← Product Code [QAU](/submissions/SU/subpart-e%E2%80%94surgical-devices/QAU) · K202929

# Nexpowder (K202929)

_Nextbiomedical Co., Ltd. · QAU · Sep 16, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QAU/K202929

## Device Facts

- **Applicant:** Nextbiomedical Co., Ltd.
- **Product Code:** [QAU](/submissions/SU/subpart-e%E2%80%94surgical-devices/QAU.md)
- **Decision Date:** Sep 16, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4456
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Nexpowder™ is used for hemostasis of non-variceal, upper gastrointestinal bleeding.

## Device Story

Nexpowder is a prescription, single-use endoscopic hemostatic device. It consists of a hemostatic powder agent (succinic anhydride, oxidized dextran, etc.) and a battery-powered delivery system (spray body, connector, catheter). The physician inserts the catheter through an endoscope working channel to the bleeding site. The battery-powered pump generates air pressure to propel the powder through the catheter. Upon contact with bodily fluids at the target site, the powder absorbs water, crosslinks, and gels to form a mechanical barrier, achieving hemostasis. The device is used in clinical settings by physicians. The powder is excreted via peristalsis within three days. It provides a physical, mechanical barrier to control bleeding without requiring human/animal proteins.

## Clinical Evidence

Evidence includes two porcine models (n=15 total) assessing hemostatic efficacy and safety, showing no perforation, obstruction, or gas embolism. Clinical evidence includes OUS data from 315 patients across multiple studies and a retrospective survey of 50 patients. Hemostasis success rates ranged from 94-100%. Re-bleed rates were <23% at 30 days. Adverse events were minimal (e.g., 2.7% fever in one study).

## Technological Characteristics

Hemostatic agent: succinic anhydride, oxidized dextran, PVP, Brilliant Blue FCF, L-HPC, lactose monohydrate, magnesium stearate. Delivery system: battery-powered air pump, 7.5Fr catheter (220cm). Sterilization: Gamma radiation (powder), EtO (catheter/connector). Biocompatibility: ISO 10993-1 compliant. Electrical safety: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6.

## Regulatory Identification

A hemostatic device for intraluminal gastrointestinal use is a prescription device that is endoscopically applied to the upper and/or lower gastrointestinal tract and is intended to produce hemostasis via absorption of fluid or by other physical means.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Performance data must support the sterility and pyrogenicity of the device.
(3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(4) In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The testing must evaluate the following:
(i) The ability to deliver the hemostatic material to the bleeding site;
(ii) The ability to achieve hemostasis in a clinically relevant model of gastrointestinal bleeding; and
(iii) Safety endpoints, including thromboembolic events, local and systemic toxicity, tissue trauma, gastrointestinal tract obstruction, and bowel distension and perforation.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
(i) Materials characterization of all components must demonstrate the device meets established specifications, which must include compositional identity and purity, characterization of impurities, physical characteristics, and reactivity with fluids.
(ii) Performance testing must demonstrate the mechanical integrity and functionality of the system used to deliver the device and demonstrate the device meets established specifications, including output pressure for propellant-based systems.
(6) Labeling must include:
(i) Information identifying and explaining how to use the device and its components; and
(ii) A shelf life.

## Predicate Devices

- Hemospray® Endoscopic Hemostat ([DEN170015](/device/DEN170015.md))

## Reference Devices

- PerClot® Topical ([K132105](/device/K132105.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 16, 2022

Nextbiomedical Co., Ltd. % Kyungyoon Kang CEO K-Bio Solutions 201 South 4th St, Suite 727 San Jose, California 95112

Re: K202929

Trade/Device Name: Nexpowder Regulation Number: 21 CFR 878.4456 Regulation Name: Hemostatic Device For Intraluminal Gastrointestinal Use Regulatory Class: Class II Product Code: QAU Dated: September 24, 2020 Received: September 29, 2020

Dear Kyungyoon Kang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Deborah Fellhauer, RN, BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202929

Device Name Nexpowder™

Indications for Use (Describe)

Nexpowder™ is used for hemostasis of non-variceal, upper gastrointestinal bleeding.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary: Nexpowder™

