← Product Code [OLP](/submissions/SU/subpart-e%E2%80%94surgical-devices/OLP) · K142246

# LightStim for Acne/ LightStim for Acne Mini (K142246)

_Led Intellectual Properties, LLC · OLP · Jan 29, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OLP/K142246

## Device Facts

- **Applicant:** Led Intellectual Properties, LLC
- **Product Code:** [OLP](/submissions/SU/subpart-e%E2%80%94surgical-devices/OLP.md)
- **Decision Date:** Jan 29, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Pediatric

## Intended Use

LightStim for Acne/LightStim for Acne Mini, model: LS4A/LS4AM is intended for Over-The-Counter use for the treatment of mild to moderate acne.

## Device Story

Hand-held device utilizing low-intensity light-emitting diodes (LEDs) at 415nm and 630nm wavelengths; intended for OTC use by patients for treatment of mild to moderate acne. Device consists of hand piece, LEDs, PCB, on/off switch, and external 9V AC/DC power supply. User applies device directly to skin; treatment duration (3-4 minutes) controlled manually by user. Light exposure targets acne; no complex processing or automated algorithms involved. Benefits include home-based management of acne symptoms.

## Clinical Evidence

Usability study conducted on predicate device (K131461) with 40 participants (ages 11-61). Results demonstrated participants could correctly determine device suitability for their acne level and comprehended instructions, warnings, and precautions.

## Technological Characteristics

Hand-held device; LEDs at 415nm and 630nm; 9V AC/DC power supply; PCB-based control; manual operation. No complex software or connectivity.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Ultra Renew Plus ([K132833](/device/K132833.md))
- LightStim for Acne Mini ([K131461](/device/K131461.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing right, with flowing lines above and below them, resembling a bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 29, 2015

LED Intellectual Properties, LLC Mr. Steve Marchese Chief Executive Officer 16552 Von Karman Avenue Irvine, California 92606

Re: K142246

Trade/Device Name: LightStim for Acne/ LightStim for Acne Mini Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OLP Dated: December 23, 2014 Received: December 29, 2014

Dear Mr. Marchese:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Jennifer R. Stevenson -A

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K142246

Device Name

LightStim for Acne/ LightStim for Acne Mini

Indications for Use (Describe)

LightStim for Acne/ LightStim for Acne Mini is an over-the-counter hand-held device intended for the use in the treatment of mild to moderate acne.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary of Safety & Effectiveness

LED Intellectual Properties, LLC.

Device: LightStim for Acne & LightStim for Acne Mini

### 1. General Information

Submitter: LED Intellectual Properties, LLC 16552 Von Karman Ave. Irvine, CA 92606

Contact Person: Steve Marchese Office: (949) 502-4088 Mobile: (949) 394-2427

Date Prepared: August 1, 2014

### 2. Names and Code

Device Name: LightStim for Acne/LightStim for Acne Mini, model: LS4A/LS4AM

Classification Name: Laser Instrument, Surgical Powered - General and Plastic Surgery - Class II, OLP

Although this device is not a laser and is intended for OTC use, the manufacturer thinks this is the closet applicable classification name.

#### 3. Predicate Devices

Ultra Renew Plus (K132833) and LightStim for Acne Mini (K131461

## 4. Device Description

LightStim for Acne / LightStim for Acne Mini, model: LS4A/LS4AM is a hand-held device consisting of low intensity light emitting diodes (LED's) that provide illumination which comes in contact with the skin. The device components include LED's of 415nm and 630nm, a hand piece, a printed circuit board, an on/off switch, a resistor, a receiver jack in the hand piece to plug a power supply into and a separate AC to DC (9 

{4}------------------------------------------------

volt) power supply. Treatment time is recommended to be three to four minutes and is controlled by the user.

#### 5. Indications for Use / Intended Use

LightStim for Acne/LightStim for Acne Mini, model: LS4A/LS4AM is intended for Over-The-Counter use for the treatment of mild to moderate acne.

#### 6. Performance Data

A usability study was performed on the LightStim for Acne Mini K131461. which is the exact same device but with 415nm Blue LEDs only. This study was conducted with 40 participants. The participants ranged in age from 11 to 61. The results showed that the participants adequately decided whether or not to use the device for their level of acne, and showed that users comprehend risks, warnings, cautions, precautions, and proper use of the device, from the Instruction Manual.

Taking into consideration the (attached) table for substantial equivalence evidencing both predicates being handheld, OTC, utilizing blue light and K132833 also utilizing red light, which poses no new issues of safety or effectiveness as the active wavelength in the treatment of mild to moderate acne is in the blue region of light, thus evidencing the primary objective of acne treatment are common in all of theses devices, and after an analysis of safety, indications, intended uses, performance, features, technological properties and methods of operations. LED Intellectual Properties, LLC believes that LightStim for Acne/LightStim for Acne Mini has shown substantial equivalence to the predicate devices Ultra Renew Plus (K132833) and LightStim for Acne Mini (K131461.

The conclusion drawn by LED Intellectual Properties, LLC is that LightStim for Acne/ LightStim for Acne Mini raise no new issues of safety and are substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OLP/K142246](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OLP/K142246)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
