← Product Code [OMP](/submissions/SU/subpart-e%E2%80%94surgical-devices/OMP) · K142626

# Invia Liberty Negative Pressure Wound Therapy System, Invia Liberty Tubing Set, Y-connector, Drain Adaptor, Invia Liberty Canisters (K142626)

_Medela AG · OMP · Jun 12, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OMP/K142626

## Device Facts

- **Applicant:** Medela AG
- **Product Code:** [OMP](/submissions/SU/subpart-e%E2%80%94surgical-devices/OMP.md)
- **Decision Date:** Jun 12, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4780
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Medela® Invia Liberty Negative Pressure Wound Therapy System is indicated to help promote wound healing, through means including drainage and removal of infectious material or other fluids, under the influence of continuous and/or intermittent negative pressures, particularly for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.

## Device Story

Invia Liberty NPWT System is a portable, battery-operated suction pump for negative pressure wound therapy; used in hospitals or home settings by medical personnel or trained lay users. System inputs include wound exudate and air; device applies continuous or intermittent negative pressure to the wound site via a double-lumen tubing set and canister. The pump features an optical and acoustic status display to alert users to variances from set values or faults. Healthcare providers use the device to manage wound drainage and promote healing. The system includes a pump, canister/tubing set, power supply, and carrying case. It is compatible with Medela gauze kits and Mölnlycke foam kits. The device benefits patients by facilitating wound healing through controlled suction.

## Clinical Evidence

No clinical data. Bench testing only, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), home use safety (IEC 60601-1-11), usability (IEC 60601-1-6, IEC 62366), biocompatibility, and performance testing of the pump with gauze and foam kits.

## Technological Characteristics

Powered suction pump; ABS housing; double-lumen tubing with quick-connectors; rechargeable battery; LCD display with 5-button keypad; audio/visual alarms. Suction range: -40 to -200 mmHg. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-6, IEC 62366. Sterilization for sterile accessories. Firmware-controlled operation.

## Regulatory Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

## Predicate Devices

- Invia Wound Therapy ([K080357](/device/K080357.md))

## Reference Devices

- Invia Motion ([K113678](/device/K113678.md))
- Avance NPWT kits ([K141847](/device/K141847.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three overlapping faces.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 12, 2015

Medela AG % Ms. Adrienne Lenz Pathways Regulatory Consulting, LLC W324 S3649 County Road East Dousman, Wisconsin 53118

Re: K142626

Trade/Device Name: Invia Liberty Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: May 11, 2015 Received: May 12, 2015

Dear Ms. Lenz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincerely yours,

# David Krause -S

- Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K142626

#### Device Name

Invia Liberty Negative Pressure Wound Therapy System

Indications for Use (Describe)

The Medela® Invia Liberty Negative Pressure Wound Therapy System is indicated to help promote wound healing, through means including drainage and removal of infectious material or other the influence of continuous and/or intermittent negative pressures, particularly for patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.

| Type of Use (Select one or both, as applicable)                                                                               |                                                                                |
|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

DATE: June 11, 2015

### SUBMITTER:

Medela AG Lättichstrasse 4b 6341 Baar / Switzerland Phone +41 (0)41 769 51 51 Fax + 41 (0)41 769 51 00

PRIMARY CONTACT PERSON:

Adrienne Lenz, RAC Member Pathway Regulatory Consulting, LLC T 262-290-0023

SECONDARY CONTACT PERSON:

Orlando Atunes Vice President Regulatory Affairs Medela AG

#### DEVICE:

TRADE NAME: Invia Liberty Negative Pressure Wound Therapy System

COMMON/USUAL NAME: Negative Pressure Wound Therapy Pump and Accessories

CLASSIFICATION NAMES: 21 CFR 878.4780 Powered Suction Pump

PRODUCT CODE: OMP

## PREDICATE DEVICE(S):

K080357 Invia Wound Therapy

{4}------------------------------------------------

# DEVICE DESCRIPTION:

The Medela Invia Liberty Negative Pressure Wound Therapy (NPWT) System is comprised of the Invia Liberty NPWT Pump, canister/tubing set, power supply, carrying case, patient and user instructions, and Invia NPWT kits. The Invia Liberty is also compatible with Avance NPWT kits manufactured by Mölnlyke Healthcare.

The Invia Liberty NPWT pump is a suction pump for Negative Pressure Wound Therapy with an optical and acoustic status display. Acoustic and optical signals are triggered for variances from the set values as well as for faults. The Invia Liberty NPWT pump provides continuous or intermittent operation.

The Invia Liberty NPWT system may be used in a home or other health care facility by medical personnel or trained lay users adhering to the instructions for use. The user may not be hard of hearing or deaf and must have normal visual acuity. The Invia Liberty NPWT pump is portable and can be operated independent of the electrical power supply via its rechargeable battery.

# INTENDED USE:

The Medela® Invia Liberty Negative Pressure Wound Therapy System is indicated to help promote wound healing, through means including drainage and removal of infectious material or other fluids, under the influence of continuous and/or intermittent negative pressures, particularly for patients with chronic, acute, traumatic, subacute and dehisced wounds, partialthickness burns, ulcers (such as diabetic or pressure), flaps and grafts.