# l. Applicant/Manufacturer Information

Type of 510(k) Submission: Traditional 510(k) Premarket Notification

# Applicant's Information

Name of Sponsor: NEXTBIOMEDICAL CO., LTD. Address: 6, Venture-ro 100beon-gil, Yeounsu-gu, Incheon, Republic of Korea Contact Name: Eunhye Lee Telephone No.: 82-32-880-0820 Email Address: ehlee@nextbiomedical.co.kr

### Correspondent's Information

Company Name: K-Bio Solutions Correspondent Name: Mr. Kyungyoon Kang Email Address: kyungyoon.kang@kbiotechsolutions.com

Date Prepared: September 9th, 2022

### II. Regulatory Information

### Proposed Device:

- Trade/Proprietary Name: Nexpowder™ ●
- Classification Name: Hemostatic device for intraluminal gastrointestinal use
- Regulatory Class: Class II, Regulation Number: 878.4456, ●
- Product Code: QAU
- 510(K) number: K202929

#### Predicate Device:

- Trade/Proprietary Name: Hemospray® Endoscopic Hemostat ●
- Classification Name: Hemostatic device for intraluminal gastrointestinal use .
- Product Code: QAU ●
- De Novo Clearance number: DEN170015 ●
- . 510k Submitter: Wilson-Cook Medical, Inc.

#### Reference Device:

- Trade/Proprietary Name: PerClot® Topical ●
- Classification Name: Dressing, Wound, Drug
- . Product Code: FRO
- 510(K) number: K132105
- . 510k Submitter: CRYOLIFE, INC.

#### III. Device Description of the Nexpowder™

The Nexpowder™ is used for hemostasis of non-variceal, upper gastrointestinal bleeding. The Nexpowder™ is a prescription only, single-use device provided with a pre-packaged powder, a vial, and a Delivery System, which consists of a Spray body, a Connector and a Delivery Catheter. The powder vials

{4}------------------------------------------------

are provided in a sterile condition with gamma radiation sterilization and the non-sterile external Spray Body delivery system with its sterile Connector and sterile Delivery Catheter.

The hemostatic powder agents of the Nexpowder™ are primarily composed of succinic anhydride ({-poly-L-lysine) and oxidized dextran and are endoscopically applied through a catheter channel of the delivery system to control qastrointestinal bleeding in the upper qastrointestinal tract. Utilizing the installed battery power, air pressures are generated from the air pump placed in the spray body of Nexpowder™ delivery system to provide effective physical force to move the hemostatic powder agent into the delivery catheter. The hemostatic agents of the Nexpowder™ ultimately get sprayed onto the hemostasis target site in the gastrointestinal tract. Nexpowder is excreted from the patient's gastrointestinal or digestive system primarily by peristalsis of the human digestive system within the three days.

### IV. Intended Use/ Indications for Use

Nexpowder™ is used for hemostasis of non-variceal, upper gastrointestinal bleeding.

