K223263 · 3M · QFC · Feb 13, 2023 · General, Plastic Surgery
Device Facts
Record ID
K223263
Device Name
Prevena Plus 125 Therapy Unit
Applicant
3M
Product Code
QFC · General, Plastic Surgery
Decision Date
Feb 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4783
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Prevena Plus 125 Therapy Unit is indicated for use with both the Prevena™ Dressings and compatible V.A.C.® Dressings. The Prevena Plus 125 Therapy Unit when used with the Prevena™ Dressings (Prevena Plus Incision Management System), manages the environment of closed surgical incisions and removes fluid away from the surgical incision via the application of -125mmHg continuous negative pressure. When used with legally marketed Prevena Dressings for up to seven days, Prevena Plus 125 Therapy Units are intended to aid in reducing the incidence of seroma and, in patients at high risk for post-operative infections, aid in reducing the incidence of superficial surgical site infection in Class I and Class II wounds. The Prevena ™ Plus 125 Therapy Unit, when used with compatible V.A.C.® Dressings on open wounds (Prevena Plus Negative Pressure Wound Therapy System), is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts.
Device Story
Prevena Plus 125 Therapy Unit is a negative pressure wound therapy (NPWT) pump; delivers continuous -125mmHg pressure. Used with Prevena Dressings for closed surgical incisions or compatible V.A.C. Dressings for open wounds. Operates in acute, extended, and home care settings; managed by healthcare providers or patients. Input: negative pressure maintenance; Output: fluid/exudate removal from wound site. Benefits: reduced seroma, reduced surgical site infection, promoted granulation tissue formation, reduced edema, wound bed preparation for closure. Device functions as a portable suction unit.
Clinical Evidence
Bench testing only. Testing demonstrated the Prevena Plus Therapy Unit maintains negative pressure within specifications when used with small and medium V.A.C. Dressings. Human factors evaluation was conducted. No clinical data was required to demonstrate equivalence.
Technological Characteristics
Negative pressure pump; delivers -125mmHg continuous pressure. Compatible with Prevena and V.A.C. dressings. Used in acute, extended, and home care settings. Duration of use up to 7 or 14 days.
Indications for Use
Indicated for patients with closed surgical incisions (Class I/II wounds) to reduce seroma and superficial surgical site infection risk, and patients with open wounds (chronic, acute, traumatic, subacute, dehisced, burns, ulcers, flaps, grafts) to promote healing via secondary/tertiary intention.
Regulatory Classification
Identification
A negative pressure wound therapy device for reduction of wound complications is a powered suction pump intended for wound management and reduction of wound complications via application of negative pressure to the wound, which removes fluids, including wound exudate, irrigation fluids, and infectious materials. This device type is intended for use with wound dressings classified under § 878.4780. This classification does not include devices intended for organ space wounds.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical data must demonstrate that the device performs as intended under anticipated conditions of use and evaluate the following:
(i) Wound complication rates; and
(ii) All adverse events.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the labeled shelf life.
(5) Usability testing must demonstrate that intended users can correctly use the device, based solely on reading the instructions for use.
(6) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested in a worst-case scenario for the intended use life:
(i) Ability to maintain pressure levels at the wound site under a worst-case scenario for the intended use life;
(ii) Fluid removal rate consistent with the wound types specified in the indications for use; and
(iii) Timely triggering of all alarms.
(7) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of the device.
(8) Software verification, validation, and hazard analysis must be performed.
(9) Labeling must include the following:
(i) Instructions for use;
(ii) A summary of the device technical specifications, including pressure settings, modes (
*e.g.,* continuous or intermittent), alarms, and safety features;(iii) Compatible components and devices;
(iv) A summary of the clinical evidence for the indications for use;
(v) A shelf life for sterile components; and
(vi) Use life and intended use environments.
(10) For devices intended for use outside of a healthcare facility, patient labeling must include the following:
(i) Information on how to operate the device and its components and the typical course of treatment;
(ii) Information on when to contact a healthcare professional; and
(iii) Use life.