# DETERMINATION OF SUBSTANTIAL EQUIVALENCE:

## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE

The Invia Liberty NPWT System uses the same fundamental technology as Invia Wound Therapy (K080357) for most features. The Invia Liberty NPWT System is identical to the Invia Wound Therapy in its indications for use and contraindications. The main differences between the Invia Liberty NPWT System and the predicate Invia Wound Therapy are:

- New double lumen pump tubing with quick connector coupling. The smaller lumen (air flush tubing) flushes air to remove exudate from the larger lumen. Larger lumen (suction tubing) removes the fluid from the wound into the canister. The quick connector will ensure that the connection is easy to connect / disconnect and tight.
- . New Drain Adapter to connect quick connect tubing to drains
- New Y-Connector for quick connect compatibility
- Firmware change, including improved flush algorithm
- Housing material change to ABS to improve the stability and cleanability
- Additional IFU for patient in home care environment

{5}------------------------------------------------

The table below summarizes the key specifications of the Invia Liberty NPWT System and the predicate devices.

|                                   | Invia Liberty Negative<br>Pressure Wound<br>Therapy System                                                                                                                                                                                                | Invia Wound Therapy<br>(K080357)                                                                                                                                                                      | Discussion                                                                                                                                                                                                                                         |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient<br>Population             | Adults                                                                                                                                                                                                                                                    | Adults                                                                                                                                                                                                | Identical. Pumps are<br>reusable and should be<br>cleaned and<br>disinfected between<br>patients.                                                                                                                                                  |
| Useful Life                       | 4000 hours                                                                                                                                                                                                                                                | 4000 hours                                                                                                                                                                                            | Identical                                                                                                                                                                                                                                          |
| Environment of<br>Use             | Hospital and Home                                                                                                                                                                                                                                         | Hospital and Home                                                                                                                                                                                     | Identical                                                                                                                                                                                                                                          |
| User Interface                    | Five button keypad<br>(power and arrow keys<br>to navigate menus),<br>LCD Display, audio<br>indicators.                                                                                                                                                   | Five button keypad<br>(power and arrow keys<br>to navigate menus),<br>LCD Display, audio<br>indicators.                                                                                               | Identical                                                                                                                                                                                                                                          |
| Accessories                       | • Disposable Canister<br>0.8 l, 0.3 l<br>• Quick connect<br>double lumen tubing<br>set<br>• Drain Adapter<br>• Y-connector<br>• Mains adapter<br>• Docking station<br>• Carrying Case<br>• Holder with<br>standard rail<br>• NPWT kits sold<br>separately | • Disposable Canister<br>0.8 l, 0.3 l<br>• Tubing set<br>• Y-connector<br>• Mains adapter<br>• Docking station<br>• Carrying Case<br>• Holder with<br>standard rail<br>• NPWT kits sold<br>separately | Equivalent. The tubing<br>set is modified with a<br>double lumen and<br>Quick Connector<br>coupling between the<br>pump tubing and the<br>drain adapter, which<br>now connects to the<br>Dressing Kits. All other<br>accessories are<br>identical. |
| Specifications                    |                                                                                                                                                                                                                                                           |                                                                                                                                                                                                       |                                                                                                                                                                                                                                                    |
| Suction capacity<br>l/min         | 5 l/min                                                                                                                                                                                                                                                   | 5 l/min                                                                                                                                                                                               | Identical                                                                                                                                                                                                                                          |
|                                   | Invia Liberty Negative<br>Pressure Wound<br>Therapy System                                                                                                                                                                                                | Invia Wound Therapy<br>(K080357)                                                                                                                                                                      | Discussion                                                                                                                                                                                                                                         |
| Max. vacuum<br>mmHg/kPa           | - 200mmHg<br>-27kPa                                                                                                                                                                                                                                       | - 200mmHg<br>-27kPa                                                                                                                                                                                   | Identical                                                                                                                                                                                                                                          |
| Min. vacuum<br>mmHg/kPa           | - 40mmHg<br>-5.3kPa                                                                                                                                                                                                                                       | - 60mmHg<br>-8kPa                                                                                                                                                                                     | Equivalent. Minimum<br>vacuum level changed<br>to -40mmHg to enable<br>therapy for sensitive<br>wounds (i.e. burns)                                                                                                                                |
| Therapy modes                     | Continuous &<br>Intermittent                                                                                                                                                                                                                              | Continuous &<br>Intermittent                                                                                                                                                                          | Identical                                                                                                                                                                                                                                          |
| Canister capacity<br>[ml]         | 300/800ml                                                                                                                                                                                                                                                 | 300/800ml                                                                                                                                                                                             | Identical                                                                                                                                                                                                                                          |
| Weight [kg]                       | 1.0kg                                                                                                                                                                                                                                                     | 1.0kg                                                                                                                                                                                                 | Identical                                                                                                                                                                                                                                          |
| Dimensions<br>mm                  | 290x95x235mm<br>(0.3l canister)                                                                                                                                                                                                                           | 290x95x235mm<br>(0.31 canister)                                                                                                                                                                       | Identical                                                                                                                                                                                                                                          |
| Standard Safety<br>device         | Bacteria / overflow /<br>protection filter                                                                                                                                                                                                                | Bacteria / overflow /<br>protection filter                                                                                                                                                            | Identical                                                                                                                                                                                                                                          |
| NPWT Kits                         | Multiple available<br>including the following                                                                                                                                                                                                             | Multiple available<br>including the following                                                                                                                                                         |                                                                                                                                                                                                                                                    |
| Dressing (to pack<br>the wound)   | Kerlix AMD Gauze or<br>Molnlyke foam                                                                                                                                                                                                                      | Kerlix AMD Gauze                                                                                                                                                                                      | Equivalent. Gauze Kits<br>used with Invia Liberty<br>and Invia Motion<br>pumps are cleared in<br>K113678. Molnlycke<br>foam kits are cleared in<br>K141847.                                                                                        |
| Sterile Saline                    | Winchester<br>Laboratories Sodium<br>Chloride 0.9%w/v                                                                                                                                                                                                     | Winchester<br>Laboratories Sodium<br>Chloride 0.9%w/v                                                                                                                                                 | Identical                                                                                                                                                                                                                                          |
| No sting barrier<br>film          | 3M Cavilon No Sting<br>Barrier Film                                                                                                                                                                                                                       | 3M Cavilon No Sting<br>Barrier Film                                                                                                                                                                   | Identical                                                                                                                                                                                                                                          |
|                                   | Invia Liberty Negative<br>Pressure Wound<br>Therapy System                                                                                                                                                                                                | Invia Wound Therapy<br>(K080357)                                                                                                                                                                      | Discussion                                                                                                                                                                                                                                         |
| Contact layer                     | 3M Tegaderm Contact<br>(OEM labeled)                                                                                                                                                                                                                      | 3M Tegaderm Contact<br>(Medela labeled)                                                                                                                                                               | Equivalent.  Only the<br>labeling has changed.<br>The kits now include<br>original 3M labeling.                                                                                                                                                    |
| Dressing to seal<br>the wound     | 3M Tegaderm Film<br>(OEM labeled)                                                                                                                                                                                                                         | 3M Tegaderm Film<br>(Medela labeled)                                                                                                                                                                  | Equivalent.  Only the<br>labeling has changed.<br>The kits now include<br>original 3M labeling.                                                                                                                                                    |
| Additional drain<br>seal material | Coloplast Stoma Paste                                                                                                                                                                                                                                     | NA                                                                                                                                                                                                    | Equivalent. The<br>Coloplast Stoma Paste<br>is used to fill crevices<br>or uneven surfaces to<br>improve seal in difficult<br>to dress areas.                                                                                                      |
| Tape                              | HyTape                                                                                                                                                                                                                                                    | HyTape                                                                                                                                                                                                | Identical                                                                                                                                                                                                                                          |
| Drains                            | Round channel (15 Fr),<br>Flat (10 mm), External<br>Suction Interface (ESI)                                                                                                                                                                               | Round channel (15 Fr),<br>Flat (10 mm)                                                                                                                                                                | Equivalent. Drains,<br>including ESI, are<br>cleared via K113678 for<br>use with Invia Motion<br>and Invia Liberty<br>pumps.  Selection of<br>drain is made by the<br>health care provider.                                                        |