### V. Substantial Equivalence Comparison with Predicate Device

| Comparison<br>Item                    | Proposed Device                                                                                                                                                                                                                                                                                                                                                         | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                              | Assessment of Substantial<br>Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade/Device<br>Name                  | Nexpowder™                                                                                                                                                                                                                                                                                                                                                              | Hemospray®<br>Endoscopic Hemostat                                                                                                                                                                                                                                                                                                                                                                             | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Common Name                           | Hemostatic Device<br>for intraluminal<br>gastrointestinal use                                                                                                                                                                                                                                                                                                           | Hemostatic Device<br>for intraluminal<br>gastrointestinal use                                                                                                                                                                                                                                                                                                                                                 | Identical to the predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| 510k Number                           | K202929                                                                                                                                                                                                                                                                                                                                                                 | DEN170015                                                                                                                                                                                                                                                                                                                                                                                                     | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Manufacturer                          | NEXTBIOMEDICAL<br>Co., Ltd                                                                                                                                                                                                                                                                                                                                              | Wilson-Cook Medical,<br>Inc.                                                                                                                                                                                                                                                                                                                                                                                  | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Application/<br>Device<br>Overview    | The device is applied<br>by using catheter,<br>delivery system and a<br>gastrointestinal<br>endoscope.<br>The device is<br>comprised of a<br>handheld delivery<br>system that propels<br>hemostatic powder<br>through a catheter to<br>the GI bleeding site.                                                                                                            | The device is applied<br>by using endoscopic<br>catheter, delivery<br>system and a<br>gastrointestinal<br>endoscope.<br>The device is<br>comprised of a<br>handheld delivery<br>system that propels<br>hemostatic powder<br>through a catheter to<br>the GI bleeding site.                                                                                                                                    | Similar to the predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Intended Use /<br>Indications for Use | The device is intended<br>to be used for<br>hemostasis of<br>non-variceal, upper<br>gastrointestinal<br>bleeding.                                                                                                                                                                                                                                                       | The device is intended<br>to be used for<br>hemostasis of<br>non-variceal<br>gastrointestinal<br>bleeding.                                                                                                                                                                                                                                                                                                    | The indications for use are a<br>subset of the indications cleared<br>for the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Operation<br>Principle                | When the device<br>comes in contact with<br>an actively<br>bleeding site, the<br>powder absorbs water,<br>then acts both<br>cohesively and<br>adhesively, forming a<br>mechanical barrier<br>over the bleeding site.                                                                                                                                                    | When the device<br>comes in contact with<br>an actively<br>bleeding site, the<br>powder absorbs water,<br>then acts both<br>cohesively and<br>adhesively, forming a<br>mechanical barrier<br>over the bleeding site.                                                                                                                                                                                          | Similar to the predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                       | Nexpowder is an inert<br>powder developed for<br>endoscopic<br>hemostasis. The<br>powder is delivered by<br>use of a battery<br>powered delivery<br>system and through a<br>catheter inserted<br>through the working<br>channel of an<br>endoscope which<br>provides access to the<br>site of the bleed. It<br>contains no human or<br>animal proteins or<br>botanicals | Hemospray is an inert<br>powder developed for<br>endoscopic<br>hemostasis. The<br>powder is delivered by<br>use of a carbon<br>dioxide powered<br>delivery system and<br>through a catheter<br>inserted through the<br>working channel of an<br>endoscope which<br>provides access to the<br>site of the bleed. It<br>contains no human or<br>animal proteins or<br>botanicals and has no<br>known allergens. | Similar to the predicate device,<br>as for the difference in the<br>delivery system that<br>NexpowderTM is operated based<br>on the battery powered delivery<br>system, while the predicate,<br>Hemospray® is operated based<br>on the aerosol delivery system<br>does not raise different questions<br>in terms of safety and<br>effectiveness, given NexpowderTM<br>delivery system has been verified<br>to meet the electrical safety and<br>performance requirements per the<br>FDA's recognized standards of<br>IEC 60601-1, IEC 60601-1-2, and<br>IEC 60601-1-6 and special<br>controls. The same FDA's<br>recognized IEC standards have<br>been applied for the electrical<br>safety and performance<br>evaluation of the predicate,<br>Hemospray Endoscopic Hemostat<br>(DEN170015, Wilson-Cook<br>Medical, Inc.). |
| Product<br>Design                     | The device system<br>consists of hemostatic<br>powder agent and a<br>delivery system.                                                                                                                                                                                                                                                                                   | The device system<br>consists of hemostatic<br>powder agent and a<br>delivery system.                                                                                                                                                                                                                                                                                                                         | The overall device system with<br>hemostatic agent and delivery<br>system is substantially<br>equivalent/similar to the predicate<br>device in terms of the use of<br>hemostatic agent and a delivery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                       |                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                               | system under gastrointestinal<br>endoscopy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Hemostatic<br>Agent                   | Powder in a vial,<br>including:<br>• ε-poly-L-lysine<br>• Aldehyded dextran<br>• Polyvinylpyrrolidone<br>• Brilliant Blue FCF<br>• Low-substituted<br>Hydroxypropylcellulose<br>L-HPC<br>• Lactose Monohydrate<br>• Magnesium Stearate                                                                                                                                  | Sodium bentonite                                                                                                                                                                                                                                                                                                                                                                                              | The composition of the<br>hemostatic powder components<br>of the subject and predicate<br>devices is different.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Specifications                        | Catheter French size:<br>7.5Fr<br>Catheter Length:<br>220.0cm                                                                                                                                                                                                                                                                                                           | Catheter French size:<br>7Fr, 10Fr<br>Catheter Length:<br>220cm,                                                                                                                                                                                                                                                                                                                                              | Similar dimensional specifications<br>to the predicate device. The<br>favorable results of the design<br>verification and validation testing<br>demonstrate conformance of the<br>proposed Nexpowder™ to the<br>FDA's recognized standards and<br>the design requirements. The<br>testing results further<br>demonstrate the proposed<br>Nexpowder™ is substantially<br>equivalent to the predicate<br>device, Hemospray® Endoscopic<br>Hemostat, as the proposed<br>Nexpowder™ and predicate<br>device are subject to the same<br>recognized standards of the FDA<br>for the special controls evaluation<br>of medical device.                                                                                                                                                                                            |
| Sterilization<br>Method               | Powder: Gamma<br>radiation<br>Delivery system<br>(catheter and<br>connector): EtO<br>Sterilization<br>Spray Body: Non-sterile                                                                                                                                                                                                                                           | Product: Gamma<br>radiation                                                                                                                                                                                                                                                                                                                                                                                   | Similar Gamma radiation<br>sterilization for hemostatic agent<br>as the predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Sterilization<br>Validation<br>Target | Sterilization Assurance<br>Level of<br>10-6                                                                                                                                                                                                                                                                                                                             | Sterilization<br>Assurance Level of 10-6                                                                                                                                                                                                                                                                                                                                                                      | Identical to the predicate devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Shelf Life                            | 15 months                                                                                                                                                                                                                                                                                                                                                               | 3 Year                                                                                                                                                                                                                                                                                                                                                                                                        | Supported by 15 months shelf life<br>testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

