← Product Code [ONF](/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF) · K131859

# MULTIFUNCTIONAL SERIES (K131859)

_Beijing Honkon Technologies Co., Ltd. · ONF · Oct 18, 2013 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF/K131859

## Device Facts

- **Applicant:** Beijing Honkon Technologies Co., Ltd.
- **Product Code:** [ONF](/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF.md)
- **Decision Date:** Oct 18, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Multifunctional Series device (inclusive of the handpiece used to deliver pulsed-light energy) are indicated for use in surgical, aesthetic and cosmetic applications in hair removal, moderate inflammatory acne vulgaris, and benign cutaneous vascular lesions.

## Device Story

The Multifunctional Series (models Aeslight-S3D, HONKON-S3C, HONKON-M40e+, HONKON-M80e+) are Intense Pulsed Light (IPL) systems. The device delivers pulsed light energy (530nm-1200nm) via a sapphire handpiece to target chromophores based on the principle of selective thermolysis. By limiting pulse width below the thermal relaxation time of skin, the device causes thermal damage to targets while sparing surrounding tissue. The system is intended for use by clinicians in surgical, aesthetic, and cosmetic settings. Operators select specific filter/wavelength settings based on the condition (hair removal, acne, or vascular lesions) and skin type. The output energy and pulse parameters are controlled by the user to achieve therapeutic effects. The device benefits patients by providing non-invasive treatment for dermatologic conditions.

## Clinical Evidence

Bench testing only. Performance testing verified output energy, spot size, and compliance with electrical safety (IEC 60601-1) and electromagnetic compatibility (IEC 60601-1-2) standards. No clinical data provided.

## Technological Characteristics

IPL system; sapphire crystal delivery; wavelength ranges 530-1200nm; energy range 20-50J/cm²; pulse duration 1-25ms; pulse delay 0.1-40ms; max power 1250-2000W; spot sizes 8x40mm and 15x60mm. Complies with IEC 60601-1 and IEC 60601-1-2.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Intense Pulsed Light (IPL) Systems ([K122995](/device/K122995.md))
- IPULSELIGHT IPL SYSTEM ([K093627](/device/K093627.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

S3 510(k) Summary

#### Project #:M0382013B

# Section #3 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number: ____

- 1. Date of Prepared: 06/07/2013

**OCT 18 2013**

2. Sponsor

Beijing Honkon Technologies Co., Ltd. No.3 Building, No.11 Yard, Kangding Street, BDA, 100176,Beijing, P.R.China

Establishment Registration Number: Not yet registered

Contact Person: Li Zhao Position: Management Representative Tel: +86 10 56370050 ext. 508 Fax: +86 10 56370067 ext. 610 Email: small@honkonlaser.com

- 3. Submission Correspondent
Diana Hong & Tarzan. Wang Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net

{1}------------------------------------------------

S3 510(k) Summary

- 4. Proposed Device Identification
Proposed Device Name: Multifunctional Series Proposed Device Common Name: Intense Pulse Light (IPL)

Regulatory Information Classification Name: powered laser surgical instrument Classification: II; Product Code: ONF; Regulation Number: 21 CFR 878.4810; Review Panel: General & Plastic Surgery;

Intended Use Statement:

The Multifunctional Series device (inclusive of the handpiece used to deliver pulsed-light energy) are indicated for use in surgical, aesthetic and cosmetic applications in hair removal, moderate inflammatory acne vulgaris, and benign cutaneous vascular lesions.

| Conditions          | Filter setting<br>and<br>wavelength<br>scope | Skin type |          |          |          |          |
|---------------------|----------------------------------------------|-----------|----------|----------|----------|----------|
|                     |                                              | I         | II       | III      | IV       | V        |
| Hair<br>Removal     |                                              | 610-1200  | 610-1200 | 610-1200 | 610-1200 | 610-1200 |
| Acne                |                                              | 530-1200  | 530-1200 | 530-1200 | 530-1200 | 530-1200 |
| Vascular<br>Lesions |                                              | 585-1200  | 585-1200 | 585-1200 | 585-1200 | 585-1200 |

- 5. Predicate Device Identification
510(k) Number: K122995 Predicate Device Name: Intense Pulsed Light (IPL) Systems Manufacturer: Beijing KES Biology Technology Co., Ltd.

