PZP · Fluid Jet Removal System
General, Plastic Surgery · 21 CFR 876.4350 · Class 2
Overview
| Product Code | PZP |
|---|---|
| Device Name | Fluid Jet Removal System |
| Regulation | 21 CFR 876.4350 |
| Device Class | Class 2 |
| Review Panel | General, Plastic Surgery |
Identification
A fluid jet system for prostate tissue removal is a prescription device intended for the resection and removal of prostatic tissue for the treatment of benign prostatic hyperplasia. The device cuts tissue by using a pressurized jet of fluid delivered to the prostatic urethra. The device is able to image the treatment area, or pairs with an imaging modality, to monitor treatment progress.
Classification Rationale
Class II (special controls). The special controls for this device are:
Special Controls
In combination with the general controls of the FD&C Act, the fluid jet system for prostate tissue removal is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must evaluate the following: (i) All adverse events associated with the device, and (ii) Improvement in lower urinary tract symptoms (LUTS). (2) Physician training must be provided that includes: (i) Information on key aspects and use of the device, and (ii) Information on how to override or stop resection. (3) Animal testing must demonstrate that the device resects targeted tissue in a controlled manner without injury to adjacent non-target tissues. (4) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Measurement of targeting accuracy and reproducibility of high velocity fluid jet, and (ii) High pressure fluid jet verification testing at target and non-target tissues. (5) Software verification, validation, and hazard analysis must be performed. (6) The patient-contacting elements of the device must be demonstrated to be biocompatible. (7) Performance data must demonstrate the electrical safety and electromagnetic compatibility of the device. (8) Performance data must demonstrate the sterility of the patient-contacting components of the device. (9) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life. (10) Performance data must validate the instructions for reprocessing and reliability of reusable components. (11) Labeling must include the following: (i) A section that summarizes the clinical testing results, including the adverse event profile and improvement in LUTS; (ii) A shelf life for single use components; (iii) A use life for reusable components; and (iv) Reprocessing instructions for reusable components.
Recent Cleared Devices (9 of 9)
| Record | Device Name | Applicant | Decision Date | Decision |
|---|---|---|---|---|
| K260723 | HYDROS Robotic System (HY1000);HYDROS Handpiece (HH1000);HYDROS TRUS Probe (HU1000) | Procept Biorobotics | Jul 9, 2026 | SESE |
| K253211 | AQUABEAM Robotic System | Procept Biorobotics | Jun 23, 2026 | SESE |
| K251082 | HYDROS Robotic System (HY1000); HYDROS Handpiece (HH1000); HYDROS TRUS Probe (HU1000) | Procept Biorobotics | Oct 10, 2025 | SESE |
| K241952 | AQUABEAM Robotic System (AB2000) | Procept Biorobotics | Sep 30, 2024 | SESE |
| K240200 | HYDROS Robotic System; HYDROS TRUS Probe; HYDROS Handpiece | Procept Biorobotics | Aug 20, 2024 | SESE |
| K231024 | AquaBeam Robotic System | Procept Biorobotics, Corporation | Aug 30, 2023 | SESE |
| K212835 | AquaBeam Robotic System | Procept Biorobotics, Corporation | Oct 6, 2021 | SESE |
| K202961 | AquaBeam Robotic System | Procept Biorobotics, Corporation | Mar 11, 2021 | SESE |
| DEN170024 | AQUABEAM System | Procept Biorobotics, Corporation | Dec 21, 2017 | DENG |
Top Applicants
- Procept Biorobotics — 5 clearances
- Procept Biorobotics, Corporation — 4 clearances