Promisemed Blood Lancet, VeriFine Safety Lancet, VeriFine Mini-Safety Lancet

K192666 · Promisemed Hangzhou Meditech Co., Ltd. · FMK · Dec 2, 2019 · General, Plastic Surgery

Device Facts

Record IDK192666
Device NamePromisemed Blood Lancet, VeriFine Safety Lancet, VeriFine Mini-Safety Lancet
ApplicantPromisemed Hangzhou Meditech Co., Ltd.
Product CodeFMK · General, Plastic Surgery
Decision DateDec 2, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4850
Device ClassClass 2

Intended Use

It is intended for capillary blood sampling.

Device Story

Promisemed Blood Lancet is a manual, handheld, non-mechanical, scalpel-like instrument; VeriFine Safety and Mini-Safety Lancets are spring-loaded, single-use devices. All devices use a stainless steel needle to puncture skin for capillary blood sampling. Manual lancet requires twisting off a cap to expose the needle. Safety lancets are activated by pressing against the finger; a dual-spring mechanism drives the needle into the skin and automatically retracts it into the shield to prevent sharps injury. Used in clinical or home settings by healthcare providers or patients. Output is a small blood specimen for diagnostic testing. Benefits include ease of use, optimal blood flow, and reduced risk of accidental needle sticks.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included visual inspection, needle dimensions, chemical properties, bond strength, corrosion resistance, lancing compatibility, locking function, spring elasticity, percussive function, and penetration force. Biocompatibility testing performed per ISO 10993-1, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation and sensitization).

Technological Characteristics

Materials: Medical-grade stainless steel needles, polyethylene (PE) bodies/caps. Principle: Manual puncture (Promisemed) or spring-loaded automatic puncture/retraction (VeriFine). Form factor: Handheld, single-use. Sterilization: Terminal sterilization. Connectivity: None. Software: None.

Indications for Use

Indicated for capillary blood sampling in patients requiring small blood specimens. Suitable for use by healthcare providers or patients for self-use.

Regulatory Classification

Identification

The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.

