E>EyeC

K200616 · E-Swin · ONF · Jun 11, 2020 · General, Plastic Surgery

Device Facts

Record IDK200616
Device NameE>EyeC
ApplicantE-Swin
Product CodeONF · General, Plastic Surgery
Decision DateJun 11, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The E>Eye is a prescription device intended for the treatment of rosacea.

Device Story

E>Eye C is an intense pulsed light (IPL) system consisting of a base unit housing electrical/electronic sub-assemblies and a handheld applicator connected by a cord. The applicator contains a flashlamp source of optical radiation. Used in a clinical setting by a healthcare professional, the device delivers pulsed light energy to the skin to treat rosacea. The provider operates the device to target affected areas, with the light energy intended to provide therapeutic benefit for the condition. The device is a prescription-only system.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by non-clinical performance testing, including electrical safety (IEC 60601-1, IEC 60601-2-57), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), photobiological safety (IEC 62471), and software verification and validation.

Technological Characteristics

Intense pulsed light (IPL) system using a flashlamp energy source. Wavelength range: 580-1200nm (cutoff <580nm). Pulse duration: 170ms (multiple pulse). Energy density: 8-13 J/cm2. Spot size: 7.5cm2. Complies with IEC 60601-1, IEC 60601-2-57, IEC 60601-1-2, IEC 60601-1-6, and IEC 62471.

Indications for Use

Indicated for the treatment of rosacea in patients requiring prescription-based intense pulsed light therapy.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ June 11, 2020 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. E-swin Emeric OBIN Regulatory Affairs Manager Rue des Cotes D'orval, ZA de la prevote Houdan, Yvelines 78550 France Re: K200616 Trade/Device Name: E>EyeC Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONF Dated: April 13, 2020 Received: April 13, 2020 Dear Emeric OBIN: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Colin Kejing Chen, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K200616 Device Name E>Eye C Indications for Use (Describe) The E>Eye is a prescription device intended for the treatment of rosacea Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. {3}------------------------------------------------ ## 510(k) SUMMARY ## 1) Applicant Name: E-SWIN Rue des Côtes d'Orval, ZA de la Pévôté 78550 Houdan, France Tel: +33 1 34 87 06 49 Fax: 01 70 248 123 ## 2) Contact Person: Emeric OBIN, Regulatory Affairs Manager Rue des Côtes d'Orval, ZA de la Pévôté 78550 Houdan, France Tel: +33 1 30 46 37 65 Fax: 01 70 248 123 Mail: emeric.o@esw-vision.com ## 3) Date Prepared: June 11, 2020 4) Trade Name: E>EYE C ## 5) Regulatory Information Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Product Code: ONF Class: Class II Regulation Number: 21 CFR 878.4810 > E-SWIN – SAS au capital de 195.000 euros - RCS Versailles 501 515 423 Rue des Côtes d'Orval, ZA de la Prévôté, 78550 Houdan- FRANCE tel : 01 34 87 06 49 1/4 {4}------------------------------------------------ Panel: General & Plastic Surgery Predicate Devices: Lumenis M22 System (K170060) , DEKA LUXEA LAZUR Pulsed Light Handpiece (K192539) Indications for Use: The complete phrasing of the indications for use statement is provided in the formal Indications for Use Statement (FDA Form 3881). ## 6) Intended use: The E>Eye is a prescription device intended for the treatment of rosacea. # 7) Device Description: The E>Eye C is an intense pulsed light system composed of a base housing the electric and electronics sub-assemblies, and a handheld piece "applicator" connected to the base by a cord. The applicator contains the source of optical radiation. {5}------------------------------------------------ # 8) Technological Characteristics and Substantial Equivalence: The intended use and indications for use of the E>EYE C are the same as the selected predicate devices. In addition, the same technological characteristics and principles of operation apply for the E>EYE C and the predicate device. | | Predicate Device | Predicate Device | Device | |------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------| | Device Name | Lumenis M22 System IPL<br>Handpiece<br>(K170060) | DEKA LUXEA LAZUR<br>Pulsed Light handpiece<br>(K192539) | E>EYE C | | Energy Medium | Flashlamp | Flashlamp | Flashlamp | | Wavelength Range | 400-1200nm | 420-950nm | 580-1200nm, cutoff<br>before 580nm | | Pulse Duration | 40-100ms - multiple<br>pulse | 30, 40, 50 ms | 170ms - multiple<br>pulse | | Energy Density | 10-56J/cm2 | 3-20J/cm2 | 8-13j/cm2 | | Spot size | 6cm2 | 6.2cm2 | 7.5cm2 | | Intended Use | Benign cutaneous<br>vascular lesions,<br>including port wine<br>stains, hemangiomas,<br>facial, truncal and leg<br>telangiectasias, erythema<br>of rosacea, angiomas and<br>spider angiomas,<br>poikiloderma of Civatte,<br>leg veins and venous<br>malformations | - The treatment of<br>benign cutaneous<br>vascular lesions including<br>port wine stains,<br>hemangiomas, facial,<br>truncal and leg<br>telangiectasias, rosacea,<br>erythema of rosacea,<br>angiomas and spider<br>angiomas, poikiloderma<br>of Civatte, leg veins and<br>venous malformations. | Treatment of<br>rosacea | Minor differences can be found in spot size, wavelength and energy density. Those differences do not raise additional safety or effectiveness questions. The E>EYE C has the same indications for use, same technology, comparable pulse duration as its predicates. > E-SWIN - SAS au capital de 195.000 euros - RCS Versailles 501 515 423 Rue des Côtes d'Orval, ZA de la Prévôté, 78550 Houdan- FRANCE tel : 01 34 87 06 49 3/4 {6}------------------------------------------------ ## 9) PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination: ### NON-CLINICAL ## Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the E>EYE C, composed of its base and its applicator/handpiece. The system complies with the IEC 60601-1, IEC 60601-2-57 standards for safety and the IEC 60601-1-2 standard for EMC. ## Usability Usability testing has been conducted, and the E>EYE C complies with IEC 60601-1-6. ## Photobiological safety Safety testing has been conducted, and the E>EYE C complies with IEC 62471. ## Software Verification and Validation Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could not directly result in serious injury or death to the patient or operator. ## Clinical Data None. ## 10) Conclusions: The E>EYE C was found to perform as well as its predicate, to be as safe and effective for its intended use as its predicate, and is substantially equivalent to its predicate device without raising any new safety and/or effectiveness issues. > E-SWIN - SAS au capital de 195.000 euros - RCS Versailles 501 515 423 Rue des Côtes d'Orval, ZA de la Prévôté, 78550 Houdan- FRANCE tel : 01 34 87 06 49 4/4
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...