Browse hierarchy: [General, Plastic Surgery (SU)](/submissions/SU) → [Subpart E — Surgical Devices](/submissions/SU/subpart-e%E2%80%94surgical-devices) → [21 CFR 878.4850](/submissions/SU/subpart-e%E2%80%94surgical-devices/878.4850) → QRK — Single Use Only Blood Lancet Without An Integral Sharps Injury Prevention Feature

# QRK · Single Use Only Blood Lancet Without An Integral Sharps Injury Prevention Feature

_General, Plastic Surgery · 21 CFR 878.4850 · Class 2_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK

## Overview

- **Product Code:** QRK
- **Device Name:** Single Use Only Blood Lancet Without An Integral Sharps Injury Prevention Feature
- **Regulation:** [21 CFR 878.4850](/submissions/SU/subpart-e%E2%80%94surgical-devices/878.4850)
- **Device Class:** 2
- **Review Panel:** [General, Plastic Surgery](/submissions/SU)

## Identification

The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.

## Classification Rationale

Class II (special controls) for paragraphs (a), (b), and (c). Class III (premarket approval) for paragraph (d).

## Special Controls

*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
*Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
*Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
*Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class III (premarket approval).(3)
*Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Recent Cleared Devices (15 of 15)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K250016](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K250016.md) | droplet® personal lancets | Htl-Strefa S.A | Jul 1, 2025 | SESE |
| [K244031](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K244031.md) | Disposable Blood Lancet (Soft Pro); Disposable Blood Lancet (Softsure); Disposable Blood Lancet (Softsure Pro); Disposable Blood Lancet (Softsense) | SteriLance Medical (Suzhou), Inc. | Feb 26, 2025 | SESE |
| [K242622](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K242622.md) | Sterile Lancets for Single Use | Ningbo Caremed Medical Products Co., Ltd. | Oct 28, 2024 | SESE |
| [K241810](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K241810.md) | MICROLET®NEXT Lancet | Ascensia Diabetes Care U.S., Inc. | Aug 15, 2024 | SESE |
| [K232330](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K232330.md) | Lancing System | Ningbo Medsun Medical Co., Ltd. | Jan 4, 2024 | SESE |
| [K230712](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K230712.md) | SG Lanset I, SG Lancets, Soft Lancets | Sewon Medical Co. | Dec 6, 2023 | SESE |
| [K230310](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K230310.md) | STAT Medical Device Lancing System | Stat Medical Devices | Sep 8, 2023 | SESE |
| [K230759](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K230759.md) | SafeLan® (2 models/SafeLan 26G, SafeLan 30G), SafeLan®-Pro (1 model/SafeLan®-Pro) | Bosungmeditech Co., Ltd. | Jun 13, 2023 | SESE |
| [K222617](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K222617.md) | EasyTouch Lancing Device | Stat Medical Devices | Mar 21, 2023 | SESE |
| [K223815](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K223815.md) | ONE DROP Lancing Device | Stat Medical Devices | Feb 15, 2023 | SESE |
| [K223314](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K223314.md) | Disposable Blood Lancets | Huaian Hening Medical Instruments Co., Ltd. | Jan 25, 2023 | SESE |
| [K223480](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K223480.md) | Medipoint Blood Lancets | Medipoint Holdings, LLC | Jan 13, 2023 | SESE |
| [K221383](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K221383.md) | MedtFine Blood Lancet | Shanghai Carelife International Trading Co. , Ltd. | Sep 21, 2022 | SESE |
| [K221507](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K221507.md) | Disposable Blood Lancet | SteriLance Medical (Suzhou), Inc. | Aug 31, 2022 | SESE |
| [K221433](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK/K221433.md) | Facet 28G Universal Lancet | Facet Technologies, LLC | Jul 15, 2022 | SESE |

## Top Applicants

- Stat Medical Devices — 3 clearances
- SteriLance Medical (Suzhou), Inc. — 2 clearances
- Ascensia Diabetes Care U.S., Inc. — 1 clearance
- Bosungmeditech Co., Ltd. — 1 clearance
- Facet Technologies, LLC — 1 clearance

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QRK)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com