K123878 — PREVENA INCISION MANAGEMENT SYSTEM WITH CUSTOMIZABLE DRESSING · Kci USA, Inc. (Kinetic Concepts, Inc.) · Jan 15, 2013
DEN180013 — PREVENA 125 and PREVENA PLUS 125 Therapy Units · Kci USA, Inc. · Apr 19, 2019
K120033 — ACTIVAC THERAPY UNIT AND INFOVAC THERAPY UNIT, VAC ATS THERAPY UNIT AND VAC FREEDOM THERAPY UNIT, VAC VIA THERAPY UNIT · Kci USA, Inc. · May 31, 2012
K214112 — Invia Ease Negative Pressure Wound Therapy (NPWT) System · Medela AG · Jul 26, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 13, 2023
3M Teri Feelev Sr. Regulatory Associate 6203 Farinon Dr. San Antionio. Texas 78247
Re: K223263
Trade/Device Name: Prevena Plus 125 Therapy Unit Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP. OFC Dated: October 12, 2022 Received: October 24, 2022
Dear Teri Feeley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Julie A. Morabito -S
Julie A. Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223263
Device Name Prevena Plus 125 Therapy Unit
#### Indications for Use (Describe)
Prevena Plus 125 Therapy Unit is indicated for use with both the Prevena™ Dressings and compatible V.A.C.® Dressings.
The Prevena Plus 125 Therapy Unit when used with the Prevena™ Dressings (Prevena Plus Incision Management System), manages the environment of closed surgical incisions and removes fluid away from the surgical incision via the application of -125mmHg continuous negative pressure. When used with legally marketed Prevena Dressings for up to seven days, Prevena Plus 125 Therapy Units are intended to aid in reducing the incidence of seroma and, in patients at high risk for post-operative infections, aid in reducing the incidence of superficial surgical site infection in Class I and Class II wounds.
The Prevena ™ Plus 125 Therapy Unit, when used with compatible V.A.C.® Dressings on open wounds (Prevena Plus Negative Pressure Wound Therapy System), is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| <span> <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) </span> | <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 1. Submitter Information:
3M Health Care Business Group 6203 Farinon Dr. San Antonio, TX 78249
Contact Person: Teri Feeley Email: tfeeley@mmm.com Phone: 210-459-1952 Facsimile: 210-255-6727 Date Prepared: 08 Feb 2023
### 2. Device
Trade/Device Name: Prevena™ Plus 125 Therapy Unit Regulation Number: 21 CFR 878.4783 Regulation Name: Negative pressure wound therapy device for reduction of wound complications Regulatory Class: Class II Product Code Primary: QFC, OMP Dated: October 12, 2022 Received: October 24, 2022
#### 3. Predicate Device
Predicate Device: Prevena™ Plus 125 Therapy Unit (DEN180013) Predicate Device: Prevena™ Plus 125 Therapy Unit (K180855) Reference Device: V.A.C. VIA NPWT System (K173447)
#### 4. Device Description
The Prevena™ Plus 125 Therapy Unit may be used part of as either of these systems:
#### A) Prevena Incision Management System Consist of:
- Prevena™ Therapy Unit .
- Prevena 150mL Canister & adaptor .
- Prevena Dressing kit ●
{4}------------------------------------------------
## B) Prevena Plus Negative Pressure Wound Therapy System Consist of:
- Prevena™ Plus 125 Therapy Unit .
- Prevena 150mL Canister & adaptor .
- V.A.C. Dressing kit ●
The Prevena™ Plus Incision Management System (Prevena™ Plus Therapy Unit when used with Prevena Dressings) is designed for use over linear, non-linear, intersecting incisions.
The Prevena™ Plus Negative Pressure Wound Therapy System (Prevena™ Plus Therapy Unit when used with V.A.C. Dressings) is designed for use in a variety of open wound types such as: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts.
### 5. Intended Use / Indications for Use
Prevena™ Plus 125 Therapy Unit is indicated for use with both the Prevena Dressings and compatible V.A.C.® Dressings.