510(k) Number: K093627 Predicate Device Name: IPULSELIGHT IPL SYSTEM Manufacturer: Shanghai APOLO Medical Technology Co., Ltd

- 6. Device Description

{2}------------------------------------------------

The Multifunctional Series device are intense pulsed light system which delivers intense pulsed light at a wavelength ranging from 530nm-1200nm. Intense Pulsed Light (IPL) systems work on the principles of selective thermolysis. That is, causing thermal damage to target chromophores by using light of appropriate wavelength in pulses that exceeds the chromophores" thermal relaxation time but sparing normal skin by limiting the pulse width below the thermal relaxation time for skin.

IPL. Systems are different from lasers in that they deliver many wavelengths in each pulse of light instead of just one wavelength. Generally, IPL enhances penctration without using excessive energy levels and enables targeting of specific chromophores.

Based on this, The Multifunctional Series device (inclusive of the handpiece used to deliver pulsed-light energy) are indicated for use in surgical, aesthetic applications in hair removal, acne, and blood vessel lesions.

The proposed device includes four models as Aeslight-S3D, HONKON-S3C. HONKON-M40e+ and HONKON-M80e+.

#### 7. Non-Clinical Test Conclusion

Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- a) IEC 60601-1. Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988: Amendment i, 1991-11, Amendment 2, 1995.
- b) IEC 60601-1-2:2007. Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests.
- c) Performance Test (output energy, spot size)

#### 8. Technological Characteristics Comparison

The proposed device has the same technological characteristics with the predicate device, such as light medium, wavelength, control method and intended use.

- 9. Substantially Equivalent Conclusion Based on the comparison and analysis as following, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

