K250734 · Pollogen, Ltd. · PBX · Sep 10, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250734
Device Name
OptiLUV Device
Applicant
Pollogen, Ltd.
Product Code
PBX · General, Plastic Surgery
Decision Date
Sep 10, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The OptiLUV device is a handheld device intended to emit energy in the RF spectrum to provide topical heating for the purpose of elevating facial tissue temperature for a temporary relief of minor pain and the temporary increase in local blood circulation.
Device Story
Non-invasive, non-ablative handheld device; delivers 1 MHz bipolar RF energy to dermis to generate heat via electrical impedance. Components include device, USB charging cord, and preparation balm. Used on face; operated by clinician (Rx only). Integrated temperature sensor between electrodes monitors skin temperature; disables RF transmission at 41.5±1 °C to prevent overheating. Output provides topical heating; intended to relieve minor pain and increase local blood circulation. Benefits patient through non-invasive thermal therapy.
Clinical Evidence
Clinical study evaluated 1.8W RF output power over 10-minute treatment duration. Results confirmed ability to maintain safe, effective treatment temperature on human skin surface. No adverse events reported. Bench testing verified power accuracy and overheating safety mechanism functionality.
Technological Characteristics
Bipolar RF energy; 1 MHz frequency; 1.8W max power. 4 electrodes (2 pairs). Integrated thermal sensor for temperature monitoring (limit 41.5±1 °C). Powered by 100-240V, 50-60Hz. Complies with IEC 60601-1, 60601-1-2, 60601-2-2, 60601-1-11, ISO 10993-1, ISO 14971. Non-sterile, reusable.
Indications for Use
Indicated for adult users requiring temporary relief of minor pain and temporary increase in local blood circulation via topical heating of facial tissue.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
K232424 — CurrentBody Skin RF · El Global Trade, Ltd. · Mar 6, 2024
K170499 — sensiLift · El Global Trade, Ltd. · Jun 15, 2017
K182774 — STOP U (Packed Black USA), STOP U (Packed White USA) · Pollogen, Ltd. · Jun 19, 2019
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 10, 2025
Pollogen Ltd.
% Elissa Burg
Regulatory Consultant
BioVision Ltd.
Perach Halilach Street 183
Had Nes, 1295000
Israel
RE: K250734
Trade/Device Name: OptiLUV Device
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical cutting and coagulation device and accessories
Regulatory Class: Class II
Product Code: PBX
Dated: August 11, 2025
Received: August 11, 2025
Dear Elissa Burg:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250734 - Elissa Burg
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250734 - Elissa Burg
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin K. Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2025.09.10
14:40:45 -04'00'
Colin Kejing Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250734
Device Name
OptiLUV Device
Indications for Use (Describe)
The OptiLUV device is a handheld device intended to emit energy in the RF spectrum to provide topical heating for the purpose of elevating facial tissue temperature for a temporary relief of minor pain and the temporary increase in local blood circulation.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
Pollogen
BY LUMENIS
K250734
# 510(k) SUMMARY - K250734
# Pollogen Ltd.'s OptiLUV Device
Applicant:
Pollogen Ltd.
6 Kaufman St.
Gibor House, P.O. Box 50320
Tel Aviv
ISRAEL 6801298
Tel. (972) 3-510-4110
Fax (972) 3-510-4112
Contact Person:
Elissa Burg
Regulatory Consultant
BioVision Ltd.
Had Nes 183
Israel 1295000
Tel. (972) 526633572
Date Prepared: September 8, 2025
Device Name: OptiLUV Device
Common Name: Electrosurgical cutting and coagulation device and accessories
Classification: Product Codes: PBX, GEI
Regulation No: 21 CFR §878.4400
Class: II
Classification Panel: General and Plastic Surgery
Predicate Device Reference Device
EL Global Trade Ltd. RF-Relief device (K181095)
Pollogen Ltd. STOP U (K140255)
# Intended Use/Indications for Use
The OptiLUV device is a handheld device intended to emit energy in the RF spectrum to provide topical heating for the purpose of elevating facial tissue temperature for a temporary relief of minor pain and the temporary increase in local blood circulation.