#### Assessment Table of Substantial Equivalence

{5}------------------------------------------------

{6}------------------------------------------------

Similar to the predicate device, Nexpowder™ is an inert powder developed for endoscopic hemostasis. Nexpowder™ contains Aldehyded Dextran and Poly-L-Lysine (Polysaccharide) derived from microbial extraction. The powder is delivered by use of a battery powered delivery system and through a catheter

{7}------------------------------------------------

inserted through the working channel of an endoscope which provides access to the intended hemostasis target site. The overall hemostatic mechanism of Nexpowder™ to form a physical control barrier to achieve hemostasis is similar to the predicate, Hemospray® Endoscopic Hemostatic agent of Nexpowder™ is provided in a form of powder polymer materials, which can be readily sprayed over a bleeding site in the non-variceal upper gastrointestinal tract. The hemostatic agent of powder polymer functions as a physical, mechanical barrier over the bleeding site in the gastrointestinal tract to effectively control bleeding. From the standpoint of forming the mechanical barrier physically, the presence of blood in and of itself is not required. But any type of bodily fluids including digestive fluids excreted in the gastrointestinal tract and blood at the hemostasis target site coming into contact with the Nexpowder™ would lead to the crosslinking gelation effects following the hemostatic agent's rapid absorption of the bodily fluids.

Although the hemostatic agent may differ in chemical composition for each the predicate and subject device, in both of these products, the individual ingredients do not contain human or animal proteins or botanicals and have the same end state, which is a gel that will act as a mechanical barrier and induce hemostasis. In addition, the modified polysaccharide material that comprises the hemostatic agent ingredient of the Nexpowder™ is similar to the polysaccharide material in the reference device, PerClot® Topical (K132105). PerClot® Topical is another class II, hemostat device, intended as a topical wound dressing and used to control bleeding from the skin by rapidly absorbing water and forming a gelled adhesive matrix that provides a mechanical barrier to further bleeding. This demonstrate that polysaccharides used in hemostatic agents are not novel. Moreover, the biocompatibility profile of Nexpowder™ agents has been confirmed with comprehensive biocompatibility testing pursuant to ISO 10993-1.

