← Product Code [ONG](/submissions/SU/subpart-e%E2%80%94surgical-devices/ONG) · K091446

# MODIFICATION TO LUX1540 HANDPIECE AND LUX 1440 HANDPIECE (K091446)

_Palomar Medical Products, Inc. · ONG · Oct 27, 2009 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONG/K091446

## Device Facts

- **Applicant:** Palomar Medical Products, Inc.
- **Product Code:** [ONG](/submissions/SU/subpart-e%E2%80%94surgical-devices/ONG.md)
- **Decision Date:** Oct 27, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Lux1540 handpiece is intended for use in dermatological procedures requiring: coagulation of soft tissue, skin resurfacing procedures, melasma, acne scars, and surgical scars. The Lux 1440 handpiece is intended for use in dermatological procedures requiring: coagulation of soft tissue and skin resurfacing procedures.

## Device Story

Lux1540 and Lux1440 are fractional laser handpieces designed for attachment to StarLux Pulsed Light and Laser System consoles. System includes power supply, chiller, electronics, portable cart, and footswitch. Handpieces deliver laser energy to skin to perform coagulation and resurfacing. Used in clinical settings by trained healthcare providers. Output affects tissue via thermal coagulation; clinical benefit includes skin resurfacing and scar/melasma treatment.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and performance data comparison to predicate devices.

## Technological Characteristics

Fractional laser handpieces; connect to StarLux console (power, chiller, electronics). Energy source: laser. Form factor: handpiece. Intended for dermatological soft tissue coagulation and resurfacing.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Lux1540TM Fractional Laser Handpiece ([K080244](/device/K080244.md))
- Lux 1440TM Fractional Laser Handpiece ([K073583](/device/K073583.md))
- Fraxel Re:store™ Laser System ([K070284](/device/K070284.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

U911446 page (1) of (2)

### 510(k) Summary

This 510(k) summary is being submitted in accordance with 21 CFR 807.92

#### 1. SUBMITTER'S INFORMATION

## OCT 2 7 2009

Name: Palomar Medical Technologies, Inc. 82 Cambridge Street ADDRESS: Burlington, MA 01803 Phone: (781) 993-2300 Fax: (781) 993-2330

Sharon Timberlake, MSHS, RAC, CCRA CONTACT: Director of Regulatory Affairs

DATE PREPARED: May 14, 2009

#### 2. DEVICE INFORMATION

| TRADE/PROPRIETARY NAME: | Lux1540TM Fractional Laser Handpiece,<br>Lux1440TM Fractional Laser Handpiece                                 |
|-------------------------|---------------------------------------------------------------------------------------------------------------|
| COMMON/USUAL NAME:      | Lux 1540, Lux1440                                                                                             |
| CLASSIFICATION NAME:    | Laser surgical instrument for use in general and<br>plastic surgery and in dermatology<br>(21 CFR § 878.4810) |

PRODUCT CODE: GEX

#### PREDICATE DEVICES 3.

Palomar Medical Technologies, Inc. Lux1540TM Fractional Laser Handpiece K080244

Palomar Medical Technologies, Inc. Lux 1440TM Fractional Laser Handpiece K073583

Reliant Technologies, Inc. Fraxel Re:store™ Laser System (Formerly SR1500) K070284

{1}------------------------------------------------

091446 Page (2) of

#### 4. INTENDED USE

The Lux1540 handpiece is intended for use in dermatological procedures requiring: coagulation of soft tissue, skin resurfacing procedures, melasma, acne scars, and surgical scars.

The Lux 1440 handpiece is intended for use in dermatological procedures requiring: coagulation of soft tissue and skin resurfacing procedures.

#### ડ. DEVICE DESCRIPTION

The Lux1540 Handpiece & Lux1440 Handpiece each attach to the StarLux Pulsed Light and Laser Systems. The complete StarLux System console consists of a power supply, chiller, electronics, portable cart, and a footswitch.

#### 6. PERFORMANCE DATA

The specifications and indications for use of the Lux1540 Handpiece &Lux1440 Handpiece are substantially equivalent to its predicate devices based on the data provided in the 510(k) Premarket Notification.

#### 7. SUBSTANTIAL EQUIVALENCE

The Lux1540 Handpiece and Lux1440 Handpiece are substantially equivalent to its predicate devices when used according to its intended use. The information that is provided in this 510(k) Premarket Notification demonstrates that the Lux1540 Handpiece and Lux1440 Handpiece also share the same technological characteristics, mechanism of action, intended use and physical properties to its predicates.

{2}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

Palomar Medical Products, Inc. % Ms. Sharon Timberlake, MSHS, RAC, CCRA Director of Regulatory Affairs 82 Cambridge Street Burlington, Massachusetts 01803

OCT 2 7 2009

Re: K091446

Trade/Device Name: Lux1440 Handpiece & Lux 1540 Handpiece Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONG Dated: September 25, 2009 Received: September 28, 2009

Dear Ms. Timberlake:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Sharon Timberlake, MSHS, RAC, CCRA

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/indcx.html.

Sincerely yours.

erely yours,

For

N. M.

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

KO91446

510(k) Number (if known):

Device Name: Lux1440 Handpiece & Lux1540 Handpiece

Indications for Use:

The Palomar Lux1540 handpiece is intended for use in dermatological procedures requiring: coagulation of soft tissue, skin resurfacing procedures, melasma, acne scars, and surgical scars.

The Palomar Lux1440 handpiece is intended for use in dermatological procedures requiring: coagulation of soft tissue and skin resurfacing procedures.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nefre Dalm for mixm

(Division Sign-Of (Division Sign of Surgical, Orthopedic, and Restorative Devices

510(k) Number K04

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONG/K091446](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONG/K091446)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
