← Product Code [PBX](/submissions/SU/subpart-e%E2%80%94surgical-devices/PBX) · K230659

# Zionic Aesthetic (K230659)

_Termosalud · PBX · Jun 8, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/PBX/K230659

## Device Facts

- **Applicant:** Termosalud
- **Product Code:** [PBX](/submissions/SU/subpart-e%E2%80%94surgical-devices/PBX.md)
- **Decision Date:** Jun 8, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Zionic Aesthetic is intended to provide topical heating for the purpose of elevating tissue temperature for treatment of selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The massage device provided is intended to provide a temporary reduction in the appearance of cellulite.

## Device Story

Zionic Aesthetic is a resistive radiofrequency (RF) device used in clinical settings by healthcare professionals. It utilizes two rolling electrodes of varying sizes to deliver monopolar RF energy (up to 200W at 470kHz) to body tissues while simultaneously performing a mechanical massage. The device transforms electrical energy into electromagnetic waves that penetrate tissue, generating heat to elevate local tissue temperature. The clinician operates the device via a touchscreen interface to control treatment parameters. By maintaining a target tissue temperature (e.g., 40°C), the device aims to provide therapeutic benefits including pain relief, reduction of muscle spasms, improved local circulation, and temporary reduction in cellulite appearance. The device is designed for prescription use.

## Clinical Evidence

No clinical studies were conducted. Evidence is based on bench testing, including thermal performance validation (maintaining 40°C ± 2°C for 10 minutes) and verification of output power, frequency, and voltage. Safety testing adhered to IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2 standards.

## Technological Characteristics

Resistive monopolar radiofrequency device; 470kHz frequency; max 200W output power. Features rolling electrodes (57/90mm). Class I type BF electrical protection. Touchscreen interface. Software-controlled. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 62304, EN ISO 14971.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Nuera Tight RF ([K200359](/device/K200359.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

June 8, 2023

TermoSalud % Aubrey Thompson Regulatory Consultant Hoy and Associates Regulatory Consulting 1830 Bonnie Way Sacramento, California 95825

Re: K230659

Trade/Device Name: Zionic Aesthetic Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: PBX Dated: March 8, 2023 Received: March 9, 2023

Dear Aubrey Thompson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

| Mark<br>Trumbore -S |  |  |  | Digitally signed by<br>Mark Trumbore -S<br>Date: 2023.06.08<br>15:25:02 -04'00' |
|---------------------|--|--|--|---------------------------------------------------------------------------------|
|---------------------|--|--|--|---------------------------------------------------------------------------------|

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K230659

Device Name Zionic Aesthetic

#### Indications for Use (Describe)

The Zionic Aesthetic is intended to provide topical heating tor the purpose of elevating tissue temperature for treatment of selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

The massage device provided is intended to provide a temporary reduction in the appearance of cellulite.

| Type of Use (Select one or both, as applicable)                   |
|-------------------------------------------------------------------|
| <span>     Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
| <span>     Over-The-Counter Use (21 CFR 801 Subpart C)   </span>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary ZIONIC AESTHETIC K230659

This 510(K) Summary of safety and effectiveness for the ZIONIC AESTHETIC is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.

| Applicant                         | TermoSalud                                        |
|-----------------------------------|---------------------------------------------------|
| Address                           | Ataulfo Friera Tarfe, 8 -33211 Gijón, Spain       |
| Contact Person                    | Aubrey Thompson, Regulatory Consultant            |
| Contact Information               | aubreythompson@hoyregulatory.com<br>(323)533-8994 |
| Preparation Date                  | June 8, 2023                                      |
| Device Trade Name                 | ZIONIC AESTHETIC                                  |
| Common Name                       | Massager, vacuum, radio frequency induced heat    |
| Regulation Number                 | 878.4400                                          |
| Product Code                      | PBX                                               |
| Regulatory Class                  | II                                                |
| Legally Marketed Predicate Device | Nuera Tight RF (K200359)                          |

# Device Description:

The ZIONIC ASTHETIC is a resistive radiofrequency device which output energy can reach up to 200W. It has two different electrodes with different sizes, to adapt the treatment to the different body areas. The electrodes roll to perform a massage while applying the radiofrequency energy.

# Indications for use:

The ZIONIC AESTHETIC equipment is indicated to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

The massage device is intended to provide a temporary reduction in the appearance of cellulite.

{4}------------------------------------------------

# 510(K) Summary ZIONIC AESTHETIC K230659

# Substantial Equivalence—Indications for Use:

| Subject Device – K230659                                                                                                                                                                                                                                                                                                                                             | Predicate Device -K200359                                                                                                                                                                                                                                                                                                                                          | Comparison |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| The ZIONIC AESTHETIC is intended:<br>• to provide topical heating for the purpose of elevating<br>tissue temperature for the treatment of selected medical<br>conditions such as temporary relief of pain, muscle spasms,<br>and increase local circulation;<br>• The massage device is intended to provide a temporary<br>reduction in the appearance of cellulite. | The NuEra Tight RF Family is intended:<br>• to provide topical heating for the purpose of elevating<br>tissue temperature for the treatment of selected<br>medical conditions such as temporary relief of pain, muscle<br>spasms, and increase local circulation;<br>• to provide, with a massage device, a temporary reduction in<br>the appearance of cellulite. | Same       |