Special Controls

*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature. (B) Handwashing instructions for the user before and after use of the device. (C) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device. (E) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Discard the entire device after use.” (B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.” (b) *Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device. (B) Handwashing instructions for the user before and after use of the device. (C) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device. (E) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Discard the entire device after use.” (B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.” (c) *Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that: (A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and (B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected. (vi) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device. (B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use. (C) Handwashing instructions for the user before and after use of the device. (D) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device. (F) Instructions for the safe disposal of the device. (G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades. (H) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.” (B) “Used lancet blades must be safely discarded after a single use.” (C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.” (d) *Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class III (premarket approval).(3) *Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 2, 2019 Promisemed Hangzhou Meditech Co., Ltd. % Wei-Shan Hsu Regulatory Manager Vee Care (Asia) Limited 17th Chung Pont Commercial Building, 300 Hennessy Road Hong Kong, China Re: K192666 Trade/Device Name: Promisemed Blood Lancet, VeriFine Safety Lancet. VeriFine Mini-Safety Lancet Regulation Number: 21 CFR 878.4800 Regulation Name: Manual Surgical Instrument for General Use Regulatory Class: Class I Product Code: FMK Dated: November 15, 2019 Received: November 15, 2019 Dear Wei-Shan Hsu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K192666 Device Name Promisemed Blood Lancet VeriFine Safety Lancet VeriFine Mini-Safety Lancet Indications for Use (Describe) It is intended for capillary blood sampling. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary #### 1 Date Prepared Nov 27, 2019 #### 2 Submitter's Information # Submission Sponsor: Promisemed Hangzhou Meditech Co., Ltd. No. 12, Longtan Road, Cangqian Street, Yuhang District, Hangzhou City,311121 Zhejiang, China Contact: Zearou Yang Telephone No.: +86 571 88772985 Fax No.:+86 571 88772985 Email: zearou.yang@promisemed.ca # Submission Correspondent: Vee Care (Asia) Limited 17F Chung Pont Commercial Building, 300 Hennessy Road, Hong Kong, China Contact: Wei-Shan Hsu E-mail: ws@vee.com.hk #### ന Trade Name, Common Name, Classification # Trade/Product Name: Promisemed Blood Lancet VeriFine Safety Lancet VeriFine Mini-Safety Lancet # Common Name: Blood Lancet ### Classification name: Lancet, Blood (21 CFR 878.4800, Product code FMK) Device Class: Class I {4}------------------------------------------------ #### ldentification of Predicate Device(s) ব K113513 PLANCET blood lancet K101145 SurgiLance® Safety Lancet #### Description of the Device 5 Promosemed Blood Lancet is a sterile, handheld, sharply-pointed, nonmechanical, scalpel-like instrument intended to be used by a healthcare provider or patient self to manually puncture the skin to obtain a small blood specimen. VeriFine Safety lancet and VeriFine Mini-Safety lancet is sterile, single use, spring loaded lancets designed for capillary blood sampling. These lancets are precision sharpened designed for maximum comfort and optimal blood flow. Safety lancets are designed to make taking a blood sample simple and easy. Safety lancets are activated when you press the device against your finger. Once activated the needle retracts into the body of the device which reduces the risk of injury as the result if an exposed needle. The first spring releases the needle into the skin and the second withdraws the needle back into the shield. Both single-use Promisemed Blood Lancet and VeriFine Safety Lancet/VeriFine Mini--Safety Lancet offers different gauge (needle diameter) options, to allow to choose the lancet which meets blood volume needs. - Intended Use 6 It is intended for capillary blood sampling. - Similarities and Differences of the Proposed Devices to 7 the Predicate Devices ### Promisemed Blood Lancet ### Manufacturing aspects: For the Promisemed Blood Lancet, stainless steel needle is fed into an injection molding machine to over-mold plastic material (polyethylene) {5}------------------------------------------------ forming a body and cap, encapsulating the stainless steel needles. Terminal sterilization process is performed to ensure sterility of an entire product. The manufacturing process of the predicate device, K113513 PLANCET blood lancet, comprises injection molding of plastic needle and plastic overmolding using polyethylene (exactly the same process as involved in Promisemed Blood Lancet). The plastic needle instead of stainless steel needle is used in PLANCET blood lancet. ### Design and Functionality aspects: The Promisemed Blood Lancet comprises a stainless steel needle encapsulated with a plastic body and cap, the cap is twisted off to expose the needle for use. The predicate device, K113513 PLANCET blood lancet, has the exact same design and functionality as compared with Blood Lance and Safety Lancet. The difference will be instead of twisting off the cap to expose the needle, the predicate device requires the end user to pull off the cap. ### Material aspects: The Promisemed Blood Lancet has a needle that is made of stainless steel and a body and a cap that are made of polyethylene (PE). The stainless steel material is bio-compatible. Both the needle and the body/cap of the predicate device, K113513 PLANCET blood lancet, are made of polyethylene (PE). ### VeriFine Safety Lancet and VeriFine Mini-Safety Lancet ### Design and Functionality aspects: Both VeriFine Safety Lancet/VeriFine Mini-Safety Lancet and its predicate device, SurgiLance® Safety Lancet, are spring-loaded lancet and have the same basic technology characteristics for a lancet with sharps injury prevention. They are intended for piercing the skin and the indications for use are the same. Material aspects: {6}------------------------------------------------ The materials are comparable in that the needles all use medical grade stainless steel and the rest part are made of plastics materials. VeriFine Safety Lancet/VeriFine Mini-Safety Lance utilizes some of the same materials, specifically the use of medical grade stainless steel for the lancet needles, but may use different types or grades of plastics for the Shield, hub, caps, and triggers. All the materials are known biocompatible materials that have been used in lancets or other similar medical devices. #### Performance Testing Summary 8 The bench testing performed verifies that the performance of the subject devices are substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include: - Visual Inspection - Needle Dimensions - Chemical properties - Bond between lancet body and needle - Resistance to corrosion of the needle - Lancing device compatibility test (Promisemed Blood Lancet only) - Locking function - Spring elasticity* - · Percussive function* - Penetrate force* - (* VeriFine Safety Lancet and VeriFine Mini-Safety Lancet only) - Biocompatibility - a. ISO 10993-1:2009 Biological Evaluation of Medical Devices -- Part 1: Evaluation and testing within a risk management process - b. ISO 10993-5:2009 Biological Evaluation of Medical Devices -- Part 5: Tests for in Vitro Cytotoxicity - c. ISO 10993-10:2010 Biological Evaluation of Medical Devices -- Part 10: Tests for Irritation and Skin Sensitization {7}------------------------------------------------ The testing performed verifies that the performance of the subject devices is substantially equivalent in terms of critical performance characteristics to the predicate device. Cytotoxicity, Sensitization, Irritation, were performed to demonstrate biocompatibility of the patient contacting materials. Overall, the results are comparable to the predicate and support a determination of substantial equivalence. #### Conclusion の Promisemed Blood Lancet and VeriFine Safety Lancet/VeriFine Mini-Safety Lancet has the same intended use and technological characteristics as the predicate. Test results demonstrate that the subject devices meet their intended use and performs as well as or better than the legally marketed predicate device. It is for these reasons that the subject devices can be found substantially equivalent.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...