Traditional 510(k) Submission - OptiLUV Device - Pollogen Ltd.
{5}
Collogen
BY LUMENIS
# Device Description
The OptiLUV device is a non-invasive, non-ablative device that delivers RF energy into the skin to generate heat through electrical impedance in the dermis. The device consists of the following components and accessories: the OptiLUV device, the USB Charging Cord, and the OptiLUV Preparation Balm.
# Technological Characteristics
The OptiLUV device delivers RF energy at a frequency of 1 MHz and a maximum output power of 1.8 watts into the skin through its electrodes. The device generates heat through electrical impedance in the dermis. The temperature sensor, located between the electrodes, constantly monitors the skin temperature and disables RF transmission once the maximum allowed skin temperature (41.5±1 °C) is obtained.
# Summary of Non-clinical and/or Clinical Tests
Pollogen conducted several performance tests to demonstrate that the OptiLUV Device complies with appropriate consensus standards and that it functions as intended.
- OptiLUV’s power accuracy testing was performed to verify that the power levels are within the predefined acceptance criteria.
- OptiLUV’s overheating safety testing was performed to verify the proper function and accuracy of the safety mechanism, while the temperature sensor constantly measures the skin temperature during treatment and deactivates RF transmission once the maximal allowed skin temperature is obtained.
- The OptiLUV software was validated as required.
In all instances, the OptiLUV Device functioned as intended, and observations were as expected.
The OptiLUV Device complies with the following consensus standards:
- ISO 14971:2019, Ed.3 - Medical devices – Application of risk management to medical devices.
- IEC 60601-1:2005+ AMD1:2012 + AMD2:2020, Ed. 3.2 - Medical Electrical Equipment Part 1: General requirements for safety.
Traditional 510(k) Submission - OptiLUV Device - Pollogen Ltd.
{6}
Collogen
BY LUMENIS
- IEC 60601-1-2: 2014 + AMD1:2020, Ed. 4.1 - Medical Electrical Equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
- IEC 60601-2-2: 2017 + AMD1:2023, Ed.6.1 - Medical Electrical Equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.
- IEC TS 60601-4-2: 2024, Ed. 1.0 - Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems.
- IEC 60601-1-11: 2015 + AMD1:2020, Ed.2.1 - Medical electrical equipment Part 1-11 – General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment (.2015 and AMD2:2020 (ed.2.1))
- IEC 60601-1-6: 2010+AMD1:2013+AMD2:2020, Ed.3.2 Medical Electrical Equipment Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability.
- IEC 62366-1: 2015+AMD1:2020, Ed. 1.1 - Medical devices - Part 1: Application of usability engineering to medical devices.
- ISO 10993-1:2018, Ed.5 - Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
- IEC 62304: 2006+AMD A1:2015, Ed.1.1 - Medical device software – Software life cycle processes.
- ANSI/AAMI ST98:2022 Cleaning validation of health care products - Requirements for development and validation of a cleaning process for medical devices.
- ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems.
- ASTM D4332-22 Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing.
## Clinical Testing
The OptiLUV device was evaluated in a clinical study to demonstrate that its RF output power of 1.8 watts has the ability to maintain the desired safe and effective treatment temperature on the surface of human skin throughout a treatment duration of 10 minutes.
No adverse events were observed or reported during the course of this study.
The clinical trial results confirm that the OptiLUV device is safe and effective for its intended use.
Traditional 510(k) Submission - OptiLUV Device - Pollogen Ltd.