{5}------------------------------------------------

# K230659

| Technical Comparison  | ZIONIC AESTHETIC                                                                                                                                       | Nuera Tight RF – K200359                                                                                                                               | Comparison                                                                                                                                                                                                                                                                                      |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Modes of action       | Monopolar radiofrequency                                                                                                                               | Monopolar and bipolar radiofrequency                                                                                                                   | Same. The Nuera Tight RF device contains an additional bipolar handpiece that the Zionic does not.                                                                                                                                                                                              |
| Mechanism of action   | Electromagnetic waves penetrate in exposed tissue and produce heat, increasing the temperature on a certain part of the body for therapeutic purposes. | Electromagnetic waves penetrate in exposed tissue and produce heat, increasing the temperature on a certain part of the body for therapeutic purposes. | Same                                                                                                                                                                                                                                                                                            |
| Clinical use          | Prescription Use                                                                                                                                       | Prescription Use                                                                                                                                       | Same                                                                                                                                                                                                                                                                                            |
| Electrical Protection | Class I type BF                                                                                                                                        | Class I type BF                                                                                                                                        | Same                                                                                                                                                                                                                                                                                            |
| User Interface        | Touchscreen                                                                                                                                            | Touchscreen                                                                                                                                            | Same                                                                                                                                                                                                                                                                                            |
| Firmware Controlled   | Yes                                                                                                                                                    | Yes                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                            |
| Type of energy        | Radiofrequency waves                                                                                                                                   | Radiofrequency waves                                                                                                                                   | Same                                                                                                                                                                                                                                                                                            |
| Temperature Control   | Yes                                                                                                                                                    | Yes                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                            |
| Frequency             | 470kHz                                                                                                                                                 | 470 kHz; 1 MHz; 2 MHz; 4 MHz ; 6 MHz                                                                                                                   | ZIONIC AESTHETIC working frequency is identical to one of the working frequencies of the predicate device Nuera Tight RF.                                                                                                                                                                       |
| Output RF Power       | Max 200 W                                                                                                                                              | Max 250 W                                                                                                                                              | Nearly identical. ZIONIC AESTHETIC maximum power is 20% lower than NueraTight RF, which does not generate safety or performance issues, as an output power of 200W is enough to achieve the desired temperature for the effectiveness of the treatment, and lowering the output power makes the |
|                       |                                                                                                                                                        |                                                                                                                                                        | device safer.                                                                                                                                                                                                                                                                                   |
| Voltage               | 115VAC (with external autotransformer) or 230VAC                                                                                                       | 100-240VAC                                                                                                                                             | Same                                                                                                                                                                                                                                                                                            |
|                       | Handpieces                                                                                                                                             |                                                                                                                                                        |                                                                                                                                                                                                                                                                                                 |
|                       | ZIONIC AESTHETIC                                                                                                                                       | Nuera Tight                                                                                                                                            |                                                                                                                                                                                                                                                                                                 |
| Transfer technology   | Monopolar Resistive                                                                                                                                    | Resistive and capacitive monopolar and bipolar                                                                                                         | Predicate device includes both transfer modes, monopolar and bipolar, resistive and capacitive.                                                                                                                                                                                                 |
| Electrode area        | 57/90mm                                                                                                                                                | 20/30/40/60/80/100mm                                                                                                                                   | Very similar- the Nuera Tight device includes more variety of electrode sizes, with lower contact surface that are not included with the Zionic.                                                                                                                                                |

{6}------------------------------------------------

# 

{7}------------------------------------------------

# 510(K) Summary ZIONIC AESTHETIC K230659

#### Performance Testing

Verification and validation activities were successfully completed and establish that the ZIONIC AESTHETIC control unit performs as intended. Testing included the following:

- IEC 60601-1:2005 + Corr.1:2006 + A1:2012; Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
- IEC 60601-1-2:2014 + A1:2020; Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests;
- . IEC 60601-2-2:2017; Medical Electrical Equipment - Part 2-2: Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories;
- IEC 62304:2006+A1:2015; Medical Device Software Life Cycle Processes; ●
- EN ISO 14971:2019+A11:2021; Medical Devices Application Of Risk Management To ● Medical Devices

Software verification and validation testing was conducted, and documentation provided in accordance with FDA's Guidance or the Content of Premarket Submissions for Software Contained in Medical Devices.

In addition, tests were performed to demonstrate that the device can maintain a temperature of 40 treatment C for 10 minutes of treatment for every electrode used. Performance testing was performed to demonstrate the accuracy of the output power, output frequency, and output voltage.

Clinical Evidence – N/A. No clinical studies were conducted as part of this submission.

# Conclusion

The ZIONIC AESTHETIC and NueraTight RF are identical in their indications for use and nearly identical in their overall design. The key specification to compare for monopolar radiofrequecy transfer technology is the output power, which is slightly lower in the ZIONIC AESTHETIC. This difference in the maximum output power does not generate safety or performance issues, but increases the safety of the device. Both devices underwent thermal testing to show that they can maintain the desired temperature during treatment. Tests for the Zionic Aesthetic included two patients and three test sites – abdomen, thighs, and arms—and a temperature of 40° C (+2 °C) was maintained for 10 minutes of treatment. Test performed on the predicate device showed the same performance. The device maintained a temperature of 40°C (+2°C) for 15 minutes when test sites and 2 patients. It can be concluded, therefore, that performance between the ZIONIC AESTHETIC and the predicate device are substantially equivalent and therefore the two devices are substantially equivalent.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/PBX/K230659](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/PBX/K230659)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