The Prevena™ Plus 125 Therapy Unit when used with the Prevena Dressings (Prevena™ Plus Incision Management System), manages the environment of closed surgical incisions and removes fluid away from the surgical incision via the application of -125mmHg continuous negative pressure. When used with legally marketed Prevena Dressings for up to seven days, Prevena™ Plus 125 Therapy Units are intended to aid in reducing the incidence of seroma and, in patients at high risk for post-operative infections, aid in reducing the incidence of superficial surgical site infection in Class I and Class II wounds.
The Prevena™ Plus 125 Therapy Unit, when used with compatible V.A.C.® Dressings on open wounds (Prevena™ Plus Negative Pressure Wound Therapy System), is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts.
### 6. Comparison of Technological Characteristics
The Prevena Plus 125 Therapy Units are negative pressure pumps that deliver -125mmHg continuously for up to 7 or 14 days.
{5}------------------------------------------------
The Prevena Plus 125 Therapy Units may be used with Prevena Dressings for use over linear, non-linear, intersecting incisions on various anatomical locations as determined at the discretion of the healthcare provider.
Alternatively, the Prevena Plus 125 Therapy Units may be used with previously cleared small or medium V.A.C. Dressings for use in a variety of open wounds such as: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts.
At a high level, the subject and predicate & reference devices are based on the following same technological elements:
- Intended use
- Indicated for closed incisions when used with Prevena Dressings .
- · Indicated for open wounds when used with V.A.C. Dressings
- · Use environment is acute, extended and home care settings
- · Intended for use with Prevena Dressings and V.A.C. Dressings
The subject device indications for use, technological characteristics and principles of operation are substantially equivalent to the predicates.
{6}------------------------------------------------
A table comparing the key features of the subject and predicate devices is provided below.
| Characteristic | Subject Device: | Predicate Device: | Predicate Device: | Reference Device: V.A.C. |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Prevena ™ Plus Therapy Unit | Prevena ™ Plus Therapy Unit; DEN180013 | Prevena ™ Plus Therapy Unit; K180855 | VIA NPWT System, K173447 |
| Indications for Use | Identical to predicate DEN180013 & reference Device<br><br>Prevena Plus 125 Therapy Unit is indicated for use with both the Prevena ™ Dressings and compatible V.A.C.® Dressings.<br><br>The Prevena Plus 125 Therapy Unit when used with the Prevena™ Dressings (Prevena Plus Incision Management System), manage the environment of closed surgical incisions and remove fluid away from the surgical incision via the application of -125mmHg continuous negative pressure. When used with legally marketed Prevena Dressings for up to seven days, Prevena Plus 125 Therapy Units are intended to aid in reducing the incidence of seroma and, in patients at high risk for post-operative infections, aid in reducing the incidence of superficial surgical site infection in Class I and Class II wounds.<br><br>The Prevena ™ Plus 125 Therapy Unit, when used with | Prevena Plus 125 Therapy Units manage the environment of closed surgical incisions and remove fluid away from the surgical incision via the application of -125mmHg continuous negative pressure. When used with legally marketed compatible dressings, Prevena Plus 125 Therapy Units are intended to aid in reducing the incidence of seroma and, in patients at high risk for post-operative infections, aid in reducing the incidence of superficial surgical site infection in Class I and Class II wounds. | Prevena ™ Plus Incision Management Systems is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy. | The V.A.C.VIA Therapy System is an integrated wound management system for use in acute, extended and home care settings.<br><br>When used on closed surgical incisions, the V.A.C. VIA ™ is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy. |
| Characteristic | Subject Device: | Predicate Device: | Predicate Device: | Reference Device: V.A.C. |
