SoftSpot

K212045 · Pediametrix, Inc. · QQO · Sep 28, 2021 · General, Plastic Surgery

Device Facts

Record IDK212045
Device NameSoftSpot
ApplicantPediametrix, Inc.
Product CodeQQO · General, Plastic Surgery
Decision DateSep 28, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4800
Device ClassClass 1
AttributesSoftware as a Medical Device, Pediatric

Intended Use

SoftSpot™ is intended to be used for cranial measurement for infants from 1 month up to 2 years of age in conjunction with other clinical methods. SoftSpot™ is not intended to be used as a tool for preparing a helmet.

Device Story

SoftSpot is a software platform for cranial measurement in infants. Input: top-view videos/photos of infant head captured by parents or clinicians via smartphone/tablet. Process: data uploaded to AWS cloud; analyst uses Windows-based software to convert video to photos, standardize images, select 5 optimal frames, and perform segmentation of head contour and nose tip. Algorithm calculates Cranial Index (CI) and Cephalic Vault Asymmetry Index (CVAI) based on geometric ratios. Output: PDF report sent to physician and parent. Usage: point-of-care or home data collection; clinical analysis at remote facility. Clinical decision-making: physician reviews report to assess cranial shape and determine next steps. Benefit: provides objective cranial measurements comparable to manual calipers for monitoring infants.

Clinical Evidence

Prospective study (n=25, age 3-12 months) compared SoftSpot to manual caliper measurements; Spearman correlation 0.96 (CI) and 0.94 (CVAI); MAE 1.48±0.92 (CI) and 1.44±0.85 (CVAI). Retrospective study (n=100, age 3-20 months) compared SoftSpot 2D-rendered images to 3D head scans; Spearman correlation 0.93 (CI) and 0.94 (CVAI); MAE 2.17±1.53 (CI) and 1.62±1.26 (CVAI).

Technological Characteristics

Software-based cranial measurement tool. Operates on smartphone/tablet/PC; cloud-based processing via AWS. Uses image segmentation and geometric calculation algorithms to derive CI and CVAI from 2D images. No physical materials of construction. Standalone software application.

Indications for Use

Indicated for cranial measurement in infants aged 1 month to 2 years. Not for use as a tool for preparing a helmet.