{7}
Collogen
BY LUMENIS
# Substantial Equivalence Comparison
The following table provides a comparison of the OptiLUV Device to its predicate and reference devices, addressing intended use, technological characteristics, and principles of operation to support the substantial equivalence determination.
| Feature | OptiLUV Proposed Device | RF-Relief (K181095) Predicate Device | STOP U (K140255) Reference Device |
| --- | --- | --- | --- |
| Manufacturer | Pollogen Ltd. | EL Global Trade Ltd | Pollogen Ltd. |
| Device Class | Class II | Class II | Class II |
| Classification Panel | General and Plastic Surgery | General and Plastic Surgery | General and Plastic Surgery |
| Product Code | PBX, GEI | GEI, PBX | GEI |
| Regulation Number | 21 CFR 878.4400 | 21 CFR 878.4400 | 21 CFR 878.4400 |
| Regulation Description | • Massager, Vacuum, Radio Frequency Induced Heat • Electrosurgical cutting and coagulation device and accessories | • Massager, Vacuum, Radio Frequency Induced Heat • Electrosurgical cutting and coagulation device and accessories | • Electrosurgical cutting and coagulation device and accessories |
| Intended use | The OptiLUV device is a handheld device intended to emit energy in the RF spectrum to provide topical heating for the purpose of elevating facial tissue temperature for a temporary relief of minor pain and the temporary increase in local blood circulation. | The RF-Relief is an over the counter hand held device intended to emit energy in RF spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscular and joint pain and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local blood circulation, and temporary relaxation of muscles. | Pollogen's STOP U is intended for use in the non-invasive treatment of mild to moderate facial wrinkles and rhytides. |
| Mode of Operation | RF energy emitted from multiple bipolar electrodes via a delivery handpiece | RF energy emitted from multiple bipolar electrodes via a delivery handpiece | RF energy emitted from multiple bipolar electrodes via a delivery handpiece |
Traditional 510(k) Submission - OptiLUV Device - Pollogen Ltd.
{8}
Collogen
BY LUMENIS
| Feature | OptiLUV Proposed Device | RF-Relief (K181095) Predicate Device | STOP U (K140255) Reference Device |
| --- | --- | --- | --- |
| Treatment area | Face | Treatment area not specified | Face |
| Energy source | RF (Bipolar) | RF (Bipolar) | RF (Bipolar) |
| Frequency | 1±0.05 MHz | 1±0.05 MHz | 1±0.05 MHz |
| Number of electrodes | 4 electrodes (2 pairs) | 4 electrodes (2 pairs) | 4 electrodes (2 pairs) |
| Intended population | Adult Users | Adult Users | Adult Users |
| Use environment | Prescription Use – Rx Only | Home Use, OTC- self-operation by an untrained user. | Prescription Use – Rx Only |
| Energy source | RF (Bipolar) | RF (Bipolar) | RF (Bipolar) |
| Energy levels | 1.8 watts
1 energy level: 1.8 watts | Up to 10 watts
3 energy levels: 6, 8 or 10 watts | 5.7 watts
1 energy level: 5.7 watts |
| Safety mechanism | Thermal sensor for skin temperature measurements | Thermal sensor for skin temperature measurements | Thermal sensor for skin temperature measurements |
| Electrical requirements | 100-240 V
50-60 Hz | 100-240 V
50-60 Hz | 100-240 V
50-60 Hz |
| Safety/EMC | • IEC 60601-1
• IEC 60601-1-2
• IEC 60601-2-2
• IEC 60601-1-11 | • IEC 60601-1
• IEC 60601-1-2
• IEC 60601-2-2
• IEC 60601-1-11 | • IEC 60601-1
• IEC 60601-1-2
• IEC 60601-2-2
• IEC 60601-1-11 |
| Non-sterile reusable | Non-sterile Reusable | Non-sterile Reusable | Non-sterile Reusable |
| Biocompatibility | All parts that are in contact with user comply with the requirements of ISO-10993-1. | All parts that are in contact with user comply with the requirements of ISO-10993-1. | All parts that are in contact with user comply with the requirements of ISO-10993-1. |
| Software | The software was verified and validated according to the FDA guidance. | The software was verified and validated according to the FDA guidance. | The software was verified and validated according to the FDA guidance. |
## Conclusion
The OptiLUV and the predicate device have the same intended use/indications for use and similar technological characteristics and principles of operation. The differences between the OptiLUV Device and its predicate do not raise new concerns of safety or effectiveness, as demonstrated by comparative bench performance testing and clinical study. Based on the totality of evidence, the OptiLUV Device is found substantially equivalent to the predicate device.
Traditional 510(k) Submission - OptiLUV Device - Collogen Ltd.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.