← Product Code [PDZ](/submissions/SU/subpart-e%E2%80%94surgical-devices/PDZ) · K253960

# Medicell Mycosis Laser (MCML24004) (K253960)

_Medicell Healthcare Co.,LTD · PDZ · Mar 9, 2026 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/PDZ/K253960

## Device Facts

- **Applicant:** Medicell Healthcare Co.,LTD
- **Product Code:** [PDZ](/submissions/SU/subpart-e%E2%80%94surgical-devices/PDZ.md)
- **Decision Date:** Mar 9, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Medicell Mycosis Laser is indicated for use for the temporary increase of clear nails in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum, and T. mentagrophytes, and/or yeasts Candida albicans, etc.)

## Device Story

Medicell Mycosis Laser is a compact, floor-standing medical laser device for treating onychomycosis. It utilizes dual-wavelength diode laser technology (635nm and 405nm) to induce a photochemical reaction in the nail bed. The device features an LCD touch screen for operator control and an integrated monitor for visualizing the treatment area. It is designed to treat both feet simultaneously. The operator, typically a clinician, selects treatment times (12, 15, or 20 minutes) via the interface. The laser output is delivered in an electronically scanned line pattern in constant wave mode. By targeting the fungal infection, the device aims to increase the appearance of clear nails. The device is intended for prescription use in a clinical setting.

## Clinical Evidence

Bench testing only. No clinical data provided. Verification included electrical safety (IEC 60601-1:2020), electromagnetic compatibility (IEC 60601-1-2:2020), photobiological safety (IEC 62471:2006), and laser safety (IEC 60825-1:2014). All tests met pre-determined acceptance criteria.

## Technological Characteristics

Diode laser system; wavelengths 405nm/635nm (±10%); constant wave output; electronically scanned line pattern. Dimensions: 540mm x 304mm x 405mm; weight 7.8kg. Interface: LCD touch screen. Standards: IEC 60601-1:2020 (safety), IEC 60601-1-2:2020 (EMC), IEC 62471:2006 (photobiological), IEC 60825-1:2014 (laser safety). Class 3B laser diodes.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- AF Laser ([K221363](/device/K221363.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

March 9, 2026

Medicell Healthcare Co.,LTD
% Spencer Walker
Director of Regulatory Affairs
Center For Medical Innovation
350 E. 400 S.
Salt Lake City, Utah 84111

Re: K253960
Trade/Device Name: Medicell Mycosis Laser (MCML24004)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: PDZ
Dated: December 9, 2025
Received: December 10, 2025

Dear Spencer Walker:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K253960 - Spencer Walker
Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn

{2}

K253960 - Spencer Walker
Page 3

(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2026.03.09
22:35:47 -04'00'

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253960 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  Medicell Mycosis Laser (MCML24004)  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The Medicell Mycosis Laser is indicated for the temporary increase of clear nails in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum, and T. mentagrophytes, and/or yeasts Candida albicans, etc.)  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

MEDICELL HEALTHCARE
510(k) Summary

# 510(K) SUMMARY

K253960

(21 CFR 807.92)

## GENERAL INFORMATION

Submitter: Medicell Healthcare, Ltd.

Contact Person: Spencer Walker, MSc.
Center for Medical Innovation || University of Utah
400 South 350 East
Salt Lake City, UT 84111
(801) 708-2238
Spencer.Walker@hsc.utah.edu

Date Prepared: December 9, 2025

Trade Name: Medicell Mycosis Laser (MCML24004)

Common Name: Laser for Clear Nails

Classification Name: Laser for Clear Nails
21 CFR §878.4810, Product Code PDZ

Device Class: Class II

Predicate Device: 510(k) No.: K221363
Model: AF Laser
Manufacture: ShenB Co. Ltd
Classification: PDZ

## Device Description:

The Medicell Mycosis Laser is a medical laser used to treat nail fungus (onychomycosis). The device produces two wavelengths of light, 635nm and 405nm, to produce a photochemical reaction to achieve its therapeutic effects. The Mycosis laser is a compact, all-in-one device that is intended to be used on the floor and is operated by an intuitive touch screen for ease of use. An LCD monitor incorporated into the face of the device helps the operator check the exact location of treatment area during a procedure.

|  Table 1: Medicell Mycosis Laser Model Number  |   |
| --- | --- |
|  Model No. | Description  |
|  MCML24004 | Medicell Mycosis Laser  |

1 | Page

{5}

MEDICELL HEALTHCARE

510(k) Summary

# Indications for Use:

The Medicell Mycosis Laser is indicated for use for the temporary increase of clear nails in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum, and T. mentagrophytes, and/or yeasts Candida albicans, etc.)

# Comparative Analysis:

It has been demonstrated that the Medicell Mycosis Laser is comparable to the predicate device in intended use, fundamental scientific technology, design, principles of operation, patient/user interface and functional performance evaluations. The Medicell Mycosis Laser has been fully assessed within the Medicell Risk Management and Design Controls systems. All necessary verification steps met pre-determined acceptance criteria to confirm safety and effectiveness.