| | Prevena™ Plus Therapy Unit | Prevena™ Plus Therapy | Prevena™ Plus Therapy | VIA NPWT System, |
| | | Unit; DEN180013 | Unit; K180855 | K173447 |
| | compatible V.A.C.® Dressings on<br>open wounds (Prevena™ Plus<br>Negative Pressure Wound<br>Therapy System), is intended to<br>create an environment<br>that promotes wound healing by<br>secondary or<br>tertiary (delayed primary)<br>intention by preparing the<br>wound bed for closure, reducing<br>edema, promoting granulation<br>tissue formation and perfusion,<br>and by removing<br>exudate and infectious<br>material. Open wound<br>types include: chronic, acute,<br>traumatic, subacute and dehisced<br>wounds, partial-thickness burns,<br>ulcers (such as diabetic,<br>pressure or venous insufficiency),<br>flaps and grafts. | | | |
| Characteristic | Subject Device:<br>PrevenaTM Plus Therapy Unit | Predicate Device:<br>PrevenaTM Plus Therapy<br>Unit; DEN180013 | Predicate Device:<br>PrevenaTM Plus Therapy<br>Unit; K180855 | Reference Device: V.A.C.<br>VIA NPWT System,<br>K173447 |
| | | | | When used on open wounds,<br>the V.A.C.VIATM Therapy<br>System is intended to create<br>an environment that promotes<br>wow1d healing by secondary<br>or tertiary (delayed primary)<br>intention by preparing the<br>wound bed for closure,<br>reducing edema, promoting<br>granulation tissue formation<br>and perfusion, and by<br>removing exudates and<br>infectious material. Open<br>wound types include: chronic,<br>acute, traumatic, subacute and<br>dehisced wounds, partial-<br>thickness burns, ulcers (such<br>as diabetic, pressure, or<br>venous insufficiency), flaps<br>and grafts. |
| Indicated Wound<br>Types | Identical to Reference Device | ● closed incisions | ● closed incisions | ● chronic<br>● acute<br>● traumatic<br>● subacute<br>● dehisced wounds<br>● partial-thickness<br>burns |
| Characteristic | Subject Device:<br>Prevena™ Plus Therapy Unit | Predicate Device:<br>Prevena™ Plus Therapy<br>Unit; DEN180013 | Predicate Device:<br>Prevena™ Plus Therapy<br>Unit; K180855 | Reference Device: V.A.C.<br>VIA NPWT System,<br>K173447 |
| | | | | ulcers (such as diabetic pressure or venous insufficiency) flaps grafts closed incisions |
| Prevena Dressings | Identical to Predicates<br>DEN180013 & K180855 | Prevena Peel & Place Dressings Prevena Customizable Dressing Prevena Restor Dressings | Prevena Peel & Place Dressings Prevena Customizable Dressing Prevena Restor Dressings | V.A.C. VIA Granufoam Spiral Dressings Dermatac Drape with V.A.C. Granufoam Dressing |
| V.A.C. Dressing | Similar to Reference | N/A | N/A | V.A.C. VIA Granufoam Spiral Dressings Dermatac Drape with V.A.C. Granufoam Dressing |
| Use<br>environment/Care<br>Setting of dressing<br>kit | Identical to Reference device | operating room/surgery center<br>and then may transition home<br>with the patient | operating room/surgery center<br>and then may transition home<br>with the patient | acute, extended or homecare<br>setting |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
#### 7. Performance Data
- a. Summary of non-clinical tests conducted for determination of substantial equivalence:
- Prevena™ Plus Negative Pressure Wound Therapy System negative pressure test ● demonstrates the Prevena Plus Therapy Unit when used with small and medium V.A.C. Dressing as part of a system, maintains negative pressure within specifications.
- Human factors evaluation ●
In all instances, the Prevena™ Plus 125 Therapy Unit functioned as intended with V.A.C. Dressings and all test results observed were as expected.
- b. Clinical and Pre-clinical testing were not necessary to demonstrate equivalence.
#### 8. Conclusions
The subject device is substantially equivalate as the predicate devices. The subject device's Intended Use, indications for use, fundamental technology and principles of operation are unchanged compared to their respective predicates (DEN180013 & K180855).
The proposed modification to add the "open wound" indication to the subject devices is identical to the reference device (K173447) and does not raise different questions of safety or effectiveness of the subject device. The performance data demonstrates that the Prevena Plus 125 Therapy Unit when used with small and medium V.A.C. Dressing are substantially equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.