OKJ · Light Based Treatment For Cold Sores Herpes Simplex Virus-1

General, Plastic Surgery · 21 CFR 878.4860 · Class 2

Overview

Product CodeOKJ
Device NameLight Based Treatment For Cold Sores Herpes Simplex Virus-1
Regulation21 CFR 878.4860
Device ClassClass 2
Review PanelGeneral, Plastic Surgery

Identification

The device emits light energy at near infrared spectrum and is applied externally to the surface of herpes simplex labialis lesions on or around the lips.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the light based energy source device for topical application is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, including wavelength, treatment time, treatment area, energy density, spot size, and power, must be characterized. (2) The cleaning and disinfection instructions for the device must be validated. (3) The device must be demonstrated to be biocompatible. (4) Performance testing must validate electromagnetic compatibility (EMC), ocular safety, and electrical safety of the device. (5) Labeling must direct end-users to contact the device manufacturer and MedWatch if they experience any adverse events when using this device. (6) Labeling must include specific information pertinent to use of the device by the intended patient population and the treatment regimen. (7) Simulated use testing must include information from a usability, label comprehension and self-selection study to demonstrate that the device can be used by the intended patient population without any assistance. (8) Clinical data must show adequate reduction in time to healing and assess risks of redness, discomfort, burns, and blisters.

Recent Cleared Devices (4 of 4)

RecordDevice NameApplicantDecision DateDecision
K251973Luminance Red Cold Sore Device (TN1927G)Luminance Medical Ventures, Inc.Sep 23, 2025SESE
K241155Cold Sore Device (QPZ-03)Shenzhen Nuon Medical Equipment Co., Ltd.Sep 5, 2024SESE
K222205Cold Sore Device (Model: QPZ-01)Light Tree Ventures Europe B.V.Oct 7, 2022SESE
DEN090012VIRULITE COLD SORE MACHINEVirulite, LLCOct 18, 2012DENG

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...