K132252 · Devon Medical · OMP · Sep 27, 2013 · General, Plastic Surgery
Device Facts
Record ID
K132252
Device Name
EXTRICARE 100CC/ 250CC/ 400CC CANISTERS
Applicant
Devon Medical
Product Code
OMP · General, Plastic Surgery
Decision Date
Sep 27, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Intended Use
extriCARE® 2400 Negative Pressure Wound Therapy System is a portable, battery powered pump intended to generate negative pressure or suction to remove wound exudates, infectious material, and tissue debris from the wound bed which may promote wound healing.
Device Story
The extriCARE® Negative Pressure Wound Therapy System is a portable, battery-powered pump designed to provide negative pressure (suction) to a wound site. The system utilizes 100cc, 250cc, or 400cc canisters to collect wound exudates, infectious material, and tissue debris removed from the wound bed. By maintaining negative pressure, the device aims to promote wound healing. It is intended for use in clinical or home settings to assist in the management of wounds.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use.
Technological Characteristics
Powered suction pump; battery-operated; includes 100cc, 250cc, and 400cc collection canisters. Classified under 21 CFR 878.4780, Product Code OMP.
Indications for Use
Indicated for patients requiring negative pressure wound therapy to remove wound exudates, infectious material, and tissue debris from the wound bed to promote healing.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Related Devices
K161599 — UNO Negative Pressure Wound Therapy System · Genadyne Biotechnologies, Inc. · Apr 6, 2017
K180840 — UNO Negative Pressure Wound Therapy System · Genadyne Biotechnologies, Inc. · Nov 30, 2018
K131464 — EXTRICARE NPWT FOOT DRESSING · Devon Medical, Inc. · Aug 13, 2013
K143726 — XLR8 PLUS (XLR8+) · Genadyne Biotechnologies, Inc. · Jun 19, 2015
K090638 — GENADYNE A4-XLR8 WOUND VACUUM SYSTEM · Genadyne Biotechnologies, Inc. · Apr 29, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## September 27, 2013
Devon Medical Incorporated Ms. Ruth Wu Chief Operating Officer 1100 156 Avenue, Suite 202 King of Prussia. Pennsylvania 19406
Re: K132252
Trade/Device Name: ExtriCARE® Negative Pressure Wound Therapy 100cc, 250cc, and 400cc Canisters Regulation Number: 21 CFR 878.4780 · Regulation Name: Powered Suction Pump Regulatory Class: II Product Code: OMP Dated: August 26, 2013 Received: August 28, 2013
Dear Ms. Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enacunent date of the Medical Device Amendments, or to connered processified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act (Act that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The roundly, mercers, manys of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean r reast be devices of the Fermination that your device complies with other requirements of the Act mar i Dr. has Intact and regulations administered by other Federal agencies. You must or any I edetal station and registements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
Page 2 - Ms. Ruth Wu
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Long H. ------Chen -A - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Mark N. Melkerson, M.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): _K132252_
- Device Name: extriCARE® Negative Pressure Wound Therapy 100cc canister extriCARE® Negative Pressure Wound Therapy 250cc canister extriCARE® Negative Pressure Wound Therapy 400cc canister
Indications for Use:
extriCARE® 2400 Negative Pressure Wound Therapy System is a portable, battery powered pump intended to generate negative pressure or suction to remove wound exudates, infectious material, and tissue debris from the wound bed which may promote wound healing.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR -(Part 21 CFR 801 Subpart D)
Over-The-Counter Use_ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
## Jiyoung Dang -S
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.