Utilizing a power source, both the spray body of Nexpowder™ and the predicate provide physical force to push the hemostatic powder agent into the delivery catheter and spray the hemostasis intended site in the gastrointestinal tract. The minor difference in the power source of the delivery system does not raise different questions in terms of safety and effectiveness, since Nexpowder™ delivery system has been verified to meet the electrical safety and performance requirements per the FDA's recognized standards of IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-6.

# VI. Biocompatibility Assessment

Nexpowder™ has met the FDA's recognized standards for biocompatibility requirements identified in the FDA Guidance Table A.1, below, Biocompatibility risk assessments have been conducted per FDA Guidance, "Use of ISO 10993-1, Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing within a Risk Management Process (Issued June 16, 2016)". The following biocompatibility testing was performed on Nexpowder™:

| Biocompatibility<br>Endpoint | Method and Purpose                                                                                                              | Result         |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------|----------------|
| Cytotoxicity                 | ISO 10993-5: MEM Elution Study used to<br>evaluate device extracts for cytotoxicity risks.                                      | Non-cytotoxic  |
| Sensitization                | ISO 10993-10: Guinea Pig Maximization<br>Sensitization Test used to evaluate device<br>extracts for dermal sensitization risks. | Non-sensitizer |

## GLP Biocompatibility Testing Completed for the Hemostatic Agent Powder of Nexpowder™:

{8}------------------------------------------------

| Irritation                        | ISO 10993-10: Intracutaneous Initiation Test<br>used to evaluate device extracts for irritation<br>risks.                                                                                                 | Non-irritant                       |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| Acute Systemic Toxicity           | ISO 10993-11: Acute systemic toxicity study<br>used to evaluate device extracts for systemic<br>toxicity risks.                                                                                           | No acute systemic<br>toxicity      |
| Material Mediated<br>Pyrogenicity | ISO 10993-11: Rabbit pyrogen test used to<br>evaluate device extracts for pyrogenicity<br>risks.                                                                                                          | Non-pyrogenic                      |
| Hemocompatibility                 | Hemolytic study (ASTM: F756-17),<br>Complement Activation, Partial<br>Thromboplastin Time, Platelet and Leukocyte<br>Counts (ISO10993-4) used to evaluate<br>device extracts for hemocompatibility risks. | Hemocompatible                     |
| Genotoxicity                      | ISO 10993-3: Genotoxicity study used to<br>evaluate device extracts for genotoxicity risk.                                                                                                                | Non-genotoxic                      |
| Implantation                      | ISO 10993-6: Implantation study used to<br>evaluate device extracts for the local effects<br>risk.                                                                                                        | Non-irritant                       |
| Subchronic Toxicity               | ISO 10993-11: Implantation study used to<br>evaluate device extracts for subchronic<br>toxicity risk.                                                                                                     | No subchronic<br>systemic toxicity |

# GLP Biocompatibility Testing Completed for the Delivery System Catheter and Connector of Nexpowder™:

| Biocompatibility Endpoint      | Method and Purpose                                                                                                        | Result                     |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Cytotoxicity                   | ISO 10993-5: MEM Elution Study used to evaluate device extracts for cytotoxicity risks.                                   | Non-cytotoxic              |
| Sensitization                  | ISO 10993-10: Guinea Pig Maximization Sensitization Test used to evaluate device extracts for de1mal sensitization risks. | Non-sensitizer             |
| Irritation                     | ISO 10993-10: Intracutaneous Initiation Test used to evaluate device extracts for irritation risks.                       | Non-irritant               |
| Acute Systemic Toxicity        | ISO 10993-11: Acute systemic toxicity study used to evaluate device extracts for systemic toxicity risks.                 | No acute systemic toxicity |
| Material mediated pyrogenicity | ISO 10993-11: Rabbit pyrogen test used to evaluate device extracts for pyrogenicity risks.                                | Non-pyrogenic              |
| Endotoxin                      | USP <85>                                                                                                                  | Non-pyrogenic              |
| Hemocompatibility              | ASTM: F756-17: Hemolytic study used to evaluate device extracts for hemocompatibility risks.                              | Hemocompatible             |

