← Product Code [OMP](/submissions/SU/subpart-e%E2%80%94surgical-devices/OMP) · K111949

# AC WOUND TREATMENT SYSTEM (K111949)

_Nanovibronix, Inc. · OMP · Jun 29, 2012 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OMP/K111949

## Device Facts

- **Applicant:** Nanovibronix, Inc.
- **Product Code:** [OMP](/submissions/SU/subpart-e%E2%80%94surgical-devices/OMP.md)
- **Decision Date:** Jun 29, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4780
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

AC Wound Management System is indicated for patients who would benefit from a suction device (negative pressure) to help promote wound healing by removing fluids, including irrigation and body fluids, wound exudates, and infectious materials. Examples of appropriate wound types include: Diabetic/Neuropathic ulcers, Pressure ulcers, Chronic wounds, Acute wounds, Dehisced wound, Partial-thickness burns, and Flaps and Grafts.

## Device Story

AC Wound Management System is a powered suction pump for negative pressure wound therapy (NPWT). System components include a non-disposable pump drive, a 300cc waste canister with integrated suction tube, and an FDA-cleared generic NPWT dressing. The pump drive, powered by a rechargeable lithium-ion battery or mains power, uses a reciprocating piston to generate negative pressure (-50 to -175 mmHg). The device is used in clinical or home settings, operated by healthcare providers or patients. A vacuum level monitor continuously tracks pressure to maintain therapy settings. Exudates are drawn from the wound site through the tube into the canister, which features a hydrophobic filter. Users interact via a keyboard interface with audible/visual alarms. The system can be mounted to IV poles, bedside rails, or carried. By removing fluids and infectious materials, the device promotes wound healing. Healthcare providers use the system to manage complex wounds, with output monitored via the device's electronic display.

## Clinical Evidence

No clinical studies were conducted. Substantial equivalence is supported by comparative bench testing demonstrating stable operation and pressure accuracy (deviation ≤1%) at -50, -100, and -175 mmHg over 72-hour durations. Electrical safety and EMC compliance were verified per IEC 60601-1 and IEC 60601-1-2.

## Technological Characteristics

Powered suction pump; rechargeable lithium-ion battery; 100-240V AC input. Pressure range: -50 to -175 mmHg. Components: pump drive, 300cc canister with hydrophobic filter, suction tube. Dimensions: 190x140x70mm; weight: 0.82kg. Connectivity: standalone. Compliance: IEC 60601-1, IEC 60601-1-2. Firmware-controlled pressure regulation.

## Regulatory Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

## Predicate Devices

- Renasys™ GO Negative Pressure Wound Therapy ([K083375](/device/K083375.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary

JUN 2 9 2012

#### Date Prepared

June 25, 2012

Sponsor Information

NanoVibronix, Inc. 105 Maxess Road, Suite S124 Melville, NY 11747

807.92(a){1)

| Sponsor Contact        | Harold Jacob                                  |
|------------------------|-----------------------------------------------|
| Title                  | CEO                                           |
| Phone                  | 1-516-668-3185                                |
| Device Name/Trade Name | AC Wound Management System 807.92(a)(2)       |
| Common/Usual Name      | Negative Pressure Wound Therapy Pump          |
| Classification Name    | Powered Suction Pump                          |
| Regulation Number      | 21 CFR 878.4780                               |
| Product Code           | OMP                                           |
| Review Panel           | Surgical, Orthopedic, and Restorative Devices |
| Device Classification  | Class II                                      |

#### Predicate Device

#### 807.92(a)(3)

| Company Name         | Brand Name                                     | 510K Number |
|----------------------|------------------------------------------------|-------------|
| Smith & Nephew, Inc. | Renasys™ GO Negative<br>Pressure Wound Therapy | K083375     |

Device Description

807.92(1)(4)

The AC Wound Management System is a powered suction pump that utilizes a pump drive to generate light negative pressure. The pump drive is powered by a rechargeable battery. Optionally the unit can be connected to mains power using an included power converter.

AC Wound Management System has a negative pressure setting range of -50 mmHg to -175 mmHg which is electronically monitored and controlled. The system includes a keyboard for user interface as well as audible and visual alarm indicators. The system includes a 300cc canister. The system can be set on a countertop, carried using a shoulder bag or mounted to an IV pole or bedside rail using the AC Wound Management System's mount.

