K110078 · Devon Medical Products, Inc. · OMP · Feb 10, 2012 · General, Plastic Surgery
Device Facts
Record ID
K110078
Device Name
EXTRICARE
Applicant
Devon Medical Products, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Feb 10, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The extriCARE 2400 Negative Pressure Wound Therapy Pump System is a portable, battery powered pump intended to generate negative pressure or suction to remove wound exudates, infectious material, and tissue debris from the wound bed which may promote wound healing.
Device Story
Portable, battery-powered negative pressure wound therapy (NPWT) system; consists of vacuum pump, canister, and tubing. Connects to all-in-one wound dressing; no wound packing required. Operates by applying digitally programmed negative pressure to wound bed; removes exudates, infectious material, and tissue debris. Used in clinical or home settings; operated by healthcare providers or patients. Output allows clinicians to manage wound healing via controlled suction. Benefits patient by facilitating wound bed preparation and promoting healing.
Clinical Evidence
Bench testing only. Performance verified through functional and mechanical testing, including vacuum pressure calibration, intermittent mode treat time verification, flow rate performance, alarm/battery verification, and exudate removal. Bandage performance verified via peel strength, shear adhesion, and leakage tests. Biocompatibility testing conducted per ISO 10993-1 for prolonged contact (24h-30 days), including sensitization, cytotoxicity (MEM elution), and intracutaneous irritation; all tests passed.
Indicated for patients requiring negative pressure wound therapy to remove wound exudates, infectious material, and tissue debris from the wound bed to promote healing.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows a handwritten number and a fraction. The number "5110078" is written at the top of the image. Below the number, the fraction "1/3" is written in a larger font size.
# 510(K) SUMMARY
#### Submitter:
Devon Medical Products, Inc.
#### Contact Person:
Ruth Wu, COO 1100 First Avenue, Suite 202 King of Prussia, PA 19406 Phone: 866.446.0092 484.636.0211 Fax:
## Common Classification & Proprietary Names:
| Common Names: | Negative Pressure Wound Therapy System |
|-------------------|----------------------------------------|
| Proprietary Name: | extriCARE 2400 |
#### Date Prepared:
August 29, 2011
#### Classification
The classification name, 21 CFR Part and Paragraph number, product code and classification of the extriCARE 2400:
| | FDA | Classification | Product Code | Class |
|-----------------|-----|----------------------|--------------|-------|
| 21 CFR 878.4780 | | Powered Suction Pump | OMP | II |
## Predicate Devices:
The extriCARE 2400 Negative Pressure Wound Therapy System is substantially equivalent to the following.
| Predicate Device | Manufacturer | 510(k)# |
|-------------------------------|-----------------|---------|
| NPD 1000 | Kalypto Medical | K080275 |
| MobIVAC W.C. | Ohio Medical | K081185 |
| Renasys Gauze<br>Dressing Kit | Smith & Nephew | K110647 |
### Device Description
The system consists of a vacuum pump, canister, and tubing. In operation, the device is attached via the tubing to a Negative Pressure Wound bandage. Maximum pressure and mode selection are digitally programmable. The extriCARE pump can be sold alone or as a part of the extriCARE 2400 system with extriCARE bandages. The extriCARE bandages are an all-in-one wound dressing, with no wound packing required.
#### Intended Use:
{1}------------------------------------------------
# SECTION 5
The extriCARE 2400 Negative Pressure Wound Therapy System is a portable, battery powered pump intended to generate negative pressure or suction to remove wound exudates, infectious material, and tissue debris from the wound bed which may promote wound healing.
0078
## Technological Characteristics:
The manufacturer believes that the technological characteristic of the extriCARE 2400 are substantially similar to those of the predicate devices. The extriCARE 2400 has very similar components to its predicate devices and very similar principles of operation. The device consists of an electrically generated source of vacuum, a canister. Like the predicates, vacuum, is applied for a specified period of time and intensity, according to the physician's prescription.
## Performance Testing
To verify that the device design met it's functional and performance requirements, representative samples of the device underwent function and mechanical testing, EMC testing in accordance with IEC 60601-1-2 and electrical safety testing in accordance with UL 60601-1.
| Function Performance Tests |
|-----------------------------------------------------------------------|
| PQ-01 extriCARE 2400 Vacuum Pressure Calibration |
| PQ-02 Comparison on SE for Vacuum Pressure Calibration |
| PQ-03 Treat Time verification on Intermittent Mode for extriCARE 2400 |
| PQ-04 Flow Rate Performance on for extriCARE 2400 |
| PQ-05 Alarm Verification for extriCARE 2400 |
| PQ-06 Battery Verification for extriCARE 2400 |
| PQ-07 Exudate Removal test on extriCARE 2400 and extriCARE bandage |
| PQ-08 Peel strength test on extriCARE Bandage T-ZG-027 |
| PQ-09 Shear Adhesion test on extriCARE Bandage T-ZG-065 |
| PQ-10 Leakage test on extriCARE Bandage |
### Biocompatibility
The extriCARE 2400 system consists of both the pump and the bandage. While the pump has no direct body contact when used as indicated, the bandage does have direct body contact. Per ISO-10993 B-prolonged contact(24h - 30 days) requirements, the bandage samples were tested for the following items and all tests were successfully passed.
| | | Date | Protocol # | Result Summary |
|--------|---------------------------------------------------------------|--------------|--------------------------|------------------------------------------------------------------|
| Test 1 | ISO Guinea Pig<br>Maximization<br>Sensitization Test<br>(GLP) | Oct 8, 2010 | Wuxi AppTec<br>900850-23 | The test article<br>did not elicit<br>sensitization<br>response. |
| Test 2 | Cytotoxicity - MEM | Oct 29, 2010 | Nelson | Non-cytotoxic |
{2}------------------------------------------------
# SECTION 5
K110078
3/3
| | Elution without silver (ISO) | | 201003532 Rev 01 | |
|--------|-------------------------------------------------|--------------|----------------------------|---------------|
| Test 3 | Cytotoxicity – MEM<br>Elution with silver (ISO) | Oct 21, 2010 | Nelson<br>201003441 Rev 01 | Non-cytotoxic |
| Test 4 | Intracutaneous Irritation | Nov 11, 2010 | NLI# 550001 | Non-irritant |
## Statement of Substantial Equivalence
The extriCARE 2400 system is substantially equivalent in technology, function, operating parameters, and intended use to predicate devices that are currently commercially available and in distribution, and does not raise any new questions of safety or effectiveness.
### Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, Devon Medical Products, believes that the extriCARE 2400, is safe and effective and substantially equivalent to the predicate devices as described herein.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around it. Inside the circle is an abstract image of an eagle or bird-like figure with stylized feathers.
Devon Medical Products, Inc. % Ms. Ruth Wu. COO 1100 First Avenue, Suite 202 King of Prussia, Pennsylvania 19406
Re: K110078
Trade/Device Name: extriCARE 2400 Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: January 18, 2012 Received: January 31, 2012
Dear Ms. Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
FEB 1 0 2012
{4}------------------------------------------------
Page 2 - Ms. Ruth Woo, COO
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely y
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## SECTION 4
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): _
Device Name: extriCARE 2400
The extriCARE 2400 Negative Pressure Wound Therapy Pump System is a portable, battery powered pump intended to generate negative pressure or suction to remove wound exudates, infectious material, and tissue debris from the wound bed which may promote wound healing.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Kucera for MIXM
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110078