SLIM E30 II CO2 LASER

K110984 · Lasering S.R.L. · ONG · Jul 22, 2011 · General, Plastic Surgery

Device Facts

Record IDK110984
Device NameSLIM E30 II CO2 LASER
ApplicantLasering S.R.L.
Product CodeONG · General, Plastic Surgery
Decision DateJul 22, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The SLIM Evolution II CO2 Laser and delivery device accessories are intended for use in surgical applications that require ablation, vaporization, excision, incision and coagulation of soft tissue in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology. Dermatology, Plastic Surgery and General Surgery procedures. Laser skin resurfacing Treatment of furrows and wrinkles Removal of skin tags, actinic keratosis, acne scars, keloids, tattos, telangiectasia, squamous and basel cell carcinoma, warts and uneven pigmentation. Treatment of cysts, abscesses, hemorrhoids and other soft tissue applications. Blepharoplasty Site preparation for hair transplants MiXto fractional scanner only for treatment of wrinkles and skin resurfacing Dental procedure including. Periodontal procedures such as - gingivectomy, removal of hyperplasias, gingivoplasty (incision and excision) Oral Surgery procedures such as - aphous ulcer excision, frenectomy, benign/malignant lesion ablation, operculectomy and homeostasis Podiatry procedures. Ablation, vaporization and excision of soft tissue lesions such as ingrown nail, fungal nail, porokeratoma, matrixectomy and verrucae vulgares. Otorhinolaryngology (ENT) procedures. Treatment of leukoplakia of larynx, nasal obstruction, rhinophyma, verrucrea vulgares, choanal atresia, rhinophyma, LAUP and papillomatosis polyps. Gynecology Treatment of condyloma acuminata, cervical intraevithelial neoplasia. leukoplakia and vulvar/vaginal intraepithelial neoplasia. cervical dysplasia. Laparoscopic treatment endometrial lesions, ablation of endometriosis, fimbroplasty, Tubal microsurgery, salpingostomy, histerectomy, uterine myomas and fibroids

Device Story

SLIM Evolution II is a CO2 laser system for soft tissue surgery. Device delivers laser energy via accessories to perform ablation, vaporization, excision, incision, and coagulation. Includes MiXto fractional scanner for skin resurfacing and wrinkle treatment. Operated by physicians in clinical settings (dermatology, plastic surgery, general surgery, gynecology, podiatry, dental, ENT). Laser energy interacts with soft tissue to achieve surgical goals; provides precise tissue removal or coagulation. Benefits include controlled tissue management across multiple surgical specialties.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and intended use.

Technological Characteristics

CO2 laser system; includes delivery accessories and MiXto fractional scanner. Operates as a surgical laser for soft tissue ablation, vaporization, excision, incision, and coagulation. Class II device under 21 CFR 878.4810.

Indications for Use

Indicated for patients requiring soft tissue ablation, vaporization, excision, incision, or coagulation across dermatology, plastic surgery, general surgery, gynecology, podiatry, dental, and ENT specialties. Specific applications include skin resurfacing, lesion removal, periodontal/oral surgery, podiatric nail/lesion treatment, and gynecological/ENT surgical procedures.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three stylized lines or strokes. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Lasering S.r.l % Transamerican Technologies International Mr. Allen R. Howes 2246 Camino Ramon San Roman, California 94583 JUL 2 2 2 2011 Re: K110984 Trade/Device Name: SLIM Evolution II CO2 Laser and Delivery Device Accessories Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: ONG, GEX Dated: June 1, 2011 Received: June 22, 2011 Dear Mr. Howes: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {1}------------------------------------------------ Page 2 - Mr. Allen R. Howes or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Eunice Keith Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number K110984 Device Name: SLIM Evolution II CO2 Laser and Delivery Device Accessories. Indications For Use: The SLIM Evolution II CO2 Laser and delivery device accessories are intended for use in surgical applications that require ablation, vaporization, excision, incision and coagulation of soft tissue in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology. ### Dermatology, Plastic Surgery and General Surgery procedures. Laser skin resurfacing Treatment of furrows and wrinkles Removal of skin tags, actinic keratosis, acne scars, keloids, tattos, telangiectasia, squamous and basel cell carcinoma, warts and uneven pigmentation. Treatment of cysts, abscesses, hemorrhoids and other soft tissue applications. Blepharoplasty Site preparation for hair transplants MiXto fractional scanner only for treatment of wrinkles and skin resurfacing X Prescription Use (part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Neil R.P. Ogden for mkm Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K110984 6-1 Page 1 of 2 {3}------------------------------------------------ ## Indications for Use 510(k) Number K110984 Device Name: SLIM Evolution II CO2 Laser and Delivery Device Accessories. Indications For Use: Continued from previous page #### Dental procedure including. Periodontal procedures such as - gingivectomy, removal of hyperplasias, gingivoplasty (incision and excision) Oral Surgery procedures such as - aphous ulcer excision, frenectomy, benign/malignant lesion ablation, operculectomy and homeostasis #### Podiatry procedures. Ablation, vaporization and excision of soft tissue lesions such as ingrown nail, fungal nail, porokeratoma, matrixectomy and verrucae vulgares. #### Otorhinolaryngology (ENT) procedures. Treatment of leukoplakia of larynx, nasal obstruction, rhinophyma, verrucrea vulgares, choanal atresia, rhinophyma, LAUP and papillomatosis polyps. #### Gynecology Treatment of condyloma acuminata, cervical intraevithelial neoplasia. leukoplakia and vulvar/vaginal intraepithelial neoplasia. cervical dysplasia. Laparoscopic treatment endometrial lesions, ablation of endometriosis, fimbroplasty, Tubal microsurgery, salpingostomy, histerectomy, uterine myomas and fibroids Prescription Use (part21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE ABOVE THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Nifer Podrea for man (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 6-2 510(k) Number Page 2 of 2
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...