OR-LIGHT

K090724 · Yperion Technology · ONF · Oct 7, 2009 · General, Plastic Surgery

Device Facts

Record IDK090724
Device NameOR-LIGHT
ApplicantYperion Technology
Product CodeONF · General, Plastic Surgery
Decision DateOct 7, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Or-Light® is a non-coherent light-based device intended to carry out dermatologic and aesthetic care. It is meant to be handled by competent professionals trained in intense pulsed light technology. Two different applicators can be used on the Or-Light® device. The HR applicator (610-1000 nm) is indicated for: - Hair Removal for skintype I to IV. The SR applicator (400-1000 nm) is indicated for: - Benign pigmentary lesions for skintype I to III - Benign vascular lesions for skintype I to III - Inflammatory acne treatment for skintype I to III

Device Story

Or-Light® is a non-coherent intense pulsed light (IPL) system for dermatologic and aesthetic treatments. Device consists of a console with a power unit, electronic/software controls, and a touch-screen interface. Two interchangeable applicators (HR: 610-1000 nm; SR: 400-1000 nm) deliver light energy to the skin. Applicators feature an integrated closed-circuit water cooling system. Operated by trained skin professionals in clinical settings. Operator selects settings via touch screen to deliver light pulses. Clinical benefit includes hair removal and treatment of benign skin lesions/acne. Device output is controlled by internal software and electronics.

Clinical Evidence

Clinical data demonstrate the device can be used effectively and safely by a trained skin professional. No specific study metrics (e.g., sensitivity, specificity) provided in the summary.

Technological Characteristics

Non-coherent intense pulsed light (IPL) system. Two applicators: HR (610-1000 nm) and SR (400-1000 nm). Features closed-circuit water cooling. Controlled by internal electronics and software via touch-screen interface. Class II device.

Indications for Use

Indicated for hair removal (skin types I-IV) and treatment of benign pigmentary lesions, benign vascular lesions, and inflammatory acne (skin types I-III). Intended for use by trained skin professionals.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 090724 Page 1/2 # 510(k) SUMMARY ### Or-Light® YPERION dermatologic Intense Pulsed Light system Submitter's name, address, telephone number, contact person and Date YPERION Technology SAS 76-78 Av des Champs Elysées 75008 Paris France 00 33 1 56 89 10 40 Telephone : 00 33 1 45 27 68 33 Fax : Pascal Danet Contact person : General Manager 2. rue Beethoven 75016 Paris France 00 33 1 56 89 10 40 Telephone : 00 33 1 45 27 68 33 Fax : pdanet@ yperion.fr Email : February 26, 2009 Date prepared: #### Name of Device Device Trade Name: Or-Light® System YPL08011 Model: Intense Pulsed light System for cosmetic and dermatologic treatments Common use: Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Product code : GEX Regulation Number : 878 4810 #### Predicated device Yperion Technology : L600®; K030480 Danish Dermatologic Developpement (as DDD) : Ellipse IPPL ; K060516 #### System Description The Or-Light® is a non-coherent light based device intended to carry out dermatologic and aesthetic care on patients by skin professional. The system consists of a console containing a power unit controlled by electronic and software. Setting is selected by the operator through a display panel (touch screen). Light energy for the treatment is generated by the applicators connected to the device. Two type of applicator can be connected, one at a time. The applicators are in the waveband of OCT - 7 2009 {1}------------------------------------------------ 090572f Page 2/2 - 400 nm 1000 nm 트 - 보 610 nm - 1000 nm. The applicator is cooled by a closed water cooling circuit built in the device. #### Intended Use The Or-Light® is designed to perform : - Hair Removal for skintype I to IV ■ - Benign pigmentary lesions for skintype I to III 디 - Benign vascular lesions for skintype I to III . - Treatment of inflammatory Acne for skintype I to III 트 ### Comparing technical characteristics/ Performance Data The differences in the specification of the Or-Light® and the predicate devices do not result in different performances or raise any new question of safety or effectiveness. The clinical data demonstrate that the device can be used effectively and safely by a trained skin professionnal. #### Summary Based on the foregoing, we believe that the Or-Light® is substantially equivalent to the legally marketed predicate devices, the L600® and the Ellipse I2PL. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized symbol that resembles three wavy lines or strokes, which is the department's logo. The logo is positioned to the right of the text, creating a balanced composition. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 YPERION Technology % Pascal Danet General Manager 2, rue Beethoven 75016 Paris France OCT - 7 2009 Re: K090724 Trade/Device Name: System Or-Light® Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONF Dated: September 18, 2009 Received: September 22, 2009 Dear Pascal Danet: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical {3}------------------------------------------------ Page 2 - Pascal Danet device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indication for use Applicant : YPERION TECHNOLOGY SAS 510(k) Number (if known): K090724 Device Name: System Or-Light® Indications of Use: The Or-Light® is a non-coherent light-based device intended to carry out dermatologic and aesthetic care. It is meant to be handled by competent professionals trained in intense pulsed light technology. Two different applicators can be used on the Or-Light® device. The HR applicator (610-1000 nm) is indicated for: - Hair Removal for skintype I to IV ■ The SR applicator (400-1000 nm) is indicated for - Benign pigmentary lesions for skintype I to III ■ - Benign vascular lesions for skintype I to III 트 - Inflammatory acne treatment for skintype I to III 피 And/Or X Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) Over-The-Counter Use _ (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Neil R.P. Oakes-Forman (Division Sign-One), Orihopedic, Division of Surgices Division of Sere and Restorative Devices 510(k) Number K090724
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...