ProGear® Surgical Mask with Oxafence®, Model AV82030

K231741 · Prestige Ameritech · OUK · Sep 5, 2024 · General, Plastic Surgery

Device Facts

Record IDK231741
Device NameProGear® Surgical Mask with Oxafence®, Model AV82030
ApplicantPrestige Ameritech
Product CodeOUK · General, Plastic Surgery
Decision DateSep 5, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Intended Use

The ProGear® Surgical Mask with Oxafence® is a single-use, disposable device used to cover the nose and mouth of the wearer to protect from transfer of microorganisms, body fluids and particulates. The outer layer of the face mask is coated with methylene blue (0.47 micrograms per square cm area). The ProGear® Surgical Mask with Oxafence® inactivates of the following test viruses after five minutes of contact with the treated surface of the face mask in laboratory (in vitro) tests under a lighting intensity of 500 ± 25 lux. Influenza A/PR/8/34/H1N1 (TC adapted, ATCC VR-1469), 3-log or 99.9% inactivation Betacoronavirus 1 OC43 (ATCC VR-1558), 4-log or 99.99% inactivation Feline Calicivirus F-9 (ATCC VR- 782), 4-log or 99.99% inactivation Antiviral efficacy on other viruses has not been evaluated. Correlations between in vitro testing results and any clinical event have not been evaluated.

Device Story

ProGear® Surgical Mask with Oxafence® is a single-use, disposable surgical mask; outer layer coated with methylene blue (0.47 µg/cm²). Designed to cover nose and mouth; provides physical barrier against microorganisms, body fluids, and particulates. Methylene blue coating acts as an antiviral agent; inactivates specific viruses (Influenza A H1N1, Betacoronavirus 1 OC43, Feline Calicivirus) upon 5-minute contact under 500 ± 25 lux lighting. Used in clinical or general settings; worn by healthcare personnel or general public. No clinical data correlates in vitro inactivation to clinical outcomes; antiviral efficacy limited to tested viruses.

Clinical Evidence

Bench testing only. In vitro antiviral efficacy testing demonstrated 3-log (99.9%) inactivation of Influenza A/PR/8/34/H1N1 and 4-log (99.99%) inactivation of Betacoronavirus 1 OC43 and Feline Calicivirus F-9 after 5 minutes of contact with the treated surface under 500 ± 25 lux lighting. No clinical data provided; correlations between in vitro results and clinical events not evaluated.

Technological Characteristics

Single-use, disposable surgical mask. Outer layer treated with methylene blue (0.47 µg/cm²). Antiviral mechanism relies on light-activated (500 ± 25 lux) chemical inactivation. Form factor: standard surgical mask.

Indications for Use

Indicated for use as a single-use, disposable surgical mask to cover the nose and mouth of the wearer to protect against the transfer of microorganisms, body fluids, and particulates. Suitable for Over-The-Counter use.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. September 5, 2024 Prestige Ameritech Danica Hofheinz Vice President of Operations 7201 Iron Horse Blvd. North Richland Hills, Texas 76180 Re: K231741 Trade/Device Name: ProGear® Surgical Mask with Oxafence®, Model AV82030 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: OUK Dated: August 27, 2024 Received: August 27, 2024 Dear Danica Hofheinz: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, Image /page/2/Picture/3 description: The image shows the name "Bifeng Qian -S" in a large, clear font. The text is horizontally aligned and appears to be the primary focus of the image. The background is plain, ensuring the text is easily readable. Bifeng Qian, M.D., Ph.D. Assistant Director DHT4C: Division of Infection Control OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K231741 Device Name ProGear® Surgical Mask with Oxafence®, Model AV82030 ## Indications for Use (Describe) The ProGear® Surgical Mask with Oxafence® is a single-use, disposable device used to cover the nose and mouth of the wearer to protect from transfer of microorganisms, body fluids and particulates. The outer layer of the face mask is coated with methylene blue (0.47 micrograms per square cm area). The ProGear® Surgical Mask with Oxafence® inactivates of the following test viruses after five minutes of contact with the treated surface of the face mask in laboratory (in vitro) tests under a lighting intensity of 500 ± 25 lux. Influenza A/PR/8/34/H1N1 (TC adapted, ATCC VR-1469), 3-log or 99.9% inactivation Betacoronavirus 1 OC43 (ATCC VR-1558), 4-log or 99.99% inactivation Feline Calicivirus F-9 (ATCC VR- 782), 4-log or 99.99% inactivation Antiviral efficacy on other viruses has not been evaluated. Correlations between in vitro testing results and any clinical event have not been evaluated. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------| | <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...