Regulatory Classification

Identification

A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION". September 28, 2021 PediaMetrix, Inc. Joshua Levin Regulatory Advisor 155 Gibbs Street, Suite 537 Rockville, Maryland 20850 Re: K212045 Trade/Device Name: SoftSpot™ Regulation Number: 21 CFR 878.4800 Regulation Name: Manual surgical instrument for general use Regulatory Class: Class I Product Code: QQO Dated: June 26, 2021 Received: June 30, 2021 Dear Joshua Levin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Jay R. Gupta -S Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212045 Device Name SoftSpot™ Indications for Use (Describe) SoftSpot™ is intended to be used for cranial measurement for infants from 1 month up to 2 years of age in conjunction with other clinical methods. SoftSpot™ is not intended to be used as a tool for preparing a helmet. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"><span>☑</span><span>Prescription Use (Part 21 CFR 801 Subpart D)</span></div> | <div style="display:flex; align-items:center;"><span>☐</span><span>Over-The-Counter Use (21 CFR 801 Subpart C)</span></div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary #### A. Basic Information: The submitter's name: PediaMetrix Inc Address: 155 Gibbs St, Suite 537, Rockville, MD 20850 Telephone number: 240-670-0171 Contact: Joshua Levin, PhD. #### The date the summary was prepared: 09-26-2021 #### B. 510(k) Number: K212045 - C. Device Proprietary and Established Names: SoftSpot™ - D. Regulatory Information: - 1. Classification Name: Manual surgical instrument for general use (21 CFR 878.4800) - 2. Classification: Class I - 3. Product code: QQO, Cranial measurement software - 4. Panel: Neurology ### E. Indications for Use: - 1. Indications for use: SoftSpot™ is intended to be used for cranial measurement for infants from 1 month up to 2 years of age in conjunction with other clinical methods. SoftSpot™ is not intended to be used as a tool for preparing a helmet - 2. Special conditions for use statement(s): For prescription use only. ## F. Predicate Device: - 1. Predicate device name(s): Holtain-Kahn Caliper - 2. Predicate 510(k) number(s): 510(k) Exempt - 3. Product Code: FTY (Tape, Measuring, Rulers and Calipers) - 4. Classification Regulation: 21 CFR 878.4800 #### G. Device Description: SoftSpot™ is a software that runs on a smartphone, tablet, PC, etc. and uses top view pictures of infant's head to measure cranial index (CI) and cranial vault asymmetry index (CVAI) similar to a mechanical caliper used by physicians which is also known as craniometer. {4}------------------------------------------------ After the photo is taken, the head contour is identified and then, based on the current standards of measuring infants' head shape, characterizing parameters including the ratio of width over length (known as Cranial Index or CI) and the ratio of difference in diagonals over the larger one (known as Cephalic Vault Asymmetry Index or CVAI) are calculated (see Test Principle below). The SoftSpot platform allows different users to upload or capture top view videos and photos of the infant's head. The user can be a physician/nurse (during well visits) or parents/guardians (between the visits). Once data are collected, they are sent to PediaMetrix Amazon Web Services (AWS) cloud. At PediaMetrix, the data analyst will use a windows-based software to download and analyze the data. The analysis includes conversion of the video into photos, standardization of the images (3x4, all nose up), selection of 5 good images, segmentation of the head contour and nose tip. Then, the software will calculate Cl and CVAI. Finally, a pdf report is generated with the results and sample top view photos and sent to the physician's account. Parents will also be notified when the results are ready. Then, the parent visits the physician (either in person or online) to discuss the results and next steps. #### H. Measurand: Cranial Index (CI) and Cephalic Vault Asymmetry Index (CVAI) #### l. Measurement Principle: The SoftSpot™ app calculates characterizing parameters including the ratio of width over length (known as Cranial Index or CI) and the ratio of difference in diagonals over the larger one (known as Cephalic Vault Asymmetry Index or CVAI) are calculated. $$CI = \left(\frac{AP}{SD}\right) \times 100,$$ $$\text{A} \qquad \text{nD}$$ $$CVAI = \frac{AA - BB}{\max\left(AA, BB\right)} \times 100$$ - . AP is defined on the head contour by the line that goes through the nose tip and center of the head contour. - . SD is defined by the line perpendicular at the midpoint of AP and intersecting the headcontour. - AA and BB are defined by the intersection of the head contour and the lines that go through the center of the head contour at +/-40 degrees relative to AP. {5}------------------------------------------------ | Predicate Comparison - Similarities | | | |-------------------------------------|------------------------------------------------------------------------------|-----------------------------------| | Item | Device<br>SoftSpot | Predicate<br>Holtain-Kahn Caliper | | Regulation | 21 CFR 878.4800 | 21 CFR 878.4800 | | Product Code | TBD | FTY | | Classification | Class I | Class I | | Intended<br>Use/Indications for Use | Used for cranial measurements in<br>conjunction with other clinical methods. | Same as SoftSpot | | Analytes measured | Cranial measurements | Same as SoftSpot | #### J. Comparison of Technological Characteristics with the Predicate Device: | Predicate Comparison - Differences | | | |------------------------------------|----------------------------------------------------------------------------------|--------------------------------------------| | Item | Device<br>SoftSpot | Predicate<br>Holtain-Kahn Caliper | | Intended Use<br>Population | Infants from 1 month to 2 years of age | Not specified | | Measurement method | Software that uses pictures of the head to provide cranial measurements | Physical measurement of cranial parameters | | Prescription/OTC | For prescription use; images may be collected by clinicians or parents/guardians | Not specified | #### K. Performance Data: #### 1. Clinical Studies: #### a. Prospective For the prospective evaluation of our measurement tool, PediaMetrix compared measurements taken with the SoftSpot™ app to the measurements taken by clinicians using a standard caliper similar to the predicate device. A comparative study was performed using 25 infants (age range 3 to 12 months). Data were collected by a physical therapist in a plagiocephaly clinic at the University of Pittsburgh Children Hospital from their typical patient population. The Spearman correlation coefficient was 0.96 and 0.94 for Cl and CVAI, respectively. The mean absolute error (MAE) and standard deviation (STD) were $1.48 \pm 0.92$ and $1.44 \pm 0.85$ units for CI and CVAI, respectively. ## b. Retrospective We also compared SoftSpot measurements of 2D rendered images (from 3D head scans) with direct measurements from 3D scans of the head from patients with cranial deformities in a retrospective study using 100 infants' head measurements along with their 3D scans (in .stl format). Data were collected from patients aged 3-20 months with deformational plagiocephaly and brachycephaly (DPB) in a helmet clinic at Singapore. The Spearman correlation coefficient {6}------------------------------------------------ was 0.93 and 0.94 for Cl and CVAI, respectively. The mean absolute error (MAE) and standard deviation (STD) were $2.17 \pm 1.53$ and $1.62 \pm 1.26$ units for CI and CVAI, respectively. #### Conclusions: L. The proposed clinical studies demonstrated that the device can function as intended for the intended population (infants 1 months to 2 years of age) and that it can be used to measure Cl and CVAI with accuracy comparable to the predicate device. This supports that the device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...