K080897 · Talley Medical · OMP · Jul 15, 2008 · General, Plastic Surgery
Device Facts
Record ID
K080897
Device Name
VENTURI NEGATIVE PRESSURE WOUND THERAPY SYSTEM
Applicant
Talley Medical
Product Code
OMP · General, Plastic Surgery
Decision Date
Jul 15, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Talley Venturi™ system is intended for use for patients with acute or chronic wounds that may be benefited by the application of negative pressure therapy and the potential wound healing effects of removal of fluids including wound exudates, irrigation fluids, body fluids, and infectious materials.
Device Story
Talley Venturi™ is a powered suction pump for negative pressure wound therapy (NPWT). System applies vacuum to wounds to remove exudates, irrigation fluids, and infectious materials. Consumables include collection canister, connection tubing, drain, and wound dressings. Used in acute care settings; operated by healthcare professionals. Device monitors vacuum levels and canister status via sensors. Healthcare providers use output to manage wound healing through fluid removal. Benefits include potential promotion of wound healing in acute or chronic wound patients.
Clinical Evidence
Bench testing only. Testing included 100% production unit verification for air-tightness, canister recognition, vacuum pump flow and pressure performance, and 48-hour soak testing. All units passed defined minimum values for pressure calibration and alarm functionality.
Technological Characteristics
Powered suction pump for NPWT. Includes collection canister, tubing, drains, and wound dressings. Operates via vacuum generation. Features sensor-based canister recognition and pressure monitoring. System is designed for acute care use. No specific materials or software architecture details provided beyond functional performance specifications.
Indications for Use
Indicated for patients with acute or chronic wounds requiring negative pressure therapy for fluid removal (exudates, irrigation, infectious materials). Intended for acute care settings. Contraindicated for necrotic tissue, untreated osteomyelitis, fistula, malignant tissue, exposed vasculature, nerves, anastomotic sites, bone, or tendons, and wounds with difficult hemostasis.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
K080897
## 510(k) Summary
## Date 15 March 2008
Name: Talley Group, Ltd Premier Way, Abbey Park Romsey, Hants SO 51 9 DQ England
JUL 1 5 2008
U.S. Talley Medical 4740 Jadestone Dr. Williamston, MI 48895 Phone: 517-290-0089 Fax: 484-465-5842 Contact: Jack Van Dyke, Director US Operations
Trade Name: Venturi™ Negative Pressure Wound Therapy
Common Name: Powered Suction Pump
| Device Classification: | Class II |
|------------------------|---------------------------------------------------|
| | Product Code: JCX |
| | Regulation 878.4780 |
| | Classification Panel: General and Plastic Surgery |
| Predicate Devices: | Blue Sky Versatile 1 System K042134, K052456 |
| | Boehringer Labs Suction Pump system K061788 |
Device Description: The Talley Venturi™ system consists of a powered suction pump for the application of vacuum to wounds for fluid removal. Consumables for use with the pump include collection canister, connection tubing, drain, and wound dressing products.
Intended Use: The Talley Venturi™ system is intended for use for patients with acute or chronic wounds that may be benefited by the application of negative pressure therapy and the potential wound healing effects of removal of fluids including wound exudates, irrigation fluids, body fluids, and infectious materials.
Technological Characteristics: The Talley Venturi™ system includes the same type suction pump as the predicate devices, operating at the same pressure ranges. Consumable accessories include collection canister, tubing, drains, and wound dressing components, These accessories correspond with accessories available with the predicate devices.
Specific Performance Testing: a) Canister vacuum test is performed on 100% of production units to check for air-tightness and recognition of sensor pins. Loss of vacuum is measured over a prescribed period. Units pass or fail. b) Vacuum pump test is performed on 100% of production sub-assemblies measuring flow and pressure. Tested levels are compared with defined minimum values to determine pass or fail. c) Soak test is performed on 100% of assembled production pump units and are run for 48 hours prior to final test. d) Final system test is performed on 100% of production units. Tests include all functions and buttons, correct pressure calibration, air tightness, canister recognition, warning systems and alarms, charging system operational.
Conclusion: The Talley Venturi™ system is substantially equivalent to the predicate devices.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular fashion around the symbol. The text is in all caps and is oriented to follow the curve of the circle.
Public Health Service
'APR - 7 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Talley Medical % Mr. Jack Van Dyke 4740 Jadestone Drive Williamston, Michigan 48895
Re: K080897
Trade/Device Name: Venturi™ Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: II Product Code: OMP Dated: June 3, 2008 Received: June 10, 2008
Dear Mr. Van Dyke:
This letter corrects our substantially equivalent letter of July 15, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not
{2}------------------------------------------------
Page 2 - Mr. Jack Van Dyke
limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Rottmo
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for use
Venturi™ Negative Pressure Wound Therapy system
Indications for Use. Use of the Venturi™ Negative Pressure Wound Therapy system is indicated for use for patients with acute or chronic wounds that may be benefited by the application of neqative pressure therapy and the potential wound healing effects of removal of fluids including wound exudates, irrigation fluids, and infectious materials. Venturi is intended for use in acute care settings only.
Contraindications. The Venturi™ Negative Pressure Wound Therapy system is contraindicated in the presence of:
Necrotic tissue Untreated osteomyelitis Fistula Wounds with malignant tissue Exposed vasculature Exposed nerves Exposed anastomotic site Exposed bone or tendons Wounds with difficult hemostasis
Prescription Use × (21 CFR 801 Subpart D)
OR
Over the Counter Use (21 CFR 807 Subpart C)
PLEASE DO NOT WRITE BELOWATHIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
| Concurrance of CDRH, Office of Device Evaluation (ODE) | A market of the program and the production of the program and the many of the may be |
|--------------------------------------------------------|--------------------------------------------------------------------------------------|
|--------------------------------------------------------|--------------------------------------------------------------------------------------|
Division Sign-O Division of General, Restorative, and Neurological De
510(k) Number LC68977