{6}------------------------------------------------

{7}------------------------------------------------

# SUMMARY OF NON-CLINICAL TESTS:

The Invia Liberty Negative Pressure Wound Therapy System complies with voluntary standards for electrical safety, electromagnetic compatibility, and safety of home use devices. The following quality assurance measures were applied to the development of the system:

- Risk Analysis
- Software Validation ●
- Hardware Integration Testing
- Electrical safety and electromagnetic compatibility testing per IEC 60601-1 and IEC ● 60601-1-2 standards, respectively

{8}------------------------------------------------

- Safety testing for use in the home per IEC 60601-1-11 standard
- Usability evaluation per the IEC 60601-1-6 and IEC 62366 Standards.
- 0 Performance Testing demonstrating the functionality of the Invia Liberty pump in combination with the Medela Invia Gauze Kits and Avance Foam Kits. Testing with the gauze kits included use of the double lumen tubing and drain adaptor and testing of the foam kits also used with double lumen tubing with the Avance ViewPad.
- Testing of the quick connector demonstrating adequate tightness and tear-out force.
- Biocompatibility testing, including cytotoxicity, sensitization, irritation.
- Validation of cleaning methods for the reusable pump housing.
- Sterilization validation and confirmation of shelf life for the sterile tubing, Y-connector and drain adapter.

# SUMMARY OF CLINICAL TESTS:

The Invia Liberty Negative Pressure Wound Therapy System has not been the subject of clinical testing.

# CONCLUSION:

Medela AG considers the Invia Liberty Negative Pressure Wound Therapy System to be as safe as, as effective as, and substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OMP/K142626](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OMP/K142626)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
