steel Shell: ABS Triggering button: ABS Safety button: ABS | Different (Note 4) |
| Blade cutting depth/width | Cutting depth: 0.65±0.25 mm 0.85±0.25 mm 1.00±0.25 mm 1.50±0.3 mm Cutting width: 1.50±0.5 mm 1.75±0.5 mm 2.50±0.5 mm 2.80±0.5 mm | Cutting depth: 0.65±0.25 mm 0.85±0.25 mm 1.00±0.25 mm 1.50±0.3 mm Cutting width: 1.50±0.5 mm 1.75±0.5 mm 2.50±0.5 mm 2.80±0.5 mm | Depth*Length: 0.65*1.40 mm, 0.85*1.75 mm 1.00*2.50 mm, 1.14*2.80 mm, 2.00*3.00 mm | Similar (Note 5) |
Note 1: The indications for use of the subject device is similar as the predicate device.
They are all used for collection of capillary blood from the heel. The used patients of the subject device are neonates and infants, the used patients of the predicate device are newborn,preemie and toddler, the two descriptions are different but the essence is consistent.Therefore,the difference will not affect the safety and effectiveness of the product.
Note 2: The target patient populations of the subject device is similar as the predicate device. They are explained in the indications for use mentioned above.Therefore,the difference will
3/5
{7}
510(k) Submission Documents-Safety Heel Lancet
K253605
not affect the safety and effectiveness of the product.
Note 3: The composition of the subject device is similar as the predicate device. Although the names are different, the actual composition is consistent. Therefore, the difference will not affect the safety and effectiveness of the product.
Note 4: The materials of the subject device is different as the predicate device. However, all the materials are known biocompatible materials that have been used in the subject device. The biocompatibility report proves the safety of the material. Therefore, the difference will not affect the safety and effectiveness of the product.
Note 5: The blade cutting depth/width of the subject device is similar as the predicate device. The cutting depth is in the range of 0.65-1.50 mm and the cutting width is in the range of 1.50-2.80 mm of the subject device. They are all within the range of the predicate device (depth: 0.65-2.00 mm, width: 1.40-3.00 mm). Therefore, the difference will not affect the safety and effectiveness of the product.
## 8. Performance Testing Summary
### Performance Test
All non-clinical bench testing performed on the subject device is to demonstrate the substantial equivalence to the predicate devices. Tests setup and execution are performed in accordance with corporate standard, which is formulated according to relevant standards and product characteristics. Results of the testing are demonstrating the compliance to the standards and matching the performance of the subject device to the predicate devices.
The following items of non-clinical performance bench testing are provided in support of the substantial equivalence determination.
- Dimension
- Appearance
- Compatibility
- Connection secureness
- Blade sharpness
- Blade cutting depth/width
- Blade surface roughness
- Blade hardness
- Blade elasticity
4/5
{8}
510(k) Submission Documents-Safety Heel Lancet
K253605
×Retractability
×Sterile
×Activation of the sharps injury protection mechanism-Trigger force 5.2 of ISO 23908
×The security of safety mode protection - resistance to external 5.3 a) of ISO 23908
×The safety of safety mode protection-minimizing accidental contact 5.3 b) of ISO 23908
×The safety protected by the safe mode - free fall resistance 5.3 c) of ISO 23908
## Biocompatibility Test
The subject device meets the requirements of ISO 10993 series standards, and the following items are evaluated: In vitro cytotoxicity, skin sensitization, intracutaneous reactivity, acute systemic toxicity and pyrogen. The results show that the product has good biocompatibility. The following biocompatibility data is provided in support of the substantial equivalence.
| Items | Requirements | Result |
| --- | --- | --- |
| In vitro cytotoxicity | ISO 10993-5: 2009 | Pass |
| Skin sensitization | ISO 10993-10: 2021 | Pass |
| Intracutaneous reactivity | ISO 10993-23:2021 | Pass |
| Acute Systemic toxicity | ISO 10993-11:2017 | Pass |
| Pyrogen | ISO 10993-11: 2017
USP <151> | Pass |
## Performance of Clinical Simulated Use Testing for Sharps Injury Protection
According to ISO 23908:2011 Sharps injury protection-Requirements and test methods-Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling and Guidance for Industry and FDA Staff-Medical Devices with Sharps Injury Prevention Features, we have completed the test of clinical simulated use testing for sharps injury protection.
## 9. Conclusions
Based on device comparison information and non-clinical bench testing, the subject device is substantially equivalent to the predicate devices (K210745).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.