← Product Code [QAU](/submissions/SU/subpart-e%E2%80%94surgical-devices/QAU) · K222481

# PuraStat (K222481)

_3-D Matrix, Inc. · QAU · Feb 28, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QAU/K222481

## Device Facts

- **Applicant:** 3-D Matrix, Inc.
- **Product Code:** [QAU](/submissions/SU/subpart-e%E2%80%94surgical-devices/QAU.md)
- **Decision Date:** Feb 28, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4456
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

PuraStat is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic therapy to prevent post procedure bleeding, and for primary non-variceal gastrointestinal (GI) bleeding. PuraStat is not indicated for arterial Forrest 1a bleeding.

## Device Story

Sterile, synthetic peptide hydrogel (RADA-16) in water; provided in prefilled cyclo-olefin polymer syringes. Delivered endoscopically via catheter to bleeding site. Upon contact with physiological fluids (blood), peptides self-assemble into transparent scaffold/mechanical barrier; blocks blood flow without volume expansion. Used in clinical endoscopy settings by physicians. Acts as primary, adjunct, bridge, or rescue hemostat. Benefits include stabilization of bleeding sites, reduction of rebleeding risk, and potential avoidance of surgery.

## Clinical Evidence

Prospective, multicenter, open-label study (Branchi et al., 2021) of 111 patients (age >18) with acute non-variceal GI bleeding. Primary endpoint: procedural hemostasis success. Results: 94% initial hemostatic success for primary application; 75% success as secondary therapy. 30-day rebleeding rate 16%. No adverse events reported. Bench and biocompatibility data referenced from predicate.

## Technological Characteristics

Synthetic 16-amino acid (RADA-16) oligopeptide in sterile water. Cyclo-olefin polymer syringe, HDPE plunger, butyl rubber cap/gasket. Self-assembling hydrogel scaffold. Delivered endoscopically via polypropylene adapter and catheter. Terminally sterilized. Non-animal/non-plant derived. No preservatives.

## Regulatory Identification

A hemostatic device for intraluminal gastrointestinal use is a prescription device that is endoscopically applied to the upper and/or lower gastrointestinal tract and is intended to produce hemostasis via absorption of fluid or by other physical means.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Performance data must support the sterility and pyrogenicity of the device.
(3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(4) In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The testing must evaluate the following:
(i) The ability to deliver the hemostatic material to the bleeding site;
(ii) The ability to achieve hemostasis in a clinically relevant model of gastrointestinal bleeding; and
(iii) Safety endpoints, including thromboembolic events, local and systemic toxicity, tissue trauma, gastrointestinal tract obstruction, and bowel distension and perforation.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
(i) Materials characterization of all components must demonstrate the device meets established specifications, which must include compositional identity and purity, characterization of impurities, physical characteristics, and reactivity with fluids.
(ii) Performance testing must demonstrate the mechanical integrity and functionality of the system used to deliver the device and demonstrate the device meets established specifications, including output pressure for propellant-based systems.
(6) Labeling must include:
(i) Information identifying and explaining how to use the device and its components; and
(ii) A shelf life.

## Predicate Devices

- PuraStat-GI ([K210098](/device/K210098.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 28, 2023

3-D Matrix, Inc. % Stephen Rhodes Principal Streamline Regulatory 3502 Dundee Driveway Chevy Chase, Maryland 20815

Re: K222481

Trade/Device Name: PuraStat Regulation Number: 21 CFR 878.4456 Regulation Name: Hemostatic Device For Intraluminal Gastrointestinal Use Regulatory Class: Class II Product Code: QAU Dated: February 15, 2023 Received: February 15, 2023

Dear Stephen Rhodes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Tek N. Tek N. l amichhane -S Date: 2023.02.27 22:27:19 Lamichhane -S for Deborah Fellhauer RN, BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222481

Device Name PuraStat

### Indications for Use (Describe)

PuraStat is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic therapy to prevent post procedure bleeding, and for primary non-variceal gastrointestinal (GI) bleeding. PuraStat is not indicated for arterial Forrest 1a bleeding.

| Type of Use (Select one or both, as applicable)                               |                                                                              |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> | <div> <span> </span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the 510(k) Summary is provided below.

#### SUBMITTER 1.

3-D Matrix. Inc. 1234 Chestnut St., Suite 205 Newton, MA 02464

Contact Person: Lisa Spirio, PhD Phone: 617-875-6204 Email: lisa@3dmatrix.com

Date Prepared: February 28, 2023

#### 2. DEVICE

Name of Device: PuraStat Common Name: Hemostatic device for intraluminal gastrointestinal use Classification Regulation: 21 CFR 878.4456 Regulatory Class: II Product Code: OAU Panel: General & Plastic Surgery

#### PREDICATE DEVICE 3.

Predicate Device: 3-D Matrix, Inc.'s PuraStat-GI (K210098)

#### DEVICE DESCRIPTION 4.

PuraStat is a sterile gel composed of a synthetic peptide and sterile water for injection. It is provided as a prefilled syringe (2.5% peptide content) ready for use as a hemostat. The gel is delivered to the intended application site(s) via a commercially available endoscopic catheter that is attached to the gel syringe via the polypropylene adapter.

PuraStat is completely non-animal and non-plant derived and contains no preservatives that might present a risk of allergic reaction or skin irritation.

Exposure to physiological fluids such as blood causes the peptide solution to quickly form a transparent gel without expansion in volume. PuraStat achieves hemostatic effects by forming a hydrogel matrix barrier that blocks the flow of blood at the site of application.