{3}------------------------------------------------

・・

|                                                       |                                                                                                                                                                                                                                                                                                                                 | מערכת המועד היה היה להתנועד האור                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |               |  |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|--|
|                                                       | Proposed Device                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |               |  |
| ITEM                                                  | Multifunctional Series                                                                                                                                                                                                                                                                                                          | Intense Pulsed Light (IPL)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | IPULSELIGHT IPL SYSTEM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Remark        |  |
|                                                       |                                                                                                                                                                                                                                                                                                                                 | Systems(K 122995)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | (K093627)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |               |  |
| Product                                               |                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |               |  |
|                                                       | ONF                                                                                                                                                                                                                                                                                                                             | ONF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | ONF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | SE            |  |
| Code                                                  |                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |               |  |
| Regulation                                            | 21 CFR 878,4810                                                                                                                                                                                                                                                                                                                 | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | ટાને          |  |
| No                                                    |                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |               |  |
| Class                                                 | ll                                                                                                                                                                                                                                                                                                                              | ll                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 11                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | ટાદ           |  |
| Intended<br>Use                                       | The Multifunctional<br>Series<br>(inclusive<br>device<br>of<br>the<br>handpiece used to deliver<br>pulsed-light<br>energy)<br>are<br>indicated for use in surgical,<br>aesthetic<br>and<br>cosmetic<br>applications in hair removal.<br>moderate inflammatory acne<br>vulgaris.<br>and<br>benign<br>cutaneous vascular lesions. | آل المحافظ من المحافظ المحافظ المحافظ المحافظ المحافظ المحافظ المحافظ المحافق المحافق المحافق المحافق المحافق المحافق المحافق المحافق المحافق المحافق المحافق المحافق المحافق<br>Intense Pulsed Light (IPL) S<br>(inclusive<br>ystems<br>of<br>the<br>handpiece used to deliver<br>energy )<br>pulsed-light<br>are<br>indicated for use in surgical,<br>aestheric<br>and<br>cosmetic<br>applications<br>in<br>permanent<br>hair reduction, treatment of<br>pigmented lesions, moderate<br>inflammatory acne vulgaris,<br>ephelides<br>(freckles),<br>and<br>vascular lesions. | IPulseLight IPL System (HS 300C,<br>HS 650) are identical with regard to<br>indications for use including<br>recommended filters to be used with<br>Fitzpatrick skin type, Both models<br>are intended for medical use in the<br>treatment of the following<br>dermatologic conditions:<br>Permanent hair reduction- long-term<br>stable reduction in number of hairs<br>re-growing after a treatment regiment<br>Treatment of:<br>- Moderate inflammatory acne<br>vulgaris<br>- Benign pigmented epidermal<br>lesions including dy schromia,<br>hyperpigmentation, melasma,<br>ephelides (freckles).<br>- Cutaneous lesions including scars<br>- Benign cutaneous vascular lesions<br>including port wine stains,<br>hemangiomas, facial, truncal and leg<br>telangiectasias, erythema of rosacea,<br>leg veins, spider anglomas and<br>venous malformations.<br>The integrated thermal cooling is<br>indicated for use in cooling the<br>epidermis at the treatment site prior<br>to, during, and after light treatment<br>in general aesthetic dermatologic and<br>plastic surgery procedures<br>· Reduce pain during light<br>treatment (via partial anesthesia from<br>cooling)<br>- Reduce discomfort during and/or<br>associated with light treatment<br>- Minimize thermal injury,<br>including thermal necrosis, to<br>non-larget skin and skin structures<br>during and/or associated with light<br>treatment, thus reducing possible<br>complications such as scabbing.<br>scarring, hyper - and/or hypo<br>pigmentation<br>- Allow the use of higher light or<br>laser fluencies for light treatments<br>(such as for hair removal and the<br>treatment of vascular or pigmented<br>lesions)<br>- Reduce potential side effects of<br>light treatments (such as for hair<br>removal and the treatment of<br>vascular or pigmented lesions) | ਟ ਦ           |  |
| ITEM                                                  | Proposed Device<br>Multifunctional Series<br>(Aeslight-S3D)                                                                                                                                                                                                                                                                     | Predicate Device<br>Intense Pulsed Light (IPL)<br>Systems(K122995)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Device<br>IPULSELIGHT IPL SYSTEM<br>(K093627)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Remark        |  |
| Light source                                          | Intense pulsed light                                                                                                                                                                                                                                                                                                            | Intense pulsed light                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Intense pulsed light                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | SE            |  |
| Wavelength                                            | 610-1200nm,<br>585-1200nm,<br>530-1200nm                                                                                                                                                                                                                                                                                        | 430-1200nm, 530-1200nm,<br>640-1200nm,<br>Optional: 480-1200nm.<br>560-1200nm, 590-1200nm,<br>690-1200nm, 750 -1200nm                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 420nm-1200, 510-1200nm,<br>560-1200nm, 610-1200nm,<br>640-1200nm (Standard):<br>480nm-1200, 585-1200nm,<br>690-1200nm, 755-1200nm<br>(Options)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Analysis<br>1 |  |
| Deliver system                                        | Sapphire                                                                                                                                                                                                                                                                                                                        | Sapphire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Sapphire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | SE            |  |
| Energy Range                                          | 20-50J/cm²                                                                                                                                                                                                                                                                                                                      | 10-60J/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 10-60J/cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Analysis<br>2 |  |
| Pulse Delay                                           | 0.1-40ms                                                                                                                                                                                                                                                                                                                        | 5 - 50ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 5 - 50ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Analysis<br>3 |  |
| Pulse Duration                                        | 1-25 ms                                                                                                                                                                                                                                                                                                                         | 1-20 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 2-20 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Analysis<br>4 |  |
| Max. Power                                            | 2000 W                                                                                                                                                                                                                                                                                                                          | 2000 W                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 1200 W                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | SE            |  |
| Spot size                                             | 8mm*40mm, 15mm*60mm                                                                                                                                                                                                                                                                                                             | MED-210: 15mmX50mm<br>(optional: 12mmX33mm,<br>15mmX35mm )<br>MED-230:<br>A: 12mm X33mm;<br>B: 15mmX50mm (optional:<br>15mmX35mm)                                                                                                                                                                                                                                                                                                                                                                                                                                             | 12x35mm, 15x50mm (Standard);<br>12x12mm, 12x25mm, 12x50mm,<br>15x35mm, 15x45mm (Options)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Analysis<br>5 |  |
| Setting for Specified Indication for use:Hair Removal |                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |               |  |
| Wavelength<br>(nm)                                    | 610-1200                                                                                                                                                                                                                                                                                                                        | 640-1200/690-1200/750-1200                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 610-1200/640-1200/690-1200                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |               |  |
| Energy Range<br>(J/cm2)                               | 20-50                                                                                                                                                                                                                                                                                                                           | 10-44                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 10-46                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |               |  |
| Pulse Duration<br>(ms)                                | 3-15 ms                                                                                                                                                                                                                                                                                                                         | 3-14                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 4-15                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Analysis<br>6 |  |
| Pulse Delay<br>(ms)                                   | 15-35                                                                                                                                                                                                                                                                                                                           | 16-32                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 15-30                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |               |  |
| Spot size                                             | 8mm X 40mm; 15mm X 60mm                                                                                                                                                                                                                                                                                                         | 12mm X 33mm; 15mm X 50mm;<br>15mm X 35mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 12mm X 25mm; 12mm X 35mm;<br>12mm X 50mm; 15mm X 45mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |               |  |
| Setting for Specified Indication for use: Acne        |                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |…

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF/K131859](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/ONF/K131859)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