It has been demonstrated that the Medicell Mycosis Laser is comparable to the predicate device in the following manner:

- Same Intended Use
- Same Indications for Use
- Same Fundamental Scientific Technology
- Same or similar Operating Principles of Use
- Same Material Properties
- Same or similar Performance Specifications

|  Table 2: Substantial Equivalence Comparison Chart  |   |   |   |
| --- | --- | --- | --- |
|   | Predicate – K221363 (AF Laser Mycosis Laser) | Subject Device – (Medicell Mycosis Laser) | Comparison  |
|  Ind. for Use | The AF Laser is indicated for use for the temporary increase of clear nails in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum, and T. mentagrophytes, and/or yeasts Candida albicans, etc.) | The Medicell Mycosis Laser is indicated for use for the temporary increase of clear nails in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum, and T. mentagrophytes, and/or yeasts Candida albicans, etc.) | Substantially Equivalent  |
|  Classification Name | General Hospital – Laser for Clear Nails
21 CFR §878.4810
Product Code: PDZ Class II | General Hospital – Laser for Clear Nails
21 CFR §878.4810
Product Code: PDZ Class II | Substantially Equivalent  |
|  Laser Type | Diode Laser | Diode Laser | Substantially Equivalent  |
|  Laser Wavelength | 405nm/635nm (±10%) | 405nm/635nm (±10%) | Substantially Equivalent  |

{6}

MEDICELL HEALTHCARE
510(k) Summary

|  Table 2: Substantial Equivalence Comparison Chart  |   |   |   |
| --- | --- | --- | --- |
|   | Predicate – K221363 (AF Laser Mycosis Laser) | Subject Device – (Medicell Mycosis Laser) | Comparison  |
|  Output Power | 405nm: 23mW ± 1.85mW
635nm: 17mW ± 1.35mW | 405nm: 23mW ± 2.00mW
635nm: 17.25mW ± 1.25mW | Substantially Equivalent  |
|  Output Area | Line pattern electronically scanned over area of treatment | Line pattern electronically scanned over area of treatment | Substantially Equivalent  |
|  Output Type | Constant Wave | Constant Wave | Substantially Equivalent  |
|  Operating Time | 0-12 minutes (±5%) with 1 minute increment | 12, 15, 20 minutes (±5%) with 1 minute increment | Substantially Equivalent  |
|  Dimensions | 424mm(W) × 308mm(L) × 352mm(H) | 540mm(W) × 304mm(L) × 405mm(H) | Substantially Equivalent  |
|  Weight | 17.5kg | 7.8kg | Substantially Equivalent  |
|  Screen | LCD Touch Screen | LCD Touch Screen | Substantially Equivalent  |
|  Treatment Bed | Dual Feet Simultaneously | Dual Feet Simultaneously | Substantially Equivalent  |

## Functional/Safety Testing:

The following testing was conducted to validate and verify that the subject device was substantially equivalent to the predicate devices. All data met pre-determined acceptance criteria.

**Design Verification** – Performance bench testing was conducted to ensure that the Medicell Mycosis Laser met the applicable design and performance requirements throughout its use life, verify conformity to applicable standards, and demonstrate substantial equivalence to the predicate device. The following performance testing was performed or fulfilled with the Medicell Mycosis Laser.

- IEC 60601-1:2020 Test for Medical Electrical equipment was performed for General Requirements for basic safety and essential performance
- IEC 60601-1-2:2020 Test for Medical Equipment for General Requirements for basic safety and essential performance: electromagnetic compatibility
- IEC 62471:2006 - Photobiological safety of lamps and lamp systems
- IEC 60825-1:2014 - Safety of laser products - Part 1: Equipment classification and requirements

3 | Page

{7}

MEDICELL HEALTHCARE
510(k) Summary

|  No. | Test Performed | Test Method Applicable Standard | Acceptance Criteria | Results  |
| --- | --- | --- | --- | --- |
|  1 | Electrical Safety and Essential Performance Testing | IEC60601-1:2020 | Device is safe for human use | The subject device meets acceptance criteria  |
|  2 | Electromagnetic Compatibility (EMC) | IEC60601-1-2:2020 | Device is electro magnetically compatible | The subject device meets acceptance criteria  |
|  3 | Photobiological Safety of Lamps and Lamp Systems | IEC 62471:2008 | Device lights are Photobiologically compatible | The subject device meets acceptance criteria  |
|  4 | Laser Equipment | IEC 60825-1:2014 | Blue and Red Laser Equipment Efficiency – Class 3B Laser Diodes | The subject device meets acceptance criteria  |

Conclusion:
Based on the similarities in design between the subject and predicate devices, and the performance testing performed, the subject Medicell Mycosis Laser is substantially equivalent to the cited predicate device.

4 | Page

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/PDZ/K253960](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/PDZ/K253960)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