#### INDICATIONS FOR USE ട്.

PuraStat is intended for hemostasis of mild and moderate bleeding post ESD or EMR, as an adjunct, bridge, prophylactic or rescue therapy for intraprocedural venous bleeding or prophylactic

{4}------------------------------------------------

therapy to prevent post procedure bleeding, and for primary non-variceal gastrointestinal (GI) bleeding. PuraStat is not indicated for arterial Forrest 1a bleeding.

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS 6.

The subject device is the identical product as the PuraStat-GI predicate (K210098).

PuraStat is comprised of a synthetic repeating 16-amino acid (RADA-16) oligopeptide in sterile water for injection. In the presence of a bleeding wound, the peptides assemble into a scaffold structure that forms a mechanical barrier over the bleeding site. The hemostatic solution is sterilefiltered and filled into 5-ml syringes made of cyclo-olefin polymers with a high-density polyethylene plunger and a butyl rubber head cap and gasket. Each syringe is filled with either 1, 3, or 5 ml of gel. The device is terminally sterilized, and the resorbable gel is delivered to the intended application site(s) via a polypropylene adapter connect to a commercially available catheter. The device is delivered endoscopically via a catheter to the bleeding site.

#### 7. PERFORMANCE DATA

The subject device is identical to the predicate device. The intended use of the subject device is identical to the intended use of the predicate device. The indications for use of the subject device are identical to the predicate's indications for use, with the exception that the subject device has the following statement not found in the predicate device: "... and for primary non-variceal gastrointestinal (GI) bleeding."

Because the subject device is identical to the predicate device and has the same intended use, the bench testing, animal testing and biocompatibility testing provided in the predicate device (K210098) was referenced for the subject device.

The additional indication for primary non-variceal gastrointestinal (GI) bleeding for the subject device was addressed by clinical data in an OUS study reported in the literature.

## Branchi et al Study (2021) *

The safety and effectiveness of the use of PuraStat as a primary hemostat to control primary bleeding in patients with acute non-variceal gastrointestinal bleeding (upper and lower) was evaluated in 111 patients in a prospective, consecutive open-label, multi-center study between July 2017 and December 2018 at 15 endoscopy departments in Germany. The effectiveness of PuraStat was assessed during the procedure, at 3 days and 1 week after application.

The primary endpoint was the achievement of hemostasis during the procedure with PuraStat (procedure success). Secondary end points were the prevention of rebleeding (defined as therapy success), the documentation of risk and side effect profiles of PuraStat. Rebleeding was defined as the presence of clinical signs of gastrointestinal bleeding (hematemesis, melena, hematochezia) in association with cardiovascular instability or Hemoglobin drop. A stable clinical condition without signs of rebleeding assessed at 3 and 7 days after the application of PuraStat was defined as therapy success.

Inclusion criteria were age > 18 years, acute upper or lower gastrointestinal bleeding including active bleeding and lesions with signs of recent hemorrhage such as visible vessels (Forrest Classification IIA or adherent clots or pigmented spots (Forrest Classification IIB) according to the Forrest classification. Adverse events were recorded for 30 days post procedure.

{5}------------------------------------------------

## i. Demographics and Patient Accounting

A total of 111 patients with gastrointestinal bleeding were included and treated with PuraStat; 68 were men and 43 were women. The bleeding activity was in most cases classified as oozing bleeding (76/111, 69%), in seven cases spurting hemorrhage (6%), in 16 cases a visible vessel (14%) and in 6 cases each (5%) an adherent clot or hematin on the ground of a lesion were observed. In all seven patients with spurting bleeding, PuraStat was applied as a secondary therapy to standard techniques, thus utilizing a combination of submucosal injection, hemoclips, over-thescope clips (OTSC) and gel to reach final hemostasis. After endoscopy therapy, none of these patients required surgery which was an overwhelming positive outcome for these patients.

No patients were lost to follow-up.

After primary application of PuraStat, initial hemostatic success was achieved in 94% of patients (74/79, 95% CI 88-99%), and in 75% of the patients when used as a secondary hemostatic product, following failure of established techniques (24/32, 95% CI 59-91%). The therapeutic success rates (absence of rebleeding) after 3 and 7 days were 91% and 87% after primary use, and 87% and 81% in all study patients. Overall rebleeding rate at 30-day follow-up was 16% (18/11). In the 5 patients who finally required surgery (4.5%), PuraStat allowed temporary hemostasis and stabilization.

No adverse events due to application of PuraStat or technical failures were reported.

The largest proportion of patients in the study were patients with active oozing bleeding, in whom typically a rebleeding rate of up to 50% is reported in the literature: however, in this study we found a rebleeding rate of 12% at 7 days and of 16% at 30 days in our whole patients' collective. After primary application of PuraStat, initial hemostatic success was achieved in 94% of patients (74/79, 95% CI 88-99%). The therapeutic success rates (absence of rebleeding) after 3 and 7 days were 91% and 87% after primary use. PuraStat was safely applied and administered without complications as a primary or secondary therapy.

* Branchi et al. PuraStat in gastrointestinal bleeding: results of a prospective multicentre observational pilot study. Surgical Endoscopy. 2021 June 15: 9 Feb; 7(1): 155–162.

#### 8. CONCLUSIONS

The intended use of the subject device is identical to the predicate device's intended use. The indications for use for the subject PuraStat are modifications of the predicate device, with the difference being that the subject device has the additional indication of primary non-variceal gastrointestinal (GI) bleeding and the additional clarification that PuraStat is not indicated for arterial Forrest 1a bleeding. The addition of non-variceal primary bleeding does not change the intended use of hemostasis. Thus, PuraStat does not have a new intended use. The subject device is the identical product as the PuraStat-GI predicate (K210098).

PuraStat complies with the special controls for a hemostatic device for intraluminal gastrointestinal internal use.

In conclusion. PuraStat is substantially equivalent to the predicate PuraStat-GI (K210098).

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QAU/K222481](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QAU/K222481)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