{1}------------------------------------------------

K111949
p. 2 of 5

The AC Wound Management System includes the following three components:

- The pump drive. .
- The waste canister, with its suction tube (the two are connected . permanently).
- FDA cleared generic NPWT dressing, which will be distributed by ● NanoVibronix.

The pump drive is a non-disposable unit which contains a motor, a battery and a user interface. The drive has a piston which protrudes the drive and reciprocates to activate the suction pump, which is integral to the canister. In the drive there is also a vacuum level monitor which monitors the vacuum at all times and maintains the reguired set therapy.

The waste canister attaches to the drive. The pump creates vacuum within the waste canister for the purpose of suctioning exudates from the wound and collecting them in the canister. The canister has a hydrophobic filter.

The suction tube connects the waste canister to the wound dressing applying the vacuum generated within the waste canister to the wound site. The tube also transfers the exudates from the wound to the waste canister by maintaining flow from the wound site towards the waste canister at all times.

#### Indications for Use

807.92(a)(5)

AC Wound Management System is indicated for patients who would benefit from a suction device (negative pressure) to help promote wound healing by removing fluids, including irrigation and body fluids, wound exudates, and infectious materials. Examples of appropriate wound types include: Diabetic/Neuropathic ulcers, Pressure ulcers, Chronic wounds, Acute wounds, Dehisced wound, Partial-thickness burns, and Flaps and Grafts.

#### Predicate Product Comparison Table

807.92(1){6}

| Parameters                            | NanoVibronix, Inc.<br>AC Wound Management<br>System                                                                                                                                                                                                                                                                        | Smith & Nephew, Inc.<br>Renasys Go                                                                                                                                                                                                                                                       |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                         |                                                                                                                                                                                                                                                                                                                            | K083375                                                                                                                                                                                                                                                                                  |
| Indications for Use                   | AC Wound Management<br>System is indicated for<br>patients who would benefit<br>from a suction device<br>(negative pressure) to help<br>promote wound healing by                                                                                                                                                           | Renasys Go is indicated for<br>patients who would benefit<br>from a suction device<br>(negative pressure) to help<br>promote wound healing by<br>removing fluids including                                                                                                               |
|                                       |                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                          |
|                                       | removing fluids, including<br>irrigation and body fluids,<br>wound exudates, and<br>infectious materials.<br>Examples of appropriate<br>wound types include:<br>Diabetic/Neuropathic ulcers,<br>Pressure ulcers, Chronic<br>wounds, Acute wounds,<br>Dehisced wound, Partial-<br>thickness burns, and Flaps<br>and Grafts. | irrigation and body fluids,<br>wound exudates and<br>infectious materials.<br>Examples of appropriate<br>wound types include: chronic,<br>acute, traumatic, sub-acute<br>and dehisced wounds, ulcers<br>(such as pressure or diabetic),<br>partial-thickness burns, flaps<br>and grafts. |
| Mode of Operation                     | Continuous or Intermittent                                                                                                                                                                                                                                                                                                 | Continuous or Intermittent                                                                                                                                                                                                                                                               |
| Maximum negative<br>pressure          | -175mmHg                                                                                                                                                                                                                                                                                                                   | -200 mmHg                                                                                                                                                                                                                                                                                |
| Negative Pressure<br>Therapy Settings | -50mmHg, -75mmHg,<br>-100mmHg, -125mmHg,<br>-150mmHg, -175mmHg                                                                                                                                                                                                                                                             | From -40 mmHg to<br>-200mmHg                                                                                                                                                                                                                                                             |
| Power Requirements                    | Input: 100-240V AC 50/60Hz<br>0.6A<br>Output: 12V DC, 2.08A, 5W                                                                                                                                                                                                                                                            | Input: 100-240V AC 50/60Hz<br>0.9A<br>Output: 21V DC, 1.71A, 36W                                                                                                                                                                                                                         |
| Battery Type                          | Lithium ion rechargeable                                                                                                                                                                                                                                                                                                   | Lithium ion rechargeable                                                                                                                                                                                                                                                                 |
| Battery Operating Time                | ~20 hours (therapy)                                                                                                                                                                                                                                                                                                        | ~ 20 hours (therapy)                                                                                                                                                                                                                                                                     |
| Dimensions                            | 190x140x70mm<br>(7.4 x 5.5 x 2.7 inches)                                                                                                                                                                                                                                                                                   | 175x210x85mm<br>(7 x 8.3 x 3.5 inches)                                                                                                                                                                                                                                                   |
| Weight                                | 0.82kg (1.8 lbs)                                                                                                                                                                                                                                                                                                           | 1.1kg (2.4lbs)                                                                                                                                                                                                                                                                           |
| Operating Environment                 | 5-35°C (41-95°F)<br>30 to 70% RH<br>700 to 1060 mbar<br>atmospheric pressure                                                                                                                                                                                                                                               | 5-35°C (41-95°F)<br>30 to 70% RH<br>700 to 1060 mbar<br>atmospheric pressure                                                                                                                                                                                                             |
| Canister Capacity                     | 300 ml                                                                                                                                                                                                                                                                                                                     | 300 ml                                                                                                                                                                                                                                                                                   |

