PQI · Wound Retraction And Protection System
General, Plastic Surgery · 21 CFR 878.4371 · Class 2
Overview
| Product Code | PQI |
|---|---|
| Device Name | Wound Retraction And Protection System |
| Regulation | 21 CFR 878.4371 |
| Device Class | Class 2 |
| Review Panel | General, Plastic Surgery |
| 3rd-Party Reviewable | Yes |
Identification
An irrigating wound retractor device is a prescription device intended to be used by a surgeon to retract the surgical incision, to provide access to the surgical wound, to protect and irrigate the surgical wound, and to serve as a conduit for removal of fluid from the surgical wound.
Classification Rationale
Class II (special controls). The special controls for this device are:
Special Controls
In combination with the general controls of the FD&C Act, the Irrigating Wound Retractor Device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible and evaluated for particulate matter. (2) Performance data must demonstrate the sterility and pyrogenicity of the patient-contacting components of the device. (3) Performance data must support shelf life by demonstrating continued functionality and sterility of the device over the identified shelf life. (4) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must: (i) Characterize the tear resistance, tensile strength, and elongation properties of the barrier material; (ii) Demonstrate that the liquid barrier material is resistant to penetration by blood, and is non-flammable; (iii) Characterize the forces required to deploy the device; (iv) Characterize the device's ranges of operation, including flow rates and maximum suction pressures; (v) Demonstrate the ability of the device irrigation apparatus to maintain a user defined or preset flow rate to the surgical wound; and (vi) Demonstrate the ability of the device to maintain user defined or preset removal rates of fluid from the surgical wound. (5) The labeling must include or state the following information: (i) Device size or incision length range; (ii) Method of sterilization; (iii) Flammability classification; (iv) Non-pyrogenic; (v) Shelf life; and (vi) Maximum flow rate and suction pressure.
Recent Cleared Devices (2 of 2)
| Record | Device Name | Applicant | Decision Date | Decision |
|---|---|---|---|---|
| K172132 | CleanCision Wound Retraction and Protection System | Prescient Surgical, Inc. | Oct 25, 2017 | SESE |
| DEN150038 | CleanCisionTM Wound Retraction and Protection System | Prescient Surgical | Dec 16, 2016 | DENG |
Top Applicants
- Prescient Surgical, Inc. — 1 clearance
- Prescient Surgical — 1 clearance