{9}------------------------------------------------

# VII. Design Verification and Validation Testing

Design Verification and Validation (DV&V) testing were performed to verify that the proposed Nexpowder™ meets the pre-determined requirements for design verification, validation, and FDA's special control requirements. Testing was also conducted to verify the effectiveness of the implemented risk control measures to mitigate the risks identified within the risk management process per ISO 14971: Medical Devices-Application of Risk Management to Medical Devices. The following design verification or performance testing of Nexpowder have been completed with favorable test results, meeting the applicable ISO standards and FDA's recognized standards pertaining to evaluations of Nexpowder™.

# Design Verification Performance Testing

# T=0 Baseline Sample Testing, T=2 Years, and T=3 Years Accelerated Aging Testing of Finished Hemostatic Powder Agents of Nexpowder™:

- Appearance Test ●
- Weight Test .
- Absorption Test
- Water Content Test ●
- Adhesion Test ●

# T=0 Baseline Sample Testing, T=3 Years Accelerated Aging Testing of Catheter and Connector of Nexpowder™ Delivery System:

- Appearance Test
- Dimension Verification Test ●
- Tensile Strength Test ●
- Air Leak Test
- Spray Test

# T=0 Baseline Sample Testing, T=3 Years Accelerated Aging Testing of Spray Body of Nexpowder™ Delivery System:

- Appearance Test
- Dimension Verification Test ●
- Air Pressure Test of the Outlet Port
- Power Switch Function Test (Battery Function and Vibration operation Test) ●
- . Battery Protection Film Test
- Battery Cap test ●
- . Spray test

{10}------------------------------------------------

# Design Validation Testing

- Animal Study ●

| Test                | Purpose                                                         | Method                                                                                                                                                                                                                | Results                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|---------------------|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Porcine<br>Study I  | Assess<br>hemostatic<br>efficacy and<br>safety of the<br>device | Design: Endoscopic<br>submucosal dissection<br>(ESD) based bleeding<br>models<br>Animals: 6 Nexpowder™<br>treated, 6 Hemospray®<br>treated, 3 sham controls<br>Duration: 30-day follow-<br>up followed by termination | Initial hemostasis was achieved in all<br>animals (6/6). None of the animals in the<br>treatment group showed re-bleeding up<br>to the 30-day follow-up time point. There<br>was no gastrointestinal perforation,<br>obstruction, or gas embolism caused by<br>the device. Histopathological evaluation<br>of the device application site did not<br>reveal any important histological<br>differences between treatment group and<br>control group animals. |
| Porcine<br>Study II | Assess<br>hemostatic<br>efficacy and<br>safety of the<br>device | Design: Gastroepiploic<br>Artery (Forrest 1a)<br>Bleeding Model<br>Animals: 6 Nexpowder™<br>treated, 3 Hemospray®<br>treated<br>Duration: 30-day follow-<br>up followed by termination                                | The rate of achieving initial hemostasis<br>was 67% (4/6) in the treatment group.<br>None of the animals in the treatment<br>group showed re-bleeding up to the 30-<br>day follow-up time point. There were no<br>gastrointestinal perforation, obstruction,<br>and gas embolism which caused by<br>device. Histopathological evaluation of<br>the device application site did not reveal<br>any important histological findings.                           |

In vivo animal studies have been conducted to support the substantial equivalence claims of our product as compared to the predicate device Hemospray®. The objective of the study was to validate the efficacy and safety of Nexpowder™ (compared to the predicate device) in a porcine model for upper gastrointestinal bleeding control. The results demonstrated that the proposed is substantially equivalent to the predicate.

- Simulation System Test
- Usability Engineering Evaluation
- Clinical data

Real-world evidence of device safety in nonvariceal upper gastrointestinal bleeding in the OUS market was provided, including data from 315 patients with an additional 50 patients from a Retrospective Aggregated Data Collection survey (see table below). Hemostasis was achieved in 94-100% of patients with a re-bleed rate of <23% at 30 days.