{2}------------------------------------------------

### Comparative Bench Testing

#### 807.92(b)(1)

A comparative bench test was performed to determine the substantial equivalence of the AC Wound Management System to the Renasys™ Negative Pressure Wound Therapy System.

- 1. The results of the tests show that both AC NPWT system and predicated device Renasys NPWT system - provide stable operation.
3

{3}------------------------------------------------

K111949
p. 4 of 5

- 2. The results were obtained for three negative pressure levels: -50mmHg, -100mmHg and -175mmHg. Test duration was 72 hours for each level..
- 3. For both systems the deviation between determined negative pressure level and measured average negative pressure level was no more then 1% ( in four measurement points).
- 4. We conclude that AC NPWT system overcomes the acceptance criteria

## Electrical and Bench Tests

AC Wound Management System complies with IEC 60601-1 and IEC 60601-1-2 Standards.

| Bench test                       | Results                                                                               |
|----------------------------------|---------------------------------------------------------------------------------------|
| Electrical Safety                | Test results demonstrated compliance<br>with the standard.                            |
| EMC Compatibility                | Test results demonstrated compliance<br>with the standard.                            |
| Firmware verification            | Test results demonstrated compliance<br>with labeling and the standard.               |
| Functionality test               | Test results demonstrated compliance<br>with labeling and the standard.               |
| Pressure stability test for pump | Test results positive                                                                 |
| Battery test                     | Test results demonstrated 20 hours work<br>therapy (max pressure, continuous<br>mode) |
| Labeling                         | Device comply with standards                                                          |

{4}------------------------------------------------

Device comply with requirements

### Clinical Tests

807.92(b)(2)

No clinical studies were conducted.

Conclusion

807.92(b)(3)

5

The AC Wound Management System and the Renasys™ Negative Pressure Wound Therapy System have the same indications for use and modes of operation. Both products use rechargeable battery, have same canister capacity, use similar pressure settings and have similar dimensions & weight. Therefore, the AC Wound Management System does not raise any new issues of safety and effectiveness. The AC Wound Management System is substantially equivalent to the Renasys™ marketed device and is safe and effective for the intended use.

{5}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

AC Wound Treatment System % Smith Associates ·Mr. EJ Smith 1468 Harwell Avenue Crofton, Maryland 21114

JUN 2 9 2012

Re: K11949

> Trade/Device Name: AC Wound Management System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered sunction pump Regulatory Class: Class II Product Code: OMP Dated: June 15, 2012 Received: June 15, 2012

Dear Mr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2- Mr. EJ Smith

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

# < 111949

# Indications for Use

510(k) Number (if known):

Device Name: AC Wound Management System

Indications for Use:

AC Wound Management System is indicated for patients who would benefit from a suction device (negative pressure) to help promote wound healing by removing fluids, including irrigation and body fluids, wound exudates, and infectious materials. Examples of appropriate wound types include: Diabetic/Neuropathic ulcers, Pressure ulcers, Chronic wounds, Acute wounds, Dehisced wounds, Partial-thickness burns, and Flaps and Grafts.

Prescription Use _ V (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David Kane for MMM
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K11949

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OMP/K111949](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OMP/K111949)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