{11}------------------------------------------------

| Clinical Study Title                                                                                                                                        | Number<br>of<br>Patients | Hemostasis (%)                                            | Adverse<br>Events/<br>Complication<br>Rates                   | Re-Bleeding<br>rate (%)                              | Bleeding Types<br>of Enrolled Subjects                                                                  |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|-----------------------------------------------------------|---------------------------------------------------------------|------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| A. Prospective, Multicenter, Single-blind (subject)<br>Controlled Clinical Trial to<br>Confirm the Efficacy and<br>Safety of the Wound<br>Dressing 'UI-EWD' | 76                       | 100%: Test<br>Group with<br>Nexpowder: 37<br>out of 37    | 2.7%<br>(1 Adverse<br>event) out of<br>37 Subjects<br>(Fever) | 8.11% within<br>3 days                               | Post-ESD (86.49%,<br>32/37)<br>Forrest Class Ib<br>(75.68%, (28/37)                                     |
| B. Novel hemostatic<br>adhesive powder for<br>nonvariceal upper<br>gastrointestinal bleeding                                                                | 56                       | 96.4% (54 out<br>of 56 cases with<br>use of<br>Nexpowder) | Zero adverse<br>events                                        | 3.7%<br>Within 30<br>days                            | Post-procedure<br>(97.80%, 45/56)<br>Forrest Class Ib<br>(97.20%, (28/56))                              |
| C. Efficacy of a novel<br>hemostatic adhesive<br>powder in patients with<br>refractory upper<br>gastrointestinal bleeding: a<br>pilot study                 | 17                       | 94% (16 out of<br>17 cases with<br>use of<br>Nexpowder)   | Zero adverse<br>events                                        | 19%<br>Within 30<br>days                             | Refractory bleeding<br>Forrest la (11.8%,<br>2/17)<br>Forrest Ib (88.2%,<br>15/17)                      |
| D. Efficacy of a novel<br>hemostatic adhesive<br>powder in patients with<br>upper gastrointestinal<br>tumor bleeding                                        | 41                       | 97.5% (40 out<br>of 41 cases with<br>use of<br>Nexpowder) | Zero adverse<br>events                                        | 4.3% within 7<br>days<br><br>22.5% within<br>28 days | Forrest Ib (100.0%)                                                                                     |
| E. Post Market Clinical<br>Aggregated Data Collection<br>Survey on Nexpowder™ for<br>Safety and Performance<br>Evaluation                                   | 50                       | 100% (50 out of<br>50 cases with<br>use of<br>Nexpowder)  | Zero adverse<br>events                                        | 2.0% within 3<br>days<br><br>4.0% within<br>30 days  | Peptic ulcer (24%,<br>12/50)<br>Tumor bleed (8.0%,<br>4/50)<br>Prevent delayed<br>bleeding (30%, 15/50) |

# VIII. Sterilization Validation and Verification

The following sterilization and verification testing was performed:

- Gamma Sterilization Validation Testing of Hemostatic Powder Agent
- . Eto Sterilization Validation Testing of Delivery Catheter/Connector
- Sterility Test of Nexpowder, Catheter and Connector both with the baseline time-zero samples and three-year aged samples

# IX. Packaging Validation, Verification and Distribution Test

The following packaging validation, verification and distribution testing was performed:

- . Packaging Validation Testing of Hemostatic Powder Agent
- . Package Seal Peeling Test and Dye Penetration Test of Nexpowder™ Catheter and Connector both with the baseline time-zero samples and three-year aged samples
- Package Seal Peeling Test and Dye Penetration Test of Nexpowder™ Spray Body both with the baseline time-zero samples and three-year aged samples

{12}------------------------------------------------

- Distribution Transportation Test of Nexpowder™ .
# X. Conclusion

Nexpowder™ functions as intended, and is substantially equivalent to the predicate, Hemospray® Endoscopic Hemostat (DEN170015). Nexpowder™ has the similar intended use and indications for use, and similar key technological and design characteristics and mechanism of action compared to the predicate device. The minor differences between the subject and predicate devices do not raise different questions of safety or effectiveness as the design verification and performance testing data support that the Nexpowder™ is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QAU/K202929](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/QAU/K202929)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
